Quick start guide

• These operating instructions describe the basic handling of the device.
• Please refer to the operating instructions available at www.sauermanngroup.com for safe use of the product and detailed information.
• Do not give this device to a child.
• The magnets in the back of the instrument can damage credit cards, hard drives, mechanical watches, pacemakers, defibrillators and other devices proven sensitive to magnetic fields. It is recommended to keep the instrument at a distance of at least 25 cm (10”) away from any of these devices.
• These gas analysers are designed to measure flue gases exhausting from combustion equipment. They are not designed for continuous operations and/or use as a control/alarm system

The Si-CA 130 and Si-CA 230 analysers are touchscreen devices with just the on/off button:

Before Using the Flue Gas Analyser

WARNING!

  • Be sure all connections are tight to assure accurate sampling.
  • When conducting measurements, the water trap/filter assembly MUST be in a VERTICAL position.

When testing is completed, always drain any accumulated condensation inside the water trap (after EVERY test!).
The condensate be slightly acidic, so avoid contact with skin and make sure condensate does not leak out of water trap.

Analysers description

Analysers description

  1. Min-DIN connection for CO and CO2 probe
  2. Thermocouple connection for air temperature
  3. Thermocouple connection for flue gas temperature
  4. P- connection
  5. P+ connection
  6. Gas connection
  7. USB connection
Operating temperature, protection of the instruments and information

about storage

  • Operating temperature: from -5 to 45 °C (23 to 113 °F)
  • Protection: IP42

If the instrument is stored outside the operating temperature (for example in a van, a warehouse,…) please let the analyser acclimate to room temperature for at least 15 minutes before starting and operating the analyser.
If the ambient temperature is lower than 10 °C (50 °F), the analyser should be acclimated to ambient temperature before performing measurements.

First start-up

DO NOT insert the gas probe in the flue/chimney/stack.
Keep pressed for 3 seconds to turn on the analyser.

Tap “Start”.

First start-up 01

Select your country then tap “Next”.

First start-up 02

Select your language then tap “Next”.

First start-up 03

Set the date then tap “Next”.

First start-up 04

Set the time then tap “Next”.

First start-up 05

Set your time zone then tap “Next”.

First start-up 06

Select if you want your pump turned ON or OFF after Auto-Zero then tap “Save”.

First start-up 07

Autozero is in progess.

First start-up 08

The probe can be inserted in the chimney. The analyser displays the measurements.

First start-up 09

ON / OFF

During autozero DO NOT insert the gas probe in the chimney.
First start-up 09

  • Keep pressed for 3 seconds to turn on the analyser.
    Once the autozero is finished, insert the gas probe in the chimney and gas measurements are displayed.

  • Keep pressed for 3 seconds to turn off the analyser. Before powering off, disconnect hoses from the analyser and empty and dry the water trap.

Stop / Restart the Pump

It is possible to turn off the pump during a gas analysis.

Tap

  • Pump is off and values are on hold.
  • Pump is on and measurements re-start.

Combustion Analysis, Saving & Printing

Before proceeding with the combustion analysis, remember to set the proper fuel.

Tap “Settings”.Combustion Analysis, Saving & Printing
01

Tap “Analysis”.

Combustion Analysis, Saving & Printing 02

Tap “Fuels”.

Combustion Analysis, Saving & Printing 03

Tap the required fuel.

Combustion Analysis, Saving & Printing 04

Once the required fuel is selected, tap then tap “Gas analysis”.

Combustion Analysis, Saving & Printing 05 .Measurements of the gas analysis are displayed.Combustion Analysis,
Saving & Printing 06
You can also quickly access the fuel selection menu via the fuel shown at the top of the combustion gas analysis screen.

  • To view the different measured parameters, just scroll the screen.
  • To save combustion data, first tap on the Save icon then select or create a folder and tap on Save. The Sauermann Combustion mobile phone app and/or PC software can also be used to save data.
  • To prints the results, use the Wireless Printer available as option.

The Sauermann Combustion app lets you control your Si-CA analyser remotely, view measurement results in real time, and customise them before exporting in the format of your choice.
QR CODE 01

Download app

Graph view

Documents / Resources

| sauermann SI-CA 130 Two Gas Commercial Combustion Analyzer Kit [pdf] User Guide
SI-CA 130, SI-CA 230, Two Gas Commercial Combustion Analyzer Kit
---|---
| sauermann SI-CA 130 Two Gas Commercial Combustion Analyzer Kit [pdf] User Guide
SI-CA 130, SI-CA 230, SI-CA 130 Two Gas Commercial Combustion Analyzer Kit, SI-CA 130, Two Gas Commercial Combustion Analyzer Kit, Commercial Combustion Analyzer Kit, Combustion Analyzer Kit, Analyzer Kit

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals