Features

2 pole EURO plug , Class Ⅱ power unit

  • Medical safety approved (2 × MOPP) according to EN60601-1/-1-11
  • Extremely low leakage current
  • No load power consumption<0.1W
  • Energy efficiency Level
  • Ⅵ and meet CoC version 5
  • ( except 5~9V for Level Ⅴ)

Applications

  • Blood glucose meter
  • Blood pressure meter
  • Nebulizer
  • Inhaler
  • Portable medical device

GTIN CODE

MW Search: https://www.meanwell.com/serviceGTIN.aspx

Description

GSM18E is a highly reliable, 18W wall-mounted style single-output green medical adaptor series. This product is equipped with a 2-pin (no FG) standard European power plug, adopting the input range from 80VAC to 264VAC. The entire series supplies different output voltages between 5VDC and 48VDC that can satisfy the demands for various kinds of miniature medical devices. The circuitry design meets the international medical standards (2 × MOPP), having an ultra low leakage current (<50μA), fitting the medical devices in direct electrical contact with the patients.
With the efficiency up to 88% and the extreme low no-load power consumption below 0.1W, GSM18E is compliant with EU ErP and meet Code of Conduct(CoC) Version 5; the supreme feature allows the adaptor to save the energy when it is either under the operating mode or the standby mode. The entire series utilizes the 94V-0 flame retardant plastic case, providing the double insulation that effectively prevents electrical shock. GSM18E is approved with the international medical safety certificates.

Model Encoding

SPECIFICATION

ORDER NO.| GSM
18E05-
P1J| GSM1
8E07
P1J| GSM1
8E09
P1J| GSM1
8E1
2-
P1J| GSM1
8E15
P1J| GSM1
8E1
8-
P1J| GSM1
8E2
4-
P1J| GSM
18E48-
P1J
---|---|---|---|---|---|---|---|---







OUTPUT

| SAFETY MODEL NO.| GSM18E05| GSM18E07| GSM18E09| GSM18E12| GSM18E15| GSM18E18| GSM18E24| GSM18E48
DC VOLTAGE **Note.2| 5V| 7.5V| 9V| 12V| 15V| 18V| 24V| 48V
RATED CURRENT| 3A| 2A| 2A| 1.5A| 1.2A| 1A| 0.75A| 0.375A
CURRENT RANGE| 0 ~ 3A| 0 ~ 2A| 0 ~ 2A| 0 ~ 1.5A| 0 ~ 1.2A| 0 ~ 1A| 0 ~ 0.75A| 0 ~ 0.375A
RATED POWER (max.)| 15W| 15W| 18W| 18W| 18W| 18W| 18W| 18W
RIPPLE & NOISE (max.) Note.3| 60mVp-p| 80mVp-p| 80mVp-p| 120mVp-p| 120mVp-p| 150mVp-p| 180mVp-p| 240mVp-p
VOLTAGE TOLERANCE Note.4| ±5.0%| ±5.0%| ±5.0%| ±3.0%| ±3.0%| ±3.0%| ±2.0%| ±2.0%
LINE REGULATION ****Note.5| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%| ±1.0%
LOAD REGULATION| ±5.0%| ±5.0%| ±5.0%| ±3.0%| ±3.0%| ±3.0%| ±2.0%| ±2.0%
SETUP, RISE TIME ****Note.6| 500ms, 30ms / 230VAC 1000ms, 30ms / 115VAC at full load
HOLD UP TIME (Typ.)**| 16ms / 230VAC         16ms / 115VAC at full load




INPUT

| VOLTAGE RANGE **Note.7| 80 ~ 264VAC  113 ~ 370VDC
FREQUENCY RANGE| 47 ~ 63Hz
EFFICIENCY (Typ.)| 80%| 83%| 84%| 85%| 85.5%| 86%| 87%| 88%
AC CURRENT (Typ.)| 0.5A / 115VAC         0.25A / 230VAC
INRUSH CURRENT (Typ.)| 55A / 230VAC          30A / 115VAC
LEAKAGE CURRENT(max.)**| Touch current < 50μA/264VAC



PROTECTION

| OVERLOAD| 105 ~ 170% rated output power
Protection type : Hiccup mode, recovers automatically after fault condition is removed
OVER VOLTAGE| 5.25 ~ 7.5V| 7.88 ~ 10.13V| 9.45 ~ 12.6V| 12.6 ~ 17.2V| 15.75 ~ 20.25V| 18.9 ~ 24.3V| 25.2 ~ 32.4V| 50.4 ~ 64.8V
Protection type : Shut down o/p voltage, re-power on to recover



