PCE Instruments PCE-TG 75 Wall Thickness Meter User Manual
- October 30, 2023
- PCE Instruments
Table of Contents
PCE Instruments PCE-TG 75 Wall Thickness Meter
Product Information
The product manual can be found by searching for the product on our website: www.pce-instruments.com. The latest version of the manual was updated on January 4, 2022 (v1.0). Below is a summary of the contents:
- Safety notes
- Specifications
- System description
- Getting started
- Menu
- Submenu
- Measurement
- Calibration (Cal)
- Maintenance
- Warranty
- Disposal
Getting Started
Before using the product, ensure it is connected to a power source as described in section 4.1 of the manual. Follow the instructions in section 4.2 for first-time use.
Measurement
To take measurements using the product, refer to section 7 of the manual.
Calibration
To calibrate the product, refer to section 8 of the manual. This section includes general calibration instructions as well as specific instructions for sound velocity calibration (V. Cal).
Maintenance
To keep the product in good condition, refer to section 9 of the manual. This section includes instructions for cleaning and troubleshooting.
Safety notes
Please read this manual carefully and completely before you use the device for the first time. The device may only be used by qualified personnel and repaired by PCE Instruments personnel. Damage or injuries caused by non- observance of the manual are excluded from our liability and not covered by our warranty.
- The device must only be used as described in this instruction manual. If used otherwise, this can cause dangerous situations for the user and damage to the meter.
- The instrument may only be used if the environmental conditions (temperature, relative humidity, …) are within the ranges stated in the technical specifications. Do not expose the device to extreme temperatures, direct sunlight, extreme humidity or moisture.
- Do not expose the device to shocks or strong vibrations.
- The case should only be opened by qualified PCE Instruments personnel.
- Never use the instrument when your hands are wet.
- You must not make any technical changes to the device.
- The appliance should only be cleaned with a damp cloth. Use only pH-neutral cleaner, no abrasives or solvents.
- The device must only be used with accessories from PCE Instruments or equivalent.
- Before each use, inspect the case for visible damage. If any damage is visible, do not use the device.
- Do not use the instrument in explosive atmospheres.
- The measurement range as stated in the specifications must not be exceeded under any circumstances.
- Non-observance of the safety notes can cause damage to the device and injuries to the user.
- We do not assume liability for printing errors or any other mistakes in this manual.
- We expressly point to our general guarantee terms which can be found in our general terms of business.
- If you have any questions please contact PCE Instruments. The contact details can be found at the end of this manual.
Specifications
Technical specifications
Specification | Value |
---|---|
Model | PCE-TG 75 |
Measurement range | 1.00 … 225.0 mm |
Operating memory | 500 readings |
Probe selection | no |
Operating frequency | 5 MHz |
Accuracy | ±(0.5 % H*+0.05) mm |
Sound velocity setting
range
| 1000 … 9999 m/s
Operating temperature| 0 … 40 °C / 32 … 104 °F
Pow er supply| 3 x 1.5 V AA battery
Dimensions| 163 x 82 x 38 mm / 6.42 x 3.23 x 1.50“
Weight| 320 g / 0.70 lbs
Delivery contents
Model | Delivery contents |
---|
PCE-TG 75
| 1 x thickness gauge PCE-TG 75
1 x coupling gel TT-GEL 3 x 1.5 V AA battery
1 x probe PCE-TG 5M10d
1 x user manual
PCE-TG 150
| 1 x thickness gauge PCE-TG 150
1 x coupling gel TT-GEL 3 x 1.5 V AA battery
1 x probe PCE-TG 5M10d
1 x probe PCE-TG 5M6d
1 x user manual
PCE-TG 150 2.5F
| 1 x thickness gauge PCE-TG 150
1 x coupling gel TT-GEL 3 x 1.5 V AA battery
1 x probe PCE-TG 2.5M
1 x user manual
PCE-TG 150 HT
| 1 x thickness gauge PCE-TG 150
1 x coupling gel TT-GEL 3 x 1.5 V AA battery
1 x probe PCE-TG HT
1 x user manual
Models
Model | Included probe |
---|---|
PCE-TG 75 | PCE-TG 5M10d |
PCE-TG 150 | PCE-TG 5M10d PCE-TG 5M6d |
PCE-TG 150 2.5F | PCE-TG 2.5M |
PCE-TG 150 HT | PCE-TG HT |
Probes
Probe | Specifications |
---|
PCE-TG 5M10d
| ****
Application
| Standard probe for flat
materials, pipes w ith a radius
10 mm
Frequency| 5 MHz
Diameter| 10 mm / 0.29“
Temperature range| -10 … 60 °C / 14 … 140 °F
PCE-TG 5M6d
| ****
Application
| Standard probe for flat
materials, pipes w ith a radius
