PT3SBT Wireless No-Touch Forehead Thermometer
INTRODUCTION
Thank you for choosing our product.
This product is a high-tech infra-red (IR) thermometer designed to take human
body temperature by measuring the energy of IR emitted from the forehead and
it can transmit the temperature to a smart device with Bluetooth. The product
will help you to assess you and your family members’ health conditions easily
and quickly.
Product Name: Wireless No-Touch Forehead Thermometer
Product Model: PT3SBT
Indication for use
The Wireless No-Touch Forehead Thermometer is intended for the intermittent measurement of body temperature from the forehead on people of all ages. It can transmit the temperature to a smart device with Bluetooth and can be used by consumers in the household environment and by healthcare providers.
Safety Precautions
Warning
- Use of this thermometer is not intended as a substitute for consultation with your physician. Please consult your doctor if you have any doubt about the temperature reading.
- Keep the thermometer out of reach of children. For accidental swallowing of the battery or other components, please contact emergency services immediately.
- Batteries must not be thrown into an open fire or short circuited.
Caution
(1) Measurements
- Thermometer readings should be regarded as a reference. Do not attempt self-diagnostics or self-treatment using the temperature readings. Please seek professional medical advice when necessary.
- There is no absolute standard for human body temperature. Knowing your own normal body temperature range is important to accurately determine if you have a fever.
- Make sure that the forehead of the subject is free from sweat, cosmetics, dirt, or grease before measuring.
- Patients should not drink, eat, or be physically active before/while taking a reading. Wait 30 minutes before taking a measurement. Temperature readings taken when a body is in a state of stable equilibrium is more accurate and useful as a reference.
- Do not take temperature measurement over scar tissue, open sores or abrasions.
- Do not measure body temperature from scar tissues on the forehead as such tissues will affect temperature conduction of the body.
- If there is a temperature difference between the thermometer storage area and the new ambient environment around the subject, please let the thermometer sit within the new environment for 30 minutes before taking the measurement.
- Do not measure body temperature immediately after consuming a drug that raises body temperature. Temperature readings taken at this time will not be accurate.
- It is normal for readings taken from continuous measurements to fluctuate within a small range. During continuous measurements, the subject’s body temperature may be transmitted to the thermometer, affecting measurement accuracy. We recommend taking only up to 3 continuous readings within a short period.
- Do not directly face the sun or an air outlet of an air conditioning or radiator device during the measurement as this will cause changes to the forehead temperature. Measurements should be taken in a stable environment where possible.
- Do not measure body temperature in an environment with strong EM interference (examples include places close to a working microwave, induction cooker, or cellphone in-use) as EM interference may cause errors in the reading or even device failure.
- Portable RF communications equipment (including peripherals such as antenna cables and external antennas) should be used no closer than 30 cm (12 inches) to any part of the product. Otherwise, degradation of the performance of this equipment could result.
- This product should be considered a personal device. Clean and sanitize the product properly to prevent cross contamination.
- To clean a dirty thermometer probe, gently swipe the probe using a cotton swab dipped in 70% isopropyl. Let the cleaned thermometer sit for at least 15 minutes before taking more measurements.
- This infrared thermometer meets requirements established in ISO 80601-2-56:2017/AC:2018. It displays subject’s temperature over a range of 89.6~109.4(32~42.9). Full responsibility for the conformance of this product to the standard is assumed by (manufacturer: ANDON HEALTH CO., LTD. Add: No. 3 Jinping Street, Ya An Road, Nankai District, Tianjin 300190, China).
(2) About the product
- This product is a precision device. Return the product to its original packaging for proper storage after use. To ensure accurate measurements, avoid the device or probe contacting any liquid or droplets. Avoid tiny particles (such as dust or powder) falling into the probe.
- Avoid dropping or subjecting the product to external forces. Do not disassemble or re-assemble the product on your own.
- Do not directly touch the probe with your fingers or blow on it. Measurements taken using a damaged or dirty IR probe may be inaccurate.
- Keep the product at a place inaccessible to children to prevent children from swallowing the batteries or small parts.
- Do not throw the thermometer or batteries into fire to prevent explosions.
- Remove the batteries from the thermometer if the device will not be used for more than one month.
- If you are allergic to plastic/rubber, please don’t use this device.
- The materials (ABS, TPU, metal) of expect contact with patient had passed the ISO 10993-5 and ISO 10993-10 standards test, no toxicity, allergy and irritation reaction. They are in compliance with the MDD requirements. Based on the current science and technology, other potential allergic reactions are unknown.
- The patient can be an intended operator. All functions of the unit can be used safely by patient.
- This product should be considered a personal device. Clean and sanitize the product properly to prevent cross contamination.
If performance of the instrument may be adversely affected should one or more of the following occurred:
- Operation outside of the manufacturer-specified subject temperature range.
- Operation outside of the manufacturer-specified operating temperature and humidity ranges.
- Storage outside of the manufacturer-specified ambient temperature and humidity ranges.
- Mechanical shock.
- Manufacturer-defined soiled or damaged infrared optical components.
-
The patient is an intended operator.
-
This product contains batteries and recyclable electronic waste. To
protect the environment, do not dispose of it in the household waste, but take it to appropriate local collection points. -
This equipment complies with FCC RF radiation exposure limits set forth for an uncontrolled environment.
-
This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) This device may not cause harmful interference, and
(2) This device must accept any interference received, including interference that may cause undesired operation. -
The manufacturer is not responsible for any radio or TV interference caused by unauthorized modifications to this equipment. Such modifications could void the user’s authority to operate the equipment.
-
This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference in a residential installation. This equipment generates uses and can radiate radio frequency energy and, if not installed and used in accordance with the instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
– Reorient or relocate the receiving antenna.
– Increase the separation between the equipment and receiver.
-Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
-Consult the dealer or an experienced radio/TV technician for help.
Please take attention that changes or modification not expressly approved by the party responsible for compliance could void the user’s authority to operate the equipment. -
This device complies with Industry Canada license-exempt RSS standard(s). Operation is subject to the following two conditions: (1) this device may not cause interference, and (2) this device must accept any interference, including interference that may cause undesired operation of the device. Under Industry Canada regulations, this radio transmitter may only operate using an antenna of a type and maximum (or lesser) gain approved for the transmitter by Industry Canada. To reduce potential radio interference to other users, the antenna type and its gain should be so chosen that the equivalent isotropically radiated power (e.i.r.p.) is not more than that necessary for successful communication. Le présent appareil est conforme aux CNR d’Industrie Canada applicables aux appareils radioexempts de licence.
Overall Description
The thermometer is mainly comprised of a plastic casing, IR temperature
sensor, MCU, vibration motor, LED display screen and batteries.
LED screen instructions
Device dimensions: 5.39 in x 1.34 in x 1.54 in
Screen dimensions: 1.01 in x 0.75 in
Product weight: 90 g
Product performance
-
Measurement position: Centre of the forehead surfac
-
Measurement distance: 1.18 in(3 cm)
-
Power source: DC 3V: 2 x 1.5V size AAA batteries
-
Measurement range: 89.6-109.2(32-42.9)
-
Measurement precision: ±0.4(±0.2) within 93.2-107.6(34 -42), and±0.5(±0.3) for other temperature ranges.
-
Resolution: 0.1(0.1)
-
Clinical reproducibility: Within ±0.5(±0.3)
-
Operating conditions:
Temperature: 32-104(0-40)
Humidity: 15-95%RH, non-condensing
Atmospheric Pressure: 70~106kP -
Operation mode:
Adjusted mode: forehead -
Expected service life: 5 year
-
Reference Body Site: Oral
-
Wireless communication: Bluetooth low energy
-
Software version: V1.0
ASTM laboratory accuracy requirements in the display range of 37 to 39°C (98 to 102°F) for IR thermometers is ± 0.2 °C ( ± 0.4°F), whereas for mercury-in- glass and electronic thermometers, the requirement per ASTM Standards E667-86 and E1112-86 is ±0.1 °C (±0.2°F).
Expected use and scope of applications
This product mainly uses infrared sensors to measure the temperature of the
human forehead and transmit the temperature to a smart device with Bluetooth.
It can be used for babies, children, and adults. Infants and children cannot
operate the thermometer; it is recommended that adults take the measurement.
Reminder: Temperature readings may differ according to skin tone and
measurement distance.
Contraindications
It is not recommended for people whose measuring part has local lesions, such as inflammation, trauma, postoperative, etc.
Instruction For Use
The patient is an intended operator. The patient can measure, transmit data and change battery.
(1) Installing the product
Insert two batteries into the battery compartment on the back of the device. The product will initiate a self-check and then enter the standby mode (if the device indicates low battery power, then replace the batteries to ensure ample power supply).
(2) Measurement process
① Aim the thermometer probe at the center of the forehead and keep the probe
1.18 in (3cm) away from the forehead (the optimal distance is about the width
of an adult’s index finger). Do not touch the forehead with the probe. The
infrared thermometer measurement window should be held perpendicular to the
forehead and the subject should remain steady.
② Gently press the measurement
button to start the measurement.
③ The device will vibrate once a reading is obtained. Once the measurement
is successful, the reading will be displayed on the screen. The whole time
is about 1s. If the measurement is failed, the device will not vibrate the
screen but display instead.
If the [ ] icon on the thermometer screen is always on, the measured
temperature data can be uploaded to the smart device.
Possible reasons for measurement errors include:
A. Measurement distance is greater than 1.18 in (3cm);
B. Environment temperature does not meet measurement requirements or there is
an excessively large temperature difference;
C. Target temperature exceeds measurement range.
(3) Power off
If no more measuring is required, simply let the device sit idle for 8 seconds to power of automatically.
(4) Unit switching
- When the device is powered OFF, press and hold the measurement button for 8 seconds to enter unit switching mode. Under this mode, both and should flash at the same time.
- Under this mode, press the measurement button to switch to Fahrenheit or Celsius. The corresponding symbol of the selected unit after switching will flash.
- After selecting the unit, press and hold the measurement button for 8 seconds to leave the unit switching mode.
Reminder: If you do not press and hold the measurement button for 8 seconds to confirm the unit selected before the screen switches of, the system will consider the switch as a failure and retain the original unit for temperature measurements.
(5) Installing and replacing batteries
Each time the thermometer is started, the battery will be automatically detected. If battery capacity is low but adequate for measurements, the low power symbol [ ] will be displayed with the measurement results. However, if the battery capacity is too low for measurements, the screen will display a single, flashing [ ] icon and automatically switch OFF after 8 seconds. To continue using the device, old batteries must be replaced.
(6) Replacing batteries
- Press down and hold the battery cover with the finger and apply some force to slide the cover backwards to open the battery compartment.
- Remove the old batteries and install the new batteries.
- Refer to the battery polarity symbols to orient the batteries properly during installation. Make sure that the new batteries are tightly inserted into the battery compartment and make sure that the polarity is not reversed during installation.
- Return the battery cover to close the battery compartment.
- Comply with relevant national laws and regulations when disposing of the used batteries.
- Do not dispose of batteries directly into the trash bag.
- Remove the batteries if the device will not be used more than one month.
- When using, shall not touch battery and the patient simultaneously.
- Do not throw batteries into fire.
- The typical service life of the new and unused batteries is 300 measurements for the operation time are 60s.
SYNC TIME AND RESULTS VIA BLUETOOTH
Prior to first use, scan the QR code to install the iHealth App.
Note: The App works with both iOS and Android device. For a complete list of
compatible devices, visit our support on page on www.ihealthlabs.eu.
Connect to iOS Device via Bluetooth
- Launch the iHealth App from your iOS device.
- Enable Bluetooth on your iOS device: Slide gently upward the iPhone screen from to open the Control Center, and click on the .
- Once the thermometer is off, press the button to start one measurement or sync the results.
- When a connection has been established, the Bluetooth indicator will always light up, if not, the Bluetooth indicator will be off.
