AMPROBE ACD-14-PRO Dual Display Clamp Multimeter User Manual
- May 15, 2024
- AMPROBE
Table of Contents
HARD AT WORK SINCE 1948.
ACD-14-PRO
ACD-14-PRO-EUR
Dual Display 600 A TRMS Clamp Multimeter
ACD-14-PRO
ACD-14-PRO-EUR
Dual Display 600 A TRMS
Clamp Multimeter
User Manual
ACD-14-PRO Dual Display Clamp Multimeter
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for
one year from the date of purchase unless local laws require otherwise. This
warranty does not cover fuses, disposable batteries or damage from accident,
neglect, misuse, alteration, contamination, or abnormal conditions of
operation or handling. Resellers are not authorized to extend any other
warranty on the behalf of Amprobe. To obtain service during the warranty
period, return the product with proof of purchase to an authorized Amprobe
Service Center or to an Amprobe dealer or distributor. See Repair Section for
details. THIS WARRANTY IS YOUR ONLY REMEDY.
ALL OTHER WARRANTIES – WHETHER EXPRESS, IMPLIED OR STATUTORY – INCLUDING
IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE
HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL,
INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY
CAUSE OR THEORY. Since some states or countries do not allow the exclusion or
limitation of an implied warranty or of incidental or consequential damages,
this limitation of liability may not apply to you.
Repair
All Amprobe returned for warranty or non-warranty repair or for calibration
should be accompanied by the following: your name, company’s name, address,
telephone number, and proof of purchase. Additionally, please include a brief
description of the problem or the service requested and include the test leads
with the meter. Non-warranty repair or replacement charges should be remitted
in the form of a check, a money order, credit card with expiration date, or a
purchase order made payable to Amprobe.
In-warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting
repair. During the warranty period, any defective test tool can be returned to
your Amprobe distributor for an exchange for the same or like product. Please
check the “Where to Buy” section on www.Amprobe.com
for a list of distributors near you. Additionally, in the United States and
Canada, in-warranty repair and replacement units can also be sent to an
Amprobe Service Center (see address below).
Non-warranty Repairs and Replacement – United States and Canada
Non-warranty repairs in the United States and Canada should be sent to an
Amprobe Service
Center. Call Amprobe or inquire at your point of purchase for current repair
and replacement rates.
USA: | Canada: |
---|---|
Amprobe | Amprobe |
Everett, WA 98203 | Mississauga, ON L4Z 1X9 |
Tel: 877-AMPROBE (267-7623) | Tel: |
Non-warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe distributor for a
nominal charge. Please check the “Where to Buy” section on www.beha-
amprobe.com for a list of distributors near you.
Amprobe Europe
Beha-Amprobe
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 – 0
www.beha-amprobe.com
(Correspondence only – no repair or replacement available from this address. European customers please contact your distributor.)
ACD-14-PRO / ACD-14-PRO-EUR
Dual Display 600 A TRMS Clamp Multimeter
1 Jaw
2 Tactile Barrier
3 Jaw Release
4 Display
5 Clamp-on Current / Amp-Tip / PEAK-RMS Button
6 SELECT/Backlight Button
7 Sensor for Non-Contact Voltage Detection| 8 Precise Low Current Measuring
Location (Amp-Tip)
9 Indicator of the Jaw Center for Current Measurement
10 Rotary Switch
11 Data Hold Button
12 Input Terminals
---|---
13 Auto-ranging
14 Alternative Current (AC)
15 P-RMS: PEAK-RMS mode (inrush current) is active
16 Precise low current measurement mode
17 Low battery indicator
18 Auto-ranging
19 ** Direct Current (DC)
Alternative Current (AC)
VFD** Variable Frequency Dive
20 Negative reading
21 Diode test mode is active| 22 Continuity buzzer is active
23 Lower display: Reading for V, Hz, Ω, μF, temperature and μA
24 μA: Microamps
μF: Microfarads
25 V: Volts
Hz: Hertz
26 Ω: Ohms
kΩ: KiloOhms
27 Data hold
28 A: Amps
29 Upper display:
Reading for AC current
---|---
SYMBOLS
| Application around and removal from hazardous live conductors is permitted.
---|---
| Caution! Risk of electric shock.
| Caution! Refer to the explanation in this manual.
|
The equipment is protected by double insulation or reinforced insulation.
| Earth (Ground).
