SKC 225-3841 Preloaded Disposable Parallel Particle Impactors Instruction Manual
- June 5, 2024
- SKC
Table of Contents
225-3841 Preloaded Disposable Parallel Particle Impactors
Instruction Manual
863 Valley View Road, Eighty Four, PA 15330
724-941-9701
SKC Preloaded Disposable Parallel Particle Impactors (PPI)
The patented†
impaction-based SKC Disposable Parallel Particle Impactor (PPI®) Samplers are
designed to match precisely the collection efficiency curves for respirable
and thoracic dust specified by ISO 7708/CEN and adopted by ACGIH, CEN, and
other occupational hygiene organizations. The performance of the respirable
PPI samplers also meets the ISO 7708/CEN criteria included in the OSHA final
silica rule. The thoracic model meets the requirements of compounds with
ACGIH® thoracic TLV®s. Disposable PPI Samplers offer the convenience of
single-use to eliminate sampler assembly and cleaning; small size for worker
comfort, even under helmets or other PPE; and a choice of flow rates for
maximum flexibility in pump options, sample duration, and contaminant
concentration. These instructions are for PPI models pre-assembled with
filter, support, and impaction substrates.
Preloaded filters in some models are available preweighed by SKC for
gravimetric analysis. See the table below.
Respirable PPIs | Preloaded Filter |
---|
225-3841
225-3851
225-3871| 5.0-μm PVC filter for chemical analysis, not preweighed. For
gravimetric analysis, contact your laboratory for pre and post-weight filter.
225-3841-PW
225-3851-PW
225-3871-PW| 5.0-μm PVC filter preweighed by SKC for gravimetric analysis
Preweighed at 70 F ± 3 F and 50% ± 5% relative humidity
Thoracic PPIs| Preloaded Filter
225-3861| 0.8-μm MCE fi lter, not preweighed
225-3862| Quartz fi lter, not preweighed
† U.S. Patent No. 7,073,402
Performance Profile
Sampling Rate: 2 L/min respirable or thoracic and 4 or 8 L/min respirable
Sample Pump:
- Universal or AirChek® Series for 2 and 4 L/min
- Leland Legacy for 8 L/min
Sample Time: Dependent on the method used. Note: SKC tests indicate
that a particulate mass of up to 6.8 mg on the four impaction substrates would
not affect PPI performance. This amount is equivalent to sampling for 6 hours
at 4 L/min in environments where respirable mass concentration is 4.76 mg/tit’
and equals 50% of total dust. However, labs have reported to SKC that they
prefer no more than 2 mg on the filter for analytical reasons. Therefore, SKC
recommends that you work with your lab to determine optimum sample times for
your unique sampling conditions.
Sample Media:
Respirable models: 37-mm, 5.0-μm PVC filter and cellulose support pad
Thoracic model: 37-mm, 0.8-μm MCE filter and cellulose support pad or
37-mm Quartz filter and cellulose support pad
Impaction Substrate: Four 3/8-in diameter pre-oiled porous plastic discs
(Assembled in all Disposable PPIs)
Analysis: Chemical or gravimetric
Body Material: Conductive ABS plastic
Dimensions:
Height: 4.25 in (10.8 cm) – clip to exhaust
Diameter: 1.8 in (4.6 cm)
Depth: 1.2 in (3 cm)
Weight: 1.1 oz (31.2 gm)
Shelf-life: 18 mos from date of manufacture
Principle of Operation
SKC Disposable PPI Samplers are impaction-based filter samplers that perform precise size-selection for either thoracic or respirable dust, depending on the model. PPI Samplers contain four small impactors in the inlet section of the device. Each impactor features a unique 50% cut-point to target a specifi c one-quarter segment of the ISO/CEN curve resulting in a precise fi t along the entire curve. A sample pump operating at 2, 4, or 8 L/min (2 L/min only for thoracic) pulls air through the inlet nozzle of each impactor in the inlet plate. Particles larger than each impactor’s 50% cut-point are scrubbed and retained by impaction onto the porous oiled impaction substrate contained in each impactor. Smaller particles continue to the standard 37-mm collection filter for analysis.
PPI Performance
Collection efficiency of the 2, 4, and 8 L/min respirable PPI Samplers compared to the ISO respirable curve
Collection efficiency of the 2 L/min thoracic PPI Sampler compared to the ISO thoracic convention
Sampler Preparation
SKC Preloaded Disposable PPIs are supplied assembled and ready to use.
For gravimetric analysis: PPIs 225-3841-PW, 225-3871-PW, and 225-3851-PW
are preloaded with a preweighed 5.0-μm PVC filter. If using PPIs preloaded with
a 5.0-μm PVC filter that has not been preweighed (Cat. Nos. 225-3841, 225-3871,
and 225-3851), contact your laboratory to pre and post weight filter.
- Remove shrink wrap.
- Write sample ID on sampling label. Adhere the sampling label to the bottom of the base plate.
- Remove the protective end cap from the exhaust.
Calibration and Sampling
**As the particle load on the filter increases during sampling, the
pressure drop will also increase. Therefore, use a compensating sample pump
such as the AirChek Series or Leland Legacy depending on flow rate
requirements.
Calibration
Calibrate pump flow rate with a representative Disposable PPI (support and
filter loaded) in line. Note: If using SKC High Flow chek-mate Calibrator,
Pulsation Dampener Cat. No. 375-150 is also required in line. See pump and
calibrator operating instructions.**
- Ensure the pump has run for 5 minutes before calibrating.
- Ensure the representative impactor is loaded with a support and collection filter and that it is fully assembled. See Sampler Preparation above.
- Align the bottom of the calibration adapter with the inlet plate of an assembled, loaded representative Disposable PPI and press down firmly until the calibration adapter’s O-ring is engaged and creates an even seal.
- Use flexible tubing to connect the exhaust of the Disposable PPI to the inlet of a sample pump.
- Use flexible tubing to connect the inlet of the calibration adapter to the suction port (outlet) of a calibrator.
- Calibrate to 2 L/min for the 2 L/min respirable and thoracic model PPIs, 4 L/min for the 4 L/min respirable model, or 8 L/min for the 8 L/min respirable model. Follow the calibration instructions in the pump and calibrator operating instructions.
- When calibration is completed, disconnect the tubing from the calibrator and calibration adapter.
- Grasp the Disposable PPI with one hand and the calibration adapter with the other hand. Firmly pull/twist to remove the calibration adapter from the Disposable PPI inlet plate.
- Replace the representative Disposable PPI used to set the flow with a new loaded Disposable PPI for sample collection.
Using excessive pressure to seal the calibration adapter to the Disposable PPI may make the calibration adapter difficult to remove.
Sampling
SKC tests indicate that a particulate mass of up to 6.8 mg on the four
impaction substrates would not affect PPI performance. This amount is
equivalent to sampling for 6 hours at 4 L/min in environments where respirable
mass concentration is 4.76 mg/m3 and equals 50% of total dust. However, labs
have reported to SKC that they prefer no more than 2 mg on the filter for
analytical reasons. Therefore, SKC recommends that you work with your lab to
determine optimum sample times for your unique sampling conditions.
- As per good industrial hygiene practice, replace the representative sampler used for calibration with a new, unused preloaded sampler.
- Record sample starts time on the label.
- Clip Disposable PPI onto a worker’s collar or lapel in the breathing zone or in the area to be sampled.
- Clip sample pump at the worker’s waist or close to the Disposable PPI.
- Use fl exible tubing to attach the Disposable PPI exhaust to the inlet of the sample pump.
- Turn on the pump and record pertinent sample data.
- After the desired sample time has elapsed, turn off the pump and unclip the sampler from the sampling location.
- Apply a Post-use label to the inlet plate to seal the sampler.
- Record sample stop time on a label on the bottom of the sampler.
