AMPROBE IR-750 50:1 Infrared Thermometer
Limited Warranty and Limitation of Liability
Your Amprobe product will be free from defects in material and workmanship for one year from the date of purchase, unless local laws require otherwise. This warranty does not cover fuses, disposable batteries or damage from accident, neglect, misuse, alteration, contamination, or abnormal conditions of operation or handling. Resellers are not authorized to extend any other warranty on Amprobe’s behalf. To obtain service during the warranty period, return the product with proof of purchase to an authorized Amprobe Test Tools Service Center or to an Amprobe dealer or distributor. See Repair Section for details. THIS WARRANTY IS YOUR ONLY REMEDY. ALL OTHER WARRANTIES – WHETHER EXPRESS, IMPLIED OR STATUTORY – INCLUDING IMPLIED WARRANTIES OF FITNESS FOR A PARTICULAR PURPOSE OR MERCHANTABILITY, ARE HEREBY DISCLAIMED. MANUFACTURER SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT, INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES, ARISING FROM ANY CAUSE OR THEORY. Since some states or countries do not allow the exclusion or limitation of an implied warranty or of incidental or consequential damages, this limitation of liability may not apply to you.
Repair
All test tools returned for warranty or non-warranty repair or for calibration
should be accompanied by the following: your name, company’s name, address,
telephone number, and proof of purchase. Additionally, please include a brief
description of the problem or the service requested and include the test leads
with the meter. Non-warranty repair or replacement charges should be remitted
in the form of a check, a money order, credit card with expiration date, or a
purchase order made payable to Amprobe.
In-Warranty Repairs and Replacement – All Countries
Please read the warranty statement and check your battery before requesting
repair. During the warranty period any defective test tool can be returned to
your Amprobe distributor for an exchange for the same or like product. Please
check the “Where to Buy” section on www.amprobe.com for a list of distributors
near you. Additionally, in the United States and Canada In-Warranty repair and
replacement units can also be sent to a Amprobe® Test Tools Service Center.
Non-Warranty Repairs and Replacement – US and Canada
Non-warranty repairs in the United States and Canada should be sent to a
Amprobe Service Center. Call Amprobe or inquire at your point of purchase for
current repair and replacement rates.
- In USA In Canada
- Amprobe Test Tools Amprobe Test Tools
- Everett, WA 98203 Mississauga, ON L4Z 1X9
- Tel: 877-AMPROBE (267-7623) Tel: 905-890-7600
Non-Warranty Repairs and Replacement – Europe
European non-warranty units can be replaced by your Amprobe distributor for a
nominal charge. Please check the “Where to Buy” section on www.amprobe.com for
a list of distributors near you.
European Correspondence Address
Amprobe® Test Tools Europe
Beha-Amprobe GmbH
In den Engematten 14
79286 Glottertal, Germany
Tel.: +49 (0) 7684 8009 – 0
www.amprobe.eu
(Correspondence only – no repair or replacement available from this address. European customers please contact your distributor.)
- Laser aperture
- Trigger
- Battery cover
- Display
- USB connector
- K-type thermocouple connector
- Power adaptor connector (for optional accessory)
- Laser “ON” symbol
- Display backlight
- Measurement lock (continuous measurement)
- Temperature unit (Celsius / Fahrenheit)
- 8 seconds auto display hold
- Making measurement (pulling the trigger)
- Built in memory up to 99 points
- Programmable high and low alarm
- Low battery indicator
- Primary display
- Secondary display
- Emissivity (adjustable from 0.10 to 1.00)
- MAX, MIN, DIF, AVG temperature values HAL, LAL temperature setting
DATA point selection T-C K-type thermocouple measurement
SYMBOLS
- Caution! Refer to the explanation in this Manual.
- Warning! Laser light. Do not stare into laser beam.
- Celsius
- Fahrenheit.
- Battery indication.
- Complies with European Directives.
- Do not dispose of this product as unsorted municipal waste. Contact a qualified recycler.
SAFETY INFORMATION
The instrument complies with:
EN 61010-1 General Safety
EN 60825-1 Laser Safety
EN 61326-1 Electromagnetic Emissions and Susceptibility
Warning
-
Do not stare into laser beam.
-
Do not point laser directly at eye or indirectly off reflective surfaces.
