SKC 225-49 IFV Pro Sampler Instruction Manual

June 3, 2024
SKC

SKC 225-49 IFV Pro Sampler Instruction Manual
SKC 225-49 IFV Pro Sampler

Description

The SKC IFV Pro Sampler collects mixed-phase (aerosol and vapor) contaminants simultaneously. The stainless steel sampling head with an IOM-style inhalable inlet holds a filter cassette with a recommended 25-mm filter to
collect the aerosol phase; the attached tube holder (rubber sleeve) holds the method-specified sorbent tube to collect the vapor phase. The sorbent tube is protected by an open-ended plastic cover (Figure 1). IFV Pro is used with a sample pump capable of maintaining a constant flow rate of 1 L/min.

Product Overview
Figure 1. IFV Pro Sampler: Sampling Head and Sorbent Tube Holder

Performance Profile

  • Flow Rate : 1 L/min

  • 50% Cut-point: 100 μm at 1 L/min inhalable fraction

  • Material: Sampling head – stainless steel front plate and filter cassette
    with 10.6-mm diameter inlet*; aluminum housing and outlet
    Tube holder – rubber
    Tube cover – plastic

  • Maximum

  • Operating

  • Temperature: 392 F (200 C) suitable for autoclaving and solvent washing

  • Media: Filters – 25-mm glass fiber or quartz filter; or coated filter. Dependent upon compound of interest. Sorbent Tubes – 8 x 110-mm size. Dependent upon compound of interest.
    See page 8 for Selection Guide.

  • Tubing: ¼-in ID

  • Weight: 2.4 oz (68 gm)

  • Dimensions: 8.5 x 1.2 in (21.6 x 3.17 cm)

Designed to meet the inhalable convention based on the scaling down technique described in the Final Report on research carried out under NIOSH-CDC Grant No. RO1-OH 03687-03, “Development of New Personal Aerosol Samplers,” Vincent, J.H., et al., 2003.

Inhalable Sampling Head Components and O-ring

Placement (Figure 2)

Head Components

Figure 2. Configuration of Sampling Head and Filter Cassette

Preparation and Assembly

Wear powder-free gloves when handling cassettes and use forceps when working with filters to prevent the transfer of moisture, dust, or other contaminants onto the sampling media.

Disassembling Sampling Head

  1. Remove the white cap from the sampling head inlet and set it aside.
  2. Unscrew and remove the front plate of the sampling head.
  3. Remove the filter cassette.

Filter Cassette: Opening, Installing Filter, and Closing

Tip: Place cap on filter cassette inlet for better grasp when opening cassette

  1. To open the cassette, twist the cassette top in the cassette bottom until the notch in the top (Figure 3) is aligned with one of the two bumps on the cassette bottom; tilt the notched side up and lift.
  2. Using forceps, place a 25-mm filter on the support grid in the cassette bottom.
  3. Align the notch in the cassette top with one of the two bumps on the cassette bottom and insert the opposite edge of the top under the opposite bump in the cassette bottom. Set the top into the bottom and twist to secure the cassette (Figure 3).
    Installing Filter
    Figure 3. Exploded Cassette

Transporting Loaded Cassettes If not ready to sample immediately or if transporting loaded cassettes to a sampling location, place the filter cassette (with cap in place on the inlet) in the transport container. Ensure that the foam insert is in the container lid (Figure 4).

Figure 4. Cassette in Transport Container

Installing Filter Cassette in Sampling Head

  1. Ensure that the O-ring is positioned correctly in the top of the housing (Figure 2) and place the loaded filter cassette on top of the O-ring (Figure 5).
  2. Replace the front plate on the housing and turn clockwise until tight. Tighten securely to achieve a good seal.
  3. Install the cap on the inlet until ready to sample
    Installing Filter
    Figure 5. Installing Loaded Cassette in Sampling Head

Assembling Sorbent Tube and Cover with Sampling Head (Figure 6)

  1. Unscrew the tube cover from the sampling head.
  2. Install a new tube holder (rubber sleeve Cat. No. P3022A) on the sampling head outlet. Note: It is important to install a new tube holder before each new sampling operation.
  3. Break the ends off the sorbent tube using a tube breaker.
  4. Insert the opened sorbent tube into the tube holder on the sampling head base. Note: The arrow printed on the tube should be pointing away from the rubber sleeve and toward the pump.
  5. Attach flexible tubing to the other end of the sorbent tube.
  6. Insert the open end of the flexible tubing through the threaded end of the protective tube cover; slide the tube cover up and over the sorbent tube to the base of the sampling head.
  7. Thread the tube cover onto the base and turn until snug

Figure 6. Assembling Sorbent Tube and Protective Tube Cover with Sampling Head

Calibration

Calibrate pump flow rate using a calibrator, calibration adapter, and a representative loaded sampling head and sorbent tube in line. See the calibration train in Figure 7.

  1. Following best practice, ensure that the sample pump has run for 5 minutes after charging to ensure a more stable flow.
  2. Remove the cap from the cassette inlet and set it aside.
  3. Place the calibration adapter outlet on the sampling head inlet and press firmly.
  4. Using flexible tubing, connect the calibration adapter inlet to a calibrator.
  5. Attach flexible tubing from the sorbent tube to the sample pump inlet.
  6. Calibrate the flow rate to 1 L/min. See pump and calibrator operating instructions.
  7. Disconnect the calibrator and calibration adapter. Remove the representative filter cassette and sorbent tube. Note: Keep representative media to verify flow rate later.Calibration Train
    Figure 7. Calibration Train

Sampling

Note: Before each new sampling operation, install a new tube holder (rubber sleeve Cat. No. P3022A) on the sampling head outlet.

