- Product Information
- Product Usage Instructions
- Models
- Package Contents
- Safety Information
- Installation Instructions
- Gateway Setup
- Positioning the Tablet Gateway
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
ELEMENTAL MACHINES EB1 Element-B Thermo Fisher IsoTemp CO2 Incubator
Product Information
- Product Name: Element-B
- Models: EB1, EB2
- Compatible with: Thermo Fisher IsoTemp CO2 Incubator
Applicable Thermo Fisher IsoTemp CO Incubator
- Models: FB3530WJ, FB3531WJ
Package Contents:
- Element-B
- Connection Cable
Safety Information
Review and abide by the safety information found in the Element-B and Thermo
Fisher IsoTemp CO Incubator user manuals. Tidy connection cable to prevent
tripping hazards.
Product Usage Instructions
Receiving the Shipment
-
If this is your first time setting up the Elemental Machines system, you will receive an email for your dashboard account verification. Save this email for when the devices arrive.
-
When devices are shipped, they will be added to your Elemental Insights DashboardTM with default names. When you first log in you will see all your devices with a ‘disconnected’ status.
-
When devices arrive, carefully remove the contents from the packaging. If any of the contents look damaged, please send a picture to help@elementalmachines.com.
-
If you have ordered multiple Element-B’s, ensure the Element-B being used is configured as Normally Closed (NC). This can be verified by checking the device label, the dashboard under ‘Machine Model’, or the packing list sent with your shipment.
Attaching Connection Cable to Asset
Follow these steps to attach the connection cable to the incubator: -
Locate the RJ-11 telephone-style jack on the rear panel of the incubator (see Figure 1).
-
Plug the RJ-11 connector end (shown in Figure 2) of the connection cable into the RJ-11 port on the incubator.
Refer to Figure 1 and Figure 2 for visual reference.
Attaching Connection Cable to Element
Follow these steps to attach the connection cable to the Element-B: -
Per your discretion, write the name and/or location of the equipment being monitored on the Element-B. This can be particularly helpful if multiple devices are being installed. Use of a Sharpie is recommended.
-
Plug the thermocouple end (Figure 3) of the connection cable into the Element-B.
Note: Ensure the (-) and (+) prongs of the thermocouple end are properly matched. Refer to Figure 3 for visual reference.
Positioning the Element-B -
Determine where the Element-B should be placed so that it is not blocked, which allows a better wireless connection, and so that the connection cable is tidy.
-
Securely position the Element-B in the chosen location. The Element-B has magnets which make it easily mountable to magnetic surfaces.
Note: If applicable, an adhesive metal mounting plate available from EM may be used.
Programming Alerts
Program alerts on the incubator. Refer to the Thermo Fisher IsoTemp CO
Incubator user manual for which alarm conditions the incubator can detect and
how to set up alerts.
Models
Applicable Thermo Fisher IsoTemp CO₂ Incubator Models
- FB3530WJ
- FB3531WJ
Package Contents
- Element-B
- Connection Cable
Safety Information
Review and abide by the safety information found in the Element-B and Thermo Fisher IsoTemp CO₂ Incubator user manuals. Tidy connection cable to prevent tripping hazards.
Installation Instructions
Receiving the Shipment
-
If this is your first time setting up the Elemental Machines system, you will receive an email for your dashboard account verification. Save this email for when the devices arrive.
-
When devices are shipped, they will be added to your Elemental Insights Dashboard™ with default names. When you first log in you will see all your devices with a
‘disconnected’ status. -
When devices arrive, carefully remove the contents from the packaging. If any of the contents look damaged, please send a picture to help@elementalmachines.com.
-
If you have ordered multiple Element-B’s, ensure the Element-B being used is configured as Normally Closed (NC). This can be verified by checking the device label, the dashboard under ‘Machine Model’, or the packing list sent with your shipment.
- Element-B’s can either be configured as Normally Closed (NC) or Normally Open (NO) by our team here at Elemental Machines. Different assets require different configurations. If you have ordered Element-B’s for multiple assets, you may have a mix and need to be aware of attaching the correctly configured device to the correct asset.
-
Ensure there is an Elemental Machines Gateway set up in the vicinity of the Element-B.
- Depending on your location and application, your Gateway type may vary. If you have a tablet Gateway, brief setup instructions are given in Appendix 1: Gateway Setup below. If you have a Gateway Model GW2 or GW3, please follow the setup instructions in your Gateway, GW2 or GW3 User Manual
- Note: ‘Vicinity’ is within wireless connection range. This is usually up to 30 meters but can depend on the lab’s layout.
Attaching Connection Cable to Asset
-
Locate the RJ-11 telephone-style jack on the rear panel of the incubator (see Figure 1).
-
Plug the RJ-11 connector end (shown in Figure 2) of the connection cable into the RJ-11 port on the incubator.
Attaching Connection Cable to Element -
Per your discretion, write the name and/or location of the equipment being monitored on the Element-B. This can be particularly helpful if multiple devices are being installed. Use of a Sharpie is recommended.
-
Plug the thermocouple end (Figure 3) of the connection cable into the Element-B.
- Note Ensure the (-) and (+) prongs of the thermocouple end are properly matched.
-
Pull the battery tab on the Element-B.
Positioning the Element-B -
Determine where the Element-B should be placed so that it is not blocked, which allows a better wireless connection, and so that the connection cable is tidy.
-
Securely position the Element-B in the chosen location. The Element-B has magnets which make it easily mountable to magnetic surfaces.
- Note: If applicable, an adhesive metal mounting plate available from EM may be used.
Programming Alerts
- Note: If applicable, an adhesive metal mounting plate available from EM may be used.
-
Program alerts on the incubator. Refer to the Thermo Fisher IsoTemp CO₂ Incubator user manual for which alarm conditions the incubator can detect and how to set up alerts.
-
Set up alert rules on the Elemental Insights Dashboard™ for the Element-B. Guidance on how to do so is available in the Support section of the dashboard.
NOTE: Alerts must be programmed on both the asset and the Elemental Insights Dashboard™ to receive remote alerts through the Elemental Machines platform.
Gateway Setup
Elemental Machines provides multiple styles of Gateways. If you have a Tablet Gateway (Model GW1), please follow the setup instructions below. For Gateway-2, please follow the setup instructions in 771-00021 Gateway (Model GW2) User Manual. For Gateway-3, please follow the setup instructions in 771-00034 Gateway (Model GW3) User Manual.
Tablet Gateway (Model GW1) Software Setup
To power on the Gateway, press and hold the upper right-hand button on the
device:
- Wait for the main screen to come up
- Press the ‘home’ button when you see the Elemental Machines Logo
- Click on the blue circle in the upper right corner (if the circle is not visible, press the hardware home button on the bottom of the tablet to make it appear, or swipe up if you do not have a hardware button):
- Click on the settings icon:
- Select WiFi from the list:
Positioning the Tablet Gateway
Tablet Gateways collect data from the Elements, collating it and transmitting it across the internet to Elemental Machines’ Cloud. The tablet Gateways’ default is to transmit by Wi-Fi; for added reliability they fall back to Cellular connection when Wi-Fi connection drops out. There is a danger of data delay or even loss if all connection is lost, so tablet Gateways should be positioned where they are getting good Wi-Fi and Cellular connection.
The strength of the Wi-Fi and Cellular connection are displayed by the Wi-Fi icon using Cellular bar icons. These icons are displayed on the tablet Gateway to the left of the battery percentage.
- 4 or more bars for both Wi-Fi and cell indicate good connectivity
- 2 bars for both Wi-Fi and cell runs an increased risk of some data delay or loss
- <2 bars for cell or Wi-Fi carry a danger of significant data delay or loss
772-00003 rev 01 Element-B, EB1 EB2 Thermo Fisher IsoTemp CO₂ Incubator Installation Guide.