CAT III| Measurement Category lll is for equipment intended to form part of a
building wiring installation. Such equipment includes
socket outlets, fuse panels, and some mains installation control equipment.
| Alternating Current (AC).
| Direct Current (DC).
| Battery.
| Underwriters Laboratories. [Note: Canadian and US.]
****| Complies with European Directives.
| Conforms to relevant Australian standards.
| Do not dispose this product as unsorted municipal waste.
Contact aqualified recycler.
SAFETY INFORMATION
The meter complies with:
- UL/IEC/EN 61010-1, CAN/CSA C22.2 No. 61010-1, Pollution Degree 2, Measurement category III 600 V
- IEC/EN 61010-2-033
- IEC/EN 61010-2-032
- IEC/EN 61010-031 (test leads)
- EMC IEC/EN 61326-1
Measurement Category III (CAT III) is for equipment intended to form part of a
building wiring installation. Such equipment includes socket outlets, fuse
panels, and some mains installation control equipment.
CENELEC Directives
The instruments conform to CENELEC Low-voltage directive 2006/95/EC and
Electromagnetic compatibility directive 2004/108/EC.
Warning: Read Before Using
To avoid possible electric shock or personal injury:
- Use the meter only as specified in this manual or the protection provided by the meter might be impaired.
- Avoid working alone so assistance can be rendered.
- Do not use the meter in wet or dirty environments.
- Do not use the meter if it appears damaged. Inspect the meter before use. Look for cracks or missing plastic. Pay particular attention to the insulation around the connectors.
- Inspect the test leads before use. Do not use them if insulation is damaged or metal is exposed.
- Check the test leads for continuity. Replace damaged test leads before using the meter.
- Have the meter serviced only by qualified service personnel.
- Use extreme caution when working around bare conductors or bus bars. Contact with the conductor could result in electric shock.
- Do not hold the meter anywhere beyond the tactile barrier.
- When measuring current, center the conductor in the clamp.
- Do not apply more than the rated voltage, as marked on the meter, between the terminals or between any terminal and earth ground.
- Remove test leads from the meter before opening the Meter case or battery cover.
- Remove the jaw from all conductors before opening the Meter case or battery cover.
- Never operate the meter with the battery cover removed or the case open.
- Never remove the battery cover or open the case of the meter without first removing the test leads or the jaws from a live conductor.
- Use caution when working with voltages above 30 V AC rms, 42 V AC peak, or 60 V DC. These voltages pose a shock hazard.
- Do not attempt to measure any voltage that might exceed the maximum range of the meter.
- Use the proper terminals, function, and range for your measurements.
- Do not operate the meter around explosive gas, vapor, or dust.
- When using probes, never touch the probe beyond the barrier.
- When making electrical connections, connect the common test lead before connecting the live test lead; when disconnecting, disconnect the live test lead before disconnecting the common test lead.
- Disconnect circuit power and discharge all capacitors before testing resistance, continuity, capacitance or diodes
- Use only 1.5V AAA batteries, properly installed in the meter case, to power the meter.
- To avoid false readings that can lead to electrical shock and injury, replace the battery as soon as the low battery indicator ( ) appears. Check meter operation on a known source before and after use.
- When servicing, use only specified replacement parts.
- Adhere to local and national safety codes. Individual protective equipment must be used to prevent shock and arc blast injury where hazardous live conductors are exposed.
- Only use the test lead provided with the meter or UL Listed Probe Assembly rated CAT III 600V or better.
UNPACKING AND INSPECTION
Your shipping carton should include:
1 Clamp multimeter
1 Test leads
1 K-type thermocouple
2 1.5 V AAA batteries (installed)
1 User manual
1 Carrying case
If any of these items are damaged or missing, return the complete package to
the place of purchase for an exchange.
MEASUREMENTS
Warning
To avoid electric shock or personal injury:
- When measuring current, center the conductor in the clamp.
- Keep fingers behind tactile barrier.
- Use the proper function and range for measurements.
- Disconnect circuit power and discharge all capacitors before testing resistance, continuity, capacitance or diodes.
- When using probes, keep fingers behind the finger guards.
- Connecting test leads:
– Connect the common (COM) test lead to the circuit before connecting the live lead;
– After measurement, remove live lead before removing the common (COM) test lead from the circuit.
Button | Description |
---|---|
SELECT / | Press SELECT button to select the alternative measurement function |
on the rotary switch.
Backlight: Press SELECT button > one second to turn ON LCD backlight. LCD
backlight automatically turns
OFF after approximately 20 minutes.
When LCD backlight is ON, press SELECT button > one second manually turn OFF
the backlight.
HOLD| Press HOLD to freeze the display reading ( is displayed) and press HOLD
a second time to release the reading.
Warning
To avoid possible electric shock or personal injury, when Display HOLD is
activated, be aware that the
display will not change when you apply a different voltage.
P-RMS| Press button to toggle between AC A and
Amp-Tip (precise low current mode). For low current on small diameter wires (<
10mm), the best accuracy is
specified near the jaw tip area.
Press P-RMS button > one second to enter P-RMS mode ( P-RMS is displayed) to
capture inrush
current (80 ms). Press a second time > one second to exit.
Note: Auto Power Off is automatically disabled under P-RMS mode.
Measuring AC and DC Voltage
To measure AC or DC voltage:
- Turn the rotary function switch to or .
- Connect the black test lead to the COM terminal and the red test lead to the V terminal. Measure the voltage by touching the probes to the desired test points of the circuit.
- View the reading on the lower display.
- When measuring AC voltage, press SELECT button to view the frequency reading on the lower display. (SELECT button: toggle among , Hz, “EF-H” and “EF-L” modes).
AC current can be measured at the same time by using the jaws while the V/COM terminals are measuring voltage. Upper display shows AC current measurement. Also see Measuring AC Current and Precise Low Current Measurement sections. Note: AC V and Hz function are equipped with digital low pass filter, and are capable of dealing with VFD (Variable Frequency Drive) signals. It also improves AC V reading stability in noisy electrical environments.
Voltage Detection (NCV)
Non-Contact Voltage Detection:
Warning
To avoid electrical shock and injury:
- Do not hold the meter anywhere beyond the tactile barrier.
-
Turn the rotary switch to NCV(EF) and press SELECT button to toggle to NCV(EF) mode. Two selectable sensitivity modes:
• “EF-H” indicating high sensitivity detection mode (“EF-H” is displayed) for the application of better detecting circuit with low electric field signals.
• “EF-L” indicating low sensitivity detection mode (“EF-L” is displayed) for the application where the electric field is too strong from the circuit under testing.
Press SELECT button to toggle among , Hz, “EF-H” and “EF-L” modes. -
The voltage detection sensor is located along the top-right end of the stationary clamp jaw for detecting electric fields surrounding energized conductors.
-
Detected electric field signal strength is indicated by a series of bargraph segments on the lower display and beeper. The stronger the electric field detected, the more bar-graph segments are displayed and the more intense the beep sounds.
Measuring AC Current
Warning
To avoid electrical shock and injury:
- Do not hold the meter anywhere beyond the tactile barrier.
- Do not use the meter to measure currents above the maximum rated frequency (400Hz). Circulating currents may cause the magnetic circuits of the jaws to reach hazardous excessive temperatures.
Voltage can be measured at the same time by using the V/COM terminals while the jaws are measuring current. Lower display shows voltage measurement. Also see Measuring AC and DC Voltage section.
To measure AC current:
- Turn the rotary switch to any function to power on the meter. The default current measurement mode is AC A (upper display).
- Open the clamp by pressing the jaw release and insert the conductor to be measured into the clamp. Ensure the jaws are firmly closed.
- Center the conductor using the jaw alignment marks.
- View the current reading on the upper display.
Caution
During current measurement keep the jaws away from other current-carrying
devices such as transformers, motors or energized wires, as they may
negatively influence accuracy of the measurement.
Precise Low-Current Measurement
Warning
To avoid electrical shock and injury:
- Do not hold the meter anywhere beyond the tactile barrier.
- Do not use the meter to measure currents above the maximum rated frequency (400Hz). Circulating currents may cause the magnetic circuits of the jaws to reach hazardous excessive temperatures.
Voltage can be measured at the same time by using the V/COM terminals while
the jaws are measuring current. Lower display shows voltage measurement. Also
see Measuring AC and DC Voltage section.
To measure AC low-current for small conductors:
-
Turn the rotary switch to any function to power on the meter. The default current measurement mode is AC A (upper display). Press button to switch between AC A and Amp-Tip mode ( is displayed).
-
Open the clamp by pressing the jaw release and insert the conductor to be measured into the clamp.
Ensure the jaws are firmly closed. -
Align the conductor at the specified jaw tip area for low-current measurement (Amp-Tip).
-
View the current reading on the upper display.
Caution
During current measurement keep the jaws away from other current-carrying
devices such as transformers, motors or energized wires, as they may
negatively influence accuracy of the measurement.
Microamps μA Measurement
The μA DC ( ) function on the meter is primarily for HVAC flame sensor testing.
To test a heating system flame sensor:
- Turn the heating unit off and locate the wire between the gas-burner controller and the flame sensor.
- Disconnect one of the flame sensor wires.
- Turn the rotary switch on the meter to .
- Connect the black test lead to the COM terminal and the red test lead to the μA terminal.
- Connecting the meter in series by attaching one test lead to the disconnected flame sensor probe and the other test lead to the disconnected control- module terminal.
- Turn heating unit on and check the reading on the Meter.
- Refer to the heating unit documentation for what the correct reading should be.
Measuring Resistance and Continuity
Warning
To avoid electrical shock and injury:
- To avoid false readings and electrical shock and injury, de-energize the circuit before taking the measurement.
- To avoid electrical shock when testing resistance/continuity in a circuit, make sure the power to the circuit is turned off and all capacitors are discharged. Use DC voltage function to check the capacitors are discharged.
- Connect the black test lead to the COM terminal and the red test lead to the Ω terminal.
- Turn the rotary switch to Ω (default mode is Ω).
- Press SELECT button switch between Ω and function.
- Connect the probes across the circuit or component to be tested.
- View the reading at the lower display
Resistance measurement: The resistance reading shows on the lower display. If
the circuit is open or resistance exceeds the meter’s range, the display reads
“OL”.
Continuity measurement: The resistance reading shows on the lower display. If
the circuit is shorted, the meter beeps (Beeper ON ≤ 10 Ω, OFF >250 Ω). If the
circuit is open or resistance exceeds the meter’s range, the display reads
OL.
Measuring Capacitance and Diode
Warning
To avoid electrical shock and injury:
- To avoid false readings and electrical shock and injury, de-energize the circuit before taking the measurement.
- To avoid electrical shock when testing capacitor/diode in a circuit, make sure the power to the circuit is turned off and all capacitors are discharged.
Use DC voltage function to check the capacitors are discharged.
- Connect the black test lead to the COM terminal and the red test lead to the terminal.
- Turn the rotary switch to (default mode is ).
- Press SELECT button switch between and function.
- Connect the probes across the circuit or component to be tested.
- View the reading at the lower display.
Capacitance: When measuring, be sure to note the correct polarity of the
capacitor.
Testing diode: When testing diode, normal forward voltage drop (forward
biased) for a good silicon diode is between 0.400V to 0.900V. A reading higher
than that indicates a leaky diode (defective). A zero reading indicates a
shorted diode (defective). Display reads “OL” indicates an open diode
(defective).
Reverse the test leads connections (reverse biased) across the diode. The
display reads “OL” if the diode is good. Any other readings indicate the diode
is resistive or shorted (defective).
Measuring Temperature
Warning
To avoid electrical shock and injury:
- When measuring temperature, DO NOT apply the temperature probe to any live conductive parts.
The meter measures temperature in either Celsius (°C) or Fahrenheit (°F).
- Connect the plug of the type-K temperature probe to the meter’s input terminal. Consider correct polarity of the probe.
- Turn the rotary switch to .
- Press SELECT button to select °C or °F. The display reflects the chosen temperature mode (°C or °F).
- Position the probe to take the measurement. The reading appears on the lower display.
Note: Type-K mini plug temperature probes can also be used with a plug
adaptor with 4mm pins to type-K socket. Backlight
Press SELECT button > one second to turn ON LCD backlight. LCD backlight
automatically turns OFF after approximately 20 minutes.
When LCD backlight is ON, press SELECT button > one second manually turns OFF
the backlight.
Auto Power OFF
The meter turns OFF if there is no button pushed, rotary function switch
operation for 32 minutes and/or no specified activities below, where
applicable:
- Significant measuring readings of above 8.5% of ranges
- Non-OL readings for Resistance, Continuity or Diode function
- Non-zero readings for Hz function
The meter will not enter auto power OFF mode when under normal measurements. To turn the meter back ON, press the SELECT button and release, or turn the rotary switch OFF and ON to restart the meter.
SPECIFICATIONS
Display | 3-5/6 digits 6000 counts; dual display |
---|---|
Sensing | True RMS |
Polarity | Automatic |
Update rate | 5 per second nominal |
Operating temperature | 32 °F to 104 °F (0 °C to 40 °C) |
Relative humidity | Maximum relative humidity 80% for temperature up |
to 31°C, decreasing linearly to 50% relative humidity at 40 °C
Storage temperature| -4 °F to 140 °F (-20 °C to 60 °C), < 80% R.H.
(with battery removed)
Measuring category| CAT III 600 V
Pollution degree| 2
Operating altitude| £ 2000 m
Temperature coefficient| nominal 0.15 x (specified accuracy)/ O
C @(0 O C to 18 O C or 28 O C to 40 O C), or otherwise
specified
Transient Protection| 6.0 kV (1.2/50 μs surge)
Overload protections| Current via jaws: 600 A AC rms continuous V
and COM terminals: 600 V AC/DC rms
**E.M.C.| Meets EN61326-1:2013
Temperature function at 80MHz ~ 150MHz, in an RF field of 1V/m: Total Accuracy
= Specified Accuracy + 25 digits
Other functions, in an RF field of 3V/m:
Total Accuracy = Specified Accuracy + 20 digits
Agency approval| **
Power supply| Two 1.5V AAA size battery
Power consumption| 6.2mA typical (backlight OFF), 62mA typical
(backlight ON)
Low battery indication| Approx. 2.85 V for Capacitance & Hz;
approx. 2.5 V for other functions
Auto power OFF| Idle for 32 minutes (approx.)
Auto power OFF power consumption| ** 5μA typical
Dimension (L x W x H)| 8.62 x 3.03 x 1.46 in (219 x 77 x 37 mm)
Weight| 208 g (0.46 lb) with batteries installed
Jaw opening & conductor diameter| ** 1.18 in (30 mm)
ELECTRICAL SPECIFICATIONS
Accuracy is ± (% reading digits + number of digits) or otherwise specified at
23°C ± 5°C.
Maximum Crest Factor < 2.5:1 at full scale and < 5:1 at half scale or
otherwise specified, and with frequency spectrum not exceeding the specified
frequency bandwidth for non-sinusoidal waveforms.
DC Voltage
Range | Accuracy |
---|---|
600.0V | ± (1.0 % + 5 LSD) |
Input Impedance: 10 MΩ, 100 pF nominal
AC Voltage (with Digital Low-Pass Filter)
Range | Accuracy |
---|---|
600.0V | ± (1.0 % + 5 LSD) |
Frequency: 50 Hz to 60 Hz
Input Impedance: 10 MΩ, 100 pF nominal
Continuity
Audible Threshold: ON at ≤ 10 Ω ; OFF at > 250 Ω
Response time: 32ms approx.
Resistance
Range | Accuracy |
---|---|
600.0 Ω, 6.000 kΩ, 60.00 kΩ | ± (1.0 % + 5 LSD) |
600.0 kΩ1)
, 6000 kΩ2)| ± (1.2 % + 5 LSD)
Open Circuit Voltage: 1.7VDC typical
- Test Current: 2 μA typical
- Test Current: 0.2 μA typical
Capacitance
Range | Accuracy 1) |
---|---|
200.0 μF, 2500 μF | ± (2.0 % + 4 LSD) |
- Accuracy with film capacitor or better
Diode
Range | Accuracy |
---|---|
3.000 V | ± (1.5 % + 5 LSD) |
Test Current: 0.3mA typically
Open Circuit Voltage: < 3.5 V DC typical
DC μA
Range | Accuracy | Burden Voltage |
---|---|---|
200.0 μA, 2000 μA | ± (1.0 % + 5 LSD) | 3.5 mV/μA |
Temperature
Range | Accuracy |
---|---|
– 40.0 °C to 99.9 °C | ±(1% + 0.8 °C) |
100 °C to 400 °C | ±(1% + 1 °C) |
-40.0 °F to 99.9 °F | ±(1% + 1.5 °F) |
100 °F to 752 °F | ±(1% + 2 °F) |
K-type thermocouple accuracy tolerances not included
Precise Low Current AC (Amp-Tip)
Range | Accuracy **1) 2) 3) 4)** |
---|---|
60.00 A | ± (1.5 % + 5 LSD) |
Frequency: 50 Hz to 60 Hz
- Induced error from adjacent current-carrying conductor: < 0.06 A/A
- Induced error from AC V measurement < 0.60A /kV @ 50/60 Hz
- Add 10 LSD to the specified accuracy @ < 6 A
- Unspecified at currents < 0.2A if function continuity or EF-Detection (NCV) is selected in the lower display.
