PeakTech 3280 Resistance Decade Box User Manual

October 27, 2023
PeakTech

PeakTech 3280 Resistance Decade Box

Safety Precautions

This product complies with the requirements of the following directives of the European Union for CE conformity: 2014/35/EU (low voltage), 2011/65/EU (RoHS). Pollution degree 2.

To ensure safe operation of the equipment and eliminate the danger of serious injury due to short-circuits (arcing), the following safety precautions must be observed.

Damages resulting from failure to observe these safety precautions are exempt from any legal claims whatever.

  • Do not use this instrument for high-energy industrial installation measurement.
  • Do not exceed the maximum permissible input ratings (danger of serious injury and/or destruction of the equipment).
  • Check test leads and probes for faulty insulation or bare wires before connection to the equipment.
  • To avoid electric shock, do not operate this product in wet or damp conditions. Conduct measuring works only in dry clothing and rubber shoes, i. e. on isolating mats.
  • Comply with the warning labels and other info on the equipment
  • Do not subject the equipment to direct sunlight or extreme temperatures, humidity or dampness.
  • Do not subject the equipment to shocks or strong vibrations.
  • Do not operate the equipment near strong magnetic fields (motors, transformers etc.).
  • Keep hot soldering irons or guns away from the equipment.
  • Allow the equipment to stabilize at room temperature before taking up measurement (important for exact measurements).
  • Periodically wipe the cabinet with a damp cloth and mid detergent. Do not use abrasives or solvents.
  • The instrument is suitable for indoor use only
  • Do not store the meter in a place of explosive, inflammable substances.
  • Do not modify the equipment in any way
  • Do not place the equipment face-down on any table or work bench to prevent damaging the controls at the front.
  • Opening the equipment and service – and repair work must only be performed by qualified service personnel
  • Measuring instruments don’t belong to children hands.

Features

  • Applications:
  • General troubleshooting,
  • Maintenance Education and Vocational training
  • Production line testing
  • Radio and TV services
  • Working standards
  • Research design and develop
  • Physics laboratory work
  • Offering accurate, reliable performance
  • Slide switches that allow the user to simply add or subtract for desired value
  • Terminals with multi way binding posts, one to switch shield case
  • ABS plastic housing case, rugged components

PeakTech®3280

  • 1 to 11,111 MOhm wide range and high resolution (1 Ohm per step)
  • With seven decades of resistance

PeakTech®3285

  • 100 pF to 11,111 µF, wide range and high resolution (100 pF per step)
  • With five decades of capacitance

PeakTech® 3290

  • 10 µH to 111,1 mH, wide range and high resolution (10 µH per step)
  • With four decades of inductance

Specifications

Specifications P 3280
Range 1 to 11.111.110 Ohm (1 Ohm per step)
Accuracy 1 % resistors used throughout
Wattage 2 W
Max. Input Voltage 250 V AC/DC
Internal Stray Resistance Max. 0,3 Ohm
Operating Temperature 0°C to 50°C (32°F to 122°F) ,< 80 % RH
Weight 400 g
Dimensions (WxHxD) 140 x 190 x 80 mm
Accessories Operation Manual

Front Panel Description


3-1 Range Select Switch
3-2 Resistance Output Terminal

Testing Procedure

  1. Start with all switches up (0) for 0 ohms.
  2. Switch down (1) to add Resistance value

WARNING!!

  • Do not add the Wattage more than 2 W to the “Resistance Output Terminals” (3-2)
  • Do not add the voltage over AC/DC 250V to the “Output Resistance Terminals” even its stand Wattage is less than 2W.
Specifications P 3285
Range 100 pF to 11,111 µF (100 pF per step)
Accuracy 5 % capacitors used throughout < 1 µF, 1 kHz test frequency > 1 µF,

100 Hz test frequency
Voltage| 50 V DC, non-polarized capacitors
Internal Residual Capacitance| 50 pF max.
Operating Temperature| 0°C to 50°C (32°F to 122°F) < 80% RH
Weight| 350 g.
Dimensions (WxHxD)| 140 x 190 x 80 mm
Accessories| Operation Manual

Front Panel Description


3-1 Range Select Switch
3-2 Capacitance Output Terminal

Testing Procedure

  1. Start with all switches up (0) for min. capacitance.
  2. Switch down (1) to add Capacitance value.

WARNING!!
Do not add the voltage more than 50 V DC to the “Output Capacitor Terminals” (3-2)

Specifications P 3290
Range 10 µH to 111,1 mH (10 µH per step)
Accuracy 5 % inductors used throughout at 23°C +/- 5°C at 1 kHz test

frequency
Max. Rating Current| 100 mA DC or AC
Internal Residual Inductance| Approx. 0,5 µH
Operating Temperature| 0°C to 50°C (32°F to 122°F) < 80 % RH
Weight| 450 g.
Dimensions (WxHxD)| 140 x 190 x 80 mm
Accessories| Operation Manual

Front Panel Description


3-1 Range Select Switch
3-2 Inductance Output Terminal

Testing Procedure

  1. Start with all switches up (0) for min. inductance.
  2. Switch down (1) to add Inductance value.

WARNING!!
Do not add the current more than 100 mA (AC or DC) into the “Output Inductor Terminals” (3-2)

All rights, also for translation, reprinting and copy of this manual or parts are reserved.

Reproduction of all kinds (photocopy, microfilm or other) only by written permission of the publisher.

This manual considers the latest technical knowing. Technical changes which are in the interest of progress reserved.

We herewith confirm, that the units are calibrated by the factory according to the specifications as per the technical specifications.

We recommend to calibrate the unit again, after 1 year.

©PeakTech® 05-2021 MP/SM/EHR

PeakTech Prüf- und Messtechnik GmbH – Gerstenstieg 4 –
DE-22926 Ahrensburg / Germany
+49-(0) 4102-97398 80 +49-(0) 4102-97398 99
[email protected] www.peaktech.de