x-rite 301-30 Transmission Densitometer Instruction Manual

May 15, 2024
x-rite

301
Black and White Transmission Densitometer
Operator Manualx-rite 301-30 Transmission
Densitometer

301-30 Transmission Densitometer

Advanced T est Equipment Corp.
www.atecorp.com
800-404- A TEC (2832)
Fcc
This equipment has been tested and found to comply with the limits for a Class A digital device, pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses, and can radiate radio frequency energy and, if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
Canada
CAN ICES-3 (A).’ NMB-3 (A)
NOTE: Shielded interface cables must be used in order to maintain compliance with the desired FCC and European emission requirements.
For 230V~ replace with 400 mA, 5 by 20 mm fuse only
For 115V~ replace with 500 mA, 5 by 20 mm fuse only.
Replace with only Same Type Lamp Assembly, X-Rite P’N 301-21.
Replace with only Same Type Fluorescent Lamp Assembly, Type F6T5/D.
Consult this documentation in all cases where the Attention symbol appears.
This symbol is used to inform you of any potential HAZARD or actions that may require your attention.
CE DECLARATION
** Hereby, X-Rite, Incorporated, declares that this 301 instrument is in compliance with the essential requirements and other relevant provisions of Directive(s) EMC 2004/108/EC and LVD 2006/95/EC.
Equipment Information
Use of this equipment in a manner other than that specified by X-Rite, Incorporated may compromise design integrity and become unsafe.
WARNING: This instrument is not for use in explosive environments.
Discontinue use if AC cord is damaged.
Instructions for disposal: Please dispose of Waste Electrical and Electronic Equipment (WEEE) at designated collection points for the recycling of such equipment.
Notice**
This product may or may not be supplied with a detachable power cord. Use an IEC320/C7 or [EC320/C 13 detachable power cord from an approved agency in your region to meet your power requirements.

Relamping Instructions

NOTE: Disconnect Power Before Relamping
Fluorescent Bulbs(s) Replacement
Replace the Fluorescent Bulb(s) as Follows:

  1. Remove the four (4) screws on the corners of the light table and remove the light table.
  2. Remove the faulty bulb(s) and replace with type F6T5/D fluorescent bulb(s).
  3. Locate the light table and refasten with the four (4) screws.

