HASWILL ELECTRONICS Hastat Software for Digital Data Loggers User Manual
- June 5, 2024
- HASWILL ELECTRONICS
Table of Contents
HASWILL ELECTRONICS Hastat Software for Digital Data Loggers
Interface Language
The default language of this datalogger software is Chinese; just need a few
clicks to show it in English; the steps as follow:
Only Chinese and English available.
Latest Data
Once running this tool, it will detect the USB Datalogger and read the data, and plot a curve automatically. This data processing may spends 10s to 5 minutes according to the size of the data volume. Once data ready, you will find 3 tabs that show the latest data
Latest Data – Summary Info
You will find the below content from the above interface
- Device Information
- Configuration Information
- Logging Summary
Latest Data – Curve Graph
Scroll the mouse to zoom in/out the graph.
Latest Data – original data
Max records 48,000 Groups Default Records each minitue 5 / 11
Configuration
With this software, you may change the default parameters, the frequently-used parameters in the “Basic Mode” and others in advance mode.
Basic Configuration Mode
Items | Description |
---|---|
Normal Record Interval | min 10s, max 24 hours, step 10s. |
Over Limit Record Interval | The record period once finds temperature over the |
limitation, min 10s, max 24 hours, step 10s.
Temperature Upper Limit| To define the range in which means acceptable, out of
which adopts the “Over Limit Record Interval.”
Temperature Lower Limit
Advanced Configuration Mode
Section 1
Cange the temperature unit.
✓ Give a trip name
✓ Select the start mode:
✓ Manual start
■ Key time-delay start (start delay time editable)
■ Timing start
■ Always start
There are 4 sections on this viewport
✓ Set the stop mode:
■ Manual Stop
■ Timing Stop
✓ Set the “Normal Record Interval time “ and the “Over Limit Record Interval
time” like what you can do on the basic configuration mode, and there is an
extra option that only exists here, that is “Alarm Delay time.
Section 2
Set the upper and lower temperature limits to trigger the visible alert, which displays an icon.
Section 3
Set the language and the file format of the temperature record report. It is
the Report Language here, not indicates the software interface language.
Section 4
Items | Options & Description |
---|---|
Circular log | Stops or Circular when reach max records |
Multiple start | Disable (Default)/Enable |
LCD always on | Always on or Timing off, default auto-dim in 25s. |
Clear saved record | No clear or Clear |
Clear data by long press | Disable/Enable |
All Data Records
This tab shows all the saved data in a datalogger.
All Records – Temperature Curve/Chart
Just double-click one of the records from the left column, then you will
get the corresponding statistics, curve graph, and the original detailed data
list.
All Records – Temperature Statistic
All Records – Temperature Deta List / Orignal Data
Thanks!
https://www.thermo-hygro.com
Documents / Resources
|
HASWILL ELECTRONICS Hastat Software for Digital Data
Loggers
[pdf] User Manual
HDT-U114B, HDT-U114E, HDT-U115B, HDT-U115E, HDT-U135B, HDT-U135E, Hastat
Software for Digital Data Loggers
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>