APERA LabSen 851-1 Viscous pH Electrode Electrochemical Sensors User Manual
- June 13, 2024
- APERA
Table of Contents
LabSen 851-1 Viscous pH Electrode Electrochemical Sensors
User Manual LabSen 851-1 Viscous pH Electrode
User Manual
LabSen 851-1 Viscous pH Electrode Electrochemical Sensors
LabSen electrochemical sensors are premium pH electrode with manufacturing
technology and key components imported from Switzerland. LabSen851-1 is
suitable for cosmetics and viscous samples.
This probe has following features:
- Impact-resist membrane (see the right picture), there is no danger of electrode breakage during normal use.
- Electrode structure, which is suitable for viscous samples: conical low resistance membrane, Protelyte reference solution, and junction of 3 ceramic cores.
- Blue gel inner solution, does not flow and will not cause bubble.
- Silver ion trap reference prevents the contamination of the junction when testing sample contains sulfides and proteins.
Technical Data
Measuring Range | (0 — 14) pH | Electrolyte | Protelyte |
---|---|---|---|
Temperature Range | (0 — 100)°C | Soaking Solution | Protelyte |
Shaft Material | Lead-free Glass | Electrode Dimension | (012×120) mm |
Membrane Shape | Conical | Connector | BNC |
Reference | Silver Ion Trap | Cable | 03x1m |
Junction | Ceramic x 3 |
Usage and Maintenance
2.1 When measuring, please unscrew the bottle cap, pull out the electrode and rinse it with deionized water. After using, please put the electrode back into the bottle and screw tight the cap.
Before pulling out or putting back the electrode, make sure to fully loosen the blue cap on the storage sleeve so that the electrode can move in and out smoothly.
2.2 Prior to measurement, remove the rubber plug to maintain pressure of the
reference solution, keep consistent flow rate of reference solution and stable
potentials of junction.
2.3 After a period of usage, the reference solution will running low. Whenever
the level falls below 1/2 height of the electrode, add Protelyte solution to
the refilling hole by using a syringe or pipette.
2.4 The connector of the electrode should be clean and dry. If being
contaminated, please clean it with medical cotton and absolute alcohol and
blow dry to prevent the short circuit of the electrode and slow reaction of
electrode.
2.5 The electrode’s measuring tip should be soaked in the soaking bottle
containing Protelyte solution to keep the membrane hydrated and junction
unblocked. Clean the bottle and replace the storage solution if the storage
solution gets turbid and mildewed. The electrode should never be stored in
pure water or buffer solution for long.
2.6 After 1-year of use, we recommend replacing the electrode to ensure the
best accuracy.
Warranty
We warrant this electrode free from defects in material and workmanship and
agrees to repair or replace free of charge, at option of APERA INSTRUMENTS any
malfunctioned or damaged product attributable to responsibility of APERA
INSTRUMENTS for a period of six months. Warranty period is the time limit to
provide free service for the products purchased by customers, not the service
life of the tester or electrodes.
This limited warranty does not cover any damages due to: transportation;
storage; improper use; failure to follow the product instructions or to
perform any preventive maintenance; modifications; combination or use with any
products, materials, processes, systems or other matter not provided or
authorized in writing by us; unauthorized repair; normal wear and tear; or
external causes such as accidents, abuse, or other actions or events beyond
our reasonable control.
APERA INSTRUMENTS (Europe) GmbH
Address: Wilhelm-Muthmann-Str.18,
42329 Wuppertal, Germany
Phone: +49 202 51988998
Website: www.aperainst.de