LIMITED WARRANTY

X-Rite warrants this Product against defects in material and workmanship for a period of twelve (12) months from the date of shipment from X-Rite’s facility, unless mandatory law provides for longer periods. During such time, X-Rite will either replace or repair at its discretion defective parts free of charge.
X-Rite’s warranties herein do not cover failure of warranted goods resulting from: (i) damage after shipment, accident, abuse, misuse, neglect, alteration or any other use not in accordance with X-Rite’s recommendations, accompanying documentation, published specifications, and standard industry practice; (ii) using the device in an operating environment outside the recommended specifications or failure to follow the maintenance procedures in X-Rite’s accompanying documentation or published specifications; (iii) repair or service by anyone other than X-Rite or its authorized representatives; (iv) the failure of the warranted goods caused by use of any parts or consumables not manufactured, distributed, or approved by X-Rite; (v) any attachments or modifications to the warranted goods that are not manufactured, distributed or approved by X-Rite. Consumable parts and Product cleaning are also not covered by the warranty.
X-Rite‘s sole and exclusive obligation for breach of the above warranties shall be the repair or replacement of any part, without charge, which within the warranty period is proven to X-Rite‘s reasonable satisfaction to have been defective. Repairs or replacement by X-Rite shall not revive an otherwise expired warranty, nor shall the same extend the duration of a warranty.
Customer shall be responsible for packaging and shipping the defective product to the service center designated by X-Rite. X-Rite shall pay for the return of the product to Customer if the shipment is to a location within the region in which the X-Rite service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations. Proof of purchase in the form of a bill of sale or receipted invoice which is evidence that the unit is within the Warranty period must be presented to obtain warranty service. Do not try to dismantle the Product. Unauthorized dismantling of the equipment will void all warranty claims. Contact the X-Rite Support or the nearest X-Rite Service Center, if you believe that the unit does not work anymore or does not work correctly.
THESE WARRANTIES ARE GIVEN SOLELY TO BUYER AND ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION, AND NON-INFRINGEMENT. NO EMPLOYEE OR AGENT OF X-RITE, OTHER THAN AN OFFICER OF X-RITE, IS AUTHORIZED TO MAKE ANY WARRANTY IN ADDITION TO THE FOREGOING.
IN NO EVENT WILL X-RITE BE LIABLE FOR ANY OF BUYER’S MANUFACTURING COSTS, OVERHEAD, LOST PROFITS, GOODWILL, OTHER EXPENSES OR ANY INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES BASED UPON BREACH OF ANY WARRANTY, BREACH OF CONTRACT, NEGLIGENCE, STRICT TORT, OR ANY OTHER LEGAL THEORY. IN ANY EVENT OF LIABILITY, X- RITE’S MAXIMUM LIABILITY HEREUNDER WILL NOT EXCEED THE PRICE OF THE GOODS OR SERVICES FURNISHED BY X-RITE GIVING RISE TO THE CLAIM.
Proprietary Notice
The information contained in this manual is derived from patent and proprietary data from X-Rite, Incorporated. This manual has been prepared solely for the purpose of assisting operation and maintenance personnel in their use of the X-Rite 301 Instrument.
The information contained in this manual is derived from patent and proprietary data from X-Rite, Incorporated. This manual has been prepared solely for the purpose of assisting operation and maintenance personnel in their use of the X-Rite 301 Instrument.
The contents of this manual are the property of X-Rite, Incorporated and are copyrighted. Any reproduction in whole or part is strictly prohibited.
Publication of this information does not imply any rights to reproduce it or use it for any purpose other than installing, operating, or maintaining the equipment described herein.
This instrument is covered by the following U.S. and foreign patents:
US. Patent #4,080,075 and other patents pending.
Copyright ©2015 by X-Rite, Incorporated.
“ALL RIGHTS RESERVED.”

INTRODUCTION

The X-Rite 301 Black and White Densitometer is a highly accurate and reliable instrument designed for ease of operation. This unit features push button nulling and will compute density comparisons and subtractions. Voltage and temperature compensation provides drift-free operation.
SPECIFICATIONS
Size: ……………………………. 10.25W x 15L x 5.25H
Weight: ……………………………. 8.5lbs.
Range: ……………………………. 0 to 5.0D w/2mm and 3mm apertures
……………………………. 0 to 4.0D with 1mm aperture
Accuracy: ……………………………. ±0.02D
……………………………. ±1.5%D (1mm aperture, 3.5D
……………………………. to 4.0D)
Repeatability: ………………………. ±0.01D
Operating Temp. Range: ……….. +10° to +40°C (+50° to 104°F)
Operating Relative Humidity: … ≤76%
Voltage Range: ……………………. 301 (Domestic) 100Vac to
……………………………. 130Vac, 60Hz – 301X (Export)
……………………………. 200Vac to 240Vac, 50Hz
Power Requirement: ……………… 80VA maximum
Warm-up Time: ……………………. 60 seconds
Scale Factor Stability: …………… ±1.0% per six months
Null Drift: …………………………… ±0.03D max., ±0.01D typical
Safety ……………………………. UL 3101-01
……………………………. C22.2, No. 1010-1-92
Altitude ……………………………. 2000m
IEC 664 ……………………………. Pollution Degree 2
Installation Type ………………….. Category 2
Usage ……………………………. Indoor only
FCC ……………………………. Part 15, Class A, Digital Device
Industry Canada …………………… ICES-003 Issue 2, Revision 1
International ………………………… EN50081-1:1992
……………………………. Class B Generic Emmision Standard
……………………………. EN50082-1:1992
……………………………. Generic Immunity Standard
NOTE: Due to heat generated by the fluorescent lamps, accuracy to the 301 Densitometer is improved with the lamps off.
UNPACKING
Remove the instrument from shipping carton. Inspect for possible damage. If any damage is noted, contact the transportation company immediately. Do nothing more until the carrier’s agent has inspected the damage.
If damage is not evident, make sure the following items are included.

