PeakTech 6060 Regulated Double Laboratory Power Supply User Manual
- June 4, 2024
- PeakTech
Table of Contents
6060 Regulated Double Laboratory Power Supply
User Manual
Safety Precautions
This plies with the requirements of the following directives of the European
Union for CE conformity: 2014/30/EU (electromagnetic compatibility),
2014/35/EU (low voltage), and 2011/65/EU (RoHS).
To ensure the safe operation of the equipment and eliminate the danger of
serious injury due to short circuits (arcing), the following safety
precautions must be observed. Damages resulting from failure to observe these
safety precautions are exempt from any legal claims.
-
Prior to connecting of the equipment to the mains outlet, check that the available mains voltage corresponds to the voltage setting of the equipment.
-
Connect the mains plug of the equipment only to a mains outlet with an earth connection.
-
Do not place the equipment on damp or wet surfaces.
-
Do not subject the equipment to direct sunlight or extreme temperatures, humidity or dampness.
-
Replace a defective fuse only with a fuse of the original rating. Never shortcircuit fuse or fuse holding.
-
Do not exceed the maximum permissible input ratings.
-
To avoid electric shock, do not operate this product in wet or damp conditions.
Conduct measuring works only in dry clothing and rubber shoes, i. e. on isolating mats. -
Comply with the warning labels and other info on the equipment.
-
Do not cover the ventilation slots of the cabinet to ensure that air is able to circulate freely inside.
-
Do not insert metal objects into the equipment by the way of ventilation slots
-
Do not place water-filled containers on the equipment (danger of short-circuit in case of knock over of the container)
-
Do not operate the equipment near strong magnetic fields (motors, transformers, etc.).
-
Do not subject the equipment to shocks or strong vibrations.
-
Allow the equipment to stabilize at room temperature before taking up measurements (important for exact measurements).
-
Do not modify the equipment in any way
-
Do not operate the meter before the cabinet has been closed and screwed safely as the terminal can carry voltage.
-
Periodically wipe the cabinet with a damp cloth and mid-detergent. Do not use abrasives or solvents.
-
The meter is suitable for indoor use only
-
Do not store the meter in a place of explosive, inflammable substances.
-
Opening the equipment and service – and repair work must only be performed by qualified service personnel
-
Measuring instruments don’t belong to children’s hands.
Cleaning the cabinet
Prior to cleaning the cabinet, withdraw the mains plug from the power outlet.
Clean only with a damp, soft cloth and a commercially available mild household
cleanser. Ensure that no water gets inside the equipment to prevent possible
shorts and damage to the equipment.
Introduction
The model PeakTech
6060 high-precision DC regulated power supply has two ways adjustable output.
The two adjustable outputs can also be selected for constant voltage or
constant currant, designed in high stability and performance circuit. In a
constant voltage state, the output voltage can be arbitrarily from 0 V on the
nominal range; and in the state of constant current, the output current can be
adjustable from 0 A on the nominal range. The two outputs can be connected in
parallel or in series, while the master used for voltage or current
adjustment. The maximum output voltage in series is double of independent’s,
and the maximum output current in parallel is double too.
There are volt and Amp meters (or 3-digit LCD) for indicating each of the two
outputs with high accuracy. The one fixed way output 5 V voltage. Due to the
single chip integrated regulator, this output has good stability and ripple
factor and has reliable overload protection to protect the unit against being
damaged whenever overload or short circuit.
The units features in small size, good performance, a novel appearance and
etc, it is the ideal power supply unit for science investigation, college,
factory, electronic appliance maintenance and etc. This power source can be
fitted into 19″-racks and can be used as a “stand-alone”-unit.
Permanent operations
This model of the power supply provides 2 adjustable outputs for a maximum
current of 10 A, which can be connected in series or parallel, which means a
maximum output of 20 A DC in parallel.
This performance is designed only for short time applications and not for
permanent operations.
For longer life, it is recommended to use up to 75% of the maximum values for
permanent operation.
