CA-45703-P1A Pressure Injectable Arrowg Lumen CVC User Manual

June 5, 2024
Arrow

45703-P1A Pressure Injectable Arrowg Lumen CVC
**User Manual

Pressure Injectable Arrowg+ard  Blue Plus® Three-Lumen  CVC**

Pressure Injectable Arrowg Lumen CVC

1: Arrowg+ard Blue Plus., Three-Lumen Catheter: 7 Fr. (2.58818 OD) x 16 cm, Pressure Injectable
1: Spring-Wire Guide, Marked: .032″ (0.81 mm) dia. x 17-3/4″ (45 cm) (Straight Soft Tip on One End – “J” Tip on Other) with Arrow Advancer
1: Catheter: 18 Ga. x 2-1/2″ (6.35 cm) Radiopaque over 20 Ga. RW Introducer Needle
1: Pressure Transduction Probe
1: Introducer Needle: Echogenic 18 Ga. x 2-1/2″ (6.35 110(01/N and 5 mL Arrow. Raulerson Spring-Wire Introduction Syringe
1: Injection Needle: Eclipse”’. 22 Ga. x 1-1/2″ (3.81 cm) RW and 5 mL Luer- Slip Syringe
1: Injection Needle: Eclipse”’. 25 Ga. x 1″ (2.54 cm) RW and 3 mL Luer-Lock Syringe
1: Syringe: 10 mL Luer-Lock
1: Tissue Dilator: 8.5 Fr. (2.8 mm) x 10.2 cm
1: SecondSiteml Adjustable Hub: Fastener
1: SecondSiteml Adjustable Hub: Catheter Clamp
1: SharpsAway. Disposal Cup
1: SharpsAway. II Locking Disposal Cup
1: Drape: 41″ x 55″ with 4″ fenestration
1: Needle Holder
1: Safety Scalpel: #11
1: Checklist/CLIP Sheet
1: Sterile Procedure Sign
2: Gauze Pad: 2″ x , (5 cm x 5 cm)
5: Gauze Pad: 4″ x 4″ (10 cm x 10 cm)
1: Dressing: Tegaderm.3 10 cm x 12 cm
1: Suture: 3-0 Silk with Curved Needle
1: HemoHopper. Fluid Receptacle
1Licensed under US Patent No. 7,329,412.
2A trademark of Becton, Dickinson, and Company.
3A registered trademark of 3M Company.
Not made with natural rubber latex.
Store below 25°C (77°F). Avoid excessive heat above 40°C (104°F).
Contraindications: The Arrowg+ard Blue Plus antimicrobial catheter is contraindicated for patients with known hypersensitivity to chlorhexidine, silver sulfadiazine, and/or sulfa drugs.


Lumen
| *Priming Volume (mL)| Gravity Flow Rate† (mL/hr)| MAX Pressure Injection Flow Rate (mL/sec)
---|---|---|---
Distal   ( 16  G a. )
Medial  ( 18  G a. )
Proximal  ( 18  G a. )| 0.39
0.39
0.40| 2283
913
996| 10
5
5

Priming volumes are approximate and are done without accessories.
†  Flow rate values are approximate and are determined using deionized water at  100  cm head height.
**  Pressure injection flow rates are determined using media of  11.8 centipoises (cP)  viscosity,  with 152  cm administration set tubing.

LBL053589 R00 (2022-01)

Documents / Resources

| ARROW CA-45703-P1A Pressure Injectable Arrowg Lumen CVC [pdf] User Manual
CA-45703-P1A Pressure Injectable Arrowg Lumen CVC, CA-45703-P1A, Pressure Injectable Arrowg Lumen CVC, Arrowg Lumen CVC, Lumen CVC
---|---

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

Arrow User Manuals

Related Manuals