  • X-Rite 301 Operation Manual (with Warranty Card)
  • 1mm, 2mm and 3mm apertures (inside back cover of manual)
  • Standard Reference Step Tablet (inside front cover of manual)
  • 301(X) Instrument

OPERATING INSTRUCTIONS

Operation of the X-Rite 301 is very simple. Procedures of the various modes of operation are provided in the following paragraphs. When operating the instrument, use these recommended common sense suggestions to protect yourself and your instrument:

  • Avoid dangerous environments (e.g., don’t use instrument in damp, wet areas)
  • Use proper electrical connections.
  • Ensure electric cord does not interfere with work.
  • Disconnect power before servicing.

POWER APPLICATION
The 301(X) is designed to operate from a standard grounded line source.
Always plug the instrument into a 3-wire receptacle.

  1. Plug line cord into a 3-wire grounded outlet.
  2. Turn on power switch.

APERTURE REPLACEMENT
Remove the installed aperture by lifting the edge upward. Install a new or different aperture by inserting the aperture into the light source cavity, making sure that the aperture is fully seated and pressed flush against the cavity. NULLING (ZEROING) PROCEDURE
Due to the electronic memory loss caused by power removal, null must be established each time power to the instrument is removed. Null remains very stable (±0.01D) as long as power remains on.
Null the instrument as follows:

  1. Remove film from reading area.
  2. Lower the reading head by pressing the read button until it deflects totally and actuates the switch.
  3. Momentarily push the null button.
  4. Remove pressure on the read button to release the reading head.

NOTE: The read button is a read-hold activator. Compress the button with only as much pressure as necessary for total deflection.

ABSOLUTE DENSITY MEASUREMENT
Measure absolute density as follows:

  1. Null the instrument as previously described in Nulling (Zeroing) Procedure.
  2. Center the film area directly over the bright green light spot under the reading head.
  3. Lower the reading head by pressing the read button until it totally deflects and actuates the switch.
  4. Allow the instrument reading to settle to a stable number.
  5. Remove pressure on the read button to release the reading head.

NOTE: Always measure density with the film emulsion side up. When measuring density values above 2.50D, ensure that reading head light seal is completely on surface of the film being measured.

DENSITY COMPARISON MEASUREMENT
Compare density readings as follows:

  1. Place reference film over aperture. Press read button and momentarily press the null button.
  2. Place the film to be compared over the aperture and measure density.
    • This measurement is the difference between the reference film density and the compared film density.
    • A negative (–) display indicates a lower compared film density than the reference film.

NOTE: When referencing with densities greater than 3.00D, press the null button longer to allow the instrument electronics to fully stabilize.

CALIBRATION

Calibration of this instrument should remain stable over a long period of time and over an extended range of voltage and temperature. However, a calibrated step tablet is provided (inside front cover) to verify the instrument calibration. Handle the step tablet with care to prevent accumulation of dirt and fingerprints on the surface.

CALIBRATION FREQUENCY
Under normal operating conditions, the instrument calibration should be verified once a week by the following the Calibration Procedure.

CALIBRATION PROCEDURE
See illustrations on the following page for calibration adjustment locations and reference.

To verify calibration:

  1. Null the instrument as described in Nulling (Zeroing) Procedure page
  2. Center the “CAL” portion of step tablet under read head.
  3. Press read button to measure step tablet.
  4. Hold read button down until a stable reading is attained.
    • The “CAL” reading should be ±0.02D of the “CAL” value marked on the step tablet.

