KEITHLEY 4200A-SCS Automation Characterization Suite Standard Edition Software User Guide

June 16, 2024
KEITHLEY

ACS Standard Edition
Version 6.2.1 Release NotesKEITHLEY 4200A-SCS Automation
Characterization Suite Standard Edition Software Keithley Instruments
28775 Aurora Road
Cleveland, Ohio 44139
1-800-833-9200
tek.com/keithley

General information

This document describes the features added to the Keithley Instruments Automation Characterization Suite (ACS) Standard Edition software (version 6.2.1).
The Keithley Instruments ACS Standard Edition software supports component characterization testing of packaged parts and wafer-level testing using probers. ACS Standard Edition software can be installed on any computer, including Keithley Instruments Model 4200A-SCS Parameter Analyzer and Model 4200 Semiconductor Characterization System (4200-SCS).

Supported operating systems

ACS Standard Edition software is supported on the following operating systems:
Microsoft Windows®
11, 64-bit
Microsoft Windows 10, 64-bit
Microsoft Windows
10, 32-bit
Microsoft Windows 7, 64-bit (with Service Pack 1)
Microsoft Windows 7, 32-bit (with Service Pack 1)

ACS Standard revision history

Version Release date
6.2.1 March 2023
6.2 November 2022
6.1 March 2022
6 August 2021
5.4 February 2021
5.3 December 2017
5.2.1 September 2015
5.2 December 2014
5.1 May 2014
5 February 2013
4.4 December 2011
4.3.1 June 2011
4.3 March 2011
4.2.5 October 2010
4.2 June 2010

Install ACS

To install ACS software:

  1. Log in to your computer as an Administrator.
  2. Open the ACS executable file.
  3. Select Yes if you have an older version of ACS installed.KEITHLEY 4200A-SCS Automation Characterization Suite Standard Edition Software - Figure 1
  4. Follow the instructions to specify how you want to install the software on your system.

Once the updated version of ACS is installed, the older version is renamed. You can copy the projects and libraries from the previous version using the following steps.
To copy and paste folders:

  1. Find the C:\ACS_DDMMYYYY_HHMMSS\Projects\ folder; copy and paste to the current C:\ACS\Projects folder.
  2. Find the C:\ACS_DDMMYYYY_HHMMSS\library\pyLibrary\PTMLib\ folder; copy and paste to the current C:\ACS\library\pyLibrary\PTMLib\ folder.
  3. Find the C:\ACS\DDMMYYYY_HHMMSS\library\26library\ folder; copy and paste to the current C:\ACS\library\26library\ folder.

NOTE
ACS 6.2.1 is based on the Python® 3.7 programming language. If you customized your projects in a previous version of ACS you may need to change the projects created in the older version of ACS, which includes the Python language test module (PTM) script libraries. You can go to this site to review the Python changes for more detail: https://docs.python.org/3/whatsnew/3.7.html#porting- to-python-37
Install applications on a 4200A-SCS
If you are installing ACS on a 4200A-SCS Parameter Analyzer, you will see the following dialog box indicating that the applications identified are needed for installation. Make sure you select Do not close applications and Next to install (see the following figure). If you select Automatically close the applications, you must restart the computer after the installation is completed.KEITHLEY 4200A-SCS Automation Characterization Suite Standard
Edition Software - Figure 2 Install applications on a system with NI-488.2 drivers
If you are installing ACS on a system that contains NI-488.2 drivers, you will see the following dialog box indicating that the applications identified are needed for installation. Make sure you select Do not close applications and Next to install (see the following figure). If you select Automatically close the applications, you must restart the computer after the installation is completed.  KEITHLEY 4200A-SCS Automation Characterization Suite Standard
Edition Software - Figure 3

Supported models and test configurations

ACS software is used with the following Keithley Instruments in a variety of different test configurations. The ACS Fundamentals Reference Manual (part number ACS-914-01) and ACS Advanced Features Reference Manual (part number ACS-908-01) contain detailed information about the supported hardware and test configurations.

  • Perform multigroup testing with Series 2600B System SourceMeter® instruments and Series 2400 SourceMeter® instruments using ACS software installed on a personal computer or laptop.
  • Control hardware using ACS software installed on the Model 4200A-SCS Parameter Analyzer or Model 4200-SCS.
  • Perform combined group testing with a 4200A-SCS Parameter Analyzer or 4200-SCS, and Series 2600B SourceMeter® instruments using the combined test-execution engine in ACS software.
  • Control other external GPIB, LAN, or USB instruments using ACS software installed on a personal computer or laptop.

The following table summarizes the instruments supported in the ACS test libraries.

Instrument type Supported models
SMU instruments 2600B Series: 2601B-PULSE (DC only), 2601B, 2602B, 2604B,

2611B, 2612B, 2614B, 2634B, 2635B, 2636B
2600A Series: 2601A, 2602A, 2611A, 2612A, 2635A, 2636A
2400 Graphical Touchscreen Series SMU (KI24XX TTI): 2450, 2460, 2460-NFP, 2460 -NFP-RACK, 2460-RACK, 2461, 2461-SYS, 2470
2400 Standard Series SMU: 2401, 2410, 2420, 2430, 2440
2606B High Density SMU
2650 Series for High Power: 2651A, 2657A
Parameter analyzers| 4200A and the following modules: 4210-CVU, 4215-CVU 4225-PMU/4225-RPM, 4225-RPM-LR, 4200-SMU, 4201-SMU, 4210-SMU, 4211-SMU, 4200-PA, 4200A-CVIV
DMMs| DMM6500, DMM7510, 2010 Series
Switching and data acquisition systems| DAQ6510, 707A/B, 708A/B, 3700A
Pulse generators| 3400 Series

The following probers are supported in ACS:

Probers| Manual Prober Micromanipulator 8860 Prober
Suss MicroTec PA200/Cascade CM300 Prober Cascade 12000 Prober
Cascade S300 Prober Electroglas EG2X Prober Electroglas EG4X Prober TEL P8/P12 Prober TEL 19S Prober
Tokyo Semitsu TSK9 (UF200/UF3000/APM60/70/80/90) Prober Wentworth Pegasus 300S Prober with SRQ check Micromanipulator P300A Prober
Yang Sagi3 Prober with SRQ check Signatone CM500 Prober (WL250) TEL T78S/80S Prober
MPI SENTIO Prober Semiprobe SPFA Prober MJC AP-80 prober
Apollowave AP200/AP300 Prober
Vector Semiconductor AX/VX Series prober
---|---

NOTE
The graphical interactive test module (ITM) supports Series 24xx Touch Test Invent® (TTI) instruments and 26xx instruments at the same time. The 24xx instrument should be connected as the primary instrument and the 26xx connected as the subordinate.
You can control any test script processor (TSPTM) instrument using script test module (STM) scripts. You can control any instrument using the Python language test module (PTM) script, including instrumentation from other vendors.
Also, existing ACS STM and PTM libraries support specific instruments based on the library definition.

Supported communication interfaces

  • GPIB
  • LAN (Auto Scan and LAN)
  • USB
  • RS-232

If you are using a RS-232 connection, the instrument is not automatically added to the hardware configuration. You will have to add instruments connected with RS-232 manually. Change the hardware configuration file that is in the following directory on your computer:
C:\ACS\HardwareManagementTool\HWCFG_pref.ini. In this file you must change the baud rate, parity, byte, and stopBit settings. Review the following figure for details.KEITHLEY 4200A-SCS Automation Characterization Suite Standard
Edition Software - Figure 4

Software license

ACS allows you to create tests, manipulate settings, and view previous data without a license. However, you must have a license for ACS to control and retrieve data from a physical instrument. You can launch a one-time, 60-day trial for ACS after the initial installation. Once the license expires, you must purchase a full license to use the software.KEITHLEY 4200A-SCS
Automation Characterization Suite Standard Edition Software - Figure
5

License management

The ACS software license is managed using the Tektronix Asset Management System (TekAMS).
To generate a license file:

  1. You must submit your Host ID to TekAMS. For more information about TekAMS, see tek.com/products/product-license.
  2. To find the host ID, open the License Manage dialog box from the ACS Help menu. Select License > Host ID > Click to copy the Host ID.
  3. Select Install.