-
For use by competent persons only.
-
Replace the batteries as soon as the low-battery indicator appears.
-
Do not use the thermometer if it operates abnormally.
-
Do not operate the thermometer around explosive gas, vapor, or dust.
-
To avoid a burn hazard or fire, know that reflective objects may be much hotter than the indicated
temperature reading. -
Do not leave the thermometer on or near objects of high temperature.
-
If the thermometer is used in a manner not specified by this manual, the protection provided by the thermometer may be impaired or may result in hazardous laser radiation exposure.
Cautions
To avoid damaging the thermometer under measurement, protect them from the
following:
- EMF (electro-magnetic fields) from arc welders, induction heaters
- Static electricity
- Thermal shock (caused by large or abrupt ambient temperature changes — allow 30 minutes for instrument to stabilize before use)
- Do not leave the thermometer on or near objects of high temperature
UNPACKING AND INSPECTION
Your shipping carton should include:
- 1 Thermometer (IR-750 / IR-750-EUR)
- 1 USB cable
- 1 K-type thermocouple probe
- 1 Carrying case
- 1 Hard carrying case
- 1 9V battery (installed)
- 1 Users manual
If any of the items are damaged or missing, return the complete package to the place of purchase for an exchange.
FEATURES
The Amprobe IR-750 / IR-750-EUR, a precision performance 50:1 spot to distance
ratio infrared thermometer, offers unparalleled accuracy and response time
with a temperature measurement range of -58°F to 2822°F
or -50°C to 1550°C. The IR-750 / IR-750-EUR is perfect for demanding quality
and process control applications with extremely high accuracy and distance to
spot ratio. The IR-750 / IR-750-EUR also features 99 data points recognition,
thermocouple inputs, and USB download for advanced HVAC/R, electrical,
industrial maintenance, automotive as well as quality control and fire
prevention applications.
- 50:1 Distance to Spot ratio
- Temperature range of 0°F to -58°F to 2822°F or -50°C to 1550°C
- Precision accuracy and rapid response time
- Laser pointer, backlit dual LCD Display
- Auto display hold and MAX/MIN memory
- Adjustable emissivity for measuring a variety of materials
- 99 Memory locations and PC download cable included
HOW THE THERMOMETERS WORK
Infrared thermometers measure the surface temperature of an object. The thermometer’s optics sense emitted, reflected, and transmitted energy, which is collected and focused onto a detector. The unit’s electronics translate the signal into a temperature reading which the unit displays .
OPERATING THE THERMOMETER
Temperature Measurement
The thermometer will turn on when you pull the trigger and also features an
auto-off function that automatically powers down the thermometer after 8
seconds of inactivity. To measure temperature, point the thermometer at an
object and pull the trigger. You can use the laser pointer to help aim the
thermometer. Pull and hold the trigger when measuring the target surface. When
release the trigger, the display will hold the reading for 8 seconds. Be sure
to consider distance-to-spot size ratio and field of view. The laser is used
for aiming only and is not related to temperature measurement.
Rotary Switch Positions
Button | Description |
---|
YELLOW
button
| Press YELLOW button to toggle between MAX, MIN, DIF, AVG, HAL, LAL, DATA and T-C options. When the Thermometer goes into sleep mode, press MODE to turn the thermometer ON again and it displays the last measurement result.
SET
| Press to enter set-up mode stepping through Emissivity, Trigger Lock and
Switching °C/°F set-up. Details refer to the below Emissivity, Trigger Lock
and °C / °F set-up.
| Press to turn the display backlight ON or OFF. Selectable two level
of display backlight to adapt different lighting conditions.
| Press to turn the laser light ON or OFF. symbol on the display
Indicates laser light is ON.
/
| When the thermometer enters the setup up mode (SET), press or to select a set-up option (Emissivity, Trigger lock, Switching °C / °F). When the thermometer enters HAL, LAL and DATA mode, press or to select a set-up option (Emissivity, Trigger lock, Switching °C / °F).
Laser
Caution
To avoid injury, do not point the laser directly at eye or indirectly off
reflective surfaces.
The thermometer is equipped with a laser used for aiming purposes only. The
laser turns off when the trigger is released.
