PeakTech 6070 Laboratory Power Supply User Manual
- June 3, 2024
- PeakTech
Table of Contents
Unser Wert ist messbar… PeakTech® 6070
Bedienungsanleitung / Operation Manual
Stabilisiertes Labornetzgerat / Laboratory Power Supply
Safety Precautions
This product complies with the requirements of the following directives of the
European Union for CE conformity: 2014/30/EU (electromagnetic compatibility),
2014/35/EU (low voltage), 2011/65/EU (RoHS).
To ensure safe operation of the equipment and eliminate the danger of serious
injury due to short circuits (arcing), the following safety precautions must
be observed.
Damages resulting from failure to observe these safety precautions are exempt
from any legal claims whatever.
- Prior to connection of the equipment to the mains outlet, check that the available mains voltage corresponds to the voltage setting of the equipment.
- Connect the mains plug of the equipment only to a mains outlet with earth connection.
- Do not practice device unattended.
- The instrument must be set up so that the power plug can be removed from the socket easily.
- do not place the equipment on damp or wet surfaces.
- do not subject the equipment to direct sunlight or extreme temperatures.
- do not subject the equipment to extreme humidity or dampness
- replace a defective fuse only with a fuse of the original rating. Never short circuit fuse or fuse housing
- do not exceed the maximum permissible input rating.
- conduct measuring works only in dry clothing and in rubber shoes, i. e. on isolating mats.
- comply with the warning labels and other info on the equipment
- do not cover the ventilation slots of the cabinet to ensure that air is able to circulate freely inside.
- do not insert metal objects into the equipment by way of the ventilation slots
- do not place water-filled containers on the equipment (danger of short-circuit in case of knock over of the container)
- do not operate the equipment near strong magnetic fields (motors, transformer etc.)
- do not subject the equipment to shocks or strong vibrations
- keep hot soldering iron or guns away from the equipment
- allow the equipment to stabilize at room temperature before taking up measurement (important for exact measurements)
- do not modify the equipment in any way
- do not place the equipment face-down on any table or work bench to prevent damaging the controls at the front.
- Do not operate the meter before the cabinet has been closed and screwed safely as terminal can carry voltage.
- Periodically wipe the cabinet with a damp cloth and mid detergent. Do not use abrasives or solvents.
- The meter is suitable for indoor use only.
- Do not store the meter in a place of explosive, inflammable substance.
- Opening the equipment and any service- and repair work must be performed by qualified service personal. Repair work should be performed in the presence of a second person trained to administer first aid, if needed.
- – Measuring instruments do not belong to children hands-
Cleaning the cabinet
Prior to cleaning the cabinet, withdraw the mains plug from the power outlet.
Clean only with a damp, soft cloth and a commercially available mild household
cleanser. Ensure that no water gets inside the equipment to prevent possible
shorts and damage to the equipment.
Introduction
Digitally controlled, high-precision DC power supply with variable voltage and
Current setting. An advanced microprocessor controls the generation, display,
control and protection of voltage, current.
This technology improves the accuracy of the generation and control, it also
facilitates the operation and the reading of the values.
Henceforth should be emphasized that the voltage can be fixed over the whole
range, making the convenience is increasing and unnecessary, inaccurate, steps
can be prevented.
Caution!
The PeakTech 6070 provides a keylock function. Please read chapter 5.3 on page
15 for more detailed information.
Caution!
Laboratory Power Supplies are not designed for charging batteries. Any use of
this type can cause serious damage to the device, which are exempt from any
legal claims whatever.
Technical Data
Input voltage: | 115/230 V AC; 50/60 Hz switchable |
---|---|
Output voltage: | 0 – 30 V adjustable |
Output current: | 0 – 5 A adjustable |
Line regulation: (0-100% load) | CV 5 0,01% + 1mV |
CC 5 0,02% + 1mA
Load regulation: (0-100% load)| CV 5 0,01% + 3mV
CC 5 0,02% + 3mA
Ripple & Noise: (100% load)| CV 5 0.5mVrms CC 5 1mArms
Protection:| Constant current and short circuit protection
Voltage indication accuracy:| ± 0,5% + 5 digits
Current indication accuracy:| ± 1,0% + 5 digits
Operation temperature:| 0°C … +40°C; < 80% RH
Dimensions (W x H x D):| 135 x 165 x 275 mm
Weight:| 5,5 kg
Accessories:| Power cable and manual
Controls and description of front-panel
1 | LED Output current indicator |
---|---|
2 | LED Output voltage indicator |
3 | Indicators for operation mode: |
– C.V.: Constant voltage indication
– C.C.: Constant current indication
– Output: Output is enabled and the set output voltage/current will be
outputted
4| Key pad for entry the output values of voltage and current
5| Decimal point
6| Power switch
7| (-) — Output socket
8| GND — Output socket
9| (+) — Output socket
10| Function keys
Output: Enable/Disable output
Del: Correction key during entry voltage — and current values with key pad
Enter: Confirmation key
11| Function key V (Voltage) and I (Current) V: Entry voltage value I: Entry
current value
Operation
5.1. Adjusting the output voltage
Direct entry
Proceed as described to adjust the output voltage:
- Switch on the device.
- Press the [V] key. The LED display indicates 00.00 and the first digit flashes.
- Press the appropriate number keys to enter the desired voltage value.
- If an incorrect entry be carried out, use the [Del] key (Correction key), the previous entry is rejected and you can repeat your entry.
- Pressing the [Enter] button will confirm and save the entry.
Example 1:
Setting 5.99V
- Press the [V]-key.
- Sequentially, press the number keys [0][5][9][9] and then [ENTER] to save the entry or …
- Press the [V]-key and then sequentially the number keys [5] [.] [9] [9] and then press the [ENTER] key to confirm and save the entry.
Example 2:
Setting 29.99 V
- Press the [V]-key.
- Sequentially, press the number keys [2][9][9][9] and then [ENTER] to save the entry or …
- Press the [V]-key and then sequentially the number keys [2] [9] [.] [9] [9] and then press the [ENTER] key to confirm and save the entry.
Entering the output voltage in 0.1V / 1V steps
Additionally, it is possible to change the output voltage using the numeric
keys 4 and 9 to change the voltage value in +/- 0.1V steps.
If the keys were pressed and hold, the output voltage will change the voltage
level continuously in 0.1V steps.
Use the number keys 5 and 0, the voltage value can change by +/- 1V.
If the keys were pressed and hold, the output voltage will change the voltage
level continuously in 1V steps.
5.2. Adjusting the output current
Direct entry
Proceed as described to adjust the output current:
- Switch on the device.
- Press the [I] key. The LED display indicates 0.000 and the first digit flashes.
- By pressing the appropriate number keys to enter the desired current value.
- If an incorrect entry be carried out, use the [Del] key (Correction key), the previous entry is rejected and you can repeat your entry.
- Pressing the [Enter] button will confirm and save the entry.
Example 1:
Setting 1.599 A
- Press the [I]-key.
- Sequentially press, the number keys [1][5][9][9] and then [ENTER] to save the entry or …
- Press the [I]-key and then sequentially the number keys [1] [.] [5] [9] [9] and then press the [ENTER] key to confirm and save the entry.
Example 2:
Setting 4.999 A
- Press the [I]-key.
- Sequentially press, the number keys [4][9][9][9] and then [ENTER] to save the entry or …
- Press the [I]-key and then sequentially the number keys [4] [.] [9] [9] [9] and then press the [ENTER] key to confirm and save the entry.
Entering the output current in 0.1A / 1A steps
Additionally, it is possible to change the output current using the numeric
keys 4 and 9 to change the current value in +/- 0.1A steps.
If the keys were pressed and hold, the output current will change the voltage
level continuously in 0.1A steps.
Use the number keys 5 and 0, the current value can change by +/- 1A.
If the keys were pressed and hold, the output current will change continuously
in 1A steps.
5.3. Additional functions
-
Output button
– Press [Output] to activate the output and outputting the set values.
– or press [Output] to disable the output. (Default setting: output is disabled after turning the unit on). -
Key lock
The device has a key lock which the accidentally change the output voltage – or prevents the output current.
– To activate the key lock, the [.] – Button for 3 seconds and hold. You will hear a confirmation tone.
– To deactivate the key lock again, – press and hold again for 3 seconds [.] until a confirmation tone is emitted. -
Correction key [Del] – If it comes to an incorrect entry, while entering the voltage – or current value and the [Enter] button is not pressed, then the actual entry could be rejected by pressing the [Del] key. The display switches back to the beginning of 0.000 (first digit flashes).
CAUTION
- In the event of a short circuit at the output the current will limit at the value set by the current controls, however the unit should be turned off and the short circuit removed before continuing use.
- The mains power must be switched off before servicing and servicing should be referred to a qualified person.
- The unit should be stored in a dry and well ventilated place and the power cord removed if storing for long periods.
- If the unit is not a long time in use, unplug the power connector from the device
6.0 Operation under Software
6.1 Driver Installation
– Virtual serial port driver installation
Choose and open the “PL2303 Driver for xxx” folder depending your Windows
system and double click the installation program.
After finishing installation, connect the power supply and the computer with USB cable and switch mains power on. Check the serial port number: enter the device manager as followed. Find the serial port. P1 shows for example that “COM4” is given the serial port number.![PeakTech 6070 Laboratory Power Supply
- fig 2](https://manuals.plus/wp-content/uploads/2022/12/PeakTech-6070 -Laboratory-Power-Supply-fig-2.png)
6.2 Communication software installation
Enter installation disk and double click the communication software
installation program and then follow the installation guide till finish
installation.
6.3 Software operation
Real-time data setup:Open the communication software and then choose your
connected model and serial port number, click the button.
Enter the value of current and voltage into the “OUTPUT SETTING” frame (see.P2) and then click the ISET/VSET to send the value to the DC power supply. For activating the output click on the button. The appearance of the button will change to and the output is switched on.
Real-time data display
Real-time voltage and current and the CC/CV condition will be indicated on
the “READING” frame (see P3).
Data graph display: The ordinate is Voltage/Current and the abscissa is the data collection points (see P4)
Data sheet check: Click the button and get the sheet as followed(P5)
Saving data
You can export the report list to an EXCEL readable file.
Click the button then enters a name and save.(P6)
The export to an EXCEL file was successful (P7)
Program Mode
Open the communication software , and choose the corresponding model and
serial port number (Shown in 6.3). Don’t click on the button.
Enter the voltage, current and time in the table (P8). Choose the value of
program steps you want to run and “single time” or “Repeat” mode.
Choose .Click the button to run the program and activate the output to
activate the output (P8).
Program mode
The program runs
To stop the program run, click on the button. The output stops at the last
program step. The output keeps power on.
To get power off, click on the button. The program keeps running. The output
is switched off.
All rights, also for translation, reprinting and copy of this manual or parts
are reserved.
Reproduction of all kinds (photocopy, microfilm or other) only by written
permission of the publisher.
This manual considers the latest technical knowing. Technical changing’s which
are in the interest of progress reserved.
We herewith confirm, that the units are calibrated by the factory according to
the specifications as per the technical specifications.
We recommend to calibrate the unit again, after one year.
© PeakTech® 09/2021 Po/We/Ehr.
PeakTech Prüf- und Messtechnik GmbH
– Gerstenstieg 4 – DE-22926 Ahrensburg / Germany
+49-(0) 4102-97398 80
+49-(0) 4102-97398 99
**** info@peaktech.de
****www.peaktech.de