PeakTech 3202 Analog Voltmeter Instruction Manual

June 3, 2024
PeakTech

PeakTech-logo

PeakTech 3202 Analog Voltmeter

Safety Precautions

This product complies with the requirements of the following directives of the European Union for CE conformity: 2014/30/EU (electromagnetic compatibility), 2014/35/EU (low voltage), 2011/65/EU (RoHS). Overvoltage category III 600 V; Pollution degree 2. To ensure safe operation of the equipment and eliminate the danger of serious injury due to short circuits (arcing), the following safety precautions must be observed. Damages resulting from failure to observe these safety precautions are exempt from any legal claims.

  • Do not use this instrument for high-energy industrial installation measurement.
  • Do not place the equipment on damp or wet surfaces.
  • Do not operate the equipment near strong magnetic fields (motors, transformers etc.).
  • Do not exceed the maximum permissible input ratings (danger of serious injury and/or destruction of the equipment).
  • Do not operate the meter before the cabinet has been closed and screwed safely as terminal can carry voltage.
  • Use caution when working with voltages above 35V DC or 25V AC. These Voltages pose shock hazard.
  • To avoid electric shock, do not operate this product in wet or damp conditions. Conduct measuring works only in dry clothing and rubber shoes, i. e. on isolating mats.
  • Never touch the tips of the test leads or probe.
  • Disconnect test leads or probe from the measuring circuit before switching modes or functions.
  • Check test leads and probes for faulty insulation or bare wires before connection to the equipment.
  • Comply with the warning labels and other info on the equipment.
  • Always start with the highest measuring range when measuring unknown values.
  • The measurement instrument is not to be to operated unattended.
  • Do not subject the equipment to direct sunlight or extreme temperatures, humidity or dampness.
  • Do not subject the equipment to shocks or strong vibrations.
  • Allow the equipment to stabilize at room temperature before taking up measurement (important for exact measurements).
  • The meter is suitable for indoor use only
  • Do not store the meter in a place of explosive, inflammable substances.
  • Opening the equipment and service – and repair work must only be performed by qualified service personnel
  • Do not place the equipment face-down on any table or work bench to prevent damaging the controls at the front.

Cleaning the cabinet
Periodically wipe the cabinet with a damp cloth and mid detergent. Do not use abrasives or solvents. Ensure that no water gets inside the equipment to prevent possible shorts and damage to the equipment.

Features

  • Analogue mirror scale with point bearing moving-coil.
  • Easy operation, compact size
  • Measuring ranges: DCV: 100mV ~ 500V in 6 ranges
    • ACV: 10 V ~ 500V in 4 ranges
  • Overload protection for each range
  • Application: Education, Maintenance, Production line, School, Laboratory, Industrial and Quality control.

Specifications

General specifications

Display Analogue display
overload protection 0,5 A / 500V; 6,3x32mm
operation temperature 0°C to +40°C; < 75% RH
storage temperature -10°C to +50°C; < 70% RH
dimensions (WxHxD) 105 x 150 x 45 mm
weight 300g
Technical specifications

DCV (Direct Current)

Range Accuracy Input Resistance
0,1 V +/- 5,0 %

full scale

| ****


20 kW / V

2,5 V| ****

+/- 3,0 %

full scale

10 V
50 V
250 V
500 V

ACV (Alternating Current)

Range| Accuracy| Input

Resistance

| Frequency-

Range

---|---|---|---
10 V| ****

+/- 4,0 %

full scale

| ****

9 kW / V

| ****

50 ~ 400kHz

50 V
250 V
500 V| 50 ~ 60 Hz

Front Panel Description

  1. Display
  2. Range Switch
  3. Zero Adjust of pointer
  4. Input Terminal „V“
  5. Input Terminal „COM“

Measuring Procedure

DC V (DC Voltage Measurements)
Note:
Always start with the highest measuring range.

  1. Select with the rotary switch the corresponding measuring range (0,1V ~500V DCV).
  2. Insert the black test lead into the “COM” socket and the red test lead into the “V” socket.
  3. Connect the test leads in parallel to the circuit/component to be measured.
  4. Read the measured value from the analogue display.

AC V (AC Voltage Measurements)
Note:
Always start with the highest measuring range.

  1. Select with the rotary switch the corresponding measuring range (10V ~500V ACV).
  2. Insert the black test lead into the “COM” socket and the red test lead into the “V” socket.
  3. Connect the test leads in parallel to the circuit/component to be measured.
  4. Read the measured value from the analogue display.

Replacement of the fuse

WARNING!
To avoid electric shock, disconnect all the test probes before removing the fuse. Replace only with the same type of fuse. Not note remove the top cover. Service should be performed only by qualified personnel.
CAUTION!
For continued protection against fire or other hazard, replace only with fuse of the specified voltage and current ratings. Follow these steps to replace the fuse:

  1. Disconnect all the test probes.

  2. Remove the protecting holster and then remove the back
    cover by unscrewing the four screws and pulling off the
    meter’s cover.

  3. Remove the blown fuse.

  4. Install the new fuse in the fuse compartment with same type
    and dimensions.

  5. Replace the cover and secure it with the screws.

Specifications of fuse:
0,5 A / 500 V FF; 6,3x32mm
WARNING!
Do not operate your meter until the back cover is in place and fully closed. All rights, also for translation, reprinting and copy of this manual or parts are reserved. Reproduction of all kinds (photocopy, microfilm or other) only by written permission of the publisher. This manual considers the latest technical knowing. Technical changings which are in the interest of progress reserved. We herewith confirm, that the units are calibrated by the factory according to the specifications as per the technical specifications. We recommend to calibrate the unit again, after one year. © PeakTech® 09/2021 Po/Ehr

PeakTech Prüf- und Messtechnik GmbH – Gerstenstieg 4 –
DE-22926 Ahrensburg / Germany
+49-(0) 4102-97398 80 +49-(0) 4102-97398 99
info@peaktech.de www.peaktech.de

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals