HASWILL ELECTRONICS Hastat Software for Digital Data Loggers User Manual

June 5, 2024
HASWILL ELECTRONICS

HASWILL ELECTRONICS Hastat Software for Digital Data Loggers

Interface Language

The default language of this datalogger software is Chinese; just need a few clicks to show it in English; the steps as follow:

Only Chinese and English available.

Latest Data

Once running this tool, it will detect the USB Datalogger and read the data, and plot a curve automatically. This data processing may spends 10s to 5 minutes according to the size of the data volume. Once data ready, you will find 3 tabs that show the latest data

Latest Data – Summary Info

You will find the below content from the above interface

  • Device Information
  • Configuration Information
  • Logging Summary
Latest Data – Curve Graph


Scroll the mouse to zoom in/out the graph.

Latest Data – original data


Max records 48,000 Groups Default Records each minitue 5 / 11

Configuration

With this software, you may change the default parameters, the frequently-used parameters in the “Basic Mode” and others in advance mode.

Basic Configuration Mode
Items Description
Normal Record Interval min 10s, max 24 hours, step 10s.
Over Limit Record Interval The record period once finds temperature over the

limitation, min 10s, max 24 hours, step 10s.
Temperature Upper Limit| To define the range in which means acceptable, out of which adopts the “Over Limit Record Interval.”
Temperature Lower Limit

Advanced Configuration Mode
Section 1

Cange the temperature unit.
✓ Give a trip name
✓ Select the start mode:
✓ Manual start
■ Key time-delay start (start delay time editable)
■ Timing start
■ Always start

There are 4 sections on this viewport

✓ Set the stop mode:
■ Manual Stop
■ Timing Stop
✓ Set the “Normal Record Interval time “ and the “Over Limit Record Interval time” like what you can do on the basic configuration mode, and there is an extra option that only exists here, that is “Alarm Delay time.

Section 2

Set the upper and lower temperature limits to trigger the visible alert, which displays an icon.

Section 3

Set the language and the file format of the temperature record report. It is the Report Language here, not indicates the software interface language.

Section 4
Items Options & Description
Circular log Stops or Circular when reach max records
Multiple start Disable (Default)/Enable
LCD always on Always on or Timing off, default auto-dim in 25s.
Clear saved record No clear or Clear
Clear data by long press Disable/Enable

All Data Records

This tab shows all the saved data in a datalogger.

All Records – Temperature Curve/Chart

Just double-click one of the records from the left column, then you will get the corresponding statistics, curve graph, and the original detailed data list.

All Records – Temperature Statistic

All Records – Temperature Deta List / Orignal Data

Thanks!
https://www.thermo-hygro.com

Documents / Resources

| HASWILL ELECTRONICS Hastat Software for Digital Data Loggers [pdf] User Manual
HDT-U114B, HDT-U114E, HDT-U115B, HDT-U115E, HDT-U135B, HDT-U135E, Hastat Software for Digital Data Loggers
---|---

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

HASWILL ELECTRONICS User Manuals

Related Manuals