To enable or disable the laser:
- Press button to enable or disable the laser. Symbol appears on the display when laser is enabled.
Emissivity set-up
- Press SET button to select Emissivity set-up, icon is blinking on the display
- Press to increase the value by 0.01. Press and hold for quick setting. The maximum value is 1.00.
- Press to decrease the value by 0.01. Press and hold for quick setting. The minimum value is 0.01.
- Press MODE button to complete the setting and exit Emissivity set-up, or press SET button to complete the setting
and continue setting for Trigger Lock.
Table of Surface Emissivity
Measure Surface| Switch Setting|
---|---|---
METALS|
Aluminum|
Oxidized| 0.2 – 0.4|
Alloy A3003|
Oxidized| 0.3|
Roughened| 0.1 – 0.3|
Brass|
Burnished| 0.3|
Oxidized| 0.5|
Cooper|
Oxidized| 0.4 – 0.8|
Electrical Terminal Blocks| 0.6|
Haynes|
Alloy| 0.3 – 0.8|
Inconel|
Oxidized| 0.7 – 0.95|
Sandblasted| 0.3 – 0.6|
Electoropolished| 0.15|
Iron|
Oxidized| 0.5 – 0.9|
Rusted| 0.5 – 0.7|
Iron Cast|
Oxidized| 0.6 – 0.95|
Unoxidized| 0.2|
Molten| 0.2 – 0.3|
Iron Wrought|
Dull| 0.9|
Lead|
Rough| 0.4|
Oxidized| 0.2 – 0.6|
Molydbenum|
Oxidized| 0.2 – 0.6|
Nickel|
Oxidized| 0.2 – 0.5|
Platinum| |
Black| 0.9|
Steel|
Cold-Rolled| 0.7 – 0.9|
Ground Sheet| 0.4 – 0.6|
Polished Sheet| 0.1|
Zinc|
Oxidized| 0.1|
NON- METALS
Asbestos| 0.95
Asphalt| 0.95
Basalt| 0.7
Carbon
Unoxidized| 0.8 – 0.9
Graphite| 0.7 – 0.8
Carborundum| 0.9
Ceramic| 0.95
Clay| 0.95
Concrete| 0.95
Cloth| 0.95
Glass
Plate| 0.85
Gravel| 0.95
Gypsum| 0.8 – 0.95
Ice| 0.98
Limestone| 0.98
Paper (any colour)| 0.95
Plastic
Opaque| 0.95
Soil| 0.9 – 0.98
Water| 0.93
Wood, (natural)| 0.9 – 0.95
Trigger Lock
The thermometer trigger can be locked on for continuous measurement.
To lock the trigger:
- Press SET button to select Trigger Lock set-up, icon is blinking on the display
- Press or to select ON or OFF.
- Press YELLOW button to complete the setting and exit Trigger Lock set-up, or press SET button to complete the setting and continue setting for °C / °F.
°C / °F Set-up
1. Press SET button to select °C / °F set-up, icon °C or °F is blinking on
the display
2. Press or to select °C or °F.
3. Press YELLOW button to complete the setting and exit °C / °F set-up.
MAX, MIN, DIF, AVG
The thermometer can measure maximum (MAX), minimum (MIN), differential (DIF) or average (AVG) temperatures each time a reading is taken. These values are available for infrared measurements.
- Press YELLOW button until MIN or MAX or AVG or DIF appears on the display
- Pull and hold the trigger while aiming at the target surface.
- The value will show on secondary display.
High Alarm (HAL) and Low Alarm (LAL)
The thermometer has a programmable high and low alarm to designate high or low
readings depending on the thresholds entered. When the alarm level is reached,
an alarm will sound. This feature is not available when measuring by
thermocouple.
To set either the high or low alarm:
- Press YELLOW button to select HAL or LAL option.
- Press or to adjust the threshold value.
- Press SET button to complete the setting. Symbol or will appear on the display.
To turn off the high or low alarm:
1. Press YELLOW button to select HAL or LAL option.
2. Press SET button. Disappearance of Symbol or on the display indicates high
or low alarm is off.
DATA Memory
The thermometer has 99 points data storage and the stored data can be downloaded via IR-750 / IR-750-EUR software. To save the data:
- Press YELLOW button to select DATA option.
- Press or to select data record number
- Pull and hold the trigger while aiming at the target surface, press DATA button to save the measurement value.
To erase the saved data, when the thermometer is in standby mode (screen is
off), press and hold “DATA” and
“ ” button while pulling the trigger until a beep occurs. The data records on
screen displays “ – – – – “, all saved data is cleared.
K-Type Thermocouple Measurement
X To avoid electrical shock or personal injury, do not connect the
thermocouple contact probe to live
electrical circuits.
The thermometer comes with a bead K-type thermocouple probe.
K-type thermocouple is connected to the thermometer via TC-K input connector
located on right side of the thermometer. The probe can be used simultaneously
while the thermometer is taking non-contact measurements.
- Press YELLOW button to enter T-C mode. The display shows OL before the measurement is made.
- Connect the K-type thermocouple to the thermometer via TC-K input connector located on right side of the thermometer. Press the trigger to start measuring. Secondary display shows the measurement reading.
- Release the trigger, the reading is kept on display (display hold: 8 seconds
Locating a Hot or Cold Spot
To find a hot or cold spot, aim the thermometer outside the target area. Then, slowly scan across the area with an up and down motion until you located the hot or cold spot.
Distance and Spot Size
As the distance (D) from the target being measured increases, the spot size (S) of the area measured by the instrument becomes larger. The spot size indicates 90% encircled energy.
Field of View
Make sure that the target is larger than the spot size. The smaller the target, the closer you should be to it.
Emissivity
Emissivity describes the energy-emitting characteristics of materials. Most
organic materials and painted or oxidized surfaces have an emissivity of about
0.95.
If possible, to compensate for inaccurate readings that may result from
measuring shiny metal surfaces, cover the surface to be measured with masking
tape or flat black paint (<150°C / 302°F ) and use the high emissivity
setting. Allow time for the tape or paint to reach the same temperatures as
the surface beneath it. Measure the temperature of the tape or painted
surface.
If you cannot use paint or use tape, then you could improve the accuracy of
your measurements with the emissivity selector. Even with emissivity selector,
it can be difficult to get a completely accurate infrared measurement of a
target with a shiny or metallic surface.
The thermometer allows you to adjust the emissivity for the type of surface
you’re trying to measure the temperature of. Use a different setting for your
application by referring to the Table of Surface Emissivity.
Reminders
-
Changes of surrounding ambient temperature can result in inaccurate reading, allow time for the instrument to adopt the change of ambient temperature before use. Specified accuracy applies after 30 minutes when the instrument changes to a different temperature environment an inaccurate.
-
The instrument cannot measure through transparent surfaces such as glass. It will measure the surface temperature of the glass instead.
-
See Table of Surface Emissivity for use in measuring shiny or polished metal surfaces (stainless steel,
aluminum, etc.). -
Steam, dust, smoke, etc., can prevent accurate measurement by obstructing the instrument’s optics.
TYPICAL MEASUREMENTS
This section describes a variety of measurements often performed by
technicians.
Reminder:
- User could select to turn on or off the backlight and laser whenever you are making readings with the thermometer.
- Relatively high emissivity normally means emissivity setting of about 0.95.
- Relatively low emissivity normally means emissivity setting of about 0.30.
- When user cannot identify the emissivity of the object to be measured, user could cover the surface to be measured (temperature >150°C) with black electric tape (emissivity of about 0.95). Allow time for the tape to reach the same temperature as the object to be measured. Measure and record the temperature of the tape.
Aim the thermometer to the object to be measured, adjust the emissivity
setting to make it as the same temperature as the tape. At this time, the
thermometer emissivity setting is close to the emissivity of the object to be
measured, measurement can be taken.
Testing Contactors (Starters)
- Press SET to select emissivity. Press / to select relatively low emissivity for bright contacts, or 0.7 mid level for darkened contacts.
- Press YELLOW button to select MAX.
- Measure line and load side of one pole without releasing trigger.
- A temperature difference between the line and load sides of a pole indicate increased resistance of one point and a contactor may be failing.
Testing Enclosed Relays
- Press SET and then press / to set emissivity to relatively low for un-insulated connectors or relatively high for plastic encased relays or for Bakelite enclosed relays or insulated connectors.
- Press YELLOW button to select MAX.
- Start to scan.
