Hygiena KIT230034 Star Prep One 8 Strip Kit Instruction Manual
- June 1, 2024
- Hygiena
Table of Contents
- Hygiena KIT230034 Star Prep One 8 Strip Kit
- Product Information
- Product Usage Instructions
- What This Product Does
- How to Use This Product
- Troubleshooting
- Additional Information on this Product
- Supplementary Information
- Change Index
- Contact Information
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Hygiena KIT230034 Star Prep One 8 Strip Kit
Product Information
What This Product Does
The foodproof Escherichia coli and Shigella Detection Kit is designed for the
rapid detection of E. coli and Shigella DNA isolated from enrichment cultures
prepared by valid methods and inoculated with potentially contaminated foods.
It should not be used for diagnostic procedures.
-
Number of Tests
The kit contains reagents for multiple tests. Refer to the kit contents for detailed information. -
Storage and Stability
Ensure proper storage conditions as indicated in the kit contents for optimal stability. -
Kit Contents****
Vial/Label | Cap Color | Contents/Function/Storage |
---|---|---|
1 | Yellow cap | foodproof E. coli and Shigella Enzyme Solution |
2 | Red cap | foodproof E. coli and Shigella Internal Control |
-
Additional Equipment and Reagents Required
Refer to the manual for information on additional equipment and reagents required for optimal use of the kit. -
Applicability Statement
Follow the provided instructions carefully for the intended use of the kit to detect E. coli and Shigella DNA from enrichment cultures.
Specifications
- Product Name: foodproof Escherichia coli and Shigella Detection Kit
- Product No.: KIT230034
- Intended Use: For food testing purposes. FOR IN VITRO USE ONLY
- Manufacturer: www.hygiena.com.
Product Usage Instructions
How to Use This Product
Before You Begin
Precautions
Detection of E. coli and Shigella DNA requires PCR amplification under
nuclease-free conditions. Follow precautions to avoid contamination.
- Avoid light exposure for the foodproof Escherichia coli and Shigella Master Mix.
Waste Disposal
Dispose of waste potentially contaminated with pathogenic bacteria properly
under local regulations.
FAQ
Q: Can this kit be used for diagnostic procedures?
A: No, the kit is not intended for diagnostic procedures.
foodproof® Escherichia coli and Shigella Detection Kit
Revision A, January 2024
PCR kit for the qualitative detection of Escherichia coli and Shigella spp.
DNA using the LightCycler® Carousel-Based or 480 II System.
- Product No. KIT230034
- Kit for 96 reactions
- Store the kit at -15 to -25 °C
What This Product Does
Number of Tests
The detection kit is designed for 96 reactions with a final reaction volume of
20 µL each. Up to 30 samples (single sample preparation) plus positive and
negative control reactions can be analyzed per LightCycler® Carousel-Based
System run (i.e., the complete detection kit allows analysis of a maximum of
90 samples). Using the LightCycler 480 System, up to 94 samples can be
analyzed.
Storage and Stability
- Store the kit at -15 to -25 °C through the expiration date printed on the label.
- Once the kit is opened, store the components as described in the following Contents table:
Kit Contents
Vial / Cap Color | Label | Contents / Function / Storage |
---|---|---|
1 – yellow cap | foodproof® E. coli and Shigella – Master Mix | · 3 x 420 µL |
· Ready-to-use primer and Hybridization Probe mix – specific for E. coli and
Shigella
· DNA and E. coli and Shigella-specific Internal Control (IC).
· For amplification and detection of E. coli- and Shigella-specific sequences
· Store at -15 to -25 °C.
· Avoid repeated freezing and thawing!
· Protect from light!
2 – red cap| foodproof E. coli and Shigella – Enzyme Solution| • 3 x 32 µL
• Contains DNA-free Taq DNA Polymerase and Uracil-DNA Glycosylase (heat labile) for prevention of carry-over contamination
• Store at -15 to -25 °C
3 – white cap| foodproof E. coli and Shigella – Internal Control| • 3 x 32 µL
• Contains a stabilized solution of plasmid DNA.
• For use as an internal amplification control.
• Store at -15 to -25°C.
• After first thawing, store at 2 to 8 °C for up to one month.
4 – purple cap| foodproof E. coli and Shigella – Control Template| • 1 x 50 µL
• Contains a stabilized solution of plasmid DNA.
• For use as a PCR run positive control.
• Store at -15 to -25°C.
• After first thawing, store at 2 to 8 °C for up to one month.
5 – clear, colorless cap| H2O, PCR-grade| • 1 x 1 mL
• Nuclease-free, PCR-grade H2O.
• For use as a PCR run negative control.
• Store at -15 to -25°C.
Additional Equipment and Reagents Required
-
LightCycler Carousel-Based System (LightCycler 1. x, 2.0 Instrument)
-
LightCycler 20 µL Capillaries
-
Standard benchtop microcentrifuge containing a rotor for 2.0 mL reaction tubes
Note: The LightCycler Carousel-Based System provides adapters that allow LightCycler Capillaries to be centrifuged in a standard microcentrifuge rotor. or -
LC Carousel Centrifuge 2.0 for use with the LightCycler 2.0 Sample Carousel (optional).
If you use a LightCycler Instrument version below 2.0, you also need the LC Carousel Centrifuge 2.0 Bucket 2.1. To adapt the LightCycler 2.0 Sample Carousel to the former LC Carousel Centrifuge, you need the LC Carousel Centrifuge 2.0 Rotor Set. or
-
LightCycler 480 II System
-
LightCycler 480 II compatible PCR plate and sealing foil
-
LightCycler Color Compensation Set
-
foodproof StarPrep One Kit (Product No. KIT230175)
or -
foodproof ShortPrep II Kit (Product No. KIT230094)
-
Nuclease-free, aerosol-barrier pipette tips
-
Pipettes
-
Sterile reaction tubes for preparing PCR mixes and dilutions
Applicability Statement
The foodproof Escherichia coli and Shigella Detection Kit is intended for the
rapid detection of E. coli and Shigella DNA isolated from enrichment cultures
prepared by various valid methods and inoculated with a variety of foods that
are potentially contaminated with E. coli or Shigella spp.
The kit must not be used in diagnostic procedures.
- The user should take care with enrichment cultures as they may contain high numbers of pathogenic organisms. It is strictly recommended to follow all necessary precautions for handling pathogenic organisms during the whole analysis.
- The kit described in this Instruction Manual has been developed for the LightCycler Carousel-Based System and the LightCycler 480 System II (96-well block type). The LightCycler 480 I System should not be used for this kit.
How to Use This Product
Before You Begin
Precautions
Detection of E. coli and Shigella DNA using the foodproof Escherichia coli and
Shigella Detection Kit requires DNA amplification by PCR. The kit provides all
the reagents required for the PCR. However, to achieve reliable results, the
entire assay procedure must be performed under nuclease-free conditions.
Follow the instructions below to avoid nuclease-, carry-over-, or cross-
contamination:
- Prepare appropriate aliquots of the kit solutions and keep them separate from other reagents in the laboratory.
- Use nuclease-free labware (e.g., pipettes, pipette tips, reaction vials).
- Wear gloves when performing the assay.
- To avoid cross-contamination of samples and reagents, use fresh aerosol-preventive pipette tips.
- To avoid carry-over contamination, transfer the required solutions for one experiment into a fresh tube, rather than directly pipetting from stock solutions.
- Do not touch the surface of the capillaries. Always wear gloves when handling the capillaries.
- Physically separate the workplaces for DNA preparation, PCR set-up, and PCR to minimize the risk of carry-over contamination.
- To avoid cross-contamination, close all capillaries that contain sample DNA and negative controls before pipetting positive controls.
Keep the foodproof Escherichia coli and Shigella Master Mix (vial 1, yellow cap) away from light.
-
Waste Disposal
Place any waste and biohazard material potentially contaminated with pathogenic bacteria in an appropriate plastic contaminated waste bag and label as follows: “CONTAMINATED waste, room number, date and initials.” The bag should be autoclaved and then disposed of according to local regulations. -
Sample Material
Use any sample material suitable for PCR in terms of purity, concentration, and absence of inhibitors. For the preparation of genomic DNA from raw material or from food enrichments, refer to the corresponding product package inserts of a suitable sample preparation kit (see ‘Additional Equipment and Reagents Required”). -
DNA Extraction
Hygiena Diagnostics provides sample preparation kits for all kinds of foods and raw materials (see “Additional Equipment and Reagents Required”). For more product information, please see www.hygiena.com. -
Positive Control
Always run a positive control with the samples. To prepare a positive control, replace the template DNA with the provided control DNA [foodproof E. coli and Shigella Control Template (vial 4, purple cap)] or with a positive sample preparation control DNA. Always close capillaries with template DNA and negative controls before adding positive control DNA. -
Negative Control
Always run a negative control with the samples. To prepare a negative control, replace the template DNA with H2O, PCR-grade water (vial 5, colourless cap). Include a negative control during sample preparation to monitor reaction purity and cross-contamination. This extraction control can be used as an additional negative control reaction. -
Colour Compensation
The use of a previously generated colour compensation object is a prerequisite for the unambiguous discrimination of E. coli/Shigella DNA and internal control (IC) DNA amplification in this dual-colour experiment. A suitable colour compensation object can be generated using dedicated reagents available as LightCycler Color Compensation Sets. As colour compensation is instrument- specific, it is necessary to generate a CC object for every LightCycler Instrument. A new object has to be created after the optical system has been repaired. For additional information on colour compensation, please refer to the manual for the respective LightCycler Instrument. -
PCR Procedure
The following procedures are optimized for the LightCycler 480 II System and the LightCycler 2.0 Carousel-Based System. Program the LightCycler Systems before preparing the reaction mixes. The protocols contain the following programs:- Pre-incubation to prevent carry-over contamination (UNG), to activate Taq polymerase and for DNA-denaturation
- Amplification of the target DNA
- Cooling of the LightCycler System
LightCycler 480 II System Protocol
The following procedure is optimized for use with the LightCycler 480 II
System. Program the LightCycler before preparing the reaction mixes. Use the
following LightCycler 480 System PCR program for the foodproof Escherichia
coli and Shigella Detection Kit (for details on how to program the
experimental protocol, see the LightCycler 480 II System Operator’s Manual):
Setup
Detection Format| Block Type| Reaction Volume
Multi Color HybProbe| 96| 20 µL
Filter Setting| Dynamic mode, LC480 II: Fluos (465-510, Red 640 (498-640) and
Cy 5/ Cy 5.5 (498-660)
Programs
Program Name| Cycles| Analysis Mode
Pre-Incubation| 1| None
Amplification| 45| Quantification
Cooling| 1| None
Temperature Targets
| Target [°C]| Acquisition Mode| Hold [hh:mm: ss]| Ramp
Rate [°C/s]| Secondary Target Temperature [°C]| Step Size [°C]|
Step Delay [cycles]
Pre-Incubation
Segment 1| 37| None| 00:02:00| 4.4| 0| 0.0| 0
Segment 2| 95| None| 00:10:00| 4.4| 0| 0.0| 0
Amplification
Segment 1| 95| None| 00:00:05| 4.4| 0| 0.0| 0
Segment 2| 59| Single| 00:00:35| 2.2| 0| 0.0| 0
Segment 3| 72| None| 00:00:15| 4.4| 0| 0.0| 0
Cooling
| 40| None| 00:00:30| 2.2| 0| 0.0| 0
LightCycler Carousel-Based System Protocol
The following procedure is optimized for use with the LightCycler Carousel-
Based System. Program the LightCycler Carousel-Based System before preparing
the reaction mixes. A LightCycler Carousel-Based System protocol that uses the
foodproof Escherichia coli and Shigella Detection Kit contains the following
programs (for details on how to program the experimental protocol, see the
LightCycler Instrument Operator’s Manual):
Pre-incubation
Programs/Cycle Program Data| Value
Cycles| 1
Analysis Mode| None
Temperature Targets| Segment 1| Segment 2
Target/Target Temperature [°C]| 40| 95
Hold/Incubation Time [h:min:s]| 00:02:00| 00:10:00
Ramp Rate/Temperature Transition Rate [°C/s]| 20| 20
Sec Target/Secondary Target Temperature [°C]| 0| 0
Step Size [°C]| 0.0| 0.0
Step Delay [cycles]| 0| 0
Acquisition Mode| None| None
Amplification
Programs/Cycle program data| Value
Cycles| 45
Analysis Mode| Quantification
Temperature Targets| Segment 1| Segment 2| Segment 3
Target/Target Temperature [°C]| 95| 59| 72
Hold/Incubation Time [h:min:s]| 00:00:00| 00:00:30| 00:00:05
Ramp Rate/Temperature Transition Rate [°C/s]| 20| 20| 20
Sec Target/Secondary Target Temperature [°C]| 0| 0| 0
Step Size [°C]| 0.0| 0.0| 0.0
Step Delay [cycles]| 0| 0| 0
Acquisition Mode| None| Single| None
Cooling
Programs/Cycle program data| Value
Cycles| 1
Analysis Mode| None
Temperature Targets| Segment 1
Target/Target Temperature [°C]| 40
Hold/Incubation Time [h:min:s]| 00:00:30
Ramp Rate/Temperature Transition Rate [°C/s]| 20
Sec Target/Secondary Target Temperature [°C]| 0
Step Size [°C]| 0.0
Step Delay [cycles]| 0
Acquisition Mode| None
Fluorescence and Run Setup Parameters
Parameter | Setting |
---|
All LightCycler Software Versions
Seek Temperature| 30°C
LightCycler Software (Before Version 3.5)
Display Mode| Fluorescence channel F2 or F3
Fluorescence Gains| Fluorimeter| Gain Value
Channel 1 (F1)| 1
Channel 2 (F2)| 15
Channel 3 (F3)| 30
LightCycler Software Version 3.5
Display Mode
• during run
• for analysis
| • Fluorescence channel F2 or F3
• F2/Back-F1 or F3/Back-F1
Fluorescence Gains| · Not required
In data created with LightCycler Software Version 3.5, all fluorescence values are normalized to a fluorescence gain of “1”.
This produces a different scale on the Y-axis than that obtained with previous LightCycler Software Versions.
This difference does not affect the crossing points nor any calculated concentrations obtained.
LightCycler Software Version 4.x
Default channel
• during run
• for analysis
| • Fluorescence channel 640 or 705
• 640/Back 530 or 705/Back 530
---|---
Fluorescence Gains| · Not required
“Max. Seek Pos”| Enter the number of samples, including controls
“Instrument Type”| “6 Ch.”: for LightCycler 2.0 Instrument (selected by
default)
“3 Ch.”: for LightCycler 1.5 Instrument and instrument versions below
“Capillary Size”| Select “20 µL” as the capillary size for the experiment. (For the “6 Ch.” instrument type only).
Preparation of the PCR Mix
Proceed as described below to prepare a 20 µL standard reaction.
Do not touch the surface of the capillaries. Always wear gloves when handling the capillaries. For LightCycler 480 users, do not touch the upper surface of the PCR multiwell plate.
- Depending on the total number of reactions, place the required number of LightCycler Capillaries in centrifuge adapters or a LightCycler Sample Carousel in an LC Carousel Centrifuge Bucket. For LightCycler 480 instruments, use a suitable multiwell plate.
- Thaw the solutions and, for maximal recovery of contents, briefly spin vials in a microcentrifuge before opening. Mix carefully by pipetting up and down.
- In a 1.5 mL reaction tube, prepare the PCR Mix for one 20 µL reaction by adding the following components in the order mentioned below, then mix gently by pipetting up and down.
The volumes indicated below are based on a single 20 µL standard reaction. Prepare the PCR mix by multiplying the amount in the “Volume” column by the number of reactions plus one positive and one negative control to be cycled plus one or two additional reactions to cover pipetting losses. Component| Volume
---|---
foodproof E. coli and Shigella Master Mix, (vial 1, yellow cap)| 13 µL
foodproof E. coli and Shigella Enzyme Solution (vial 2, red cap)| 1 µL
foodproof E. coli and Shigella Internal Control (vial 3, white cap)| 1 µL
Total volume| 15 µL - Mix carefully by pipetting up and down. Do not vortex.
- Pipet 15 µL PCR mix into each LightCycler capillary or plate well respectively.
- For the samples of interest, add 5 µL sample DNA to a capillary or a well (LC 480), and seal the capillary with a stopper.
- For the negative control, add 5 µL H2O, PCR-grade (vial 5, colourless cap); seal the capillary with a stopper.
- For the positive control, add 5 µL foodproof E. coli and Shigella Control Template (vial 4, purple cap); seal the capillary with a stopper.
- For LightCycler Carousel Based System: Place the adapters (containing the capillaries) in a standard benchtop microcentrifuge. (Place the centrifuge adapters in a balanced arrangement within the centrifuge.)
- Centrifuge at 700 x g for 5 s (3,000 rpm in a standard benchtop microcentrifuge).
- Alternatively, use the LC Carousel Centrifuge for spinning the capillaries.
- Transfer the capillaries to the LightCycler.
- For LightCycler 480 II System: Seal the plate accurately with an optical sealing foil. Place the plate in a swing bucket centrifuge and centrifuge at 1,500 x g for 30 s.
- Cycle the samples as described above.
Analysis
Analyze real-time PCR results in channels F2/Back-F1 and F3/Back-F1
(LightCycler Software 3.5 and software versions below), in channels 640/Back
530 and 705/Back 530 (LightCycler Software 4. x) or channels Red 640 (498-640)
and Cy 5/ Cy 5.5 (498-660) (LightCycler 480 Software 1.5) respectively, using
the Quantification module (LightCycler Software 3.5 and software versions
below), the Qualitative Detection module (LightCycler Software 4. x) or the
Abs Quant/2nd Derivative Max analysis type (LightCycler 480 Software 1.5) of
the LightCycler Analysis Software. Check for a positive result of the Internal
Control (visible signal in channel F3, 705 or Cy5 / Cy 5.5 (498-660)) for each
sample that is negative for E. coli and Shigella DNA (no signal in channel F2,
640 or Red 640 (498-640)). Compare the results from channel F2, 640 or Red 640
(498-640) (E. coli and Shigella) and channel F3, 705 or Cy 5 / Cy 5.5
(498-660) (Internal Control) for each sample, and interpret the results as
described in the table below:
E. coli and Shigella
Channel F2/Back-F1, Channel 640/Back 530
or Red 640 (498-640)
| Internal Control
Channel F3/Back-F1, Channel 705/Back 530
or Cy 5 / Cy 5.5 (498-
660)
| Result Interpretation
---|---|---
Positive| Positive| Positive
Negative| Positive| Negative
Positive| Negative| Positive
Negative| Negative| Invalid
Notes:
For LightCycler 480 II System: Use the “High Sensitivity“ setting of the
LightCycler Software to calculate results. For LightCycler 2.0 and LightCycler
480 II System: Always verify the software results (“positive”, “negative”,
“uncertain”) for plausibility by inspection of the amplification curves.
Troubleshooting
Observation
| Possible Cause|
Recommendation
---|---|---
No signal increase is observed, even with
positive controls.
| An incorrect detection channel has been chosen.| · Set Channel Settings to F2/Back-F1 [640/Back 530, Red 640 (498-640)] or F3/Back-F1 [705/Back 530, Cy 5 / Cy 5.5 (498-660)].
· For the carousel-based systems, fluorescence data is acquired for all channels during the run, regardless of the channel settings. If the incorrect channel is selected, there is NO need to abort and redo the run.
Pipetting errors or omitted reagents.| · Check for the correct pipetting scheme and reaction setup. Repeat the PCR run.
· Always run a positive control along with your samples.
No data acquisition programmed.| · Check the cycle programs.
· Select acquisition mode “single” at the end of each annealing segment of the PCR program.
No signal increase in Channel F3/ Back-F1 [705/Back 530,
Cy 5 / Cy 5.5
(498-660)] is
observed.
| Inhibitory effects of the sample material (e.g., caused by insufficient purification).| · Use the recommended DNA sample preparation kit to purify template DNA.
· Dilute samples or pipet a lower amount of sample DNA (e.g., 2.5 µL instead of 5 µL).
· Perform a sub-cultivation of the enrichment culture (e.g., 1:10 in BPW) to dilute the portion of food matrix in the sample.
Fluorescence intensity is too low.
| Inappropriate storage of kit components.| · Store the foodproof E. coli and Shigella Master Mix (vial 1, yellow cap) at -15 °C to -25°C, protected from light.
· Avoid repeated freezing and thawing.
foodproof E. coli and Shigella Master Mix (vial 1, yellow cap) are not
homogeneously mixed.| · Mix the foodproof E. coli and Shigella Master Mix
(vial 1, yellow cap) thoroughly before pipetting.
The low initial amount of target DNA.| · Increase the amount of sample DNA.
Depending on the chosen DNA isolation method, inhibitory effects may occur.
Negative control samples are positive.| Carry-over contamination.| ·
Exchange all critical solutions.
· Repeat the complete experiment with fresh aliquots of all reagents.
· Always handle samples, kit components and consumables by commonly accepted practices to prevent carry-over contamination.
· Add positive controls after sample capillaries and negative control capillaries have been sealed with stoppers.
Fluorescence intensity varies.| Insufficient centrifugation of the
capillaries or PCR plate.
| · Always centrifuge capillaries or PCR plates (loaded with the PCR mix) as
described.
---|---|---
The outer surface of the capillary tip or the sealing foil is dirty (e.g., by
direct skin contact).| · Always wear gloves when handling the capillaries or
the sealing foil.
Additional Information on this Product
How this Product Works
The foodproof Escherichia coli and Shigella Detection Kit provides primers and
Hybridization Probes (for sequence-specific detection), convenient premixed
reagents, and a control template for reliable interpretations of results. To
ensure maximum reliability of the kit and to prevent misinterpretation of
negative results due to inhibition of the amplification, an Internal Control
(IC) is supplied with the kit (vial 3, white cap). The IC has to be added to
each reaction. Hybridization Probes were designed to bind specifically the IC,
allowing detection in channel F3 (LightCycler Software 3.5 and versions
below), 705 (LightCycler Software 4. x) or Cy 5 / Cy 5.5 (498-660)
(LightCycler 480 Software 1.5), whereas the E. coli and Shigella DNA is
detected in channel F2 (LightCycler Software 3.5 and versions below), 640
(LightCycler Software 4. x) or Red 640 (498-640) (LightCycler 480 Software
1.5). In the case of a negative result due to inhibition of the amplification
by the sample DNA of interest, the amplification of the IC is suppressed as
well. Therefore, a negative result for the sample DNA of interest and
amplification of the IC indicates the absence of E. coli and Shigella DNA in
the sample. The foodproof Escherichia coli and Shigella Detection Kit
minimizes contamination risk and contains all reagents (except for template
DNA) needed for the detection of E. coli and Shigella DNA. The kit is
specifically adapted for PCR using the LightCycler System. Primers and
Hybridization Probes provide specific detection of E. coli and Shigella DNA in
food preparations. The kit described in this Instruction Manual has been
developed for the LightCycler System.
Test Principle
- Using the kit’s supplied sequence-specific primers in a polymerase chain reaction (PCR), the LightCycler Carousel-Based System and its associated reagents amplify and simultaneously detect fragments of E. coli and Shigella genomic DNA.
- The LightCycler Carousel-Based System detects these amplified fragments in real-time through fluorescence generated by their corresponding pair of sequence-specific Hybridization Probes. For each amplicon, one probe is labelled at the 5´-end with an acceptor fluorophore and, to avoid extension, is modified at the 3´-end by phosphorylation. The other oligonucleotide probe is labelled at the 3´-end with a donor fluorophore.
- During the annealing phase of each PCR cycle, these probes hybridize to an internal amplicon sequence. Only while hybridized near each other do these probes result in fluorescence resonance energy transfer (FRET) between the two fluorophores. During FRET, the light source of the LightCycler Carousel-Based System excites the donor fluorophore and part of the excitation energy is transferred to the acceptor fluorophore.
- The LightCycler Instrument measures the emitted fluorescence of the acceptor fluorophore.
Prevention of Carry-Over Contamination
The heat-labile Uracil-DNA Glycosylase (UNG) is suitable for preventing carry-
over contamination between PCRs. This technique relies on the incorporation of
deoxyuridine triphosphate (dUTP) during all amplification reactions, and the
pretreatment of all successive PCR mixtures with the heat-labile UNG. The UNG
cleaves DNA at any site where a deoxyuridine residue has been incorporated.
The resulting abasic sites are hydrolyzed due to the high temperatures during
the initial denaturation step, and can no longer serve as PCR templates. The
heat-labile UNG is inactivated during the initial denaturation step. Native
DNA (e.g., the isolated E. coli and Shigella genomic DNA) does not contain
uracil and is therefore not degraded by this procedure. Since dTTP is replaced
with dUTP and UNG is included in the foodproof Escherichia coli and Shigella
Detection Kit, decontamination can be achieved with the provided reagents.
Quality Control
The foodproof Escherichia coli and Shigella Detection Kit is function-tested
using the LightCycler Carousel-Based System.
Supplementary Information
Ordering Information
Hygiena Diagnostics offers a broad range of reagents and services. For a
complete product overview and more information, please visit our website at
www.hygiena.com.
License Notice
The purchase price of this product includes limited, nontransferable rights
under U.S. Patent No. 7,687,247 owned by Life Technologies Corporation to use
only this amount of the product to practice the claims in said patent solely
for activities of the purchaser for bioburden testing, environmental testing,
food testing, or testing for genetically modified organisms (GMO) by the
instructions for use accompanying this product. No other rights are conveyed,
including no right to use this product for in vitro diagnostic, therapeutic,
or prophylactic purposes. Further information on purchasing licenses under the
above patent may be obtained by contacting the Licensing Department, Life
Technologies Corporation, 5791 Van Allen Way, Carlsbad, CA 92008. Email:
outlicensing@lifetech.com.
Trademarks
foodproof®, microproof®, vetproof®, ShortPrep®, StarPrep®, RoboPrep® and
LyoKit® are registered trademarks of Hygiena Diagnostics GmbH. Hygiena® is a
registered trademark of Hygiena. Other brand or product names are trademarks
of their respective holders.
Contact and Support
If you have questions or experience problems with this or any other product of
Hygiena Diagnostics GmbH, please contact our Technical Support staff
(www.hygiena.com/support). Our scientists
commit themselves to providing rapid and effective help. We also want you to
contact us if you have suggestions for enhancing our product performance or
using our products in new or specialized ways. Such customer information has
repeatedly proven invaluable to us and the worldwide research community.
Reference Number
The reference number and original Hygiena Diagnostics GmbH article number: R
300 09
Change Index
-
Version 1, August 2009
The first version of the package insert. -
Version 2; March 2017
License Notice changed. -
Version 3; September 2017
License Notice changed. -
Revision A, January 2024
- Rebranding and new layout.
- R 300 09 20 -> INS-KIT230034-RevA
Contact Information
Hygiena®
- Camarillo, CA 93012 USA
- diagnostics.support@hygiena.com
Manufactured by Hygiena Diagnostics GmbH
- Hermannswerder 17 14473 Potsdam Germany
- www.hygiena.com.
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>