ENVIRONMENT

| WORKING TEMP.| -25 ~ +60℃ (Refer to “Derating Curve”)
WORKING HUMIDITY| 20% ~ 90% RH non-condensing
STORAGE TEMP., HUMIDITY| -40 ~ +85℃, 10 ~ 95% RH non-condensing
TEMP. COEFFICIENT| ±0.03% / ℃ (0~40℃)
VIBRATION| 10 ~ 500Hz, 2G 10min./1cycle, period for 60min. each along X, Y, Z axes











SAFETY & EMC

(Note. 8)

| SAFETY STANDARDS| TUV EN60601-1 / EN60601-1-11, EAC TP TC 004 approved
ISOLATION LEVEL| Primary-Secondary: 2xMOPP
WITHSTAND VOLTAGE| I/P-O/P:4KVAC
ISOLATION RESISTANCE| I/P-O/P:100M Ohms / 500VDC / 25℃/ 70% RH




EMC EMISSION

| Parameter| Standard| Test Level / Note
Conducted emission| EN55011 (CISPR11)| Class B
Radiated emission| EN55011 (CISPR11)| Class B
Harmonic current| EN61000-3-2| Class A
Voltage flicker| EN61000-3-3| —–








EMC IMMUNITY

| EN60601-1-2, EN61204-3
Parameter| Standard| Test Level / Note
ESD| EN61000-4-2| Level 4, 15KV air ; Level 4, 8KV contact
RF field susceptibility| EN61000-4-3| Level 3, 10V/m( 80MHz~2.7GHz ) Table 9, 9~28V/m( 385MHz~5.78GHz )
EFT bursts| EN61000-4-4| Level 3, 2KV
Surge susceptibility| EN61000-4-5| Level 3, 1KV/Line-Line
Conducted susceptibility| EN61000-4-6| Level 3, 10V
Magnetic field immunity| EN61000-4-8| Level 4, 30A/m
Voltage dip, interruption| EN61000-4-11| 100% dip 1 periods, 30% dip 25 periods,

100% interruptions 250 periods


OTHERS

| MTBF| 3888.9K hrs min.      Telcordia SR-332 (Bellcore) ; 794.6K hrs min.        MIL-HDBK-217F (25℃)
DIMENSION| 795433mm (LWH)
PACKING| 200g ; 60pcs / 13Kg / CARTON
CONNECTOR| PLUG| See page 4~5 ; Other type available by customer requested
CABLE| See page 4~5 ; Other type available by customer requested





NOTE

| 1. All parameters are specified at 230VAC input, rated load, 25℃ 70% RH ambient.

2. DC voltage: The output voltage set at point measure by plug terminal & 50% load.

3. Ripple & noise are measured at 20MHz by using a 12″ twisted pair terminated with a 0.1μF & 47μF capacitor.

4. Tolerance: includes set up tolerance, line regulation, load regulation.

5. Line regulation is measured from low line to high line at rated load.

6. Length of set up time is measured at first cold start. Turning ON/OFF the power supply may lead to increase of the set up time.

7. Derating may be needed under low input voltage. Please check the derating curve for more details.

8. The power supply is considered as an independent unit, but the final equipment still need to re-confirm that the whole system complies with the EMC directives. For guidance on how to perform these EMC tests, please refer to “EMI testing of component power supplies.”

(as available on https://www.meanwell.com//Upload/PDF/EMI_statement_en.pdf )

※ Product Liability Disclaimer:For detailed information, please refer to https://www.meanwell.com/serviceDisclaimer.aspx

Derating Curve

Static Characteristics

Mechanical Specification

DC output plug

Standard plug: P1J

DC plug changeable through:

  • Customization of the standard part with an optional DC plug according to the table (MOQ applicable)
  • Quick adapter accessory (sold separately without MOQ)
  • Please refer to below table and online selection guide : https://www.meanwell.com/upload/pdf/DC_plug.pdf

Example quick adapter accessory


Installation Manual

Please refer to : http://www.meanwell.com/manual.html

Read More About This Manual & Download PDF:

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

MEAN WELL User Manuals

Related Manuals