X-Rite i1Basic Pro 3 Plus PANTONE User Manual
Corporate Headquarters USA
4300 44th St. SE, Grand Rapids, Ml 49512 United States Telephone 888-800-9580
X-Rite Europe GmbH
Althardstrasse 70, Regensdorf, 8105 Switzerland Telephone +800 700 300 01
X-Rite (Shanghai) Color Management Co., Ltd.
6/F, Building #1, IBP Phase 2, 518 Fuquan North Road, Changning District, Shanghai, China 200335 Telephone +86 021 3279 6666
Please visit www.xrite.com for a local office near you. (Rev. 10/2019)
Important Notices
CE Declaration
Hereby, X-Rite, Incorporated, declares that this device is in compliance with
the essential requirements and other relevant provisions of Directive(s) EMC
2014/30/EU, LVD 2014/35/EC, and RoHS EU 2015/863.
FEDERAL COMMUNICATIONS COMMISSION NOTICE
NOTE: This equipment has been tested and found to comply with the limits for a
Class B digital device, pursuant to Part 15 of the FCC Rules. These limits are
designed to provide reasonable protection against harmful interference in a
residential installation. This equipment generates, uses and can radiate radio
frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications.
However, there is no guarantee that interference will not occur in a particular installation. If this equipment does cause harmful interference to radio or television reception, which can be determined by turning the equipment off and on, the user is encouraged to try to correct the interference by one or more of the following measures:
- Reorient or relocate the receiving antenna.
- Increase the separation between the equipment and receiver.
- Connect the equipment into an outlet on a circuit different from that to which the receiver is connected.
- Consult the dealer or an experienced radio/TV technician for help.
INDUSTRY CANADA COMPLIANCE STATEMENT
This Class B digital apparatus complies with Canadian ICES-003.
NOTE: USB interface cable shipped with this device must be used in order
to maintain compliance with the desired CE mark requirements, FCC Part 15
Rules, and Canadian ICES-003.
Safety Information
EQUIPMENT INFORMATION
Use of this equipment in a manner other than that specified by X-Rite, Incorporated may compromise design integrity and become unsafe.
WARNING: This instrument is not for use in explosive environments.
If this product is used in a manner not specified by the instruction, the safety protection provided by the device may be impaired or become inoperable.
Package Contents
- Tripod Holder (optional)
- Ambient Light Measuring Head (optional)
- Ambient Light Measuring Head Case (optional)
- Display Holder
- Positioning Target
- Calibration Plate
- i1Pro 3 Device
- USB Cable
- Backer Board (optional)
- Ruler (optional)
- ColorChecker Proof (optional)
- ColorChecker Classic (optional)
- User Manual
- Certificate of Performance (not shown)
- Carrying Case (not shown)
Device Overview
- Status indicator
- Measurement button
- Limiting aperture
- Type label with serial number
- Positioning sensor
- Reference gap for ruler mount
- Mounting interface for accessories
- USB connector port
Installation
Before you can use your device or see its status, you need to install the software application on your computer. Please do not plug the device‘s USB cable into your computer until after you have installed the software application.
-
Visit https://www.xrite.com/i1profiler to download the i1Profiler application and
install the software application on your computer. -
Review all instructions in this user manual before you start using your new device.
-
After you have completed the software installation process, attach your i1Pro 3 device to your computer‘s USB port. If connected successfully the status indicator lights up solid white.
The X-Rite i1Pro 3 will not work if it is connected to the USB port on your keyboard or to a USB hub without external power supply. i1Pro 3 power rating: 5 V 500 mA.
Device Status Indicator
The i1Pro 3 device is equipped with a device status indicator on its top housing. The device status indicator provides feedback on the current status of the device, and guides you through the measurement process.
General device status:
To avoid interference of the device status indicator with the measurement process, the device status indicator is turned off during measurement.
Calibrating the Device
If you own multiple devices make sure that the serial number on the front and back of the calibration plate matches the serial number of your i1Pro 3 device. Refer to the user guide of your measurement software for additional information on the calibration process.
