RENISHAW H-1000-3323 RVP Vision Probe for REVO-2 Instructions
- June 5, 2024
- RENISHAW
Table of Contents
RENISHAW H-1000-3323 RVP Vision Probe for REVO-2
RVP vision probe for REVO-2
The REVO vision probe (RVP) provides non-contact vision measurement on an
infinitely positioning, 5-axis platform.
Now, tactile scanning measurement, surface finish analysis and non-contact
inspection can all be performed on one CMM.
The RVP system is a breakthrough in technology in the world of non-contact CMM
inspection. The combination of non-contact vision measurement and a 5-axis,
infinitely positioning platform makes the RVP system truly unique.
RVP further expands the range of applications for the REVO-2 system with a
non-contact edge detecting sensor automatically interchangeable with all other
REVO probe options. As a result, the system provides the optimum tool to
measure multiple features, and a wider variety of parts can be inspected.
System components
The RVP system comprises a vision probe body, a number of vision modules, an
angle change mirror, rack ports and a calibration artefact. The image capture
and processing components of the system are held inside the vision probe body
and include an industry standard, robust, CMOS sensor for reliable image
capture.
The vision modules allow a range of features with different sizes and shapes
to be inspected. The ACM accessory provides enhanced access allowing vision
inspection of bore surfaces and other features previously inaccessible to RVP.
All vision modules contain integral LED lighting to achieve a sharp contrast
between holes and part material. Background feature enhancement is also
available using backlighting combined with bespoke part fixturing.
Benefits
Added part inspection capability
Small features and delicate or flexible parts that cannot be measured using
tactile probes can now be inspected on a 5-axis non-contact measurement
platform.
Added value multi-sensor system
A machine equipped with REVO-2 can be used for a large range of inspection
applications including scanning measurement, surface finish analysis and now,
non-contact inspection.
High speed data collection
5-axis motion between part features and real time image processing
dramatically increase data collection rates.
Innovations
5-axis movement, infinite positioning
RVP benefits from REVO’s infinite positioning and 5-axis movement, which
ensure accurate and unrestricted placement in front of features at any angle.
Automatically interchangeable sensors
The RVP non-contact probe is automatically interchangeable with all other
probe options available for REVO-2, and data from multiple sensors is
automatically referenced to a common datum.
3D feature reconstruction
5-axis motion and infinite positioning provide access to the data required for
3D feature reconstruction.
System specification
Probe specification | RVP |
---|---|
Weight | 551g |
Dimensions | 158 mm x 131 mm x 134 mm |
Rack port compatibility | VPCP (heated) |
Sensor type | 1/1.8” CMOS sensor |
Head compatibility | REVO-2 |
Operating temperature | +10 °C to +40 °C |
Software compatibility | UCC suite 5.0 onwards |
MODUS 1.7 onwards
Integration| I++ DME protocol
Module type
| Feature size error (circle diameter) 1
---|---
Ø25 mm| Ø10 mm| Ø2 mm
VM10| 3 µm| 2 µm| 2 µm
VM11-2| n/a2| 1 µm| 1 µm
VM12| n/a2| n/a2| TBC
- Feature size diameter error measured using calibrated glass artefact (uncertainty ±0.5 μm) with backlighting.
- Circle diameter not applicable to lens specification.
Tests completed using REVO-2 with RVP and VM10 and VM11-2 modules.
Test machine specification: MPE(E150) 3 µm + L / 333 mm MPE(PFTU) 1.6
µm
Module specification | VM10 | VM11- 2 | VM12 |
---|---|---|---|
Weight | 153 g | 134 g | 138 g |
Dimensions | 86 mm × 39 mm | 74 mm × 67 mm | 144 mm × 67 mm |
Rack port compatibility | VMCP (heated) | VMCP (heated) | VMCP (heated) |
Measurement range | 1 mm diameter or greater | 0.4 mm diameter or greater |
0.05 mm diameter or greater
Field of view| 50 mm × 40 mm| 12.5 mm × 10 mm| 3.1 mm × 2.5 mm
Stand-off| 80 mm| 120 mm| 50 mm
Depth of field| 5 mm| 5 mm| 0.2 mm
Resolution| 40 μm| 20 μm| 2.5 µm
LED illumination| 24 LEDs| 10 LEDs| 10 LEDs
Backlight compatibility| Yes| Yes| Yes
ACM compatible| No| Yes| No
+44 (0) 1453 524524
uk@renishaw.com
© 20172022 Renishaw plc. All rights reserved. RENISHAW® and the probe symbol
are registered trade marks of Renishaw plc. Renishaw product names,
designations and the mark `apply innovation’ are trade marks of Renishaw plc
or its subsidiaries. Other brand, product or company names are trade marks of
their respective owners. Renishaw plc. Registered in England and Wales.
Company no: 1106260. Registered office: New Mills, Wotton-under-Edge, Glos,
GL12 8JR, UK. WHILE CONSIDERABLE EFFORT WAS MADE TO VERIFY THE ACCURACY OF
THIS DOCUMENT AT PUBLICATION, ALL WARRANTIES, CONDITIONS, REPRESENTATIONS AND
LIABILITY, HOWSOEVER ARISING, ARE EXCLUDED TO THE EXTENT PERMITTED BY LAW.
Part no.: H-1000-3323-05-A Issued: 05.2022