sauermann Class 320 Kistock Data Loggers User Guide
- June 5, 2024
- sauermann
Table of Contents
- sauermann Class 320 Kistock Data Loggers
- [Quick start guide
- Display
- [Dimensions (mm)
- [Connections
- Recorder function
- Mounting
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
sauermann Class 320 Kistock Data Loggers
Quick start guide [
](https://manuals.plus/wp- content/uploads/2022/07/sauermann-Class-320-Kistock-Data-Loggers-FIG6.jpg)
Operating temperature, protection of the instruments and information about storage and the battery
-
*Operating temperature:** KT320: from -40 to +70 °C; KTT320: from -20 to +70 °C; KCC320,
KPA320, KP320 & KP321: from 0 to 50 °C -
KT320: IP65; KTT320: IP54; KCC320, KPA320: IP40, KP320, KP321: IP65*
-
Storage temperature: from -40 to +85 °C
Battery power supply: 2 x lithium AA 3.6 V batteries
Display
- °C: Temperature in °Celsius.
- °F: Temperature in °Fahrenheit
- %RH: Relative humidity
Dimensions (mm) [
](https://manuals.plus/wp- content/uploads/2022/07/sauermann-Class-320-Kistock-Data-Loggers-FIG2.jpg)
Connections [
](https://manuals.plus/wp- content/uploads/2022/07/sauermann-Class-320-Kistock-Data-Loggers-FIG5.jpg)
Recorder function
Five recording modes KISTOCK can record in 5 different ways:
-
”Immediate” mode records values according to a predefined interval.
-
”Minimum”, ”Maximum” and ”Average” record automatically the calculation of minimum, maximum or average of measured values during an interval of recording.
-
”Monitoring” mode allows to get an accurate history report during error events to help troubleshooting, without stopping the measurement logging. To proceed this way, you just have to define:
- a record interval to be used whilst the readings are beyond the setpoints
- a record interval for the values measured during each reading beyond the setpoints
-
Furthermore, you can also let your KISTOCK record non-stop (”loop” recording option).
-
Four types of dataset start
Once your recording mode has been set, you can launch your dataset: -
With a delayed start (with predefined date and time)
-
With the software
-
With push-button
-
With ”Online” option. In this case, your datasets are directly sent, saved and displayed on your PC in real time.
-
Six types of dataset stop
You can stop your dataset: -
According to a date and time (if it was started the same way)
-
According to a period
-
According to a predefined number of recording points
-
Once the storage capacity is full
-
With ”Stop” option of the software
-
By holding ”OK” key during 5 seconds, if this function has been previously activated by the software
Mounting
Safety lock wall mount with padlock
- Present the KISTOCK data logger on the support starting with the inferior part
- Clip the KISTOCK on the support by falling back the superior part
- Insert the padlock to ensure the safety lock function
- The padlock can be replaced by a fail-safe sealed.
- The data logger can be placed on the screw-mount without the safety lock function.
- To remove the data logger from the support, proceed on reverse order.
Dimensions of the wall mount (mm) Accessories
Please refer to the data sheet to get more information about available
accessories.
Maintenance: please avoid any aggressive solvent. Please protect the
device from any cleaning product containing formalin, that may be used for
cleaning rooms or ducts.
Precautions for use: please always use the device in accordance with its
intended use and withinparameters described in the technical features in order
not to compromise the protection ensured by the device.