WINNCARE PA0010115 Viscoelastic Foam Mattress User Manual
- June 12, 2024
- WINNCARE
Table of Contents
WINNCARE PA0010115 Viscoelastic Foam Mattress User Manual
DEVICE INTENDED USE
Viscoelastic foam static mattress, called “memory foam” is intended to be used in combination with a medical bed, for adults or children, for specific medical purposes due to a loss of autonomy and/or deterioration of health status resulting in reduced mobility and the risk of pressure ulcers.
INDICATIONS
For bedridden adults or children patients with mobility impairment with “Medium to High” risk of pressure ulcers (according to the Braden scale or any other validated scale and on medical advice). Assists in the treatment of stage 1to 2 pressure sores according to EPUAP, up to stage 4 outside the support area.
CONTRAINDICATIONS
Patient weight level below or above the limits defined in the technical data table. For use on medical imaging table.
ADVERSE SIDE EFFECTS
Possible increase in certain sensory impairments in people with multiple sclerosis. Immobility syndrome in the elderly suffering from loss of independence. Instability of seated transfers (excluding mattresses with VMABS stabilizing side borders). Any serious incident occurring in connection with the device must be notified to the manufacturer and the competent authority of the Member State in which the user and/or patient is established. Inform the competent authority if you consider or have reason to believe that the device poses a serious risk or has been tampered with.
MEDICAL DEVICE COMPOSITION/TECHNICAL DATA
References Models and versions*| Products for adult medical beds (height
146 cm)| For Junior children’s beds
---|---|---
VMA62
VSA6ST+
VMS| VMA63
VSA7ST +
VMS| VMA64| VMA64S| VMA70
VMA80| VMA I I 7XURC
VMA137XURC| VMA I 20BAR
VMA I 40BAR| VMA82-CIC| VMA/PEDIA/M
Sizes (cm)| I 96x87x I 4| 196x87x I 4| 196x87x14| I 95x85x I 4| I 96x70x I 4
I 96x80x I 4| 195x117x17
195x137x17| 196x117x17
196x137x17| I 70x82x I 4| I 60x80x I 4
Specification Foam Sup| PU VE: 75kg/m31.8kPa| PU VE: 75kg/m3 I .8kPa| PU VE:
75kg/m31.8kPa| PU VE:
fl5kg/m3 1.8
kPa| PU VE:
fl5kg/m3 1.8
kPa| PU VE:
85 kg/m31.7 kPa| PU VE :
80kg/m3 2.7kPa| PU VE:
85 kg/m31.7 kPa| PU VE:
85 kg/m31.7 kPa
Specification| PU:18 kg/m33,0 kPa| PU:
25kg/m3 3,4kPa| PU: 34kg/m33,4 kPa| Foam laPU:34kg/m33,4kPa| PU:
34kg/m33,4kPa| PU :
25kg/m3 3,4 kPa,| PU :
40 kg/m3 6 kPa| PU:n5kem3 3,4
kPa| PU:25kg/m3 3,4
kPa
Patientweight min and max| 30 1 120 kg| 30 a 150 kg| 30 1 180 kg| 30 a 180 kg| 30 1 135 kg| 30 1 150 kg| 135 I 270 kg| IS to 70 kg
Size: from 125
to 145 cm| 15 to 70 kg
Size: from 125
to 145 cm
Foam material| 100% PU
Cover material| PES mesh (except CIC with PE mesh) – PU induction
The references mentioned can be completed with a suffix to constitute a commercial reference according to configuration (type of protective cover, packaging) and/or distributor (examples: VMA62/TH ; VMA64CIC; VMA63BOX-OXP…)
CLINICAL BENEFIT, DEVICE PERFORMANCE, MECHANISM OF ACTION DEVICE
PERFORMANCE CHARACTERISTICS: By immersion and enveloping, this support
relieves pressure on the body by improving blood circulation and consequently
reducing the risk of pressure ulcers in the supporting areas of the body.
Integrated heel support.
EXPECTED CLINICAL BENEFITS: Maintenance of tissue oxygenation at the
cutaneous and subcutaneous tissue level in contact with the support.
INFORMATION FOR HEALTHCARE PROFESSIONALS: Monitor the condition of the
patient’s skin in contact with the support several times a day. Use pressure
relief devices or positioning systems in patients with pressure ulcer(s).
Cover| | | | | |
---|---|---|---|---|---|---
Moderate washing up to 90°C| Maximum chlorine concentration of 5000ppm| Do not
iron| Do not dry clean| Dry with reduced thermal restriction| Use an approved
surface disinfectant/detergentb
I NFORMATION ON APPROPRIATE PROCESSES TO ALLOW ITS REUSE
Before re-using for another patient, the product must exhibit physical and
bacteriological cleanliness after application of the above procedures. Do not
use the mattress without a cover. Item treated with a biocide substance at no
risk to users. www.winncare.fr Avoid extended wet
contact with the foam. Always protect the mattresses with the supplied
rigorously dry mattress protectors. Prohibit all scouring, stripping products
or solvents and “piercing-sharp” objects in direct contact with the cover.
WARNINGS, PRECAUTIONS FOR USE, MEASURES REQUIRED
PRECAUTIONS FOR USE
Non-stabilized bone and/or muscle injuries in contact with the support. COPD-
type bronchopulmonary disorders with severe reduction in respiratory muscle
tone. A bedsore is a lesion of varying depth in the skin, associated with
excessive and prolonged compression of tissue between the body and a support
surface. This excessive pressure can prevent blood circulation and lead to
pressure ulcers, a dangerous lesion because it develops from the inside of the
body to the outside, and is therefore initially invisible. A pressure sore can
have several forms: a simple persistent rash for more than a day, hardening of
the skin, a wound of varying depth that can in severe cases affect muscles or
underlying bone.
SUITABILITY FOR USE | Max weight User | Head | Buttocks | Heel |
---|---|---|---|---|
VMA Mattress + Promust PU/PUHD/CIC Cover | 150 kg | 89 | 60 | 75 |
BAR Mattress + Pro must PU Cover | 270 kg | 130 | 85 | 93 |
WARNINGS
(Re) assess the risks of patient entrapment in the non-moving parts of the
medical bed associated with “therapeutic mattresses and accessories and
articulated positions of the bed base” according to standard EN 60601-2-52 in
adults and EN 50637 in adults. child. The use of segmented or full-length side
rails should be prescribed.
External cardiac massage not compatible without a board inserted between the
thorax and the upper surface of the mattress
STORAGE, HANDLING, DISPOSAL
STORAGE AND USAGE CONDITIONS
The supports must ideally be stored flat, away from direct light and excessive
humidity.
| Use| Storage
---|---|---
Temperature range| +15°C to + 40°C / +59°F to +104°F| -25°C to 70°C / +13°F to
+158°F
Humidity range| 30% – 70%| 30% – 95%
Atmospheric pressure| 50kPa – 106 kPa // 7.25 Psi / 15.37 Psi
ROLLED MATTRESSES
ALOVA: 2 mn → 10 mn: This time may be increased if the storage
temperature is lower than 15°C and/or in line with the storage period (0 to 6
months). The return to a normal visual appearance of the mattress under normal
usage temperature conditions may take up to 24 hours for ALOVA, APLOT and
MORFÉA mattresses but this does not present any risk for their use. Small
creases may be visible on the lower surface of the CIC cover after unrolling.
They present no risk to the patient and do not affect the efficiency of the
final product.
SERVICE LIFE
Viscoelastic foam (overlay) mattresses have an estimated service life of 6
years.
The rolled-compressed mattress life span is 6 months (see date on package
label).
PRODUCT DISPOSAL
Do not dispose of the product outside of the dedicated area. Respect the
established recycling systems in your country.
GUARANTEE
The guarantee duration for ALOVA (overlay) mattresses is 3 years. This
guarantee starts from the date of purchase of the product from your
distributor. In the case of single patient use, the guarantee relates to the
support and its protective cover. In the case of multi-patient use, the
duration of guarantee for the protective cover is 2 years. This guarantee does
not cover normal wear and tear of the product and its protective cover and is
not a substitute for legal guarantees. Please contact your distributor about a
defective product. They will carry out the necessary procedures with our
company to proceed with either repair or exchange.
Class 1 medical device according to Regulation (EU) 2017/745.
Management systems: Quality: ISO 13485:2016 – Environmental: ISO
14001:2015
Technical compliance certificates (FCBA) : on request or downloadable
from our website
Contents of package
1 mattress
1 protective cover
1 user manual
Création le 30/12/2012. Mise à jour le 14/03/2023 **Version numéro 22
**
WINNCARE FRANCE : – 4, Le Pas du Château – 85670 Saint-PaulMont-Penit –
France
Tél. +33 (0)2 51 98 55 64 – Fax +33 (0)2 51 98 59 07
Site de production ASKLE SANTE 200 Rue Charles Tellier – 30034
Nîmes Tél. : +33 (0)4 66 02 15 15 –
Fax : +33 (0)4 66 02 15 00
www.winncare.com
References
- Winncare Group: Manufacturer of medical beds, treatment and prevention of bedsores: therapeutic mattresses and cushions, lifts and furniture for healthcare institutions. - Winncare Group
- Winncare Group: Manufacturer of medical beds, treatment and prevention of bedsores: therapeutic mattresses and cushions, lifts and furniture for healthcare institutions. - Winncare Group
- Groupe Winncare: Fabricant de lits médicalisés, matelas et coussins thérapeutiques, lèves-personnes et mobilier pour collectivités - Winncare France
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>