2500/1370 Series Type K Thermocouple Meters
- 2500 Meter Range -100° to 2500°F
- 1370 Meter Range -70° to 1370°C
- Meter options for °F or °C range, display type and power
- Meter accuracy is ±0.2% of reading ±1 digit
- Maxi-Temp®automatically holds peak reading
- Accepts all type K probes with ANSI mini-connectors
- Microprocessor circuitry for thermocouple linearization
Heat-Prober®Thermocouple Thermometers and Probes are rugged and reliable
instruments made for process and maintenance temperature measurements.
The pocket-sized microprocessor-based Heat-Prober Type K thermometer updates
readings three times per second with an accuracy of ±0.2% of reading ±1 digit
in 1°F or °C resolution.
The precision cold junction is compensated with a unique, permanently
calibrated circuit.
The Wahl Maxi-Temp peak hold switch, available on all models, permits the
Heat-Prober to hold and display the highest temperature reached. This function
is ideal where the operator must concentrate on the probe when making the
measurement and where an instantaneous transient peak is to be recorded.
Maxi-Temp meters also can be switched to normal operation.
Advanced electronics are housed in ABS plastic, thoroughly tested for
environmental integrity and subjected to 4-foot drop tests, battery life and
running life tests. All units are calibrated with simulated EMF inputs and
powered during a hightemperature burn-in cycle.
Model 2500M (-100° to 2500°F) features a bright .33″ high red LED display
for normal to low lighting conditions. It operates on a rechargeable 6.25V
NiCad battery pack. Includes battery recharger/AC line adapter and Wrist
Strap.
Model 1370M as above with -70° to 1370°C range.
Model TCMR-K °F/°C switchable.
Model 2500MX (-100° to 2500°F) features a .5″ high red LCD display for
normal to bright lighting conditions. It operates on a rechargeable 6.25V
NiCad battery pack. Includes battery recharger/AC line adapter and Wrist
Strap.
Model 1370MX as above with -70° to 1370°C range.
Model TCMR-KX °F/°C switchable.
Model 2500MVX (-100° to 2500°F) features a .5″ high red LCD display for
normal to bright lighting conditions and is powered by a replaceable single
standard 9V battery. Includes Wrist Strap.
Model 1370MVX as above with -70° to 1370°C range.
Model TCMR-KVX °F/°C switchable.
See page 12-13 for 2500/1370 Series meter, probe, and accessory money-saving
kits.
See pages 14-17 for interchangeable thermocouple probes.
See Meter specifications on page 12. NIST TRACEABLE THERMOCOUPLE CALIBRATORS
TA-70AFK Precision Calibrator in use with 2500MX Thermocouple Meter
NIST Certification is optional with Calibrator
This thermocouple meter accessory allows calibration over the meter’s full
range.
Useful for quick field diagnostics, it assures meter accuracy whenever a
measurement is in question.
Accuracy is ±0.3°F/°C or 0.2% of reading, whichever is greater.
User sends the calibrator only back to the factory for NIST Certification.
One Calibrator maintains NIST traceability for all in-house meters.
THERMOCOUPLE CALIBRATOR MODELS
Model | Type | Calibration Points |
---|---|---|
TA30AFK | K | Discontinued |
TA50AFK | K | 100°F – 155°F – 200°F – 250°F – 300°F |
TA70AFK | K | 130°F – 650°F – 1300°F – 1900°F |
TA7OACK | K | 70°C – 400°C – 800°C – 1200°C |
TA70AFT | T | Discontinued |
Thermocouple Meters TM410 TM500
This product has been Discontinued
TM410 Thermocouple Food Service Thermometer
TM410 with TCL329K Extension Handle allows use of interchangeable probes with
one handle and cord set. See page 15 for probe descriptions.
The Wahl TM410 Type K Thermocouple Digital Heat Prober® Thermometer is
economical and compact. It provides the highest accuracy at the lowest price
of any Thermocouple Instrument on the market today.
- Accuracy of ±1.0°F (±0.5°C) in an operating range of 32° to 200°F (0° to 93°C)
- Meets and exceeds FDA Requirements (Food Code requires ± 2°F)
- Optional probes plug into the top of TM410 for easy single-handed use
- Optional TCL329K Extension Handle for extended reach
- Built in pocket/belt clip on back of unit
- Reduced dirt-harboring crevices for easy cleaning
- °F and °C Switchable
TM410 shown in use with NIST Certified TA50AFK Calibrator, page 10.
Wahl TP Probes (shown left with TM410, described on page 15) are specially manufactured with food service in mind to give fast and accurate results. They can be inserted into the TM410 for single handed operation and are durable enough to withstand the rigors of a busy kitchen. Probes can be safely passed through the dishwasher for ease of cleaning. Accuracy is maintained regardless of which probe is attached, assuring greater certainty of safe food temperature, and tighter control over the quality of the finished product.
TM500 Thermocouple Meter
The Wahl TM500 Heat Prober Meter offers value with exceptional features. The TM500 has advanced microprocessor functions, dual display and dual probe capability for more advanced testing in critical process and laboratory applications.
- TM500 Type J Range of -328°F to 1922°F (-200°C to 1050°C).
- TM500 Type K Range of -328°F to 2498°F (-200°C to 1370°C).
- Large LCD and Dual probe input accepts one or two Type J or Type K Thermocouple probe(s) with standard ANSI mini-connectors.
- Selectable temperature scale, Fahrenheit or Celsius readings.
- Main LCD displays the temperature reading from either selected probe or the temperature difference between the two.
- Third LCD displays time or timing duration (hours, minutes, seconds) of MAX/MIN or AVERAGE temperature events.
- “Recording” mode allows data storage for MAX, MIN, and AVERAGE temperatures reached over a time period.
- “Set” mode allows entry of reference and alarm points.
- Supplied with water and shock resistant rubber boot, built-in easel stand, 9V battery, and two Type K test probes.
See Specifications for both Meters above on page 12.
See page 13 for Thermocouple Meter, probe, and accessory money-saving kits.
See pages 14-17 for interchangeable thermocouple probes. TM500 CALIBRATOR
Type K Range | -328° to 2498°F and -200° to 1370°C |
---|---|
Type J Range | -328° to 1922°F and -200° to 1050°C |
Resolution | 0.1°, °F or °C |
Meter Accuracy | ±0.05% reading ±0.6°F (0.3°C) |
@ -58°/2498°F (-50° to 1370°C)
±0.05% reading ±1.4°F (0.7°C)
@ -58°/328°F (-50 to 200°C)
Power| One standard 9V, NEDA 1604 or equiv.
Display| 5 Digit LCD
Meter Size| 7-5/8″ x 3-5/8″ x 2-1/8″ (190 x 92 x 54mm)
Thermocouple Meter Specifications & Accessories
Thermocouple Meter Specifications
See our TM602 Thermocouple Thermometer and TM630 Thermocouple and RTD
Thermometer featured on page 3
Thermocouple Meter Specifications
Model | 2500M | 2500MX1 | 2500MVX | 1370M | 1370MX | 1370MVX | TM410 | TM500 |
---|---|---|---|---|---|---|---|---|
Range | 100°/2500°F | -70°/1370°C | -40°/1999°F |
-40°/1092°C
328°/2498°F| –
-200°/1370°C
Digital Display| LED| LCD| I LCD| LED| LCD| LCD| LCD| Dual LCD
F/C Switch| Optional| Optional| Yes| Yes
Resolution| 1°| 1°| 1.0° or 0.1°| 0.1°
Meter Accuracy| ±0.2%
±1 digit| ±0.2%
±1 digit| ±1.0°F @ 32°/200°F
(±0.5°C @ 0°193°C)| ±0.05% reading
±0.6°F (0.3°C)
@ -58°/2498°F (-5071370°C)
0.3% reading, ±2°F
@ -40°/32°F (-40°/0 C)
±0.05% reading
±1.4°F (0.7°C)
(a3 -58°/-328°F (-50/-200°C)
0.3% reading, ±2°F
@ 200°/1999°F
±0.4°C 4, 93°/1092°C
Thermocouple Type| K| K| K| J, K
Peak Hold| Yes| Yes| No| Yes
Ambient Operating
Range| -20°/120°F
-29°/49°C| 0°/120°F
-18°/49°C| -20°/120°F
-29°/49°C| 0°/120°F
-18°/49°C| 32°/122°F
0°/50°C| 32°/122°F
0°/50°C
Power| 6.25V NiCad
Recharge Pk| 9V| 6.25V NiCad
Recharge Pk| 9V| Three 1.5V AAA| 9V
AC Power| Yes| No| Yes| No| No| No
Battery Life| 8 hours| 50 hours| 50 hours| 8 hours| 50 hours| 50 hours| 140
hours| 100 hours
Low Battery Ind| Yes| Yes| Yes| Yes
Recharger Included| 7.5V @ 100 mA
110 or 220 V AC| N/A| 7.5V @ 100 mA
110 or 220 VAC| N/A| N/A| N/A
Size| 6″ x 3″ x 1′| 6″ x 3″ x 1″| 6′ x 2″ x 3/4′| 7.5″ x 4″ x 2′
Weight| 10.5 oz. (298 grams)| 10.5 oz. (298 grams)| 3.6 oz. (102 grams)| 15.5
oz. (439 grams)
THERMOCOUPLE METER ACCESSORIES
DA-4 | Silicone Paste |
---|---|
DA-6 | Belt-Clip Meter Case with Hand Strap |
DA-10 | Nylon Safety Wrist Strap |
MA-150 | Battery Recharger, 110V AC 50/60Hz |
MA-150E | Battery Recharger, 220V AC 50/60Hz, European connector |
TA-60 | Shock-Proof attache-style Instrument Case |
11681-1 | Spare NiCad Battery Pack |
12232 | Standard 9V Alkaline Battery |
12234-02 | 3V, Lithium, coin Cell Battery for TA30 and TA50 |
12235 | 7.5V, Alkaline, Multi-Cell Battery for TA50 and TA70 |
METER OPTIONS
NISI-12| Certificate with 3 Standard Calibration Points between 32°F (0°C) &
500°F (250°C). Meter calibration includes one probe.
NIST-12-1| Certificate with 3 User-specified Calibration Points between -40°F
(-40°C) & 600°F (315°C). Meter calibration includes one probe.
Sample Meter Kit sold on pages 7 and 13. Kit includes Meter, Carrying Case, Calibrator, Wrist Strap, Recharger and Probe.
170 Years of Continued Innovation
800-421-2853
FAX828-658-0728
www.palmerwahl.com
Documents / Resources
|
PALMER Wahl 2500MX Heat
Prober
[pdf] Owner's Manual
2500M, 27504, 2500MX, 2500MX Heat Prober, 2500MX, Heat Prober, Prober
---|---
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>