SR9313B uses the standard USB, and easy access to PLC, DCS, and other instruments or systems for monitoring temperature, and humidity state quantities. The internal use of a high-precision sensing core and related devices to ensure high reliability and excellent long-term stability can be customized RS232, RS485, CAN,4-20mA, DC0~5V\10V, ZIGBEE, Lora, WIFI, GPRS and other output methods.
Technical Parameters
Technical parameter | Parameter value |
---|---|
Brand | SONBEST |
Temperature measuring range | -30℃~80℃ |
Temperature measuring accuracy | ±0.5℃ @25℃ |
Humidity measuring range | 0~100%RH |
Humidity accuracy | ±3%RH @25℃ |
Communication Interface | USB |
Default baud rate | 9600 8 n 1 |
Power | DC3.3~4.2V |
Storage | 512000bits |
Running temperature | -40~80°C |
Working humidity | 5%RH~90%RH |
Product Size
Application solution
KEY POSITION & SCREEN DISPLAY DESCRIPTION
- The upload interval can be set after connecting to the computer The storage interval can be set in PAGE The upper limit of the storage interval is 43199s
- The measurement record counts down, and the current measurement value will be recorded once after the countdown is over
- Display the current number of records
SOFTWARE INSTRUCTIONS
A: After opening the software, enter the main interface, select the
corresponding product model at the time of purchase, and save the settings
after selection, in case you need to reselect it after opening it next time.
B: Use the USB converter to insert the computer’s USB2.0 interface, the
driver will be installed automatically, If the installation fails, you can
install it in the driver wizard software. After connecting to the computer,
Select the serial port number: USB-SERIAL CH340 (COM5), set the baud rate
Click to open the serial port to complete the connection.
C D: After opening the serial port, click online collection (C) to see
the real-time measurement data (D).
E: Enter the parameter setting interface
F: A variety of parameters can be set in the parameter setting, such as
temperature and humidity correction, baud rate, upload time and other
parameters. For the parameters on the right, you can click on “read all” to
set the parameters on the recorder.
G: Enter the
record operation interface
H: Click “Read Data” to start transferring the data on the recorder to
the software.
I: After the step (H), the number of read data, the total number of data,
and the number of pages will be displayed at the top of the reading.
J: The read data will be displayed on it
K: You can export the read data in Excel or txt format
How to use?
Disclaimer
This document provides all information about the product, does not grant any
license to intellectual property, does not express or imply, and prohibits any
other means of granting any intellectual property rights, such as the
statement of sales terms and conditions of this product, other issues. No
liability is assumed. Furthermore, our company makes no warranties, express or
implied, regarding the sale and use of this product, including the suitability
for the specific use of the product, the marketability or the infringement
liability for any patent, copyright or other intellectual property rights,
etc. Product specifications and product descriptions may be modified at any
time without notice.
Contact Us
Company: Shanghai Sonbest Industrial Co., Ltd
Address:Building 8,No.215 North east road,Baoshan District,Shanghai,China
Web: http://www.sonbest.com
Web: http://www.sonbus.com
SKYPE: soobuu
Email: sale@sonbest.com
Tel: 86-021-51083595 / 66862055 / 66862075 / 66861077
References
- Home-上海æœåšå®žä¸šæœ‰é™å…¬å¸
- Home-SONBEST -SONBUS.COM| THE SENSOR COMPANY
- Home-SONBEST -SONBUS.COM| THE SENSOR COMPANY
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>