6 mm
Frequency| 5 MHz
Diameter| 6 mm / 0.24“
Temperature range| -10 … 60 °C / 14 … 140 °F
PCE-TG 2.5M
| Application| Thickness measurement of cast
materials
Frequency| 2.5 MHz
Diameter| 12 mm / 0.47“
Temperature range| -10 … 60 °C / 14 … 140 °F
PCE-TG HT
| Application| Measurement of the material thickness of hot components
Frequency| 5 MHz
Diameter| 10 mm / 0.29“
Temperature range| -10 … 350 °C / 14 … 662 °F
Optional accessories
Item number | Description |
---|---|
TT-GEL | Ultrasonic coupling gel |
K-GEL | Coupling gel for temperatures of up to 350 °C / 662 °F (100 ml) |
PCE-TG 5M10d | Standard probe for PCE TG 75 and PCE-TG 150 series, 5 MHz, |
∅10 mm / 0.29“, 1 … 225 mm (PCE-TG 75),
1 … 300 mm (PCE-TG 150), -10 … 60 °C / 14 … 140 °F
PCE-TG 5M6d| Miniature probe for PCE-TG 150 series, 5 MHz, ∅6 mm / 0.24“,
1 … 50 mm, -10 … 60 °C / 14 … 140 °F
PCE-TG 2.5M| 2.5 MHz probe for PCE-TG 150 series, ∅12 mm / 0.47“, 1.2 … 300
mm, -10 … 60 °C / 14 … 140 °F
PCE-TG HT| High-temperature probe for PCE-TG 150 series, ∅10 mm / 0.29“,
1.2 … 300 mm, -10 … 350 °C / 14 … 662 °F
PCE-BAG M| Universal bag for meters: inside size M
(190 x 120 x 70 mm / 7.48 x 4.72 x 2.76”) black/blue
System description
The PCE-TG 75 is a thickness gauge that measures the speed of ultrasonic waves
and can thus determine the thickness of materials accurately and non-
destructively. The meter can also be used to detect corrosion damage to piping
and pressure vessels of production equipment. It can also be used in
manufacturing, metal processing and cargo inspection. The meter is suitable
for the thickness measurement of any homogeneous material that transmits
ultrasonic waves at a constant rate and reflects them off the back of the
material.
Applications
This meter can be used to measure the thickness of any good ultrasonic
conductor with parallel top and bottom, such as metal, plastics, ceramics and
glass. Materials such as aluminium, copper, gold, resin, etc. are not suitable
for this measuring instrument!
Device
Interfaces
- Probe socket
Display (in measuring condition)
- Title bar
- Battery level
- Sound velocity
- Coupling icon
- Selected memory method
- Selected probe
- Selected file
- Reading
- Unit
- Information about the measurement
Getting started Power supply
Three 1.5 V AA batteries are required for the power supply. The battery
compartment is located on the back of the instrument and is secured with two
screws. Before replacing the batteries, switch off the device. Loosen the
screws, lift off the cover, insert the batteries as marked, and close the
battery compartment again by tightening the screws.
First use
Before switching on the instrument, insert the probe into the probe socket of
the instrument. To start the meter, press the on/off & menu key until the
display shows the name of the meter. The interface for measuring the sound
velocity will then open. The thickness of the material is determined by
measuring the sound velocity. The measurement can be started here.
Menu
The menu of the meter can be reached by pressing the menu key after start-up. With the arrow keys, you can navigate through the menu. You can confirm your selection with the OK key. To navigate one step back, press the back key .
Measure
In this menu item, make the measurement settings:
- d. Material to be measured (sound velocity)
- e. Unit of measurement m/s (unit)
- f. Resolution
Probe (PCE-TG 150 only)
In this menu item, select the probe you want to use for the measurement.
Limit
In this menu item, make the limit value settings:
- c. Set the upper and lower limit value for the measurement
- d. Enable or disable the alarm for exceeded limit values
Memory
Measurement processes can be saved and started under this menu item. If you
activate “Autosave”, all measuring processes are saved automatically.
d. Select a directory in which the measurement is to be saved.
- e. View saved data
- f. Delete the data of the selected file
Analyze
Under this menu item, you can display the saved data in the form of a graph.
- a. You can view basic information such as the data number, the maximum value, the minimum value and the average value
- b. You can enlarge the diagram
- c. Use the cursor to view a certain data value
Settings
Make further settings in the meter:
- a. Language
- b. Automatic power off time
- c. Key tone
- d. Backlight brightness
- e. Color
- f. Software version number
- g. Factory reset
Submenu
Operation
- a. Enter the submenu: Press the back key in the measuring interface to open the submenu.
