This rotameter is packed to protect against damage during shipment. If damage occurs, notify the carrier immediately. Also, check that you have received the correct model and fl ow range required for your application.
Calibration
SKC rotameters are secondary standards and should be calibrated by the user
with a primary standard calibrator at regular intervals. Since rotameters are
affected by temperature and atmospheric pressure, rotameter calibration should
be performed under the conditions of use or a mathematical correction should
be used.
Operation
Operate the rotameter in a vertical position to assure accuracy. Use it in a
manner that minimizes vibration and fl ow pulsation; both conditions may affect
the rotameter.
Flow rate is indicated by the point on the printed scale at which the center
of the ball fl oat comes to rest. The fi eld rotameter inlet must always be at
atmospheric pressure (i.e., exposed to the atmosphere). The rotameter is the
last device connected to the sampling train. Connect the inlet of the sampling
pump to the outlet of the sampling medium. Connect the inlet of the sampling
medium to the outlet of the rotameter.
CAUTION
This rotameter is designed for use with non-hazardous gases at pressures up to
100 psi and temperatures up to 65 C (150 F). Do not use hazardous gases or
liquids and do not exceed temperature or pressure limits. Use with hazardous
fluids or beyond pressure and temperature limits may cause failure, which could
result in personal injury.
Maintenance
Occasionally clean the rotameter if dirt appears in it or if fl oat movement is restricted. To clean:
- Remove the top plug, fl oat stop, and ball fl oat.
- Wash the tapered hole, ball fl oat, fl oat stop, and top plug with a mild liquid detergent and a soft brush. Rinse all parts with clean water and dry thoroughly. Avoid using solvents.
- Reinstall the ball fl oat, fl oat stop, and top plug. Use a small amount of halocarbon grease or any other compatible O-ring lubricant to ease the installation of the top plug and prolong O-ring life.
Specifications
Meas. Range (L/min)| Scale (inch)| **Accuracy***
±
| Subdivisions (L/min)| Cat. No.
---|---|---|---|---
0.05-0.5| 2| 5%| 0.02| 320-2A05
0.1-1.0| 4| 3%| 20 ◊| 320-4A1
0.4-5.0| 4| 3%| 0.1| 320-4A5
2-20| 4| 3%| 0.5| 320-4A20L#
4-50| 4| 3%| 1.0| 320-440
3-30| 4| 3%| 1.0| 320-530
3-30| 4| 3%| 1.0| 320-100†
Full Scale
- Product scale is marked in ccm (cc/min), which is equivalent to ml/min.
- Fittings adapted for use with QuickTake Sample Pumps
- Fittings adapted for use with BioLite+ Sample Pump
- Float: Black glass or stainless steel ball
- Body: Clear acrylic
- Seals: Buna N O-ring
- Pressure: 100 psi maximum
- Temperature: 150 F (65 C) maximum
- Fittings: Brass with hose connector, Tygon® tubing, and Luer taper adapter
SKC Limited Warranty and Return Policy
SKC products are subject to the SKC Limited Warranty and Return Policy, which provides SKC’s sole liability and the buyer’s exclusive remedy. To view the complete SKC Limited Warranty and Return Policy, go to http://www.skcinc.com/warranty.
References
- SKC, Inc. | Industrial Hygiene and Occupational Safety Products and Technology
- Warranty Information | Learn More About Industrial Hygiene & Occupational Safety Products at SKC, Inc.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>