AC Current
Range | Accuracy **1) 2) 3) 4)** |
---|---|
60.00 A 5) , 600.0 A | ± (1.8 % + 5 LSD) @ 50 Hz to <100 Hz |
60.00 A 5) , 600.0 A | ± (2.0 % + 5 LSD) @ 100 Hz to 400 Hz |
- Induced error from adjacent current-carrying conductor: < 0.06 A/A
- Induced error from AC V measurement < 0.60A /kV @ 50/60 Hz
- Specified accuracy is for measurements made at the jaw center. When the conductor is not positioned at the jaw center, add 2% to specified accuracy for position errors
- Unspecified at currents < 0.2A if function continuity or EF-Detection(NCV) is selected in the lower display.
- Add 10 LSD to specified accuracy @ < 6 A
PEAK-rms (for AC A function)
Response: 80 ms to > 90 %
Frequency Hz
Function | Sensitivity 1) (Sine rms) | Range |
---|---|---|
600 V | 50 V | 5.00 Hz to 999.9 Hz |
Accuracy: ± (1.0 % + 5 LSD)1)
DC-bias, if any, not more than 50% of Sine rms
Voltage Detection (NCV)
Bar-graph Indication| EF-H (High sensitivity) Typical Voltage
(Tolerance)| EF-L (Low sensitivity) Typical Voltage (Tolerance)
---|---|---
–| 10 V (5 V to 25 V)| 40 V (32 V to 70 V)
– –| 25 V (20 V to 66 V)| 110 V (55 V to 165 V)
– – –| 55 V (50 V to 125 V)| 220 V (130 V to 265 V)
– – – –| 110 V (90 V to 200 V)| 400 V (250 V to 500 V)
– – – – –| 220 V (>180 V)| 550 V (>430 V)
Indication: bar-graph segments and audible beep tones proportional to the
field strength
Detection frequency: 50/60 Hz
Detection sensor: inside the top side of the stationary jaw
MAINTENANCE AND REPAIR
If the meter fails to operate, check battery, test leads, etc., and replace as
necessary.
Double check the following:
- Replace the fuse or battery if the meter does not work.
- Review the operating instructions for possible mistakes in operating procedure.
Except for the replacement of the battery, repair of the meter should be
performed only by a Factory Authorized Service Center or by other qualified
instrument service personnel.
The front panel and case can be cleaned with a mild solution of detergent and
water. Apply sparingly with a soft cloth and allow to dry completely before
using. Do not use aromatic hydrocarbons, gasoline or chlorinated solvents for
cleaning.
BATTERY REPLACEMENT
When battery voltage drops below the value required for proper operation, the
battery symbol ( ) appears.
Warning
To avoid shock, injury, or damage to the meter, disconnect test leads before
opening case.
Replacing BATTERY follow below steps:
- Disconnect the test lead probe from all measuring circuits and/or remove the jaw from all conductors.
- Turn the meter to OFF position.
- Remove the screws from the battery cover and open the battery cover.
- Remove the batteries and replace with 1.5V AAA size (IEC R03). Observe correct polarity when installing the batteries.
- Put the battery cover back and re-fasten the screw.
Visit www.Amprobe.com for
- Catalog
- Application notes
- Product specifications
- User manuals
Amprobe®
www.Amprobe.com
info@amprobe.com
Everett, WA 98203
Tel: 877-AMPROBE (267-7623)
Beha-Amprobe
Amprobe® Europe
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 – 0
**** Please Recycle
7/2015, 6006090 A
©2015 Amprobe Test Tools.
All rights reserved. Printed in Taiwan
References
- Amprobe | Electrical Test and Measurement Tools
- Beha-Amprobe: The answer to all electronics, electrical and predictive maintenance questions
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>