- b. Select menu item: Press the up and down keys to change between the menu items
- c. Opening the selected menu item: Press OK to enter the selected menu item
- d. Leave menu item: Press the back key to leave the menu item
Functions of the submenu
- e. Save: Save current readings to the selected file
- f. Clear: Delete readings and measuring information g. Calibration (Cal): See calibration instructions for details
- h. Sound velocity calibration (V.Cal): See Chapter 8.2 Sound velocity calibration (V. Cal) for details.
Measurement
Apply the TT-Gel evenly to the surface to be measured. Place the probe onto the surface covered with the coupling agent.
-
Clean surface
Before the measurement, the surface of the object to be tested should be cleaned of dust, dirt and rust. Coatings such as paint must also be removed. -
Reducing the roughness
Rough surfaces can cause measurement errors or missing readings. Before the measurement, the material surface should be made as smooth as possible by grinding, polishing or using a high-viscosity coupling agent. -
Rough surfaces due to heavy machines
Regular fine grooves which occur, for example, during production with lathe machines and planers can also cause measurement errors. The corrective measure is the same as in chapter 7.2. In addition, a better result can be obtained by adjusting the angle using the silver marking on the centre of the probe head and the fine grooves of the material to be tested (orthogonal or parallel to the marking). -
Measurement of a round surface
To measure a round surface such as a pipe or an oil barrel, it is important to set the angle between the silver line of the probe and the axis of the material to be tested. In short, the mark in the sensor head must be kept parallel or perpendicular to the axis of the material under test. Slowly move the probe head perpendicular to the axis across the material under test and the values on the display will change regularly. The lowest reading displayed is the minimum thickness of the material being measured.
The direction in which the probe is guided depends on the curvature of the material. For large diameter pipes, the mark in the probe head should be perpendicular to the pipe axis; for smaller-diameter pipes, the mark in the probe head can be parallel or perpendicular to the material axis. The lowest reading is saved as the measured value. -
Measurement of composite material
When measuring composite shapes (such as pipe elbows), the method described in chapter 7.4 can be used. However, the measurement must be carried out twice to obtain two readings. The silver line of the probe must be vertical or parallel to the axis. The lower value is used as the measured thickness. -
Non-parallel surfaces
To obtain a satisfactory result, the surface of the material to be tested must be parallel or coaxial to the surface of the sensor, otherwise measurement errors will occur or no reading will be obtained. -
Effects of the material temperature
The material thickness and the transmission speed of ultrasonic waves are influenced by the temperature. To achieve high measurement accuracy, measure samples of the same material under the same temperature conditions. In case of steel, high temperatures cause strong measurement errors (measured values are lower than the actual data). -
Damping materials
Materials with fibres, pores and coarse particles cause strong scattering and energy attenuation of the ultrasonic waves, resulting in abnormal readings or even missing readings (usually the abnormal reading is lower than the actual thickness). In this case, the material is not suitable for testing with this thickness gauge. -
Reference examples
A material of known thickness or sound velocity is used to calibrate the meter. At least one reference sample is required to calibrate the instrument. The ultrasonic thickness gauge is equipped with a calibration block with a thickness of 4.0 mm / 0.16”. Refer to the calibration instructions in chapter 8 of this manual for the calibration method.
Calibration (Cal)
- Calibration instructions
- a. Press the back key when you are in the measuring interface, select the menu item Calibration and confirm this selection with the OK key.
- b. The display will look as shown in the image to the right.
- c. During the measurement, the sound velocity is automatically set to 5900 m/s.
- d. Press the probe onto the calibration block to calibrate the meter.
Sound velocity calibration (V. Cal)
- a. Enter the submenu and select the item Sound velocity calibration (V.Cal).
- b. The display will look as shown in the image to the right.
- c. As shown in the image, press the up and down keys to set the value.
- d. When the value has been adapted, press OK to complete the calibration.
Maintenance
Cleaning
Alcohol and thinner can damage the housing, especially the display. Therefore,
only clean the meter with a damp cloth.
Troubleshooting Very thin material
With any ultrasonic thickness gauge, measurement errors occur when the
thickness of the material being tested is smaller than the lower limit of the
probe. When measuring ultra-thin materials, an error called “double
refraction” sometimes occurs, where the measured value is twice the actual
thickness. Another error is called “pulse envelope and loop jump” where the
measured value is higher than the actual thickness. To avoid such errors,
repeat the control measurement of critically thin material. Rust stains and
corrosion Rust stains and corrosion pits on the surface to be measured will
cause the readings to change irregularly or not at all in extreme cases; in
addition, small rust stains are sometimes difficult to detect. If pits
(blowholes) are found or suspected, be careful when measuring this area and
place the silver line of the probe at different angles for several tests.
Incorrect material identification If you have calibrated the instrument with
one material and use it to test another material, an incorrect result will be
displayed. Make sure to select the correct sound velocity. If the sound
velocity in the current test has some deviation from the calibration, errors
may also occur. Make an adjustment when using it. Probe abrasion The surface
of the probe is made of acrylic resin. After prolonged use of the meter, the
roughness of the probe surface increases, resulting in a decrease in accuracy.
If the user can detect measurement errors caused by this, the probe surface
can be polished with a little bit of sandpaper or 500 grit grindstone to
smooth the surface and ensure parallelism. If the result is still incorrect,
you may need to replace the probe. Laminated material, composite material It
is not possible to measure uncoupled laminated materials because ultrasonic
waves cannot penetrate the uncoupled space. Since ultrasonic waves cannot be
transmitted at a constant velocity in composite materials, measuring
instruments that measure thickness by ultrasonic reflection are not suitable
for measuring laminated and composite materials.
Effects of oxide layers on surfaces
Some metals can form a dense oxide layer on the surface, for example aluminum.
This oxide layer is firmly bonded to the substrate without an obvious boundary
layer. However, the transmission speed of ultrasonic waves is different in
these two substances, which causes errors. Different oxide layers cause
different errors that the user should be aware of. You can use a piece of the
same material the thickness of which you have checked with a gauge to
calibrate the meter. Use and selection of the coupling agent A coupling agent
is used to transmit high-frequency ultrasonic waves between the probe and the
material being tested. The wrong coupling gel or improper use can lead to
errors or failed measurements. It should be used in appropriate amounts and
applied evenly. The coupling agent is usually applied to the surface of the
material under test. If the temperature is too high, the coupling agent is
applied to the bottom of the probe. Only PCE Instruments coupling gel should
be used.
Warranty
You can read our warranty terms in our General Business Terms which you can
find here: https://www.pce-instruments.com/english/terms.
Disposal
For the disposal of batteries in the EU, the 2006/66/EC directive of the
European Parliament applies. Due to the contained pollutants, batteries must
not be disposed of as household waste. They must be given to collection points
designed for that purpose. In order to comply with the EU directive 2012/19/EU
we take our devices back. We either re-use them or give them to a recycling
company that disposes of the devices in line with the law. For countries
outside the EU, batteries, and devices should be disposed of in accordance
with your local waste regulations. If you have any questions, please contact
PCE Instruments.
Germany
- PCE Deutschland GmbH
- Im Langel 26
- D-59872 Meschede
- Deutschland
- Tel.: +49 (0) 2903 976 99 0
- Fax +49 (0) 2903 976 99 29
- info@pce-instruments.com.
- www.pce-instruments.com/deutsch.
United Kingdom
- PCE Instruments UK Ltd
- Unit 11 Southpoint Business Park
- Ensign Way, Southampton
- Hampshire
- United Kingdom, SO31 4RF
- Tel: +44 (0) 2380 98703 0
- Fax +44 (0) 2380 98703 9
- info@pce-instruments.co.uk.
- www.pce-instruments.com/english.
Turkey
- PCE Teknik Cihazlan Ltd.Sti.
- Halkali Merkez Mah.
- Pehlivan Sok No.6/C
- 34303 Kücükcekmece – Istanbul
- Türkiye
- Tel: 0212 471 11 47
- Faks: 0212 705 53 93
- info@pce-cihazlari.com.tr.
- ww.pce-instruments.com/turkish.
User manuals in various languages can be found by using our product search on : www.pceinstruments.com. Specifications are subject to change without notice.
References
- France.fr : Actualités, destinations et infos du tourisme en France
- iberica.es
- Make an offer on the domain instruments.co.uk - Domains.co.uk
- Computer Instruments | Home
- Discover Italy: Official Tourism Website - Italia.it
- Industrial Measurement Products and Solutions | PCE Instruments
- Industrial Measurement Products and Solutions | PCE Instruments
- PCE Deutschland GmbH Prüfgeräte vom Hersteller | PCE Instruments
- PCE Brookhuis B.V. | PCE Instruments
- PCE Americas Inc. : Test Instruments | PCE Instruments
- PCE Iberica S.L. Instrumentación | PCE Instruments
- PCE Instruments France | PCE Instruments
- PCE Italia s.r.l. / Strumenti di Misura | PCE Instruments
- PCE Teknik Cihazlar Paz. Tic. Ltd.Şti. | PCE Instruments
- PCE Americas Inc. : Test Instruments | PCE Instruments
- PCE Americas Inc. : Test Instruments | PCE Instruments
- PCE Teknik Cihazlar Paz. Tic. Ltd.Şti. | PCE Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>