PCE-CMM 8 CO2 Analyser User Manual
- June 5, 2024
- PCE
Table of Contents
PCE-CMM 8 CO2 Analyser
Safety Notes
The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non-observance of the manual are excluded from our liability and not covered by our warranty.
- If used otherwise, this can cause dangerous situations for the user and damage to the meter.
- The instrument may only be used if the environmental conditions temperature, relative humidity are within the ranges stated in the technical specifications.
- Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
- Do not expose the device to shocks or strong vibrations.
- The case should only be opened by qualified PCE Instruments personnel.
- Never use the instrument when your hands are wet.
- You must not make any technical changes to the device.
- The appliance should only be cleaned with a damp cloth.
- Use only pH-neutral cleaner, no abrasives or solvents.
- The device must only be used with accessories from PCE Instruments or equivalent.
- Before each use, inspect the case for visible damage.
- If any damage is visible do not use the device.
- Do not use the instrument in explosive atmospheres.
- The measurement range as stated in the specifications must not be exceeded under any circumstances.
- Non observance of the safety notes can cause damage to the device and injuries to the user.
- We do not assume liability for printing errors or any other mistakes in this manual.
- We expressly point to our general guarantee terms which can be found in our general terms of business.
- If you have any questions please contact PCE Instruments.
Device Description
Technical Specifications
Measured variable CO 2
Measurement range Accuracy
Resolution
| 0…9999 ppm CO2 ±75 ppm or ±5 % of rdg the higher value applies
1 ppm
CO2 sensor type| NDIR sensor
Measured variable temperature
Measurement range
Accuracy Resolution
| -10… 50°C / 14…. 122°F ±1°C
0.1°C
Measured variable air humidity
Measurement range
Accuracy Resolution | 20…95 % RH
±5 % RH
1 % RH
General Technical Specifications
Display| 3.5″ touch LCD
Alarm display| Traffic light function through colour change of reading Alarm
tone
Power supply| 5 V mini USB (no rechargeable battery)
Standby power| <0.5 W
Environmental conditions| 1..85 °C / 14…122 °F
20…85 % RH
Protection class| IP21
Dimensions| approx. 96 x 86 x 32.5 mm / 3.7 x 3.3 x 1.2 mm
Weight| <200 g / <7 oz
Display / Touch Screen
- Date and time
- Set lower or upper alarm
- CO2 reading (changes ist color)
- Increase or decrease alarm value
- Lock screen
- Unit change °C / °F
- Air humidity reading in % RH
- Show progress diagram
- Temperature reading
- Settings
- Alarm message
- On & off
- Mute alarm
Dimensions
Operating Instructions
On/off
Connect the meter to the power supply. Touch to turn on the meter. Touch while the power is on to turn off the meter.
Settings
Touch to access the settings.
-
Setting the language
The following languages can be selected in the meter: English, Spanish, Italian, Russian, Polish, Czech, German, French, Slovenian, Swedish, Portuguese -
Setting the date and time
Touch the menu item Clock Setting to set the date and time. -
Setting the temperature unit
Touch the menu item Temperature Format to switch between °C and °F. -
Setting the display brightness
Touch the menu item Backlight to adjust the display brightness using the arrow keys at the bottom right of the display. If you touch the item Ambient Light Dimming, the display brightness is adjusted depending on the ambient light at the time of saving. -
Standby screen
Touch the menu item Standby Screen to set the time period when the display dims the brightness. -
Screen lock
Touch the menu item Screen Lock to set a 4-digit PIN to protect against menu access. Attention: Remember the PIN well. However, if you have forgotten your PIN, you can use the master PIN 8657. -
CO2 alarm limits
Touch the menu item to set the limits for the lower and upper alarm between 450…4500 ppm.
Calibration
-
CO2 factory calibration
Touch the menu item RESET CO2 To Factory Settings to calibrate the CO2 sensor to factory settings. The meter must be placed in a well ventilated room or placed outside and run for 15 minutes. -
Automatic calibration
The PCE-CMM 8 has an integrated self-correcting ABC algorithm (Automatic Baseline Correction) for automatic adjustment of the baseline value. This calibration is carried out regularly at intervals of 10-15 days.
For this, the PCE-CMM 8 requires a regular supply of fresh air. The measured value is automatically adjusted. The ABC function is permanently switched on (cannot be switched off) and adjusts the baseline value at regular intervals when the CO2 level is low. Do not use this meter in a closed environment where the CO2 concentration remains relatively high. This would negatively affect the self-correction of the device. We therefore recommend performing the manual calibration
Offset Settings
-
CO2 offset
Touch the menu item CO2 Sensor Calibration to set an offset for the CO2 reading. -
Temperature offset
Touch the menu item Built In Sensor to set an offset for the temperature reading. -
Humidity sensor offset
Touch the menu item Humidity Sensor to set an offset for the humidity value reading. -
Selecting the time zone
Touch this menu item to set the time difference from GMT. Press to enable synchronization of the power on/off time with the internet time. To do this, the meter must be successfully connected to the internet. -
Daylight saving time
Touch the menu item to set the daylight saving time based on the location. -
Home screen color
In the menu item Home Screen Color , select whether the background color should be blue or black. -
Alarm tone
Select the option Alarm Speaker Settings . Touch the button to enable/disable the alarm sound for the lower or upper alarm. -
Reset to factory settings
Select Reset To Factory Settings to reset the meter to its factory default settings. All previously made settings are lost.
Measurement graph/data history
-
Graph
Touch to view the measurement history of the last 7 days. Here you can view the course of CO2, temperature and humidity. To view the details of a particular day, touch the corresponding area above the date. Touch to view the previous screen. -
Save data history
Touch to view the measurement history of the last 7 days. Then select the menu item Data Log Table. In the next window, you can save the measurement data per hour in a file on the device by selecting the Save to File item. -
Read data history on the PC
Connect the meter to a PC using the USB cable. The CO2 meter is recognized as a mass storage device. In the file DATALOG.TXT, you can view the previously saved measurement data.
Alarms
Touch the number in the top right corner to set the alarm values for High and Low. The CO2 meter uses these values to change the color of the digits for the measured value and an alarm tone can be heard, if selected.
-
Set upper alarm (High)
As soon as “high” appears in the upper right corner, you can set the upper alarm value using the + and – buttons. This default setting for this alarm threshold is 1200 ppm. -
Set lower alarm (Low)
As soon as low appears in the upper right corner, you can set the lower alarm value using the + and – buttons. This default setting for this alarm threshold is 800 ppm. -
Temporarily deactivating the alarm tone
If the alarm sound has been enabled in the settings, it sounds as soon as the measured value exceeds or falls below the upper/lower alarm value. However, you can also temporarily deactivate it by touching the icon on the main screen.
Function CO2 Traffic Light
This means that the color if the reading changes, depending on the set alarm limit. The default limit values for changing colors are as follow
Air quality | CO 2 content | Color |
---|---|---|
Good | 400…800 ppm | Green |
Ok | 800…1200 ppm | Yellow |
Poor | 1200…5000 ppm | Red |
Warranty
You can read our warranty terms in our General Business Terms which you can find here: https://www.pce-instruments.com/english/terms
Disposal
- For the disposal of batteries in the EU the 2006/66/EC directive of the European Parliament applies.
- Due to the contained pollutants, batteries must not be disposed of as household waste.
- They must be given to collection points designed for that purpose.
- In order to comply with the EU directive 2012/19/EU we take our devices back.
- We either reuse them or give them to a recycling company which disposes of the devices in line with law.
- For countries outside the EU, batteries and devices should be disposed of in accordance with your local waste regulations.
- If you have any questions, please contact PCE Instruments.
PCE Instruments contact information
Germany
PCE Deutschland GmbH Im Langel 4
D-59872 Meschede Deutschland
Tel.: +49 (0) 2903 976 99 0
Fax: +49 (0) 2903 976 99 29
info@pce-instruments.com
www.pce-instruments.com/deutsch
United Kingdom
PCE Instruments UK Ltd
Unit 11 Southpoint Business Park Ensign Way, Southampton Hampshire
United Kingdom, SO31 4RF Tel: +44 (0) 2380 98703 0
Fax: +44 (0) 2380 98703 9
info@pce-instruments.co.uk
www.pce-instruments.com/english
The Netherlands
PCE Brookhuis B.V. Institutenweg 15
7521 PH Enschede Nederland
Telefoon: +31 (0)53 737 01 92
info@pcebenelux.nl
www.pce-instruments.com/dutch
United States of America
PCE Americas Inc.
1201 Jupiter Park Drive, Suite 8 Jupiter / Palm Beach
33458 FL USA
Tel: +1 561-320-9162
Fax: +1 561-320-9176
info@pce-americas.com
www.pce-instruments.com/us
France
PCE Instruments France EURL 23, rue de Strasbourg
67250 Soultz-Sous-Forets France
Telephone: +33 (0) 972 3537 17
Numéro de fax: +33 (0) 972 3537 18
info@pce-france.fr
www.pce-instruments.com/french
Italy
PCE Italia s.r.l.
Via Pesciatina 878 / B-Interno 6
55010 Loc. Gragnano Capannori (Lucca) Italia
Telefono: +39 0583 975 114
Fax: +39 0583 974 824
info@pce-italia.it
www.pce-instruments.com/italiano
China
PCE (Beijing) Technology Co., Limited 1519 Room, 6 Building
Zhong Ang Times Plaza
No. 9 Mentougou Road, Tou Gou District 102300 Beijing, China
Tel: +86 (10) 8893 9660
info@pce-instruments.cn
www.pce-instruments.cn
Spain
PCE Ibérica S.L. Calle Mayor, 53
02500 Tobarra (Albacete) España
Tel. : +34 967 543 548
Fax: +34 967 543 542
info@pce-iberica.es
www.pce-instruments.com/espanol
Turkey
PCE Teknik Cihazları Ltd.Şti. Halkalı Merkez Mah.
Pehlivan Sok. No.6/C
34303 Küçükçekmece – İstanbul Türkiye
Tel: 0212 471 11 47
Faks: 0212 705 53 93
info@pce-cihazlari.com.tr
www.pce-instruments.com/turkish
Hong Kong
PCE Instruments HK Ltd. Unit J, 21/F., COS Centre 56 Tsun Yip Street
Kwun Tong Kowloon, Hong Kong
Tel: +852-301-84912
jyi@pce-instruments.com
www.pce-instruments.cn
References
- France.fr : Explore France and its wonders - Explore France
- iberica.es
- instruments.cn - The domain is available for purchase
- Make an offer on the domain instruments.co.uk - Domains.co.uk
- Computer Instruments | Home
- Discover Italy: Official Tourism Website - Italia.it
- PCE(北京)科技有é™å…¬å¸
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>