- Please repeat these steps when you switch to another iOS device with the thermometer.
Connect to Android Device via Bluetooth
- Press the button to turn on the thermometer or start one measurement.
- Enable Bluetooth on your Android device.
- When using the thermometer for the first time, you should pair the thermometer to the Android device. Go into your device’s setting to pair it with your Android device. Check the Bluetooth menu for the model name of your thermometer (PT3SBT) to appear, and then select it to pair.
- Launch the iHealth app to sync with your thermometer.
- When a connection has been established, the Bluetooth indicator will always light up, if not, the Bluetooth indicator will be off.
- Please repeat these steps when you switch to another Android device with the thermometer.
Update the firmware of PT3SBT
If a new version of the PT3SBT’s firmware is released, the iHealth App will
inform user to update when the PT3SBT is connected to the smart device with
Bluetooth. User can choose whether to update this version of the PT3SBT’s
firmware, but it is strongly recommended to do this updating to get a better
experience.
NOTE: During the upgrade process, the Bluetooth icon on the thermometer
is always on and the three horizontal bars are displayed in cycles.
Product errors and troubleshooting
Problem | Item to detect | Solution |
---|---|---|
- Battery depleted
| Replace old batteries with new ones
- Batteries have been installed with the wrong polarity
- Batteries are not installed properly
| Take out the batteries and re-install them correctly.
****|
- Unable to carry out measurement as current battery capacity is too low.
| Replace the batteries.
****|
- Meaurement distance is too long
- Target temperature is beyond range of measurements 89.6℉ -109.2℉(32℃-42.9℃)
- Environment temperature either exceeds the design range 32℉ -104℉(0℃-40℃)or is unstable
| Follow the instruction manual and repeat the measurements
****|
- Current state: all symbols are falshing on the screen. The product is not usable.
| Please contact customer services.
****|
- Current battery capacity is too low.
| Relpace the battery as soon as possible.
The application cannot find the product.|
- There may be a problem with the Bluetooth connection
|
- Restart the device.
- Turn off and turn on the Bluetooth of the smart device.
Body temperature
-
Body temperature varies from person to person and fluctuates during the course of the day. For this reason, it is suggested to know one’s normal, healthy forehead temperature to correctly determine the temperature.
-
Body temperature runs approximately from 95.9°F to100°F(35.5℃-37.8℃).
To determine if one has a fever, compare the temperature detected with the person’s normal temperature. A rise over the reference body temperature of 1.8℉(1℃) or more is generally an indication of fever. -
Different measurement sites (rectal, axillary, oral, frontal, auricular) will give different readings. Therefore it is wrong to compare the measurement taken from different sites
-
The following are typical temperatures for adults, based on different measurement sites:
Rectal | 97.9℉-100.4℉ | 36.6℃ to 38℃ |
---|---|---|
Axillary | 94.5℉-99.1℉ | 34.7℃ to 37.3℃ |
Oral | 95.9℉-99.5℉ | 35.5℃ to 37.5℃ |
Auricular | 96.4℉-100.4℉ | 35.8℃ to 38℃ |
Cleaning and Care
The thermometer is mainly for home use, if use for multiple patients, please
clean the device in between uses with the following steps:
① Use an alcohol swab or cotton tissue moistened with alcohol (70% Isopropyl)
to clean the thermometer casing thoroughly,. The wiping process preferably
lasts 15 seconds.
② Allow at least 5 minutes drying time before taking a temperature.
③ After cleaning, if the device is not visually clean when observed with
magnification and adequate lighting, please repeat the clean steps above.
Note 1. The cleaning steps above has been validated according to the FDA
Guidance, “Reprocessing Medical Devices in Health Care Settings: Validation
Methods and Labeling”.
Note 2: Ensure that no liquid enters the interior of the device. Never use
abrasive cleaning agents, thinners or benzene for cleaning and never immerse
the device in water or other cleaning liquids.
Care the probe
- The probe (comprising a temperature sensor and a distance sensor) is the most intricate part of the thermometer, and should be kept clean and intact to acquire accurate readings. Use the following method to clean the probe: Gently swab the surface of the probe using a cotton bud soaked in > with 70% isopropyl.
- If the probe (sensor) is broken, please contact customer services.
- The product is not waterproof. Do not clean the device with detergents. Do not soak the thermometer in water or other liquids.
Maintenance
- This company has not authorized any agency or individual to carry out product repairs or maintenance. Do not attempt to disassemble or modify the thermometer if you suspect functional issues with the device.
- The IR thermometer is an extremely precise instrument. Any improper maintenance, disassembly, or modification may lead to inaccuracies of the product measurements.
- If you suspect any product issues during the warranty period, please contact customer services for subsequent handling.
Disposal
| Dispose of the battery in accordance with the regulation applicable at the
place of operation. Dispose of at public collection point in the EU countries
– 2002/96/EC WEEE Directive.
If you have any queries, please refer to the local authorities responsible for
waste disposal.
---|---
Notes
- Please act according to local laws for disposal of used batteries.
- Take out the battery if you are not going to use the unit for a long time.
| To protect the environment, dispose of empty batteries at appropriate
collection sites according to national or local regulations.
Dispose of at public collection point in the EU countries – 2006/66/EC
Directive.
---|---
Operating, shipping, and storage conditions
(1) Operating conditions
Range of operating environments
Temperature: 32℉-104℉(0℃- 40℃)
Humidity: 15-95%RH, non-condensing
Atmospheric Pressure: 70kPa~106kPa
(2) Transportation / storage conditions
Main Unit: -4℉~131℉(-20℃~ 55℃)
Humidity: 15-95%RH, non-condensing
Atmospheric Pressure: 70kPa~106kPa
Accessories
Use only accessories provided by the original manufacturer, and check for any missing accessories.
- 1 IR Thermometer
- 2 AAA batteries
- 1 Storage Box
- 1 Instruction manual
- 1 Quick user guide
Standard icons
| Symbol for “THE OPERATION GUIDE MUST BE
READ”
The sign background color: blue.
The sign graphical symbol: white.
---|---
| The batteries and electronic instruments must be disposed of in accordance
with the locally applicable regulation, not with domestic waste.
| Symbol for “MANUFACTURER”
| Symbol for “WARNING”
| Symbol for “Batch code”
| Symbol for “EUROPEAN REPRESENTATIVE”
**IP22**
| IP code of the device: this device’s grade of against ingress of solid foreign objects
--
| ≥ 12.5mm diameter (and the against access to hazardous parts with finger);
the grade of waterproof is dripping (15° tilted).
| Symbol for “COMPILES WITH MDD93/42/EEC REQUIREMENTS”
iHealth is a trademark of iHealth Labs Inc.
The Bluetooth® word mark and logos are registered trademarks owned by
Bluetooth SIG, Inc. And any use of such marks by iHealth Labs, Inc. is under
license.
Calibration
The thermometer is initially calibrated at the time of manufacture. If this thermometer is used according to the use instruction, periodic re-adjustment is not required. If any time your question the accuracy of measurement, please contact distributor or manufacturer, the contact information is on the last page.
Storage
- Do not put the thermometer under direct sunlight, high temperature, or moist environments. Do not allow it to come into contact with re or harsh vibrations.
- Take out the battery if the device is not used for a long period of time.
OTHER STANDARDS AND COMPLIANCES
The Wireless No-Touch Forehead Thermometer corresponds to the following
standards: IEC 60601-1:2005 +A1:2012(E)/EN 60601-1:2006/A1: 2013
(Medical electrical equipment — Part 1: General requirements for basic safety
and essential performance), IEC 60601-1-2:2014(Medical electrical equipment —
Part 1-2: General requirements for basic safety and essential performance
Collateral standard: Electromagnetic compatibility – Requirements and Tests),
ISO 80601-2-56:2017/AC:2018( Medical Electrical Equipment — Part 2-56:
Particular Requirements For The Basic Safety And Essential Performance Of
clinical thermometers for body temperature measurement), ASTM E1965-98:2016(
Standard Specification for Infrared Thermometers for Intermittent
Determination of Patient Temperature).
