PRAXISDIENST DZ 003 Dry Block Thermostat User Manual
- June 3, 2024
- PRAXISDIENST
Table of Contents
Dry Block Thermostat
Operating Manual
Edition 2021-01
Symbols
The following symbols and abbreviations may be found on the packaging material, the device type plate, and in the instructions for use:
| Manufactured by
---|---
****| This product meets the requirements of EC Directives 89/336/EEC
“Electromagnetic Compatibility” and 73/23/EEC “Low Voltage Directive”.
| Caution, refer to the accompanying documents!
Please follow the safety guidelines in the operating manual enclosed with this
device.
| Please consult user instructions.
| Biohazard
Samples containing material of human origin must be treated as potentially
infectious. Observe the relevant laboratory guidelines
for safe handling.
IP XO| No special protection against moisture penetration
(IP = International Protection).
REF| Order number
SN| Serial number
Safety Instructions
This device has been tested and left the factory in perfect technical condition. To maintain this condition and to ensure safe, trouble-free operation, the user must follow the instructions and warnings provided in this operating manual.
Device safety
The Dry Block Thermostat DZ 003 complies with the safety regulations according
to the standard DIN EN 61010-1.
The device meets the EMC requirements for laboratory devices according to the
standard DIN EN 61326.
Disposal
Diaglobal GmbH will take back and dispose of units that are no longer needed or cannot be repaired, free of charge.
General information about the Dry Block Thermostat
In order to carry out kinetics tests more efficiently in the laboratory, it is necessary to preheat the samples before measurement. In addition, the reagents must also be kept at incubation temperature during a series of measurements. For this purpose, a small and handy dry block thermostat was developed as a stand-alone system in addition to the Diaglobal Photometers.
The features in detail are:
- Temperature unit of 37 °C with an accuracy of ± 0.2 °C
- 12 positions for round cuvettes
- Indication of temperature control by LED
- Heating up in 5 minutes per cuvette filled with 1000 μL
- Power supply: 12 VDC / 15 VA
- AC-Adapter: 110 – 240 VAC 50/60 Hz
Delivery and Installation
Delivery
After carefully unpacking the device, check whether it is in perfect
condition.
Also, check if it is complete according to the delivery note. The box contains
the following items: Dry Block Thermostat and AC-Adapter
The packaging materials must be stored carefully in case an item needs to be
returned. Notify your dealer immediately if any defects are found.
Installation
Connect the AC-Adapter to the mains.
The operating voltage range is between 110 VAC and 240 VAC at 50/60 Hz.
Plug the connector into the socket on the backside of the Dry Block
Thermostat.
The Dry Block Thermostat DZ 003 is switched on with the red switch on the
front of the device.
After a few seconds, the green LED flashes. The red LED shows a continuous
light.
Operating Elements
Operation
After switching on the Dry Block Thermostat DZ 003, the device heats up to a
temperature of 37 °C.
The device needs about 20 minutes to heat up.
After the heat-up time, the GREEN LED shows a continuous light.
In this state, the system is ready for use.
The status of the device is shown by two LEDs (Light Emitting Diode):
GREEN LED (used for stability check)
- The LED is flashing: The temperature is not stable.
- The LED lights continuously: The temperature is stable.
RED LED (used for heating power)
- The LED is flashing: The heating is switched on and off.
- The LED lights continuously: The heating is switched on during warm-up.
Maintenance
This chapter provides necessary information concerning general maintenance by
the user.
If fault-free operation of the device cannot be achieved, customer service
should be contacted. Repairs to the device may only be carried out by
authorised specialist staff. Unauthorised repairs can endanger the operator
and also invalidate the warranty.
9.1 Cleaning Instructions
Liquid waste is potentially biohazardous. Always wear gloves when handling
these materials. Do not touch any parts of the device other than those
specified for use. Consult the laboratory protocol regarding the handling of
biohazardous materials.
Take care that no liquid enters the device! There is no protection against
penetrating liquids (Code IP X0).
Commercially available decontaminating solutions commonly used in clinical
chemistry laboratories, such as Mikrozid® AF Liquid, Bacillol® plus, 3 %
Kohrsolin® or similar, are recommended for cleaning the device and the
surface.
Before cleaning the unit with a soft cloth and the decontaminating solution,
it must be switched off and the electrical power supply must be disconnected.
9.2 Adjustment and Calibration
The Dry Block Thermostat is adjusted and calibrated at the factory on
delivery; adjustment by the customer is neither necessary nor possible.
Technical Data
10.1 Environmental Conditions
Climatic conditions for storage and transport of the packed device:
- Temperature: -25 °C to +70 °C
- Relative humidity: 20 % to 85 %
The device must be used in an environment that meets the following conditions:
- Temperature: +15 °C to +35 °C
- Relative humidity: 20 % to 85 %
- No direct exposure to sunlight or similar sources of radiant heat
- Well ventilated area
- Free from excessive dust
- Free from flammable gases
- Free from vibrations
- Free from electromagnetic wave interference
- Well distanced from any machine generating a high-frequency voltage
10.2 Type Plate
10.3 Short Specifications
MEASURING SYSTEM
- Microcontroller based temperature device
- Temperature: 37 °C with an accuracy of ± 0.2 °C
POWER SUPPLY
-
Supply voltage: 12 VDC
-
Maximum current consumption: 1.2 ADC
-
Power consumption: < 15 VA
-
External mains adapter
Output: 12 VDC
Input: 110 VAC up to 240 VAC 50/60 Hz 0.25 A max. -
Operating with a 12 VDC (car) battery is possible
DIMENSIONS: Dry Block Thermostat
- Dimensions (L/W/H): 160 mm / 100 mm / 85 mm
- Weight: 850 g
DIMENSIONS: AC-Adapter
- Dimensions (L/W/H): 95 mm / 45 mm / 30 mm
- Weight: 250 g
CONTAINER
Dry Block Thermostat DZ 003: 12 positions for round cuvettes
10.4 Certificates
EC Declaration of Conformity
Manufacturer: Diaglobal GmbH
Adress: Diaglobal GmbH
Innovationspark Wuhlheide
Köpenicker Str. 325 / Haus 41
12555 Berlin
Germany
Diaglobal GmbH declares under its sole responsibility that the product:
Product name: Dry Block Thermostat DZ 003
to which this declaration relates, complies with the requirements of the
standards and directives
EN 61326-1
EN 61010
and followed the requirements of EC Directives 89/336/EEC “Electromagnetic
Compatibility” and 73/23/EEC “Low Voltage Directive”.
This declaration describes a Dry Block Thermostat for laboratory use.
Berlin, October 2008
Diaglobal GmbH Manufacturer:
Diaglobal GmbH
Innovationspark Wuhlheide
Köpenicker Str. 325 / Haus 41
12555 Berlin
Germany
Tel: +49 (0) 30 6576 2597
Fax: +49 (0) 30 6576 2517
E-Mail: info@diaglobal.de
http://www.diaglobal.de
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>