Technical Data
Output voltage: | 2 x 0 … 30 V DC (adjustable) |
---|---|
Output current: | 2 x 0 … 10 A DC (adjustable) |
Source regulation: | Two adjustable outputs: |
1 x 10-4 + 3 mV
2 x 10-3 + 3 mA
Fixed output:| 10 mV
Load regulation:| Two adjustable outputs:
£ 1 x 10-4 + 2 mV (I £ 3 A)
£ 1 x 10-4 + 5 mV (I > 3 A)
£ 2 x 10-3 + 3 mA (I £ 3 A)
£ 2 x 10-3 + 5 mA (I > 3 A)
Fixed output:| 10 mV
Ripple and noise:
Two adjustable outputs:| CV £ 0,5 mV rms (I £ 3 A)
CV £ 1,0 mV rms (I > 3 A)
CC < 3 mA rms
Fixed output:| 10 mV
Protection:| current limit
Indication accuracy:| Volt-indication: LED ± 0,2% + 2 digits
Amp-Indication: LED ± 1% + 2 digits
Fixed output:| 10 mV
Protection:| current limit
Indication accuracy:| Volt-indication: LED ± 0,2% + 2 digits
Amp-Indication: LED ± 1% + 2 digits
Mains voltage:| 115 V AC; 60 Hz/230 V AC; 50 Hz (switchable)
Fuse:| T 8, 15 A
Dimensions:| W x H x D 482 x 140 x 430 mm
Dimensions Installation:| W x H x D 425 x 140 x 400 mm
Weight:| 18 kg
Accessories:| Operation manual, power cord, 2 x fuse
Operation
3.1. Controls and description of front-panel
- Slave constant-current indicator or two-way parallel state indication: the LED illuminates when the slave output is in the current-regulated state or the two adjustable outputs are in parallel.
- Slave constant current adjustment: adjusting slave output current value (adjusting the current-limit protection point).
- Slave constant-voltage indicator: the LED illuminates when the slave output is in a voltage-regulated state.
- Slave constant voltage adjustment: adjusting the slave output voltage.
- Amp display: indicating slave output current by analog meter or LCD.
- Volt display: indicating slave output voltage by analog meter or LCD.
- Function switch: for selecting single-use of both adjustable outputs, press the “Normal”-button
- Function switch: for selecting series-use of both adjustable outputs, press the “Series”-button
- Function switch: for selecting single-use of both adjustable outputs, press the “Parallel”-button
- Volt display: indicating master output voltage by analog meter or LCD.
- Master constant-current indicator: the LED illuminates when the master output is in a current-regulated state.
- Mater constant-voltage indicator: the LED illuminates when the master out- put is in a voltage-regulated state.
- Power switch: the unit is “ON” when this button switch is depressed, while CV LED (3) (12) or CC LED (1) (11) illuminating.
- Master constant voltage adjustment: adjusting the master output voltage.
- Fixed 5V output Terminal (+): connecting the positive terminal of the load.
- Fixed 5V output Terminal (-): connecting the negative terminal of the load.
- Master constant current adjustment: adjusting master output current value (adjusting the current-limit protection point).
- Master output Terminal (+): connecting the positive terminal load.
- Case ground: connecting the case to the ground.
- The master output terminal (-): connects the negative terminal of the load.
- The slave output terminal (+): connecting the positive terminal of the load.
- Case ground: connecting the case to the ground.
- Slave output terminal (-): connecting the negative terminal of the load.
- AMP display: indicating master output current by LED
- Output ON/OFF switch: output off, when pressing the button
- Fuse holder
- Case ground: connecting the case to ground
- Cooling Fan
- Main voltage input socket
- Control switch: to choose the input voltage 110 V or 220 V
3.2. Operating method
Independence use of two adjustable outputs.
- Press switch (7).
- When the adjustable output is used as CV output, first should rotate clockwise the CC adjustment (2) and (17) to maximum, then turn on the power switch (13), adjust CV adjustment (4) and (14) till the output voltage reach required voltage value, at this time, the CC state indicator (1) and (11) go out the CV state indicator (3) and (12) light on.
- Used as CC output, after turning on the power switch (13), first rotate clockwise the CV adjustment (4) and (14) to maximum, while rotating counter-clockwise the CC adjustment (2) and (17) to a minimum, connect the required load, again adjust clockwise adjustment (2) and (17) till output current reach the required current value. At this time, the CV state indicator (3) and (12) go out and the CC state indicator (1) and (11) light on.
- Used as the CV output, in general, the CC adjustment (2) and (17) should be set to maximum, but for this unit, the current-limiting protection point can also be set arbitrarily.
Setting procedure: turn on the power, rotate counter-clockwise the CC adjustment (2) and (17) to a minimum, then make the positive and negative output terminal in short connection and rotate clockwise the CC adjustment (2) and (17) till output current equal to the required current-limiting protection point, so the current-limiting protection point is well set.
Series using of the two adjustable outputs.
-
Press the switch (8). At this time, turn the master voltage adjustment (14) and the slave out voltage tracks strictly the master output voltage, and the output voltage can be up to double of independent´s maximum voltage (voltage between terminal (18) and (23).
-
Before the series connecting, it must be examined if the negative terminal of both master and slave output is connected to the case grounded terminal, if they are, must be disconnected, otherwise, short-circuit will be caused in the slave output when the two outputs are connected in series.
-
When the two outputs are in series, the voltage is controlled by the master output, but the current adjustment to the two outputs is still independent. Therefore, attention should be paid to the position of the CC adjustment (2) and (17). For example, knob (2) is at the position counterclockwise to the end or current of slave output excesses current limiting protection point, at this time, the voltage of slave output will not track the voltage of the master.
So knob (2) should be rotated clockwise to maximum then the two outputs are in series. -
This unit can be output from the front panel and rear panel at the same time.
When use the unit as bench type, it is better for you to use the front panel outputs, and when you use the unit as rack mount type, it is better for you to use the rear panel outputs. Of course, you can use both front panel outputs and rear panel outputs at the same time.
The output current should be the summation of both front-panel outputs and rear-panel outputs.
Parallel use of the two adjustable outputs.
- Press the switch (9).
- At this time, the two outputs are in parallel, adjust voltage adjustment (14) of the master output is, the voltage of the two ways keeps the same, and the slave output CC indicator (1) lights on.
- When the two outputs are in parallel, the CC adjustment (2) of slave output does not work. When used as CC supply, simply adjust the CC adjustment (17) of master output, at this time, the output current of both master and slave output is controlled by it and are the same, output current is up to double (20 A) of independent’s maximum current.
- While the two outputs in parallel, since the unit has relay switch control, the two positive terminals and the two negative terminals of the master, and slave output and shorted reliably and separately by the relay inside the unit.
You can connect the load directly between terminals (18) and (20), (21) and (23), (18) and (23). For output current balance, connecting the load between terminals (18) and (23) is recommended.
The LED displays are in three digits.
Caution
- This unit has excellent protection function, 5 V output has reliable protection for current-limit and short. The two adjustable outputs have current-limit protection. As there is a controlling circuit for regulating the transistor’s power loss in the circuit, when short-circuit occurs, the power loss on large power transistors is not very high, it can’t cause any damage to the unit. But there is still power loss when short-circuit, in order to reduce aging and energy consumption, so this situation should be found as soon as possible and turn off the power, then exclude the faults.
- When operating is finished, put it in a dry place of good ventilation, and keep it clean. If it is not in use for a long period, pull off the power supply plug for storage.
- For maintenance, the input voltage must be cut off.
All rights, also for translation, reprinting, and copy of this manual or parts
are reserved. Reproductions of all kinds (photocopy, microfilm or other) only
by written permission of the publisher.
This manual is according to the latest technical knowledge. Technical
alterations reserved.
Misprints and errors are reserved.
We herewith confirm that the units are calibrated by the factory according to
the specifications as per the technical specifications.
We recommend calibrating the unit again, after 1 year.
© PeakTech® 06/2022 AW./Ehr./ Lie.
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>