  1. Ensure that representative sampling media has been replaced with fresh, unused media (designated filter and sorbent tube) and that the cassette cap has been removed and set aside.

  2. Connect the flexible tubing from the sorbent tube to the pump inlet.

  3. Clip the sampler onto the worker’s clothing in the breathing zone; clip the pump onto the worker’s belt or place it in a protective pouch (Figure 8).

  4. Start the pump and sample for the time specified in the method being used. See pump operating instructions.

  5. Stop the pump. Detach the flexible tubing from the pump inlet.

  6. Remove the protective tube cover and flexible tubing from the sorbent tube.

  7. Remove the sampling media. Seal the sorbent tube ends with the caps provided. Place the filter cassette, with cap on, in the transport container; place the foam insert in the lid (Figure 4).

  8. Remove and dispose of the tube holder.

  9. Record all pertinent information.

  10. Package samples as appropriate for shipment to a laboratory for analysis. See Transporting Samples to a Laboratory.

Transporting Samples to a Laboratory

Filters
Place the cassette transport container in a plastic bag or suitable container. Package in a padded envelope along with pertinent sampling information and blanks. See Sample Blanks.

Sorbent Tubes
Pack the sample sorbent tubes, field (sample) blanks, and all pertinent information securely for shipment to a laboratory for analysis.

Sample Blanks
Load and handle the blank IFV Pro Sampler in the same way as the sampler used for sampling. Do not pull air through the blanks. For sorbent tubes, submit field blanks from the same lot as the sample tubes. Send blanks with samples to a laboratory.

Cleaning Sampling Head

  1. Disassemble the sampling head (Figure 2). Ensure that the tube holder has been removed.
  2. Place the front plate, housing, and filter cassette in an ultrasonic cleaner with water and a wetting agent such as mild soap. Components may also be cleaned with a solvent such as isopropyl alcohol. Clean the O-ring separately with water.
  3. Wipe components using a clean, lint-free paper, cloth, or soft brush. Allow components to dry completely

Ordering Information

Description: Cat. No.

IFV Pro Sampler includes sampling head (aerosol sampler body, cassette, and front plate), protective tube cover, 10 extra tube holders (rubber sleeves), cassette cap,and cassette transport container, requires 25 mm filter and 8 x 110 mm sorbent tube: 225-49
IFV Pro Sampler Kit includes sampler as described above, calibration adapter,and case, requires 25-mm filter and 8 x 110-mm sorbent : 225-49K
Filter Cassette, stainless steel, with cap and transport container, for 25 mm filter : 225-4903
Tube Holders (Rubber Sleeves), pk/25, change after each : P3022A
Accessories

Calibration Adapte: 225-361

SKC Limited Warranty and Return Policy

SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view  the complete SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/ warranty

IFV Pro Filter and Sorbent Tube Selection Guide

Compound Recommended Filter Sorbent Tube
Acrylamide 225-702 226-10-04
Alachlor 225-702 226-30-06
Aldrin 225-702 226-30-06
Azinphos-methyl 225-702 226-30-06
Butylated hydroxytoluene 225-702 226-211
Carbaryl 225-702 226-30-06
Carbofuran 225-702 226-30-06
Chlorpyrifos 225-702 226-30-06
Clopidol 225-702 226-30-06
Coumaphos 225-702 226-30-06
Cresol, all isomers 225-702 226-211
Demeton 225-702 226-30-06
Demeton S-methyl 225-702 226-30-06
Diazinon 225-702 226-30-06
Dibutyl phosphate 225-702 226-30-06
Dichlorvos 225-702 226-30-06
Dicrotophos 225-702 226-30-06
Dieldrin 225-702 226-30-06
Diesel fuel as total hydrocarbons 225-702 226-09
Diethanolamine 225-702 226-214
Dioxathion 225-702 226-30-06
Disulfoton 225-702 226-30-06
Endosulfan 225-702 226-30-06
2-Ethylhexanoic acid 225-702 226-10-04
Fenamiphos 225-702 226-30-06
Fensulfothion 225-702 226-30-06
Fenthion 225-702 226-30-06
Fonofos 225-702 226-30-06
Glyoxal 225-9036 226-119-7
Malathion 225-702 226-30-06
Maleic anhydride 225-9028 226-213
Methomyl 225-702 226-30-06
Methyl demeton 225-702 226-30-06
Methyl parathion 225-702 226-30-06
Mevinphos 225-702 226-30-06
Monochloroacetic acid 225-702 226-10-04
Monocrotophos 225-702 226-30-06
Naled 225-702 226-30-06
o-Phthalodinitrile 225-702 226-83
Parathion 225-702 226-30-06
Pentachlorophenol 225-702 226-211
Phorate 225-702 226-30-06
Propoxur 225-702 226-30-06
Ronnel 225-702 226-30-06
Sulfotepp 225-702 226-30-06
Sulprofos 225-702 226-30-06
Temephos 225-702 226-30-06
Terbufos 225-702 226-30-06
1,1,2,2-Tetrabromoethane 225-702 226-10-04
Tetraethylpyrophosphate (TEPP) 225-702 226-106A
Thiram 225-702 226-30-06
Toluene-2,4-diisocyanate 225-9035
Toluene-2,6-diisocyanate 225-9035
Xylidene isomers 225-702 226-10-04

SUPPORT

863 Valley View Road, Eighty Four, PA 15330 USA
Tel: 724-941-9701
www.skcinc.com
http://www.skcinc.com/

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals