Amprobe TMD-53 Thermocouple thermometer K-J with wireless User Manual
- June 4, 2024
- AMPROBE
Table of Contents
- Amprobe TMD-53 Thermocouple thermometer K-J with wireless
- Limited Warranty and Limitation of Liability
- TMD-53/TMD-53W
- SYMBOLS
- Federal Communications Commission
- UNPACKING AND INSPECTION
- INTRODUCTION
- SPECIFICATION
- MAINTENANCE AND REPAIR
- Installing and Replacing Battery
- Cleaning
- References
- Read User Manual Online (PDF format)
- Download This Manual (PDF format)
Amprobe TMD-53 Thermocouple thermometer K-J with wireless
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for
1 year from the date of purchase. This warranty does not cover fuses,
disposable batteries or damage from accident, neglect, misuse, alteration,
contamination, or abnormal conditions of operation or handling. Resellers are
not authorized to extend any other warranty on Amprobe’s behalf. To obtain
service during the warranty period, return the product with proof of purchase
to an authorized Amprobe Test Tools Service Center or to an Amprobe dealer or
distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY
REMEDY. ALL OTHER WARRANTIES – WHETHER EXPRESS, IMPLIED OR STAUTORY –
INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR
MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR
ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING
FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the
exclusion or limitation of an implied warranty or of incidental or
consequential damages, this limitation of liability may not apply to you.
Repair
All test tools returned for warranty or non-warranty repair or
for calibration should be accompanied by the following: your name, company’s
name, address, telephone number, and proof of purchase. Additionally, please
include a brief description of the problem or the service requested and
include the test leads with the meter. Non-warranty repair or replacement
charges should be remitted in the form of a check, a money order, credit card
with expiration date, or a purchase order made payable to Amprobe® Test Tools.
In-Warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting repair. During the warranty period any defective test tool can be returned to your Amprobe® Test Tools distributor for an exchange for the same or like product. Please check the “Where to Buy” section on www.amprobe.com for a list of distributors near you. Additionally, in the United States and Canada In- Warranty repair and replacement units can also be sent to a Amprobe® Test Tools Service Center (see address below).
Non-Warranty Repairs and Replacement – US and Canada
Non-warranty repairs in the United States and Canada should be sent to a
Amprobe® Test Tools Service Center. Call Amprobe® Test Tools or inquire at
your point of purchase for current repair and replacement rates.
In USA
Amprobe Test Tools
Everett, WA 98203
Tel: 877-AMPROBE (267-7623)
In Canada
Amprobe Test Tools Mississauga, ON L4Z 1X9 Tel:
905-890-7600
Non-Warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe® Test Tools
distributor for a nominalv charge. Please check the
“Where to Buy” section on www.amprobe.com for a list
of distributors near you.
European Correspondence Address
Amprobe® Test Tools Europe
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 – 0
(Correspondence only – no repair or replacement available from this address. European customers please contact your distributor.)
TMD-53/TMD-53W
- Power
- °C /°F
- Data Hold ,backlight
- “T1 T2/T1-T2” Main display
- “K/J” Main display
- “MAX/MIN” Main display
- “REL” Main display
- “Sec. Minu.” Third display
- “SET”
- “T1 T2 T1-T2” Second display
- “K/J” Second display
- “Hi/Lo LIMITS” Main display
- Wireless (TMD-53W only)
- Display
- T1/T2 Input
SCREEN DISPLAY
- Temperature Type
- Low battery
- MAX/MIN/AVG record mode
- Relative mode
- HOLD enabled
- Hi/Lo Limits comparative mode
- Temperature measurement units
- T1/T2 thermocouple or T1-T2 differential
- Wireless function enabled (TMD-53W only)
SYMBOLS
Caution! Refer to the explanation in this Manual | |
---|---|
Complies with European Directives | |
**** | Tested Comply With FCC Standards |
Do not dispose of this clamp meter as unsorted municipal waste. |
WARNING and PRECAUTIONS
- To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC.
- To avoid damage or burns, do not make temperature measurement in microwave ovens.
- Repeated sharp flexing can break the thermocouple leads. To prolong lead life, avoid sharp bends in the leads, especially near the connector.
Federal Communications Commission
This device TMD-53W complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) this device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation.
Note
This equipment TMD-53W has been tested and found to comply with the limits for
a Class B digital device, pursuant to Part 15 of the FCC Rules. These limits
are designed to provide reasonable protection. This equipment generates, uses
and can radiated radio frequency energy and, if not installed and used in
accordance with the instructions, may cause harmful interference to radio
communications. However, there is no guarantee that interference will not
occur in a particular installation if this equipment does cause harmful
interference to radio or television reception, which can be determined by
turning the equipment off and on, the user is encouraged to try to correct the
interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
Shielded interface cables must be used in order to comply with emission
limits.
Changes or modifications not expressly approved by the party responsible for
compliance could void the user’s authority to operate the equipment.
Wireless Note
Wireless receiver must keep a distance at least 40cm from the meter and meter
to meter distance must be at least 30cm.
UNPACKING AND INSPECTION
Your shipping carton should include:
- 1 Meter
- 4 AAA batteries
- 1 Software CD ROM (THWD-10W)
- 1 USB Wireless Antenna (THWD-10W)
- 2 K type thermocouple
- 1 Manual
- 1 Plain white box
If any of the items are damaged or missing, return the complete package to the place of purchase for an exchange.
INTRODUCTION
TThis instrument is a 5 digit, compact-sized portable digital thermometer designed to use external K-type and J-type thermocouples as temperature sensor. Temperature indication follows Reference Temperature/Voltage Tables (N.I.S.T. Monograph 175 Revised to ITS-90) for K-type and J-type thermocouples. Two K-type thermocouple are supplied with the thermometer.
Features
- Wireless two ways transmission. (TMD-53W only)
- K/J-type thermocouple selection dual input.
- °C/°F user-selective.
- MAX/MIN with elapsed time, AVG, REL, HOLD function.
- TIME setting, REL setting.
- Warning beeper with Hi/Lo setting.
- Resolution 0.1°C to 0.2°F.
OPERATING INSTRUCTIONS
Power Button
The “ ” key turns the thermometer on or off. In the SET mode cannot be powered
off. Exit SET mode to power off.
APO function mode
Press “ ” power key for more than 4 seconds to disable the auto-power
function.
The display will show “APO OFF”.
“°C/°F” Selecting the Temperature Scale
Readings are displayed in either degrees Celsius(°C) or degrees
Fahrenheit(°F). When the thermometer is turned on, it is set to the
temperature scale that was in use when the thermometer was last turned off. To
change the temperature scale, press the “°C/°F” key.
“ ” Mode (only Main display)
Press this key to enter the Data Hold mode, the “HOLD” annunciator is
displayed. When HOLD mode is selected, the thermometer holds the present
readings and stops all further measurements. Press the “ ” key again to
cancel HOLD mode and resume measurements. In the MAX/MIN recording mode, press
“ ” key to stop the recording. Press “ ****” key again to resume
recording. (Previously recorded readings are not erased).
Backlight function mode
Press the “ ****” key for more than two seconds to turn on the backlight.
Press the key again for more than two seconds to turn off the backlight. The
backlight will switch-off automatically after 30 seconds.
“T1 T2/T1-T2” Main Display Input Selection
The input selection indicates which input is selected for main display; T1
thermocouple, T2 thermocouple or the difference between the two thermocouples
(T1-T2), when the thermometer is turned on, it is set to T1.
“K/J” Main display Input Thermocouple Type Select
The “K/J” key selects the T1 thermocouple type, when the main display is
showing T1. When the thermometer is turned on, it is set to the type selected
when the thermometer was last turned off.
“MAX/MIN” With Time Record Mode (only Main display)
Press “MAX/MIN” key to enter the MAX MIN Recording mode, (displays the Maximum
reading with time, Minimum reading with time and Average reading stored in
record mode). In the this mode the automatic power-off feature is disabled and
“ POWER” key, “°C/°F” key, “REL” key, “SET” key, “Hi/Lo LIMITS” key and main
display
“T1 T2 T1-T2” key, “K/J” key are disabled. The beeper emits a tone when a new
minimum or maximum value is recorded.
Press “MAX/MIN” key to cycle through the MAX, MIN and AVG readings. If an
overload is recorded, the averaging function is stopped. In this mode, press
the
“HOLD” key to stop the recording of readings, all values are held, press again
to restart recording.
To prevent accidental loss of MAX, MIN and AVG data, this mode can only be
cancelled by pressing and holding the “MAX/MIN” key for 2 seconds. All
recorded readings are erased.
“REL” Relative Mode (only Main display)
Press the “REL” key to enter the relative mode, zero the display, and store
the displayed Reading as a reference value. REL is shown on the display. Press
“REL” key again to exit the relative mode. The relative reference value can
also be entered by the user. (See “SET mode” later in this manual). When the
desired relative value has been entered, press “REL” key to enter the relative
mode and than press “SET” key use the entered relative value as a reference
value. Press “REL” key again to exit the relative mode. In the relative mode,
the value (can not >±3000.0 counts) shown on the LCD is always the difference
between the stored reference and the present reading.
“Sec. Minu.” Selecting the Time scale
Press this key to display the elapsed time on the third display in either
hours and minutes or minutes and seconds. When the thermometer is turned on,
it is set to seconds. To change the time scale, press “Sec. Minus.” key.
Maximum elapsed time reading is 100 hours. If 100 hours is exceeded, the
elapsed time resets to zero.
“SET” Mode (Relative value set, Time set and Hi/Lo Limits value set)
- Press the “SET” key to enter relative values SET mode (Press “ENTER” key to skip setting relative value). “= = = =.=” is displayed on the main display. The relative value is entered via overlay numbers, press overlay “ENTER” key to store the relative value, and advance to elapsed time set mode.
- Elapsed time set mode, (Press “ENTER” key to skip Elapsed time set mode) “=.= = = : = =” is displayed in second and third display. Time (hours, minutes,seconds) value is entered via overlay numbers, press overlay “ENTER” key to store time value. Elapsed time starts from set time value.
- WHi Limit value set mode, “ ” is displayed (Press “ENTER” key to skip Hi Limit value set mode), “= = = =.=” is displayed in main display, Hi Limit value is entered via overlay numbers, then press “ENTER” key to store the Hi Limit value. “= = = =.=” is displayed in main display, Lo Limit value is entered via overlay numbers, then press overlay “ENTER” key to store the Lo Limit value and exit SET mode.
- When the thermometer is turned on it uses the relative value and Hi/Lo Limits values that were entered when thermometer was last in use.
“T1 T2 T1-T2” Second Display Input Selection
The input selection indicates which input is selected for second display; T1
thermocouple, T2 thermocouple or the difference between the two thermocouples
(T1-T2). When the thermometer is turned on, it is set to T2 input. To select a
different type of thermocouple use the “K/J” key.
“K/J” Second display Input thermocouple type select
The “K/J” key for T2 input selects K-type or J-type thermocouple as input,
when the second display has T2 shown. When the thermometer is turned on, it is
set to the type selected when the thermometer was last turned off.
“Hi/Lo LIMITS” Mode (only Main display)
Press “Hi/Lo LIMITS” key to enter the Hi/Lo Limits comparative mode, “ ” is
displayed. When the input temperature value exceeds the Hi or Lo Limits value,
the beeper emits a continuous or pulsed tone. Press “Hi/Lo LIMITS” key again
to exit the Hi/Lo Limits mode.
Wireless Mode (TMD-53W)
Press the “ ” key for more than two seconds to start wireless function. Press
the “ ” key again for another two seconds to stop wireless function. The
wireless mode will shut down if there is no wireless signal for two minutes.
To SET CH/ID to 00,00, press the “Hi/Lo LIMITS” key and power key for more
than 6 seconds with the meter powered down. The meter will set channel and ID
to 00,00 status. The second display will show 00, which means that the channel
and ID has been set to 00.
To check the channel and ID of the meter
When the meter is off, press “°C/°F” key and “ POWER BUTTON ” for 5 seconds,
LCD’s main display will show channel number, the second display will show ID
number.
SPECIFICATION
Electrical
Temperature Scale | Celsius or Fahrenheit user-selectable |
---|---|
Measurement Range | J-Type -200°C to 1050°C, (-328°F to 1922°F) |
K-Type -200°C to 1370°C,
(-328°F to 2498°F)
Resolution| 0.1°C or 0.2°F
Accuracy| Accuracy is specified for operating temperatures over the
range of 18°C to 28°C (64°F to 82°F), for 1 year, not including thermocouple
error.
±(0.05%rdg+0.3°C) on -50°C to 1370°C
±(0.05%rdg+0.7°C) on -50°C to -200°C
±(0.05%rdg+0.6°F) on -58°F to 2498°F
±(0.05%rdg+1.4°F) on -58°F to -328°F
Temperature Coefficient| 0.1 times the applicable accuracy specification per °C from 0°C to 18°C and 28°C to 50°C
(32°F to 64°F and 82°F to 122°F).
Input Protection| 24V dc or 24V ac rms maximum input Voltage on any
combination of input pins.
---|---
Input Connector| Accepts standard miniature thermocouple connectors
(flat blades spaced 7.9mm, center to center).
Maximum Differential Common Mode Voltage (Maximum Voltage between T1 and T2
during measurement): 1 Volt.
Environmental
Ambient Operating Ranges | 0°C to 50°C (32°F to 122°F) <80% R.H. |
---|---|
Storage Temperature | -20°C to 60°C (-4°F to 140°F) <70% R.H. |
General
Display | 5 digit liquid crystal display (LCD). |
---|---|
Overload | “—– ” or “ OL ” is display.. |
Reading Rate | 1 time per second. |
Auto power off | 30 minutes, press power key to resume operation. |
Battery | 4 X 1.5V AAA |
Battery Life | 190 hours typical with carbon zinc battery. |
Dimensions | 160mm(H) x 83mm(W) x 38mm(D); 6.3 |
in (H) x 3.3 in (W) x 1.5 in (D).
Weight| Approx. 270g (0.6lb) including batteries.
Supplied Wire| 4 feet type “ K ” thermocouple bead wire (Teflon tape
insulated). Maximum insulation temperature 260°C (500°F).
Wire accuracy ±2.2°C or ±0.75% of reading (whichever is greater) from 0°C to 800°C (32°F to 1472°F).
---|---
Wireless Features
Frequency range 910~920MHz (TMD-53W)
Low current consumption less than 1mA.
The transmitting distance can reach 25M without magnetic interference.
– EMC: Conforms to EN61326-1.
This product complies with requirements of the following European Community
Directives: 89/ 336/
EEC (Electromagnetic Compatibility) and 73/ 23/ EEC
(Low Voltage) as amended by 93/ 68/ EEC (CE Marking). However, electrical
noise or intense electromagnetic fields in the vicinity of the equipment may
disturb the measurement circuit. Measuring instruments will also respond to
unwanted signals that may be present within the measurement circuit. Users
should exercise care and take appropriate precautions to avoid misleading
results when making measurements in the presence of electronic interference.
MAINTENANCE AND REPAIR
WARNING
To avoid possible electrical shock, disconnect the thermocouple connectors
from the thermometer before removing the cover.
Installing and Replacing Battery
- A. Screw
- B. Battery Cover
- C. Battery
- Power is supplied by 4pcs 1.5V (SIZE AAA) UM-4 R03.
- The “ ” appears on the LCD display when replacement is needed. To replace battery remove screw from back of meter and lift off the battery cover.
- Remove the batteries from the battery compartment and replace them with new ones.
- When not use for long time, remove the battery.
- Do not store the unit in place with temperature and humidity beyond the recommended storage temperature.
Cleaning
Periodically wipe the case with a damp cloth and detergent, do not use abrasives or solvents.
Visit www.Amprobe.com for
- Catalog
- Application notes
- Product specifications
- User manuals
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>