The SAR limit of USA (FCC) is 1.6 W/kg averaged over one gram of tissue. Device has also been tested against this SAR limit.
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Frequency range
Bluetooth: 2402MHz~2480MHz
Max EIRP: 4.40dBm
Table 1 – Emission limits per environment
Phenomenon | Compliance | Electromagnetic environment |
---|---|---|
Conducted and radiated RF emissions | CISPR 11 Group 1, Class B | The device is |
intended to be used in home healthcare environment
Harmonic distortion| IEC 61000-3-2 NA| The device is powered by battery
Voltage fluctuations and flicker| IEC 61000-3-3 NA| The device is powered by
battery
Table 2 – Enclosure Port
Phenomenon | Basic EMC standard | Immunity test levels |
---|
Home Healthcare Environment
Electrostatic Discharge| IEC 61000-4-2| ±8 kV contact
±2kV, ±4kV, ±8kV, ±15kV air
Radiated RF EM field| IEC 61000-4-3| 10V/m
80MHz-2.7GHz
80% AM at 1kHz
Proximity fields from RF wireless communications equipment| IEC 61000-4-3|
Refer to table 3
Rated power frequency magnetic fields| IEC 61000-4-8| 30A/m 50Hz or 60Hz
Table 3 – Proximity fields from RF wireless communications equipment
Test frequency (MHz) | Band (MHz) | Immunity test levels |
---|
Professional healthcare facility environment
385| 380-390| Pulse modulation 18Hz, 27V/m
450| 430-470| FM, ±5kHz deviation, 1kHz sine, 28V/m
710| 704-787| Pulse modulation 217Hz, 9V/m
745
780
810| 800-960| Pulse modulation 18Hz, 28V/m
870
930
1720| 1700-1990| Pulse modulation 217Hz, 28V/m
1845
1970
2450| 2400-2570| Pulse modulation 217Hz, 28V/m
5240| 5100-5800| Pulse modulation 217Hz, 9V/m
5500
5785
WARRANTY
Please contact your dealer or the device center in case of a claim under the
warranty. If you have to send in the unit, enclose a copy of your receipt with
clear statement of defect description.
The warranty terms as below:
-
The warranty period for device is one year from date of delivery. In case of a warranty claim, the date of delivery has to be proven by means of the sales receipt or invoice.
-
Repairs under warranty do not extend the warranty period.
-
The following cases are excluded under the warranty
• All damage which has arisen due to improper treatment, e.g. nonobservance of the user instruction.
• All damage which is due to repairs or tampering by the customer or unauthorized third parties.
• Damage which has arisen during transport from the manufacturer to the consumer or during transport to the service center.
• Accessories which are subject to normal wear and tear. -
Liability for direct or indirect consequential losses caused by the unit is excluded even if the damage to the unit is accepted as a warranty claim.
Manufactured for iHealth Labs Inc.
120 San Lucar Ct., Sunnyvale, CA 94086, USA
+1-855-816-7705
[ **www.ihealthlabs.com**](http://www.ihealthlabs.com)
iHealthLabs Europe SAR
36 Rue de Ponthieu, 75008, Paris, France
support@ihealthlabs.eu |
www.ihealthlabs.eu
ANDON HEALTH CO., LTD.
No. 3 Jinping Street, Ya An Road, Nankai District, Tianjin 300190, China
Tel: 86-22-60526161
This product can be used across EU member states.
EU Regulatory Conformance
Hereby, iHealth Labs Inc. declares that this device is in compliance with the
essential requirements and other relevant provisions of Directive 2014/53/EU.
For the declaration of conformity, visit the Web site
https://www.ihealthlabs.com.
Date of issue: July 16, 2020
PT3SBT-SMSY02 V1.0
References
- iHealth Labs – iHealth Labs Inc
- iHealth Labs Europe - Connected Health
- iHealth Labs – iHealth Labs Inc