- Measure the relay casing, looking for hot spots.
- Measure electrical connections on relay terminals looking for hot spots.
Testing Fuses and Buss Connections
- Press SET and then press / to set emissivity to relatively high for paper covered fuse body or insulated connections.
- Press YELLOW button to select MAX.
- Scan the paper covered length of fuse.
- Without releasing the trigger, scan each fuse. Unequal temperatures between fuses may indicate voltage or amperage imbalance.
- Press SET and then press / to set emissivity to relatively low, for metal fuses and caps and insulated buss connections.
- Press MODE to select MAX.
- Scan each end cap on each fuse.
Note: Unequal temperatures or a high temperature indicates loose or corroded connection through the fuse buss spring clip
Testing Electrical Connections
-
Press SET and then press / to set emissivity to relatively low for un-insulated connectors or buss connections or relatively high for insulated connections.
Note: Conductors are typically smaller than the thermometer’s spot size.
If the spot size is bigger than the connector, the temperature reading is the average within the spot. -
Scan the conductor, moving toward direction of electrical connector (quick connector, wire nut, buss connection, etc.).
Scanning Walls for Air Leaks or Insulation Deficiencies
- Turn off heating, cooling, and blower.
- Press SET to select emissivity. Press / to select emissivity relatively high for painted surfaces or window surfaces.
- Press YELLOW button to select MIN when opposite side of wall is at lower temperature and or select MAX when opposite side of wall is at higher temperature.
- Measure an interior partition wall surface temperature.
- Do not release the trigger. Record this temperature as your baseline (or benchmark) for a “perfectly” insulated wall.
- Face the wall to be scanned. Stand 1.5m away to scan a 4cm spot on the wall.
- Scan horizontal rows of wall from top to bottom, or horizontal rows of ceiling from wall to wall. Look for greatest deviations from baseline temperature to identify problems. This completes the insulation test scan.
Turn on the blower (no heat, no cooling) and retest. If test results with the blower on are different than results with the blower off, this may indicate air leaks in conditioned envelope walls. The air leaks are caused by duct leaks that create a pressure differential across the conditioned space envelope.
Testing Bearings
Warning
To avoid injury when testing bearings:
-
Do not wear loose clothing, jewelry, or anything around neck when working around moving parts such as motors, belts, blower, and fans.
-
Make sure an electrical disconnect is within reach and operating correctly and freely.
-
Do not work alone.
Note: It works best to compare two similar motors operating similar loads -
Press SET and then press / to select relatively high emissivity.
-
Press SELECT to select MAX.
-
Enable motor and allow it to reach steady state operating temperatures.
-
Disable the motor if possible.
-
Measure the two motor bearing temperatures.
-
Compare the two motor bearing temperatures. Unequal temperatures or a high temperature can indicate a lubrication or other bearing problem that is resulting from excess friction.
-
Repeat the sequence for the blower bearings.
Testing Belts and Sheaves
- Press SET and then press / to select relatively high emissivity.
- Press SELECT to select MAX.
- Enable the motor and allow it to reach a steady state operating temperatures.
- Aim the thermometer at the surface to be measured.
- Start recording temperature.
- Slowly move the thermometer up the belt toward second sheave.
- If belt is slipping, sheave temperature will be high from friction.
- If belt is slipping, belt temperature will remain high between sheaves.
- If belt is not slipping, belt temperature will reduce between sheaves.
- If inner surfaces of sheaves are not a true “V” shape, this indicates belt slippage and will continue to operate at elevated temperatures until sheave is replaced.
- Sheaves must be properly aligned (include “pitch & yaw”) for belt and sheaves to operate at appropriate temperatures. A straight edge or taut string, can be used to check alignments.
- Motor sheave should operate at a temperature consistent with blower sheaves.
- If motor sheave is at a higher temperature at motor shaft than at outer circumference, belt is probably not slipping.
- If outer circumference of sheave is at higher temperature than sheave at motor shaft, then belt is probably
slipping and sheaves may be misaligned.
Checking Hydronic Radiant Heat Applications
- Press SET and then press / to select relatively high emissivity.
- Press YELLOW button to select MAX.
- To locate radiant heat tubes in floor, temporarily elevate the loop temperature to create hotter spots for identifying tubing runs.
