ELEMENTAL MACHINES B1 Element-B Fully Automatic Switchover Installation Guide
- June 1, 2024
- ELEMENTAL MACHINES
Table of Contents
ELEMENTAL MACHINES B1 Element-B Fully Automatic Switchover
Product Information
- Specifications:
- Model: Element -B
- Compatible Models: EB1, EB2 Compatible with Taylor Wharton 10K/24K Cryo Freezers with CS100 controllers
- Package Contents: Element-B TW-24K-CS100 Connection Cable
Product Usage Instructions
- Installation Instructions:
- Receiving the Shipment:
- Set up your Elemental Machines system and verify your dashboard account via the email received.
- Devices will be added to the Elemental InsightsTM Dashboard with default names upon shipment.
- Inspect the contents upon arrival. Report any damages to help@elementalmachines.com.
- If using multiple Element-B’s, ensure correct configuration (Normally Closed or Normally Open).
- Place the Element-B within range of the Elemental Machines Gateway.
- Attaching Connection Cable:
- Locate the ‘Remote Alarm’ port on the Cryo Freezer.
- Connect the green end of the cable to the Remote Alarm port on the Cryo Freezer.
- Write the name/location of the monitored equipment on the Element-B for easy identification.
FAQs
- Q: How do I verify if my Element-B is configured as Normally Closed or Normally Open?
- A: You can check the device label, and dashboard under ‘Machine Model’, or the packing list sent with your shipment to verify the configuration.
- Q: What should I do if my devices arrive damaged?
- A: Please send a picture of the damaged contents to help@elementalmachines.com for further assistance.
Applicable Asset Models
- TW-10K-CS100
- TW-24K-CS100
Package Contents
- Element-B
- Connection Cable
Safety Information
- Review and abide by the safety information found in the Element-B and Asset user manuals.
- Ensure compliance with your company’s safety best practices, which may include de-energizing equipment before working with electrical connections.
- Tidy connection cable to prevent tripping hazards
Installation Instructions
Receiving the Shipment
-
If this is your first time setting up the Elemental Machines system, you will receive an email for your dashboard account verification. Save this email for when the devices arrive.
-
When devices are shipped, they will be added to your Elemental Insights™ Dashboard with default names. When you first log in you will see all your devices with a ‘disconnected’ status.
-
When devices arrive, carefully remove the contents from the packaging. If any of the contents look damaged, please send a picture to help@elementalmachines.com.
-
If you have ordered multiple Element-B’s, ensure the Element-B being used is configured as Normally Closed (NC). This can be verified by checking the device label, the dashboard under ‘Machine Model’, or the packing list sent with your shipment.
- a. Element-B’s can either be configured as Normally Closed (NC) or Normally Open (NO) by our team here at Elemental Machines. Different assets require different configurations. If you have ordered Element-B’s for multiple assets, you may have a mix and need to be aware of attaching the correctly configured device to the correct asset.
-
Ensure there is an Elemental Machines Gateway set up in the vicinity of the Element-B.
- a. Depending on your location and application, your Gateway type may vary. If you have a tablet Gateway, brief setup instructions are given in the appropriate Element-B user manual. If you have a Gateway Model GW2 or GW3, please follow the setup instructions in your Gateway, GW2 or GW3 User Manual
- b. Note: ‘Vicinity’ is within wireless connection range. This is usually up to 30 meters but can depend on the lab’s layout.
Attaching Connection Cable to Asset
-
Locate the ‘Remote Alarm’ port on the back of the instrument, as shown in Figure 1.
-
Plug the green end of the connection cable, circled in Figure 2, into the Remote Alarm port. Attaching the Connection Cable to the Element
-
Per your discretion, write the name and/or location of the equipment being monitored on the Element-B. This can be particularly helpful if multiple devices are being installed. Use of a Sharpie is recommended.
-
Plug the thermocouple end of the connection cable, shown in Figure 3, into the Element-B.
- a. Note Ensure the (-) and (+) prongs of the thermocouple end are properly matched.
-
Pull the battery tab on the Element-B.
Positioning the Element-B -
Determine where the Element-B should be placed so that it is not blocked, which allows a better wireless connection, and so that the connection cable is tidy.
-
Securely position the Element-B in the chosen location. The Element-B has magnets which make it easily mountable to magnetic surfaces.
- a. Note: If applicable, an adhesive metal mounting plate available from EM may be used.
Programming Alerts
- a. Note: If applicable, an adhesive metal mounting plate available from EM may be used.
-
Refer to the asset user manual for which alarm conditions can be detected.
-
Set up alert rules on the Elemental Insights™ Dashboard for the Element-B. Guidance on how to do so is available in the Support section of the dashboard.
- NOTE: Alerts must be programmed on both the Asset and the Elemental Insights™ Dashboard to receive threshold alerts through the Elemental Machines platform.
- 772-00077 rev 01 Element-B, EB Taylor Wharton 10K/24K LN2 Freezer Installation Guide Copyright © 2024 Elemental Machines help@elementalmachines.com.
- Models: EB1, EB2
- For Taylor Wharton 10K/24K Cryo Freezers with CS100 controllers
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>