- Disconnect sampler from the pump and reinstall protective end cap on the exhaust.
- Reinstate calibration train with a representative Disposable PPI and verify flow rate. See Calibration.
Sample Transport and Analysis
Package and transport samples and blanks to an accredited laboratory for
gravimetric or chemical analysis.
References
Traumas, S., Hall, P., Personal Respirable Sampler Containing Four Impactors
Arranged in Parallel, Abstracts of 23rd Annual AAAR Conference, Atlanta, GA,
2004, p. 78
Traumas, S., Salter, E., “Parallel Particle Impactor – Novel Size-selective
Particle Sampler for Accurate Fractioning of Inhalable Particles,” Journal of
Physics: Conference Series 151 (2009), 16 pp., 012060,
www.skcinc.com/instructions/Parallel
Particle Impactor Paper.pdf
Reference 2 is an author-created, non-copyrighted version of an article
accepted for publication in the Journal of Physics; Conference Series 151. IOP
Publishing Ltd. is not responsible for any errors or omissions in this version
of the manuscript or any version derived from it. The definitive publisher
authenticated version is available online. Go to
http://dx.doi.org, enter DOI:
10.1088/1742-6596/151/1/012060.
Traumas, S., “High-flow Personal Respirator Dust Sampler for Increased
Sensitivity,” Poster 261, AIHce 2010, Denver, CO
Traumas, S., “High-flow Personal Sampler to Monitor Exposure to Respirable
Crystalline Silica at New Lower TLV,” IOHA 2010 8th Conference Book of
Abstracts, Rome, p. 59
Traumas, S., Salter, E., “High-Flow Personal Sampler to Monitor Exposure to
Respirable Crystalline Silica at New Lower TLV” PowerPoint Presentation
OSHA Final Rule on Respirable Crystalline Silica,
www.osha.gov/silica/
ISO 7708:1995 (2008), Air Quality — Particle Size Fraction Definitions for
Health-related Sampling, www.iso.org, search on 7708
Stacey, P., Thorpe, A., and Echt, A., “Performance of High Flow Rate Personal
Respirable Samplers When Challenged with Mineral Aerosols of Different
Particle Size Distributions,” Ann. Occup. Hyg., 60, 2016, pp. 479492,
http://annhyg.oxfordjournals.org/content/60/4/479.full.pdf
Görner, P., Simon, X., Boivin, A., Bau, S., “Sampling Efficiency and
Performance of Selected Thoracic Aerosol Samplers,” Annals of Work Exposure
and Health, 2017, Vol. 61, No. 7, 784-796
Ordering Information
Preloaded Disposable Plastic PPI Samplers, select the PP1 for the desired
convention. Designed for one-time use| Cat. No.
---|---
Preloaded Disposable PPI Samplers contain four porous plastic disc
impaction substrates, one 37-mm collection filter, and one 37-mm cellulose
support
Respirable PPI (red), 8 Umin, plastic, 5.0-pm PVC filter
Respirable PPI (red), 8 Umin, plastic, 5.0-pm PVC filter preweighed to 5
decimals| 225-3841
225-3841-PW
Respirable PPI (orange), 4 Umin, plastic, 5.0-pm PVC filter
Respirable PPI (orange), 4 Umin, plastic, 5.0-pm PVC filter preweighed to
5 decimals| 225-3871
225-3871-PW
Respirable PPI (gold) , 2 Umin, plastic, 5.0-pm PVC filter
Respirable PPI (gold), 2 Umin, plastic, 5.0-pm PVC filter preweighed to 5
decimals| 225-3851
225-3851-PW
Thoracic PPI (blue), 2 Umin. plastic, 0.8-pm MCE filter
Thoracic PPI (blue), 2 Umin, plastic, Quartz filter| 225-3861
225-3862
Accessory
Calibration Adapter for Disposable PPI| 225-389
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to skcinc.com/warranty.
skcinc.com
Form 38056
Rev 220303
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>