- Before releasing trigger, press YELLOW button to toggle between MIN, MAX, DIF floor temperatures and record the temperature for future comparison and trending under similar conditions.
Measuring Grill, Register, or Diffuser Discharge Temperature
- Press SET and then press / to select relatively high emissivity.
- Aim the thermometer at the discharge air grille, register, or diffuser.
- Measure discharge temperature.
- Release trigger to freeze the temperature reading for 8 seconds and record this temperature.
- Grille, register, or diffuser temperature should be equivalent to discharge temperature at the air handler.
Checking for Blockage in Air-To-Air Evaporator or Condensers
- Remove panels to gain access to coil return bends or hairpins.
- Press SET and then press / to select relatively high emissivity for copper tube.
- Start the refrigeration system.
- Aim the thermometer at coil turn bends/hairpins.
- Start recording temperature.
- Take temperature of each return bend/hairpin.
- All evaporator return bends/hairpins should be at or slightly above evaporator saturation temperature from the pressure/temperature chart.
- All condenser return bend/hairpins should be at or slightly less than condenser saturation temperature.
- If a group of return bends/hairpins do not conform to expected temperatures, that indicates a blocked or restricted distributor or distributor tube.
OPERATING IR-750 / IR-750-EUR SOFTWARE
System Prerequisites
Recommended configuration
Processor| Pentium® 4 or higher with support of all customary operating
systems
---|---
Operating system| Microsoft Windows® 2000 / XP / Vista / Win7
Memory| 512 MB RAM or higher
Monitor| VGA (1024 x 768)
Drive| CD-ROM
Input device| Mouse or compatible
Interface| USB for data transmission
Install IR-750 / IR-750-EUR Software
• Insert the provided CD disk into CD-ROM drive.
• Double-click “IR-750_VX.XX.exe” (X can be 0 to 9 or blank, indicating the
software version)
• The pop-up windows will guide you through the program setup process.
Connecting USB cable to IR-750 / IR-750-EUR Thermometer
Main Screen
- command enables you to create New, Save file and set the numbers of measurement (1000, 5000, 10000, 50000, 100000 or limitless).
- command enables you to view the software operation instructions.
- Blinking indicates IR-750 / IR-750-EUR is connected to the software
- Main display screen of measurement
- Secondary display screen for measurement data
- Measurement data in graph
Using Program
button enables you to connect to IR-750 / IR-750-EUR thermometer and start
measurement. “USB Connect” blinking indicates that the IR-750 / IR-750-EUR
Thermometer is successfully connected to the PC and that the thermometer will
automatically start taking measurements. The primary display screen shows
current measurement temperatures and the secondary display screen shows all
measurement data. Download Data enables you to download temperature
measurements saved in IR-750 / IR-750-EUR Thermometer. Tick “Download Data”
and then click “USB Connect”; the saved data will automatically upload to your
PC and display on secondary display.
Exit and close the program
- Print data
- Save data (.txt, .xls, *.xml)
- Open files (.txt, .xls, *.xml)
- Deleting current data and create a new data.
Repeat: Tick to display every measurement with the interval you set. Un-
tick “Repeat”, the secondary display will only show measurement results that
are different from the last measured temperature. The graph display will show
complete measurement results in the graphics despite whether “Repeat” is
ticked or un-ticked. Sampling Interval allows you to set the sampling
intervals of measurement ( 1 to 9999 seconds).
Note: When measurement data reaching to 10000 sets (measurement number),
a warning message will pop up and stop measurements. Save or clear the
measurement data before continuing next measurements.
- Last page
- Next page
- Current page
- Previous page
- First page
- Save graph in current page as *.bmp
- Set numbers of measurement data on graph display
Test Report
Test data can be saved as .txt, .xls, *.xml by clicking button. When opening
data report, make sure selecting font format as Amprobe-A1 or Amprobe-A2 in
“Unit” column in order to display correct temperature 0unit °C or °F.
Zoom In Selected Graph
To enlarge the particular part of the graph, use the mouse pointer to select
area from upper-left point to lower-right point.
You can use button to save the selected enlarged graph.
To resume to its original full view, use the mouse pointer to select any area
on the graph from lower-right point to upper-left point.
