oticon Charger Case 1.0 miniRITER Instructions

June 6, 2024
oticon

Instructions for use


Charger 1.0 miniRITE R

Oticon More™, Oticon Opn S™, Oticon Opn Play™
Oticon Ruby, Oticon CROS PX

Introduction to this booklet

This booklet guides you on how to use and maintain your new charger.
Ensure you read this booklet carefully, including the Warnings section. This will help you get the most benefit from your charger.
If you have additional questions about the use of your charger, contact your hearing care professional.
For your convenience, this booklet contains a navigation bar to help you navigate through the different sections.

IMPORTANT NOTICE
Please refer to your hearing aid’s Instructions for use for details on the functionality of your specific hearing aid.

Intended use

Intended use| The charger is intended to recharge the hearing aids and CROS transmitter that contain rechargeable batteries.
The product is intended to be used together with the given hearing solution.
---|---
Indications for use| There are no indications for use (diagnoses) for the charger itself.
Intended user| Person with hearing loss using a hearing aid and their caregivers. Hearing care professional responsible for adjusting the accessory.
Intended user group| Adults and children older than 36 months.
Use environment| Home setting.
Contraindications| No contraindications.
Clinical benefits| See clinical benefits of the hearing aid.

About

Package content

The packaging box consists of the following elements:

Charger

What it is and does

Temperature conditions
Temperature Charging condition
Below +5°C (+41°F) No charging due to low temperature
+10°C to +35°C (+50°F to +95°F) Best charging condition. Charging takes up to

3 hours.
+5°C to +10°C and +35°C to +40°C (+41°F to 50°F and +95°F to +104°F)| Extended charging. Charging takes up to 4 hours.
Above +40°C (+104°F)| No charging due to high temperature

IMPORTANT NOTICE
Ensure you always leave the charger connected to a power source when the hearing aid is seated in the charging port. Turning the charger OFF makes the hearing aid turn ON and start using battery power.

Start-up

Prepare charger

Connect to power source

| Charger is turned ON

---|---
1. Insert the USB plug into the power plug.
2. Insert the power plug into a power socket.
The charger turns ON automatically.| When the charger is connected to power, the GREEN LED turns ON.

Charge the hearing aids

Place hearing aids in charger

| Charging

---|---
Place the hearing aids in the charging ports as shown in the picture. Make sure that the hearing aids match the right and left colour indicators on the charger.
The RED indicator is for the right side hearing aid. The BLUE indicator is for the left side hearing aid.| The hearing aid LED lights turn ON when the hearing aids are correctly placed in the charger.
Hearing aid LED light charging status:
ORANGE / RED = Charging GREEN = Fully charged
The colour is either orange or red depending on the hearing aid family. For more information see your hearing aid’s instructions for use.

Charging time

Ensure you fully charge your hearing aid before first time use.
Charge your hearing aid every night. That ensures you start your day with a fully charged hearing aid.
Be aware that charging time may vary depending on the remaining capacity of the battery and between the left / right side hearing aid.

If your hearing aid’s battery is completely drained the normal charging time is:

3 hours 1 hour 0.5 hour
Fully charged 50% charged 25 % charged

When the battery is fully charged, the charging process stops automatically.
Leave the hearing aid in the charger when you do not use it.
Remember to keep the charger connected to a power source at all times.

Note
If you have not used your hearing aid in a long time and you then try to recharge it, it may take up to 30 seconds before the hearing aid begins charging. For more information, see your hearing aid’s instructions for use.

Handling

Power the charger using other sources

It is possible to power the charger from other sources than a power socket. This can be done by using the charger’s USB cable.
Ensure that the power source is USB 2.0 compliant, with a minimum of 500mA output.

Examples of power sources:

  • Power bank
  • Computer
  • Car

Note
When using a power bank, ensure that it does not have a power-saving or auto- shutdown mode. As this can result in your hearing aid not being fully charged or not charged at all.

IMPORTANT NOTICE
When using a power socket, use the supplied original power supply (5.0V/1A) for the charger.