If “CAL” reading is NOT within tolerance, calibrate instrument.
To calibrate instrument:

  1. Center the “CAL” portion of step tablet under the read head.

  2. Press and hold read button to measure the step tablet.

  3. Use a small screwdriver or adjustment tool into hole on the right side of instrument, to engage the calibration adjustment (potentiometer).

  4. With read button still depressed, adjust as necessary to within ±0.02D of the “CAL” value marked on the calibrated step tablet.
    • Rotate counterclockwise to increase value, clockwise to decrease.

  5. Remove step tablet and press read button to check null, re-null the instrument as necessary.

  6. Repeat until the calibration value and null are within tolerance.

NOTE: Always return the step tablet to the plastic bag on inside front cover for protection.

Calibration

x-rite 301-30 Transmission Densitometer - parts1

MAINTENANCE

The 301 Densitometer is completely covered by a one (1) year warranty and should be referred to the factory or authorized service center for repairs within the warranty period. Attempts to make repairs within the warranty period may void the warranty. If repairs are needed after the warranty period, only qualified technicians should perform such repairs.
Refer to the exploded view illustration of the 301 on page 16.
WARNING: DISCONNECT POWER BEFORE PERFORMING ANY MAINTENANCE.
CLEANING
General cleaning of the 301 Densitometer should be done with mild a soap solution.
CAUTION: Do not use alcohol based solutions to clean light table.
Clean the instrument optics (apertures and read head opal glass) with a cotton swab moistened with alcohol. Remove residue and lint with lens paper. See Figure 5.
CAUTION: To prevent contamination of optics, always cover the reading area when cleaning the opal glass.

FLUORESCENT BULB REPLACEMENT
Replace fluorescent bulbs as follows:

  1. Remove four (4) screws securing light table to the unit chassis.
  2. Remove light table.
  3. Remove the faulty bulb(s) and replace with new, type F6T5/D fluorescent bulb(s).
  4. Reinstall light table and secure to unit chassis with four (4) screws.

READING LAMP REPLACEMENT (P/N 301-21)
Replace the reading area lamp as follows:

  1. Remove the unit bottom cover by removing the four (4) screws located in the rubber feet.

  2. Remove the four (4) screws securing the Lamp P.C. board to the arm assembly at bottom of chassis.

  3. Disconnect lamp connector and remove faulty Lamp P.C. board.
    NOTE: Ensure lamp P.C. board bulb is clean. Use alcohol on a lint free cloth or cotton swab to remove smudges.
    CAUTION: The bulb on Lamp P.C. board has been prealigned for optimum optical performance. Use care not to bend or reposition the bulb when cleaning or installing.

  4. Connect the lamp connector to new Lamp P.C. board.

  5. Secure Lamp P.C. board to arm assembly with four (4) screws.

  6. Reinstall bottom cover and secure with four (4) screws.

FUSE REPLACEMENT
Replacement Fuse for 115VAC Instrument:
X-Rite P/N SE49-0050 (Time-delay 500ma-250V, 5mm x 20mm)
Replacement Fuse for 230VAC Instrument:
X-Rite P/N SE49-0040 (Time-delay 400ma-250V, 5mm x 20mm)
To replace fuse:

  1. Turn power off (O) and remove detachable line cord.
  2. Insert small flat-blade screwdriver into left edge of fuse carrier and pry out.
  3. Remove blown fuse from carrier clip and discard.
  4. Insert replacement fuse in fuse clip.
  5. Reinsert fuse carrier into fuse cavity. Make sure carrier is firmly seated.
  6. Reinstall detachable line cord.

FACTORY REPAIR
X-Rite recognizes the need to provide complete technical repair service to customers. X-Rite provides repair of any 301 submitted past warranty. The customer shall pay shipping costs and the instrument shall be submitted in its original shipping carton as a complete unaltered unit.

TROUBLESHOOTING

The following chart is included for troubleshooting your 301 instrument. Use care to avoid danger of electrical shock when troubleshooting or performing any type of maintenance. Only qualified technicians should perform repairs.
WARNING: DISCONNECT POWER BEFORE REPLACING PARTS.

PROBLEM CAUSE REMEDY

Unit keeps blowing fuses (500ma for 301) (400ma for 301X)
Unit display will not turn on| Fuse loose in holder
Short in p.c. board
Short in wiring
Short in ballast
Short in transformer
Unit not nulled
Blown fuse
Loose connector(s)
Faulty P.C. board
Faulty switch
Bad transformer winding
Open wiring| Tighten fuse holder

  • Null unit
    Replace fuse
    Re-seat connector(s)

  • Light source will not illuminate| Blown fuse
    Loose connector(s)
    Faulty Lamp P.C. board
    Faulty switch
    Bad transformer winding
    Open wiring| Replace fuse
    Re-seat connector(s)

  • Unit will not Null| Faulty light source
    Faulty P.C. board
    Faulty null switch
    Loose connector(s)
    Faulty sensor wiring
    Faulty sensor(s)| (See light source)

  • Replace switch
    Re-seat connector

  • Fluorescent bulb(s) will not illuminate| Loose bulb(s)
    Blown fuse
    Loose connector
    Faulty bulb(s)
    Faulty switch
    Faulty ballast(s)
    Open wiring| Seat bulb(s)
    Replace fuse
    Re-seat connector
    Replace bulb(s)

  • Reading will not change or hold| Loose connector(s)
    Faulty P.C. board
    Faulty switch
    Open wiring| Re-set connector

*Contact X-Rite Instrument Services or an authorized service center for repair.

OUTPUT CONNECTION (301RS)

The Model 301RS has the capability of sending measurement data to a printer, computer, or modem. This is done through the use of a RS- 232 output connection located in the rear of the instrument. Data is sent each time the READ button is released.
RS-232 CONNECTION
Use a RS-232S type cable to connect the 301RS and the receiving device. The Transmit Function Switch, located inside the instrument, must be set to DTE (Data Terminal Equipment) or DCE (Data Communication Equipment) to correspond with the receiving device. The DTE setting sends the TRANSMIT signal through pin 2 and is normally used for printers or computers outputs. The DCE setting sends the TRANSMIT signal through pin 3 and is normally used for modem outputs.
NOTE: The output connection is preset from factory to DTE operation for X-Rite 301 units and DCE operation for Fuji 301 units.
To change the Transmit Function Switch setting:

  1. Disconnect power to instrument.
  2. Turn unit upside down and remove bottom cover by the four screws in located in the rubber feet.
  3. Set the Transmit Function Switch to DTE for printer or computer connections or DCE for modem connections.
  4. Return bottom cover in place and secure with four screws.

DATA INSTRUCTION FORMAT
Data is sent in the following format:

Lead Character represents the ASCII code for a positive character (space) or negative character (–). Units, Tenths, & Hundredths are three ASCII digits (0-9) representing density values. Data is transmitted at the rate of 300 baud with one start bit and stop bit, and the parity bit set to logic zero. Output logic level is −5Vdc to +5Vdc DTE (Pin 2 = TRANSMIT) DCE (Pin 3 = TRANSMIT) Ground = Pin 7

Corporate Headquarters
X-Rite, Incorporated
4300 44th Street SE
Grand Rapids, Michigan 49512
Phone 1 800 248 9748 or 1 616 803 2100
Fax 1 800 292 4437 or 1 616 803 2705| European Headquarters
X-Rite Europe GmbH
Althardstrasse 70
8105 Regensdorf
Switzerland
Phone (+41) 44 842 24 00
Fax (+41) 44 842 22 22| Asia Pacific Headquarters
X-Rite Asia Pacific Limited
Suite 2801, 28th
Floor, AXA Tower
Landmark East, 100 How Ming Street
Kwun Tong, Kowloon, Hong Kong
Phone (852)2568-6283
Fax (852)2885 8610
---|---|---

Please visit www.xrite.com for a local office near you.

P/N 301-30
Rev. SS

References

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals