Palmer Wahl WD1120 Advanced Digital Pressure Gauges User Manual

June 5, 2024
Palmer Wahl

Palmer Wahl WD1120 Advanced Digital Pressure Gauges

Product Information

Specifications

  • Accuracy: 0.1% plus 0.005% Span
  • Pressure Units: 11 units – psi, bar, mbar, Pa, kPa, MPa, kg/cm2, inH2O, mmH2O, inHg, mmHg
  • Meter Display Update Rate: 1-second standard, selectable: 1s, 2s, 5s
  • Storage Temperature: Not specified
  • Relative Humidity: 5% to 100% RH non-condensing
  • Analog Output: 4 to 20mA or 0 to 5V
  • Digital Output: RS485 via Wahl Protocol can be integrated with existing customer software
  • Switch Voltage: 12 – 24 VDC

Product Usage Instructions

  • When you receive the product, carefully unpack and inspect it for any visible damage or missing parts.
  • Ensure all components are included before proceeding with installation.
  • The instrument is designed for pressure measurement only. Make sure to adhere to the documented technical specifications for all components to achieve accurate results.
  • Follow the operating instructions diligently for optimal performance.
  • When disposing of the instrument, ensure compliance with professional and environmental regulations in your area for electronic waste disposal.
  • Do not discard the instrument as regular trash.

FAQ

  • Q: Can I use non-original replacement parts and accessories?
  • A: It is recommended to use only original parts and accessories approved by the manufacturer. The use of other parts may compromise the operation, safety, and functionality of the instrument.
  • Q: What is the acceptable pressure range for this product?
  • A: The pressure range can vary from -15 to 15,000 PSI depending on the type of pressure being measured.
  • Q: How do I set up the digital display and alarms?
  • A: The primary display shows real-time pressure readings while the secondary display offers additional features like MAX/MIN/AVG readings and alarm settings. Use the menu settings to configure advanced visual alarms based on your preferences.

Warranty

Digi-Stem® DSPG Advanced Digital Pressure Gauge Warranty
Manufacturer warrants all Wahl Digi-Stem® DSPG Advanced Digital Pressure Gauges manufactured by us to be free from defects in material or workmanship under normal use and service. The Manufacturer agrees to repair or replace any product listed above which upon examination is revealed to have been defective due to faulty workmanship or material if returned to our factory, transportation charges prepaid, within the product-specific warranty period of two (2) years from the date of purchase. This warranty is in place of all other warranties, expressed or implied and of all obligations or liabilities on its part for damages including but not limited to normal wear and tear or consequential damages following the use or misuse of this or any instrument sold by the Manufacturer. In addition, if the product is tampered with in any way or calibrated in any way other than by the instructions supplied by Palmer Wahl, it will immediately void the warranty. No agent is authorized to assume, for the manufacturer, any liability except as set forth above. Freight cost to return item(s) for evaluation, duties and other fees are not covered by the manufacturer. The product is automatically registered for warranty by serial number at the time of purchase. The serial number must remain attached to the product upon return.

Safety Instructions

This document contains important information and should be kept with the instrument at all times during its operational life. Carefully read all instructions before using the instrument. Other users of this instrument should be given these instructions with the instrument.

  • Do not alter the instrument.
  • Use the instrument only as specified, or the protection provided by the instrument can be compromised.
  • Do not use the instrument if it operates incorrectly.
  • Do not use the instrument if it is altered or damaged.
  • Disable the instrument if it is damaged.
  • Avoid any action that can generate an electrostatic charge. Electrostatic discharge is an explosion hazard.
  • Replace the batteries when the battery status indicator appears low to prevent incorrect measurements.
  • Ensure the battery polarity is correct to prevent battery leakage.
  • If the unit is equipped with 24V DC power, the pipe to which the gauge is attached must be properly grounded.
  • If the display reads “OL”, the range limit is exceeded and the pressure source must immediately be reduced.