Caring for your charger

Always ensure that the charger is dry and clean before use. Keep the charging ports clean from dust by using a dry cloth. Do not use water or cleaning fluids to clean the charging ports.
Store your charger in a dry location, e.g. in the bedroom.
Ensure you keep the charger in an environment with a temperature between +5°C and +40°C (+41°F and +104°F).

****Warnings

For your personal safety and to ensure correct usage, you should familiarise yourself fully with the following general warnings before using your charger. Consult your hearing care professional if you experience unexpected operations or serious incidents with your charger during use or because of its use. Your hearing care professional will support you with issue handling and, if relevant, reporting to the manufacturer and/or the national authorities.

Usage of charger
The charger should be used only as directed and adjusted by your hearing care professional.

Risk of burns
Do not place anything but the hearing aids in the charger ports.

Choking hazards
Charger, hearing aids and their parts should be kept out of reach of children and anyone who might swallow these items or otherwise cause injury to themselves.
If a small part is swallowed, see a doctor or hospital immediately.

Heat
The charger must never be exposed to extreme heat, e.g. left inside a parked car in the sun.
The charger must not be dried in microwave ovens or other ovens.

Connection to external equipment
The safety of the charger, when connected to external equipment with USB cable, is determined by the external signal source. When the charger is connected to external equipment which is plugged into a power socket, this equipment must comply with IEC 62368-1 or equivalent safety standards.

Fire hazards
The power source in your charger has insufficient energy to cause fire in normal usage conditions.

More info

Troubleshooting
Symptom Possible causes Solutions
The charger LED remains turned OFF when the charger is connected to power The
charger is not turned ON Verify that the power plug is correctly connected

Ensure that charging has started, to make sure the power source delivers sufficient output for your charger
The charger LED shows long ORANGE blinks| The hearing aid or charger’s temperature is too warm| Move the charger and hearing aid to a location with a temperature between +5°C and +40°C (+41°F and +104°F)
The charger LED repeatedly shows three short ORANGE blinks| The charger’s temperature is too cold| Move the charger to a location with a temperature between +5°C and +40°C (+41°F and +104°F)
System error| Contact your hearing care professional
The hearing aid LED blinks GREEN when hearing aid is placed in the charger| Hearing aid has not been in use for an extended period of time| Depending on how depleted the battery in the hearing aid is, the hearing aid automatically resumes normal charging mode after a timeframe of up to 10 minutes
Ensure you leave the hearing aid in the charger during this process
The hearing aid LED blinks ORANGE / RED when the hearing aid is placed in the charger| The charger is not turned ON| Verify that the charger power plug is correctly connected
The hearing aid or charger’s temperature is either too warm or too cold| Move the charger and hearing aid to a location with a temperature between +5°C and +40°C (+41°F and +104°F)
Charging is incomplete. The room temperature exceeds +35°C (+95°F), which extends the charging time. The charger has stopped charging to protect the battery.| Reinsert the hearing aid into the charger. This completes the charging within approximately 15 minutes.
The hearing aid is not correctly seated in the charger| Check the charging ports for foreign objects
System error| Contact your hearing care professional

More info

Charger: Conditions of use

Operational environment conditions| Temperature: +5°C to +40°C (+41°F to +104°F)
Humidity: 5% to 93%, non-condensing
Atmospheric pressure: 700 hPa to 1060 hPa
---|---
Transportation and storage conditions| Temperature and humidity shall not exceed the below limits for extended periods during transportation and storage:
Temperature: -25°C to +70°C (-13°F to +158°F)
Humidity: 5% to 93%, non-condensing

For information about your hearing aid’s Conditions of use, see your hearing aid’s instructions for use.

Compliance statements

USA and Canada
This device complies with part 18 of the FCC Rules.
This device complies with FCC and Innovation, Science and Economic Development Canada’s RF radiation exposure limits set forth for an uncontrolled environment.
This device should be installed and operated with minimum distance of 20 centimeters (0.6 feet) between the device and your body. Use of other accessories not verified by the manufacturer may not ensure compliance with FCC and Innovation, Science and Economic Development Canada’s RF exposure guidelines.

User information
This equipment has been tested and found to comply with the limits pursuant to part 18 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful interference in a residential installation.
This equipment generates, uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.

However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to other electronic equipment, which can be determined by turning the equipment OFF and ON, the user is encouraged to try to correct the interference by one or more of the following measures:

  • Reorient or relocate the equipment
  • Increase the separation between the equipment

Changes or modifications not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment.

This medical device complies with the Medical Device Regulation (EU) 2017/745.
Declaration of Conformity is available from the headquarters:

Oticon A/S
Kongebakken 9
DK-2765 Smørum Denmark
www.oticon.global/doc

SBO Hearing A/S
Kongebakken9
DK-2765 Smorum
Denmark|
Waste from electronic equipment must be handled according to local regulations.
---|---

Australian Sponsor
Audmet Australia Pty Ltd
Suite 4, Level 4, Building B
11 Talavera Road
North Ryde NSW 2113

Description of symbols and abbreviations used in this booklet

**| Warnings
Text marked with a warning symbol must be read before using the device.
---|---
|
Manufacturer
The device is produced by the manufacturer whose name and address are stated next to the symbol. Indicates the medical device manufacturer, as defined in EU Regulations 2017/745.
|
CE mark
The device complies with all required EU regulations and directives.
| E
lectronic waste (WEEE)
Recycle hearing aids, accessories or batteries according to local regulations. Hearing aid users can also return electronic waste to their hearing care professional for disposal. Electronic equipment covered by Directive 2012/19/EU on waste and electrical equipment (WEEE).
|
Regulatory Compliance Mark (RCM)**
The device complies with electrical safety, EMC and radio spectrum requirements for devices supplied to the Australian or New Zealand market.

Description of additional symbols used on labels

| Keep dry
Indicates a medical device that needs to be protected from moisture.
---|---
| Caution symbol
Consult instructions for use for warnings and cautions.
| Catalogue number
Indicates the manufacturer’s catalogue number so that the medical device can be identified.
| Serial number
Indicates the manufacturer’s serial number so that a specific medical device can be identified.
| Medical Device
The device is a medical device.
| Power supply outlet
Indicates the rate at which power is sent from the outlet to your electrical device.

International warranty

Your device is covered by an international limited warranty issued by the manufacturer for a period of 12 months from the date of delivery to you. This limited warranty covers manufacturing and material defects in the device itself, but not in accessories such as e.g. batteries, cables, power supplies, etc. Problems arising from improper/ incorrect handling or care, excessive use, accidents, repairs made by an unauthorised party, exposure to corrosive conditions, damage due to foreign objects entering the device or incorrect adjustments are NOT covered by the limited warranty and may void it. Under this warranty, the manufacturer will choose, in its sole discretion, whether to repair the device, or replace it with an equivalent model. The above warranty does not affect any legal rights that you might have under applicable national legislation governing sale of consumer goods. Your hearing care professional may have issued a warranty that goes beyond the clauses of this limited warranty. Please consult him/her for further information.

If you need service
Take your device to your hearing care professional, who may be able to sort out minor problems and adjustments immediately. Your hearing care professional can assist you in obtaining warranty service from the manufacturer. Your hearing care professional may charge a fee for their services.

Australia only

The rights given under this warranty are in addition to other rights and remedies under a law in relation to the good or services to which the warranty relates to. Our goods come with guarantees that cannot be excluded under the Australian Consumer Law. You are entitled to a replacement or refund for a major failure and compensation for any other reasonably foreseeable loss or damage. You are also entitled to have the goods repaired or replaced if the goods fail to be of acceptable quality and the failure does not amount to a major failure.

Australian customers should contact Oticon Australia at Suite 4, Level 4, Building B, 11 Talavera Road, North Ryde NSW 2113, telephone 1800 633 738 if you have any questions regarding this warranty.

Warranty

Certificate
Name of owner: ____
Hearing care professional:

Hearing care professional’s address: ____
Hearing care professional’s phone: ____
Purchase date: ____
Model:
Serial no.: __

223138AU / 2021.05.24 / v3

Documents / Resources

| oticon Charger Case 1.0 miniRITER [pdf] Instructions
Charger Case 1.0 miniRITER
---|---

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals