PHILIPS 018OTW Digital IVUS Catheter Instruction Manual

June 5, 2024
Philips

PHILIPS 018OTW Digital IVUS Catheter

PHILIPS 018OTW Digital IVUS Catheter Instruction product
kk

Instructions for Use Reconnaissance PV Digital IVUS Catheter

PHILIPS 018OTW Digital IVUS Catheter Instruction 1

Caution

  1. U.S. Federal Law restricts this device to sale by or on the order of a physician.
  2. Prior to use, read this entire package insert.
Intended use

The Reconnaissance PV .018 OTW Digital IVUS catheter is designed for use in the evaluation of vascular morphology in blood vessels of the peripheral vasculature by providing a cross sectional image of such vessels. This device is not currently indicted for use in cerebral vessels. The Reconnaissance PV .018 OTW Digital IVUS catheter is designed for use as an adjunct to conventional angiographic procedures to provide an image of the vessel lumen and wall structures.

DESCRIPTION

The Reconnaissance PV .018 OTW Digital IVUS catheter is an over the wire intravascular imaging catheter with a digital ultrasound transducer at the distal end. The transducer utilizes a 64-element cylindrical array that radiates acoustic energy into the surrounding tissue and detects the subsequent echoes. The information from the echoes is used to generate real time images of the peripheral vessels. The Reconnaissance PV .018 OTW Digital IVUS catheter is introduced percutaneously or via surgical cutdown into the vascular system, and is designed to track over 0.014”-0.018” (0.35-0.46 mm) guide wires. A lubricious hydrophilic coating is applied externally to a distal portion of the catheter.
THE CATHETER IS FOR EXCLUSIVE USE WITH S5, CORE SERIES OF SYSTEMS AND INTRASIGHT SYSTEMS. THIS CATHETER WILL NOT OPERATE IF CONNECTED TO ANY OTHER IMAGING SYSTEM.

Contraindications

Reconnaissance PV .018 OTW Digital IVUS catheter is generally contraindicated in situations presenting a reasonable probability of tissue or organ damage. This device is not currently indicated for use in cerebral vessels.

Adverse effects

As with all catheterization procedures, complications may be encountered with the use of the Reconnaissance .018 PV Digital IVUS catheter. Possible adverse effects include, but are not limited to, the following: occlusion; vessel spasm; vessel dissection; perforation, rupture or injury; restenosis; hemorrhage or hematoma; drug reactions; allergic reaction to contrast medium; hypo/hypertension; infection; arteriovenous fistula; embolism; entry puncture site bleeding; vascular wall injury; vessel thrombosis; pseudoaneurysm at site of catheter insertion; renal failure; aneurysm; vessel trauma requiring surgical repair or intervention; death.

Reporting of a serious incident

If a serious incident has occurred in relation to the device, it should be reported to the manufacturer and the competent authority of the Member State in which the user and/or patient is established. A serious incident means any incident that directly or indirectly led, might have led or, in case of recurrence, could lead to any of the following: the death of a patient, user or other person, the temporary or permanent serious deterioration of a patient’s, user’s, fetus or other person’s state of health, or a serious public health threat.

WARNINGS
Use of the Reconnaissance PV .018 OTW Digital IVUS catheter should be restricted to specialists who are familiar with, and have been trained to perform, the procedures for which this device is intended.  The product is supplied sterile; if the pouch is opened or damaged compromising the sterile barrier, please discard the product. This product cannot be re-sterilized or re-used. The Reconnaissance PV .018 OTW Digital IVUS catheter is designed for single use only. Philips, makes no warranty, representation or condition of any kind, whether expressed or implied (including any warranty of merchantability, suitability or fitness for a particular purpose) respecting the re-use of the catheter. In addition, Philips assumes no responsibility or liability for incidental or consequential damages which may result from such re-use. Re-use including resterilization of unused product may result in, but is not limited, to the following. Potential critical harm to patient due to Device Separation, Material Deformation or Infection/Sepsi  Failure to Image or other device malfunctions. The catheter transducer is a delicate electronic assembly. Deliberate misuse by bending, twisting or any other severe physical manipulation will void the warranty. Do not use the Reconnaissance PV .018 OTW Digital IVUS catheter for purposes other than those indicated. Do not exceed maximum pressure rating of 300 psi.

Precautions

The Reconnaissance PV .018 OTW Digital IVUS catheter is a delicate scientific instrument and should be treated as such. Always observe the following precautions

  • Protect the catheter tip from impact and excessive force.
  • Do not cut, crease, knot, or otherwise damage the catheter.
  • Protect the electrical connections from exposure to fluid.
  • Do not handle the transducer.
  • The outside diameter along the entire length of the guide wire should not exceed the maximum specified
  • During use, ensure that the placement of the catheter does not preclude blood flow through the vessel
  • Clean guide wire and flush catheter thoroughly with sterile heparinized normal saline before and after each insertion
  • Keep the exterior of the catheter wiped down with sterile heparinized normal saline during prolonged use
  • When inserting the guide wire both catheter and wire must be straight with no bends or kinks, or damage to inner lumen may occur.
  • Do not advance the guide wire against significant resistance. If binding occurs between the catheter and the guide wire while inside the patient, CAREFULLY REMOVE
  • BOTH the wire and catheter and do not use. If binding occurs outside of the patient, remove the catheter and do not use.
  • When advancing or re-advancing the catheter over a guide wire and through a stented vessel, in the event that the stent is not fully apposed against the vessel wall, the guide wire and/or catheter may become entangled in the stent between the junction of the catheter and guide wire or within one or more stent struts.
  • This may result in entrapment of catheter/guide wire, catheter tip separation, and/or stent dislocation. Never use force to advance the catheter.
  • Use caution when re advancing a catheter over a guide wire and into a stented vessel, in the event that the catheter may come in contact with one or more stent struts.
  • Subsequent advancement of the IVUS catheter could cause entanglement between the catheter and the stent(s) resulting in entrapment of catheter/guide wire, catheter tip separation, and/or stent dislocation.
  • Use caution when removing the catheter over the guide wire from a stented vessel to minimize patient risk.
  • The catheter should never be forcibly inserted into lumens narrower than the catheter body or forced through a tight stenosis.
  • If resistance is encountered during pullback, remove the entire system guide wire, IVUS catheter, sheath/guide catheter at the same time.
  • The use of an anticoagulant is recommended as per the local standard angiographic protocols and at the discretion of the physician to prevent thrombus formation. Do not expose the device to a magnetic resonance environment.
  • Device may not perform as intended and may present additional risks to patient.

Instructions for use

The Reconnaissance PV .018 OTW Digital IVUS catheter may be introduced into the vascular system percutaneously or surgically and advanced to the desired location. The frequency and duration of administration is subject to the discretion of the physician and depends upon the procedure and information required. Check system operation prior to the use. Open the Reconnaissance PV .018 OTW Digital IVUS catheter packaging using sterile technique and place the hoop in the sterile field. Prepare the catheter by flushing the guide wire lumen through the port at the catheter’s Y-connector, and then wipe down the entire working length with sterile heparinized normal saline. Remove the clear/white cap from the PIM connector. Connect the PIM connector of the Reconnaissance PV .018 OTW Digital IVUS catheter to the Patient Interface Module as described in the imaging system Operator’s Manual. Verify that the device is imaging. Place the Reconnaissance PV .018 OTW Digital IVUS catheter onto the intravascular guide wire. A guide wire of 0.014” (0.35 mm) or 0.018” (0.45 mm) can be used. Activate the hydrophilic coating using sterile heparinized normal saline. Advance the Reconnaissance PV .018 OTW Digital IVUS catheter over the guide wire to the site of the vasculature to be imaged. The guide wire should always be advanced ahead of the IVUS catheter. Check the Monitor for an image. Once the image has been obtained, the catheter can be advanced over the guide wire to image additional segments of vasculature. If an image is not obtained or is not satisfactory. Once imaging has been completed, remove the Reconnaissance PV .018 OTW Digital IVUS catheter and flush thoroughly with sterile heparinized normal saline. For subsequent imaging, clean guide wire and flush catheter thoroughly with sterile heparinized normal saline before re-insertion.  When the procedure is completed, remove and discard the catheter in accordance with local regulations.

Storage and handling

Products should be stored indoors at room temperature in their original packaging.

Product specifications
  • Model: Reconnaissance PV .018 OTW Digital IVUS catheter
  • Catalog Number : 018OTW
  • Maximum shaft outer diameter: 0.068 (1.723 mm)
  • Maximum scanner diameter: 0.046 (1.17 mm)
  • Maximum guide wire: 0.018 (0.46 mm)
  • Minimum guide sheath :5F (0.074, 1.88 mm)
  • Usable length :150 cm

Acoustic Output

Parameter

| B-Mode| Chromaflo
---|---|---
I SPTA.3 (mW/cm2)| 2.93×10-3| 7.98×10-2
I SPPA.3 (W/cm2)| 7.5×10-3| 175.0×10-3
Pr.3 (MPa)| 20.0×10-3| 81.5×10-3
PD (μs)| 161.0×10-3| 125.0×10-3
PRF (Hz)| 53760| 75368
Center Freq (MHz)| 18.6| 17.9
MI*| 4.5×10-3| 1.92×10-2
TI**| 2.06×10-5| 1.56×10-4

  • Maximum overall uncertainty + 33.9%/- 30.5%
  • As estimated in tissue
  • TI: Thermal Index defined as TI = W01x1fc  210
  • W01x1: Bounded-square Output (mW)
  • fc: Center Frequency (MHz)
  • MI: Mechanical Index defined as MI= Pr.3/(fc 1/2 )
  • ISPPA.3: Derated Intensity, Spatial Peak Pulse Average (W/cm2)
  • ISPTA.3: Derated Intensity, Spatial Peak Temporal Average (mW/cm2)
  • Derated Peak Negative Pressure at a location of the maximum derated pulse intensity integral (MPa)
  • W0: Total Power (mW)
  • PD: Pulse Duration (μs)
  • PRF: Pulse Repetition Frequency (Hz)

Limited warranty
Subject to the conditions and limitations on liability stated herein, Philips warrants that the Reconnaissance .018 PV Digital IVUS catheter (the “catheter”), as so delivered, shall
materially conform to Philips’ then current specification for the catheter upon receipt for a period of one year from the date of delivery. Any liability of Philips with respect to the catheter or the performance thereof under any warranty, negligence, strict liability or other theory will be limited exclusively to catheter replacement or, if replacement
is inadequate as a remedy or, in Philips’ opinion, impractical, to refund of the fee paid for the catheter. Except for the foregoing, the catheter is provided “as is” without
warranty of any kind, expressed or implied, including without limitation, any warranty of fitness, merchantability, and fitness for a particular purpose of noninfringement.
Further, Philips does not warrant, guarantee, or make any representations regarding the use, or the results of the use, of the catheter or written materials in terms of correctness, accuracy, reliabilty, or otherwise. Licensee understands that Philips is not responsible for and will have no liability for any items or any services provided by any
persons other than Philips. Philips shall have no liability for delays or failures beyond its reasonable control.

Additionally, this warranty does not apply if
The catheter is used in a manner other than described by Philips in the Instructions.

  • For Use supplied with the catheter.
  • The catheter is used in a manner that is not in conformance with purchase specifications or specifications contained in the Instructions For Use.
  • The catheter is re-used or re-sterilized.
  • The catheter is repaired, altered, or modified by other than Philips authorized personnel or without Philips authorization.
  • If claims under this warranty become necessary, contact Philips for instructions and issuance of a Return Material Authorization number if the catheter is to be returned.
  • Equipment will not be accepted for warranty purposes unless the return has been authorized by Philips.
  • Patent: www.philips.com/patents
  • An electronic copy of these instructions can be found at: www.philips.com/IFU.
  • The summary of safety and clinical performance will be available in the European database on medical devices (EUDAMED) at the following URL
    https://ec.europa.eu/tools/eudamed

REACH declaration
REACH requires Philips Healthcare (PH) to provide chemical content information forsubstances of Very High Concern (SVHC) if they are present above 0.1% of the product
weight. The SVHC list is updated on a regular basis. Therefore, refer to the following. Philips REACH website for the most up-to-date list of products containing SVHC above
the threshold: www.philips.com/REACH
This product is licensed to the customer for single use only. Additional questions regarding this product should be directed to Philips in the U.S.A.

Legal Manufacturer

USA & Canada

International
vecomplaints@philips.com or IGTD.CustomerInquiry@philips.com

European Authorized Representative : EC REP Volcano Europe SA/NV

  • Excelsiorlaan 41
  • B-1930 Zaventem, Belgium
  • Telephone: +32.2.679.1076
  • Fax: +32.2.679.1079

Australian Sponso r

  • Philips Electronics Australia Ltd
  • 65 Epping Road
  • North Ryde, NSW Australia 2113

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Related Manuals