To avoid possible damage to the instrument or equipment under test

  • Always apply thread seal tape to the threads of the gauge. When installing in Stainless Steel, use nickel-filled PTFE Thread Sealant Tape to prevent galling.
  • Do not exceed the maximum allowed torque or 10 Newton Meters (8-foot pounds.)
  • Never tighten the gauge threads by holding the body of the gauge. This could damage the gauge threads and make the gauge inoperable. Use a wrench to tighten the gauge to the process.
  • When the access port(s) are not in use, keep the included dust cover protectors attached to prevent damage to the connectors.

After installation, long press the RIGHT ARROW button to reset the ZERO of the gauge. Future updates to this information can be found at http://www.palmerwahl.com and must be added to the original document. The instrument can only be operated by trained personnel under these instructions and local safety regulations.

Acceptable Operation
This instrument is intended only for the measurement of pressure. The instrument is appropriate for continuous use and operates reliably in demanding conditions as long as the documented technical specifications for all instrument components are adhered to. Compliance with the operating instructions is necessary to ensure the expected results.
Unacceptable Operation
Replacement Parts and Accessories: Use only original parts and accessories approved by the manufacturer. The use of other parts can compromise the operation, safety, and functionality of the instrument.
Instrument Disposal
Disposal of old instruments should be handled according to professional and environmental regulations of your area for electronic waste.
Includes
Wahl Digi-Stem® DSPG Advanced Digital Pressure Gauges come with:

  • 1 – DSPG Precision Pressure Gauge Assembly
  • 1 – Quick Start Guide
  • 1 – Lithium Thionyl Chloride Battery

Wahl Digi-Stem® DSPG Advanced Digital Pressure Gauge Specifications

DSPG

Digital Pressure Gauge with Display and Alarm
Optional Transmitter and Switch



METER SPECIFICATIONS

| Range / Scale| -15 to 15,000 PSI, depending on Pressure Type


Accuracy

| 0.1% of Full Scale @ 23°C
Optional: 0.1% at 0° to 50°C with 9-Point NIST Certificate of Conformance*
Pressure Units| 11 units: psi, bar, mbar, Pa, kPa, MPa, kg/cm2, inH2O, mmH2O, inHg, mmHg




DISPLAY SPECIFICATIONS

| ****

Meter Display

| Primary Display: 0.9″ 5-digit LCD Display of Real-Time Pressure, Readable up to 30 ft. away
Secondary Display: 0.4″ 5-digit for Secondary Pressure: MAX / MIN / AVG / Delta, Hi and Low Alarm Settings and Pressure Sensor Temperature
Display Update Rate| 1-second standard, selectable: 1s, 2s, 5s
Menu Bar| Icon Menu Settings
Advanced Visual Alarms| High (Red), Low (Blue), Normal (Green), Fully Adjustable



ENVIRONMENT

| Ambient Operating Environment| -18° to 65°C (0° to 150°F)
Storage Temperature| -40° to 82°C (-40° to 180°F)
Relative Humidity| 5% to 100% RH non-condensing


PRESSURE TRANSMITTER

(Optional with Switch)

| Accuracy| 0.1% plus 0.005% Span
Analog Output| 4 to 20mA or 0 to 5V
Digital Output| RS485 via Wahl Protocol can be integrated with existing customer software, for Transmitter & Switch option only



PRESSURE SWITCH

(Optional with Transmitter)

| Switch| Switch at High and Low Alarm

Two programmable pressure points for the ON/OFF switch

Accuracy| 0.1% plus 0.1% Span
Relays| Consistent Open Relay, Overload Protection Relay, Polarized Relay, Safety Relays, Solid- State Relay, Time Delay Relay
Voltage| 12 – 24 VDC




POWER

| Battery Internal| 1 – C size, Lithium Thionyl Chloride Battery, 3.6 V, 8.5 Ah Capacity


Ultra-Low Battery Power Consumption

| User-selectable low Refresh Rate Operation for Long Battery Life
4 years with 1 second update rate, (default) (with C-size Lithium battery)
External Power| Loop Powered, 12 – 24 VDC (Customer supplied) (Includes “C” Battery for display backup) Required for Switch and Transmitter Options




ENCLOSURE

| Case / H Frame and Window Material| Stainless Steel / Polycarbonate
EMI Interference| Designed for Maximum EMI and RFI Protection
Protection Ratings| NEMA 6P, IP68
Enclosure Dimensions / Weight| 5.3″ W x 4.3″ H x 2.7″ D (135 x 109 x 69 mm) / Weight: 2.0 lbs. (1kg)
Includes: “C” size Lithium Thionyl Chloride Battery, Quick Start Guide and User Manual as download.

  • 9-Point NIST Certificate of Conformance purchased at time of order. Includes 0%, 50% and 100% at 0°, 23° and 50°C.

    • Wahl Digi-Stem® DSPG Advanced Digital Pressure Gauge Available Ranges

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-2 Palmer-Wahl-WD1120-Advanced-
Digital-Pressure-Gauges-FIG-3

RoHS
The Wahl Digi-Stem® DSPG Advanced Digital Pressure Gauge conforms to RoHS requirements according to EU Directive (EU) 2015/863- amending Annex II to Directive 2011/65/EU for restriction of Hazardous substances.

Unpacking & Inspection
If you have any questions about the shipment, please call our Customer Service Department at 1-800-421-2853 or 828-658-3131. We can also be reached online at www.palmerwahl.com, or by email to info@palmerwahl.com.

Display and Keypad Info

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-4

Status

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-5

Menu

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-6

Display Readings

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-7 Palmer-Wahl-WD1120-Advanced-
Digital-Pressure-Gauges-FIG-8 Palmer-Wahl-
WD1120-Advanced-Digital-Pressure-Gauges-FIG-9 Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-10 Palmer-Wahl-WD1120-Advanced-
Digital-Pressure-Gauges-FIG-11 Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-12 Palmer-Wahl-WD1120-Advanced-
Digital-Pressure-Gauges-FIG-13 Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-14

Keypad

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-15

Analog Wiring Diagram

24V DC

24V GND
mA (+)
5V
Analog COM
Unused
Relay NO/NC
Relay COM

Each wire is labelled individually with its purpose.

How to Replace the Battery

  • Replace the battery only with a C size, Lithium Thionyl Chloride battery, 3.6 V, 8.5 Ah Capacity (Palmer Wahl PN: DSA3062).

DO NOT USE ALKALINE BATTERIES

  • When the battery status indicator shows low battery, replace the battery by first removing the 4 screws and lock washers on the corners of the case with a Philips head screwdriver.

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-16

  • Lift the case cover up carefully. When the PC Board is exposed, hold down the board gently while pulling the battery out of the holder.

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-17

  • Place the new battery in the holder and snap it into place.

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-18

  • When the new battery is connected the board will cycle through red, blue and green lights.
  • Gently replace the case cover and insert the screws with the lock washers back into place. Do not use a power screwdriver due to the potential to overtighten the screws.

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-19

Dimensional Drawings

Bottom Mount

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-20

All-Angle Mount

Palmer-Wahl-WD1120-Advanced-Digital-Pressure-Gauges-
FIG-21

Return for Calibration Service
To receive a NIST Traceable Certificate of Conformance or repair service on your product, go to palmerwahl.com and click on: Request an RMA (Return Material Authorization Number) and follow the instructions. You will receive your RMA number via email once your item is received at our facility. If you prefer, call Customer Service at 1-800-421-2853 for assistance with the RMA process. Material being returned to us should be packed well, preferably in the original shipping container.
Our quality management system is certified to conform to ISO 9001:2015. We maintain a calibration system in conformance with ANSI/NCSL Z-540 and MIL-STD- 45662A.
All calibrations are performed against standard instruments, traceable to NIST. Records are stored for a minimum of three years. Standards used have a test ratio of four times greater than the unit being calibrated. We can perform certified, traceable calibrations in support of companies that require meeting FAA and FDA quality standards.

CONTACT

Palmer Wahl Instruments, Inc.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>