ELEMENTAL MACHINES EB1 Element-B ThermoFisher Forma 3100 SteriCult Incubators Installation Guide

June 12, 2024
ELEMENTAL MACHINES

MACHINES EB1 Element-B Thermop Fisher Forma 3100
Sterically Incubators

Installation GuideELEMENTAL MACHINES EB1 Element-B ThermoFisher Forma
3100 SteriCult Incubators - fig

Element-B Installation Guide
Models: EB1, EB2
For ThermoFisher Forma 3100, 3950, 3951, 4100,
Teracycle, and Sterically Incubators

Applicable ThermoFisher Incubator Model Series

Thermop Forma 3100 Series
Thermop Forma 3950 and 3951 Series
Thermop Forma 4100 Series
Thermop Teracycle Series
Thermop Sterically Series

Package Contents

Element-B
Connection Cable

Safety Information

Review and abide by the safety information found in the Element-B and
ThermoFisher Forma 3100, 3950, 3951, 4100, STERIS Cycle, and STERIS Cult Incubator user manuals. Tidy connection cable to prevent tripping hazards.

Installation Instructions

Receiving the Shipment

  1. If this is your first time setting up the Elemental Machines system, you will receive an email for your dashboard account verification. Save this email for when the devices arrive.

  2. When devices are shipped, they will be added to your Elemental Insights Dashboard™ with default names. When you first log in you will see all your devices with a ‘disconnected’ status.

  3. When devices arrive, carefully remove the contents from the packaging. If any of the contents look damaged, please send a picture to help@elementalmachines.com.

  4. If you have ordered multiple Element-B’s, ensure the Element-B being used is configured as Normally Closed (NC). This can be verified by checking the device label, the dashboard under ‘Machine Model’, or the packing list sent with your shipment.
    a. Element-B’s can either be configured as Normally Closed (NC) or Normally Open (NO) by our team here at Elemental Machines. Different assets require different configurations. If you have ordered Element-B’s for multiple assets, you may have a mix and need to be aware of attaching the correctly configured device to the correct asset.

  5. Ensure there is an Elemental Machines Gateway set up in the vicinity of the Element-B.
    a. Depending on your location and application, your Gateway type may vary. If you have a tablet Gateway, brief setup instructions are given in Appendix 1: Gateway Setup below. If you have a Gateway Model GW2 or GW3, please  follow the setup instructions in your Gateway, GW2 or GW3 User Manual
    b. Note: ‘Vicinity’ is within wireless connection range. This is usually up to 30 meters but can depend on the lab’s layout.
    Attaching Connection Cable to ThermoFisher Forma 3100, 3950, 3951, 4100, SteriCycle, or SteriCult Incubator

  6.  Locate the Remote Alarm Connector (RJ-11 telephone-style jack) on the rear panel of the incubator as shown in the figures below:ELEMENTAL MACHINES EB1 Element-B ThermoFisher Forma 3100 SteriCult IncubatorsELEMENTAL MACHINES EB1 Element-B ThermoFisher Forma 3100 SteriCult Incubators - fig1ELEMENTAL MACHINES EB1 Element-B ThermoFisher Forma 3100 SteriCult Incubators - fig2ELEMENTAL MACHINES EB1 Element-B ThermoFisher Forma 3100 SteriCult Incubators - fig3

  7. Plug the RJ-11 connector end (shown in Figure 6) of the connection cable into the RJ-11 port on the incubator.ELEMENTAL MACHINES EB1 Element-B ThermoFisher Forma 3100 SteriCult Incubators - cableAttaching Connection Cable to Element

  8. Per your discretion, write the name and/or location of the equipment being monitored on the Element-B. This can be particularly helpful if multiple devices are being installed. Use of a Sharpie is recommended.

  9. Plug the thermocouple end (Figure 7) of the connection cable into the Element-B.
    a. Note Ensure the (-) and (+) prongs of the thermocouple end are properly matched.ELEMENTAL MACHINES EB1 Element-B ThermoFisher Forma 3100 SteriCult
Incubators - cable 1

  10. Pull the battery tab on the Element-B.
    Positioning the Element-B

  11.  Determine where the Element-B should be placed so that it is not blocked, which allows a better wireless connection, and so that the connection cable is tidy.

  12. Securely position the Element-B in the chosen location. The Element-B has magnets which make it easily mountable to magnetic surfaces.
    a. Note: If applicable, an adhesive metal mounting plate available from EM may be used.
    Programming Alerts

  13. Program alerts on the incubator. Refer to the ThermoFisher Forma 3100, 3950, 3951, 4100, SteriCycle, or SteriCult Incubator user manual for which alarm conditions the incubator can detect and how to set up alerts.

  14.  Set up alert rules on the Elemental Insights Dashboard™ for the Element-B. Guidance on how to do so is available in the Support section of the dashboard.

NOTE: Alerts must be programmed on both the ThermoFisher Forma 3100, 3950, 3951, 4100, SteriCycle, or SteriCult Incubator and the Elemental Insights Dashboard™ to receive threshold alerts through the Elemental Machines platform.

Appendix 1: Gateway Setup

Elemental Machines provides multiple styles of Gateways. If you have a Tablet Gateway (Model GW1), please follow the setup instructions below. For Gateway-2, please follow the setup instructions in 771-00021 Gateway (Model GW2)  User Manual. For Gateway-3, please follow the setup instructions in 771-00034 Gateway (Model GW3) User Manual.

ELEMENTAL MACHINES EB1 Element-B ThermoFisher Forma 3100 SteriCult
Incubators - Gateway Setup

Tablet Gateway (Model GW1) Software Setup
To power on the Gateway, press and hold the upper right-hand button on the device:

  1. Wait for the main screen to come up

  2. Press the ‘home’ button when you see the Elemental Machines Logo

  3. Click on the blue circle in the upper right corner (if the circle is not visible, press the hardware home button on the bottom of the tablet to make it appear, or swipe up if you do not have a hardware button):http://www.elementalmachines.com

  4. Click on the settings icon:http://www.elementalmachines.com

  5. Select WiFi from the list:

Positioning the Tablet Gateway
Tablet Gateways collect data from the Elements, collating it and transmitting it across the internet to Elemental Machines’ Cloud. The tablet Gateways’ default is to transmit by Wi-Fi; for added reliability they fall back to Cellular  connection when Wi-Fi connection drops out. There is a danger of data delay or even loss if all connection is lost, so tablet Gateways should be positioned where they are getting good Wi-Fi and Cellular connection.
The strength of the Wi-Fi and Cellular connection are displayed by the Wi-Fi icon using Cellular bar icons. These icons are displayed on the tablet Gateway to the left of the battery percentage.

  • 4 or more bars for both Wi-Fi and cell indicate good connectivity
  •  2 bars for both Wi-Fi and cell runs an increased risk of some data delay or loss
  • <2 bars for cell or Wi-Fi carry a danger of significant data delay or loss

772-00007 rev 01 Element-B, EB1 EB2 ThermoFisher
Forma 3100, 3950, 3951, 4100, SteriCycle, and SteriCult
Incubator Models Installation Guide
Copyright © 2022 Elemental Machines
help@elementalmachines.com

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>