KEITHLEY 4200A-SCS Automation Characterization Suite Standard Edition
Software - Figure 6

ACS Standard version 6.2.1

Enhancements
Hardware configuration

Issue number: ACS-652
Enhancement: Added support in ACS to configure the DMM6500 instrument and the

DAQ6510 instrument
using TSP-Link.

ACS software and libraires

Issue number: ACS-624
 Enhancement: Added hold time during testing for the sweep-step, which is

enabled for each step.
Issue number:| ACS-544
 Enhancement:| Added the option to stop testing during compliance and power compliance for each step.

Resolved issues

Issue number: ACS-667
Symptom: When running multiple group testing using TSP-Link with a 2430

instrument in Group3, there
is no output after the test has completed.
Resolution:| This issue has been corrected.
Issue number:| ACS-666
Symptom:| A hardware scan error occurs when connecting TSP-Link using 24xx TTI SMU instruments.
Resolution:| This issue has been corrected.
Issue number:| ACS-665
Symptom:| A hardware scan error occurs when connecting multiple 2410 and 2430 instruments.
Resolution:| This issue has been corrected.
Issue number:| ACS-663
Symptom:| The error code –1014 is due to the MPI prober failing when using the PrRelReturn function.
Resolution:| This issue has been corrected.
Issue number:| ACS-662
Symptom:| The voltage drain output is not correct when used with the 100 A compliance.
Resolution:| This issue has been corrected.
Issue number:| ACS-661
Symptom:| When combining the 100 A compliance for the ITM, it will not refresh with new data.
Resolution:| This issue has been corrected.
Issue number:| ACS-660
Symptom:| There is no data output when pulse transient mode for the ITM is enabled.
Resolution:| This issue has been corrected.
Issue number:| ACS-657
Symptom:| Selecting the pulse sweep function in ITM mode does not work.
Resolution:| This issue has been corrected.
Issue number:| ACS-655
Symptom:| The Sweep2V and Sweep2I from the ITM function list have been removed from ACS
software testing.
Resolution:| This issue has been corrected.
Issue number:| ACS-648
Symptom:| If the stress count is too high during a Subsite Loop, the test will not complete.
Resolution:| This issue has been corrected.
Issue number:| ACS-554
Symptom:| ACS software will crash when you select the Ron_5A_PULSE_check function.
Resolution:| This issue has been corrected.

Software compatibility

Issue number: N/A
Resolution: When you start ACS on the 4200A-SCS system with Clarius software

version 1.4 or newer (with the Windows 10 operating system), a warning message may appear indicating that Keithley External Control Interface (KXCI) did not start successfully. Select Cancel to dismiss the warning.

KEITHLEY 4200A-SCS Automation Characterization Suite Standard Edition
Software - Figure 7 To manually configure the compatibility settings:

  1. Right-click the ACS icon and select Properties.
  2. Open the Compatibility tab.
  3. Select Run this program as an administrator and select OK to save.KEITHLEY 4200A-SCS Automation Characterization Suite Standard Edition Software - Figure 8

Usage note

Issue number: N/A
Resolution: If you install a KUSB-488B GPIB driver, you will see the

following message. You must select
the Keithley Command Compatible option. Select Next to continue the installation.

KEITHLEY 4200A-SCS Automation Characterization Suite Standard Edition
Software - Figure 9

PA-1008 Rev. U March 2023KEITHLEY 4200A-SCS Automation Characterization
Suite Standard Edition Software - Bar code

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>