iHealth PO3M Wireless Pulse Oximeter Instruction Manual
- June 6, 2024
- iHealth
Table of Contents
- OPERATION MANUAL
- INTENDED USE
- PARTS AND DISPLAY INDICATORS
- PARTS AND DISPLAYS
- USING YOUR PULSE OXIMETER
- TESTING INSTRUCTIONS
- CARE AND MAINTENANCE
- SPECIFICATIONS
- TROUBLESHOOTING
- IMPORTANT INFORMATION REQUIRED BY THE FCC
- ELECTROMAGNETIC COMPATIBILITY INFORMATION
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Wireless Pulse Oximeter
OPERATION MANUAL
MODEL PO3M
Wireless Pulse Oximeter
OPERATION MANUAL
SYMBOLS
The symbols below associate with your PO3M
Symbols
|
Definition of Symbol
---|---
| Symbol for”THE OPERATION MANUAL MUST BE READ”
**| Symbol for “WARNING”
| Symbol for”Type BF Applied Part”
| Symbol for”no alarm for Sp02″
| Symbol for “ENVIRONMENT PROTECTION-Waste electrical products should not be
disposed of with household waste. Please recycle where facilities exist. Check
with your local Authority or retailer for recycling advice”.
| Symbol for”Use by date”
| Symbol for”Manufacturer”
| Symbol for”EUROPEAN REPRESENTATIVE”
SN| Symbol for “SERIAL NUMBER”
| Symbol for”KEEP DRY”
0197**| Symbol for”COMPILIES WITH MDD93/42/EEC REQUIREMENTS”
INTENDED USE
The PO3M Wireless Pulse Oximeter is a non-invasive device intended for spot-
checking of functional oxygen saturation of arterial hemoglobin (SpO2) and
pulse rate.
The wireless pulse oximeter is intended to measure blood oxygen saturation and
pulse rate of adults above 16 years old in-home and hospital environments
(including clinical
use in internist/surgery, anesthesia, intensive care, etc.).
The Wireless Pulse oximeter is not intended for continuous monitoring.
Compatibility
The Wireless Pulse Oximeter PO3M is designed for use with the following
devices:
iPod touch 5th generation
iPhone 4S
iPhone 5
New iPad
iPad 4
with an iOS version of V5.0 or higher.
iPhone and iPod touch are trademarks of Apple Inc., registered in the U.S. and
other countries. iPad is a trademark of Apple Inc.
PARTS AND DISPLAY INDICATORS
One (1) Wireless Pulse Oximeter PO3M
One (1) Lanyard
One (1) Operation Manual
One (1) Quick Start Guide
One (1) USB cable
PARTS AND DISPLAYS
DEVICE DESCRIPTION.
PO3M pulse oximeter measures the amount of oxygen in your blood and the pulse
rate. The oximeter works by shining two light beams into the small blood
vessels or capillaries of the finger; the measured signal is then obtained by
a photosensitive element and processed by the microprocessor. The oxygen
saturation (SpO2) is measured as a percentage of full capacity.
Typically, a SpO2 reading between 94%-99% is considered normal. High altitudes
and other factors may affect what is considered normal for a given individual.
Concerns about your readings should be shared with your physician or
healthcare professional.
IEC 60601-1: 2005 + CORR. 1 (2006) + CORR. 2 (2007)/EN 60601-1:2006 /AC:2010(
Medical electrical equipment-Part1: General requirements for basic safety and
essential performance)
IEC60601-1-2:2007/EN 60601-1-2:2007 /AC:2010 (Medical electrical equipment —
Part
1-2: General requirements for basic safety and essential performance –
Collateral standard: Electromagnetic compatibility – Requirements and tests)
IEC 60601-1-11 (First Edition): 2010(Medical electrical equipment-Part 1-11:
General requirements for basic safety and essential performance – Collateral
Standard:
Requirements for medical electrical equipment and medical electrical systems
used in the home healthcare environment)
ISO 80601-2-61:2011 (Medical electrical equipment-Particular requirements for
the basic safety and essential performance of pulse oximeter equipment for
medical use).
CONTRAINDICATIONS
The PO3M Wireless Pulse Oximeter cannot be used on infant babies.
WARNINGS
- This device is for use by adults only.
- Certain activities may pose a risk of injury, including strangulation if the lanyard should become wrapped around your neck. Use the lanyard with caution.
- Do not use the device in a magnetic resonance (MR) environment.
Notice
-
Do not use the device as the only basis for making medical decisions. It is intended only to be used as additional information that you can give to your licensed health care professional.
-
The device might misinterpret excessive movement as good pulse strength. Limit finger movement as much as possible when using the device.
-
Do not use the device on the same hand/arm when using a blood pressure cuff or monitor.
-
The device has no alarms of blood oxygen saturation and pulse rate, and it will not sound if the amount of oxygen in your blood is too low or your pulse rate is abnormal.
If the measurement of SpO2 and pulse rate is not in the normal range, please contact your health care professional. -
Do not place the device in liquid or clean it with agents containing ammonium chloride or products that are not listed in this Operation Manual.
-
Any of the following conditions may reduce the performance of the device:
a)Flickering or very bright light;
b)Excessive Movement;
c)Weak pulse quality (low perfusion);
d)Low hemoglobin;
e)Nail polish, and/or artificial nails;
f)Any tests recently performed on you that required an injection of intravascular dyes -
The device may not work if you have poor circulation. Rub your finger to increase circulation, or place the device on another finger.
-
The device measures oxygen saturation of functional hemoglobin. High levels of dysfunctional hemoglobin (caused by sickle cell anemia, carbon monoxide, etc.) could affect the accuracy of the measurements.
-
Do not use the device in a combustible environment (oxygen-enriched environment).
-
Do not use the device outside the specified operating temperature range, and do not store the device outside the specified storage temperature ranges
-
The materials used in the device conform to the biocompatibility and nontoxic regulations and present no hazard to the body.
-
Use in emergency vehicles with communication systems may affect the accuracy of the measurements.
-
The packaging of the device is recyclable, and it must be collected and disposed of according to the related regulation in the country or region where the package of the device or its accessories is opened.
-
Any material of the device that may cause pollution must be collected and disposed of according to local rules and requirements.
-
Any single functional tester cannot be used to assess the accuracy of a pulse oximeter.
-
Do not stare at the lighting LED, as it may irritate your eyes.
-
The device is calibrated to display FUNCTIONAL OXYGEN SATURATION
-
Do not use the device for more than 30 minutes.
-
The wavelength range of pulse oximeter can be especially useful to clinicians
-
Because the pulse oximeter measurements are statistically distributed, only about two-thirds of pulse oximeter measurements can be expected to fall within ±Arms of the value measured by an oximeter.
-
The SpO2 accuracy was tested by comparing it to a Co-oximeter and the pulse rate accuracy was tested by comparing it to a function tester.
USING YOUR PULSE OXIMETER
Before Using Pulse Oximeter
The wireless pulse oximeter may be used when the user is seated, standing, or
lying down. The user should not walk or run during measurements and should
take care of not excessively moving the finger where the device is attached
and the corresponding hand and arm.
It is recommended that the user should wash hands before use. Nail polish,
especially dark shades, may affect the accuracy of the measurement and it is
suggested that any polish be removed prior to monitoring.
The device is suitable for use on any finger excluding the thumb. It is
preferred to use the index or middle finger.
Charge the battery before first use
Link the wireless Pulse Oximeter to a USB port of an electrical power source
(maybe a
PC), and press the “start” button. Then the battery indicator “ ” will blink,
which means the battery charging is started. When the battery indicator “ ”
stops blinking, it means the battery has been fully charged.
Download App
Download the app from the App Store and install it. Search for “iHealth
MyVitals”. (Your compatible iOS device should be version 5.0 or later.)
Create User and Cloud Account
After downloading the app, register and set up your user account following the on-screen instructions.
When setting up your user account, you will also have access to a free, secure
iHealth Cloud account. Go to
www.ihealthlabs.com, then click “Sign In” to
access your cloud
account from PC or Mac.
Turn Bluetooth “On”
Turn Bluetooth “On” under the “Settings” menu of your iOS device (The date and
time of the Pulse Oximeter will be synced with your iOS device upon first
successful connection). Once your Bluetooth is on, the Pulse Oximeter will
connect automatically when the app is launched and the Bluetooth indicator on
the screen of the oximeter will
turn on.
TESTING INSTRUCTIONS
-
Open the clamp of the Pulse Oximeter PO3M, then place the middle, ring, or index finger of your left hand into the rubber opening of the oximeter with the nail side down, as pictured.
-
On the front panel, press the “Start” button once to turn the oximeter on.
-
Keep your hand still for the reading.
Note: If the pulse signal is too weak to measure, dashes (- – -) will appear on the display. -
After a few seconds, your SpO2 reading will appear on the oximeter display screen and the app if the app is turned on.
-
If the signal strength is too low, switch to another finger and perform the test again.
-
Remove the oximeter from the finger. The oximeter will automatically turn off after 8 seconds.
USING WITHOUT iOS DEVICE
After it has been used for the first time, the date and time of the Pulse
Oximeter PO3M will be synchronized with your iOS device. It can also be used
without being connected
to an iOS device. In this case, the measurement data is stored in the memory
and can be uploaded to the app when the connection is re-established. The
Pulse Oximeter PO3M
can store up to 100 measurements. When the memory is full, any new measurement
overwrites the oldest ones.
CARE AND MAINTENANCE
- Clean the device once per week or more frequently if handled by multiple users.
- Wipe the device with a soft cloth dampened with rubbing alcohol to avoid cross-infection. Do not pour the alcohol directly on or into the device. Dry with a soft cloth, or allow to air dry.
- Avoid dropping this device on a hard surface.
- Do not immerse the device in water or other liquid, as this will result in damage to the device.
- If this device is stored below 0℃, please acclimate the device to room temperature before use.
- Do not try to disassemble this device.
- The PO3M is a precision electronic instrument and must be repaired/serviced by an accredited iHealth service center.
- Incorrect replacement of battery by inadequately trained personnel could result in an unacceptable risk (e.g., excessive temperatures, fire, or explosion).
- The patient simulator « Index 2 », made by the Fluke company, can be used to verify the operation of the oximeter.
- The expected service life of the PO3M is about 5 years.
SPECIFICATIONS
- Model: PO3M
- Classification: Internally powered, type BF
- Enclosure degree of ingress protection: IPX1
- Display System: LED
- Power Source: battery 3.7V Lithium-ion 330mAh
- Peak wavelength: 660nm/880nm;
- Maximum optical output power: 1mW;
- SpO2 Measuring Range: 70-99%
- Average Root Mean Square (ARMS) of SpO2 Accuracy: 80%~99%:±2%, 70%~79%: ±3% ,<70%: no definition.
-
Range Arms 90%-100% 1.2215 80%,..,89% 1.3282 70%-79% 1.7277
The figure below shows the graphical plot of all SaO2 versus SpO2 with linear
regression fit for all the sample data in the clinical protocol.Scatter plot of SaO2 versus SpO2 with linear regression fit
The figure below shows the graphical plot of SaO2 versus error (SpO2 – SaO2)
with upper 95% and lower 95% limits of agreement:Scatter plot of the difference between methods against
the SaO2
- Pulse Rate Measuring Range: 30/min-250/min
- Pulse Rate Accuracy: 30/min ~ 99/min: ±2, 100/min ~ 250/min: ±2%.
- Data update period: 15s
- Automatic Shut-off: After 8 seconds of no indication on the sensors
- Operation Environment: 5℃-40℃; Humidity <80%; Atmospheric pressure: 700hPa-1060hPa
- Storage Environment: -20℃-55℃; Humidity <95%; Atmospheric pressure: 700hPa-1060hPa
TROUBLESHOOTING
Problem| Possible Cause|
Solution
---|---|---
SpO2 or pulse rate shows no value, or the number fluctuates.| 1. Finger
may not be inserted correctly.
2. Finger or hand may be moving.
3. The device may be damaged| 1. Remove finger and reinsert, as directed.
2. Try to keep perfectly still and test again.
3. Please contact the iHealth Customer Service
The device does not turn on.| 1. The battery may be low.
2. The device might be damaged.| 1. Charge the battery and try again.
2. Please contact the iHealth Customer Service
“El” is displayed on the screen| The sensor is damaged| Please contact the
iHealth Customer Service
The low battery indicator is blinking.| The battery is low.| Charge the
battery and try again.
The app cannot find the Wireless Pulse Oximeter PO3M.| The Bluetooth does not
work| Reestablish the Bluetooth connection. If still not successful, restart
your wireless device (iPod, iPhone, or iPad).
IMPORTANT INFORMATION REQUIRED BY THE FCC
This device complies with Part 15 of the FCC Rules. Its operation is subject
to the following two conditions:
(1) This device may not cause harmful interference.
(2) This device must accept any interference received, including interference
that may cause undesired operation.
Note: This product has been tested and found to comply with the limits
for a Class B digital device, pursuant to Part 15 of the FCC Rules. These
limits are designed to provide
reasonable protection against harmful interference in a residential
installation. This product generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may
cause harmful interference to radio communications. However, there is no
guarantee that interference will not occur in a particular installation. If
this product does cause harmful interference to radio or television reception,
which can be determined by turning the equipment off and on, the user is
encouraged to try to correct the interference by one or more of the following
measures:
—Reorient or relocate the receiving antenna.
—Increase the separation between the equipment and receiver.
—Connect the equipment into an outlet on a circuit different from that to
which the receiver is connected.
—Consult the dealer or an experienced radio/TV technician for help.
IMPORTANT INFORMATION REQUIRED BY THE INDUSTRY CANADA
Under Industry Canada regulations, this radio transmitter may only operate
using an antenna of a type and maximum (or lesser) gain approved for the
transmitter by
Industry Canada. To reduce potential radio interference to other users, the
antenna type and its gain should be so chosen that the equivalent
isotropically radiated power (e.i.r.p.) is not more than that necessary for
successful communication.
This device complies with Industry Canada license-exempt RSS standard(s).
Operation is subject to the following two conditions: (1) this device may not
cause interference, and (2) this device must accept any interference,
including interference that may cause undesired operation of the device.
Manufacturer Information
ANDON HEALTH CO., LTD.
No. 3 Jinping Street, YaAn Road, Nankai District, Tianjin 300190, China.
Tel: 86-22-60526081
IHealthLabs Europe SARL
3 rue Tronchet, 75008, Paris, France
Tel: +33 1 44 94 04 81
EC REP
Health is a trademark of iHealth Lab Inc. Bluetooth® associated logos are
registered trademarks owned by
Bluetooth SIG, Inc. and any use of such marks by iHealth Lab Inc. is permitted
under the license.
Other trademarks and trade names are those of their respective owners.
ELECTROMAGNETIC COMPATIBILITY INFORMATION
Table 1
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacture’s declaration – electromagnetic emissions
The PO3M is intended for use in the electromagnetic environment specified
below.
The customer or the user of the PO3M should assure that it is used in such an
environment.
Emissions test
| Compliance|
Electromagnetic environment – guidance
---|---|---
RF emissions CISPR 11| Group 1| The PO3M uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
RF emissions CISPR 11| Class B| The PO3M is suitable for use in all
establishments other than domestic and those directly connected to the public
low-voltage power supply network that supplies buildings used for domestic
purposes.
Harmonic emissions IEC/EN 61000-3-2| Not applicable
Voltage fluctuations/ flicker emissions IEC/EN 61000-3-3| Not applicable
Table 2
For all ME EQUIPMENT and ME SYSTEMS
Guidance and manufacturer’s declaration – electromagnetic immunity
The PO3M is intended for use in the electromagnetic environment specified
below. The customer or the user of the PO3M should assure that it is used in
such an environment.
IMMUNITY test
| IEC 60601test level| Compliance level|
Electromagnetic environment – guidance
---|---|---|---
Electrostatic discharge (ESD) IEC/EN 61000-4-2| ± 6 kV contact ± 8 kV air| ± 6
kV contact ± 8 kV air| Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative humidity should be at
least 30 %.
Power frequency (50/60 Hz) magnetic field IEC/EN 61000-4-8| 3 A/m| 3 A/m|
Power frequency magnetic fields should be at levels characteristic of a
typical location in a typical commercial or hospital environment.
NOTE: UT is the A.C. mains voltage prior to application of the test
level.
Table 3
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Guidance and manufacturer’s declaration – electromagnetic immunity
The PO3M is intended for use in the electromagnetic environment specified
below.
The customer or the user of the PO3M should assure that it is used in such an
environment.
IMMUNITY test| IEC/EN 60601 test level| Compliance level|
Electromagnetic environment – guidance
---|---|---|---
Radiated RF| 3 V/m 80 MHz to| 3 V/m| Portable and mobile RF
IEC/EN 61000-4-3| 2.5 GHz| |
communications equipment should be used no closer to any part of the [1303M], including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
| | | Recommended separation distance :
| | | d=1.21ff 80 MHz to 800 MHz
d=2.31,/TD 800 MHz to 2,5 GHz
| | |
Where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m). Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey, should be less than the compliance level in each frequency range.b Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
A) Field strengths from fixed transmitters, such as base stations for radio
(cellular/cordless) telephones and land mobile radios, amateur radio, AM and
FM radio broadcast, and TV broadcast cannot be predicted theoretically with
accuracy. To assess the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the PO3M
is used exceeds the applicable RF compliance level above, the PO3M should be
observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as re-orienting or relocating the
[PO3M].
B) Over the frequency range 150 kHz to 80 MHz, field strengths should be less
than [V1] V/m.
Table 4
For ME EQUIPMENT and ME SYSTEMS that are not LIFE-SUPPORTING
Recommended separation distances between portable and mobile RF communications
equipment and the [PO3M] The PO3M is intended for use in an electromagnetic
environment in which radiated RF disturbances are controlled. The customer or
the user of the PO3M can help prevent
electromagnetic interference by maintaining a minimum distance between
portable and mobile RF communications equipment (transmitters) and the PO3M as
recommended
below, according to the maximum output power of the communications equipment.
Rated maximum output power of transmitter
W| Separation distance according to the frequency of the transmitter
m
---|---
150 kHz to 80 MHz
d=1.23| 80 MHz to 800 MHz
d=1.2g,| 800 MHz to 2,5 GHz
d=2.W
0,01| 0.12| 0.12| 0.23
0,1| 0.38| 0.38| 0.73
1| 1.2| 1.2| 2.3
10| 3.8| 3.8| 7.3
100| 12| 12| 23
For transmitters rated at a maximum output power not listed above, the
recommended separation distance d in meters (m) can be determined using the
equation applicable to
the frequency of the transmitter, where P is the maximum output power rating
of the transmitter in watts (W) according to the transmitter manufacturer.
NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher
frequency range applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic
propagation is affected by absorption and reflection from structures, objects,
and people.
0197