DETAILED SPECIFICATIONS
Feature | IR-750 / IR-750-EUR |
---|---|
Temperature Range | -50°C to 1550°C (-58°F to 2822°F) |
Accuracy (assumes ambient operating temperature of 23°C to 25°C (73°F to
77°F))| >0°C to 1550°C (>32°F to 2822°F): ±1.8% or ±1.8°C (±4°F),
whichever is greater (Typical) >-35°C to 0°C (-31°F to 32°F): ±1.8%+1°C (2°F)
or ±2.8°C (±6°F), whichever is greater (Typical) -50°C to -35°C (-58°F to
-31°F): not specified (for reference only)
Repeatability| ±0.5% of reading or ±0.5°C (±1°F), whichever is greater
(Typical)
Display Resolution| 0.1°C / 0.1°F of reading < 999.9 0.2°C / 0.2°F
of reading below 10°C (50°F) 1°C / 1°F of reading > 999.9
Spectral Response| 8µm to14µm
Laser Sighting| Single point laser
Laser Power| Output > 1mW Class 2, wavelength 630 to 670nm
Response Time (95%)| 250ms
Distance to Spot (D:S)| 50:1
Minimum Spot Size| 19mm
Emissivity| Digitally adjustable from 0.10 to 1.00 by 0.01. Pre-set
emissivity is 0.95
Ambient Operating Temperature| 0°C to 50°C 32°F to 120°F
Relative Humidity| 0% to 75% non-condensing
Storage Temperature| -20°C to 65°C / -4°F to 150°F (Battery not
installed)
Temperature Display| °C or °F selectable
Display Hold| 8 sec
MAX/MIN Temperature Display| Ö
DIF/AVG Temperature Display| →
USB Interface| →
Data Storage| 99
Programmable High and Low Alarm| →
K-Type Thermocouple Measurement| →
Dual LCD Display| →
LCD Backlit| →
Low Battery Indication| →
Tripod Mount| →
Power| 9V 6F22 alkaline battery or by Amprobe EPS-700-US (US plug) or
EPS-700-EUR (European plug) power adaptor
Battery Life| 10 hours with laser and backlight on 30 hours with laser
and backlight off
Dimension (H x L x W)| Approx.183 x 147 x 57 mm (7.2 x 5.8 x 2.3 in)
Weight| Approx.345 g (0.76 lb) with battery installed
Optional Accessory| Power adaptor EPS-700-US (US plug), EPS-700-EUR
(European plug)
MAINTENANCE
Lens Cleaning: Blow off loose particles using clean compressed air.
Gently brush remaining debris away with a
camel hair brush. Carefully wipe the surface with a moist cotton swab. The
swab may be moistened with water
or rubbing alcohol.
NOTE: DO NOT use solvents to clean the plastic lens.
Case Cleaning: Use soap and water on a damp sponge or soft cloth.
Caution! Do not submerge the unit in water.
TROUBLE SHOOTING
Code | Problem | Action |
---|---|---|
OL | Target temperature is over range | Select target within |
specifications
– OL| Target temperature is under range| Select target within
specifications
Battery indication| Low battery| Check and/or replace battery
Blank display| Possible dead battery| Check and/or replace battery
Laser does not work
| 1. Low or dead battery
2. Ambient temperature above 40°C (104°F)
| 1. Replace battery
2. Use in area with lower ambient temperature
Beeper beeps continuously| Check if high/low alarm has been set and if the measurement value is over the limit| Reset high/low alarm setting or cancel the limit setting
BATTERY REPLACEMENT
To install or change one 9V battery (see below):
- Push the button and pull the handle downward to open battery cover.
- Install the battery noting its correct polarity.
- Re-install the battery cover.
Battery: 9V 6F22 alkaline battery or equivalent
Visit amprobe.com for
- Catalog
- Application notes
- Product specifications
- User manuals
Amprobe®
amprobe.com
info@amprobe.com
Everett, WA 98203
Tel: 877-AMPROBE (267-7623)
References
- Amprobe | Electrical Test and Measurement Tools
- Beha-Amprobe: The answer to all electronics, electrical and predictive maintenance questions
- Amprobe | Electrical Test and Measurement Tools
- Beha-Amprobe: The answer to all electronics, electrical and predictive maintenance questions
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>