-
In the center of the calibration plate is a slider to protect the white calibration tile. Open this protective slider until it clicks into place to access the white reference ceramic tile on this calibration plate.
-
Place the i1Pro 3 measurement device on the calibration plate. The device must seat firmly on the calibration plate.
-
Start the calibration mode in the software, and press the measurement button on the device or in the software. Status indicator light switches off during calibration. Once the device is successfully calibrated the status indicator light pulsates white.
If the calibration fails the status indicator light pulsates red and then returns to solid white. -
When the calibration is finished you should close the protective slider for the white reference ceramic tile.
A clean white reference ceramic tile is essential for providing accuracy of your measurements. The protection cover should always be closed when the white reference ceramic tile is not in use. If necessary you can clean the white reference ceramic tile with isopropyl alcohol and a soft clean cloth.
Scan Measurement
Virtual aperture technology
During a scan measurement the i1Pro 3 device is performing 400 measurements per second. The automatic patch detection of the device identifies useable measurements made on a patch and unusable measurements made between two patches. Valid measurements on a patch are averaged and the device reports the averaged result to the software. Thanks to this technology the virtual aperture of the i1Pro 3 device adapts to the length of a patch. For best measurement results the length of the patches on your test chart should be selected based on the resolution of your printer. In some cases, increasing the width of the patches will yield more accurate results.
Measurement conditions
The i1Pro 3 measurement device supports multiple measurement conditions. In a
single pass scan measurement, the patches are illuminated with the LEDs in the
i1Pro 3 device.
Single Pass Measurement is required for measurement condition
- ISO 13655 M0 | Tungsten filament lamp with UV included (NoFilter)
- ISO 13655 M1 | D50
- ISO 13655 M2 | Illumination with UV excluded (UVcut)
- OBC | X-Rite Optical Brightener Compensation
The status indicator on the i1Pro 3 device guides you through the measurement process.
To prepare your i1Pro 3 device for a scan measurement you need to mount the i1Pro 3 device on the ruler. Make sure your device is calibrated and the status indicator is pulsating white before you mount it.
- Place the ruler on the backer board.
- Position the i1Pro 3 device in the carriage on the ruler. Make sure that the positioning pins on the carriage slide into the reference gaps on the bottom of the i1Pro 3 device. Double-check that the i1Pro 3 device seats firmly in the carriage.
- During scanning measurement make sure that the device’s limiting aperture glides smoothly on the guide rail of the ruler.
- Align the backer board for ease of use during scanning. Only apply light pressure to the device when scanning.
The distance between the i1Pro 3 device and the surface of the printing substrate is critical for the accuracy of your measurements. If the limiting aperture of the device or the edge of the ruler‘s guide rail is worn out you should replace them. Spare parts can be ordered through X-Rite‘s Service Centers.
Single Pass Scan Measurement
- Make sure that your device is calibrated and the status indicator is pulsating white.
- Position the ruler with the device on the substrate in front of row #1 of your test chart. It does not matter if you scan the row from left to right or right to left.
- Set your software to single scanning mode.
- Press the measurement button on the device and wait for a short green pulse before you start moving the device to the opposite side of the ruler. Hold the measurement button pressed until you have reached the far end of your test chart.
The short delay before the measurement process is required to warm up the LEDs. The status indicator on the device provides feedback if the measurement was successful:
5. Position the ruler with the device in front of the next row of your test chart and measure the next row.
Spot Measurement
To prepare your i1Pro 3 device for spot measurement you need to mount the i1Pro 3 device on the positioning target. Make sure your device is calibrated and the status indicator is pulsating white before you mount the device. Refer to the user guide of your measurement software for further details on the spot measurement process.
- Mount the i1Pro 3 measurement device on the positioning target by inserting the mounting connector on the positioning target in the mounting interface at the rear of the i1Pro 3 device.
- Configure your software to take spot measurements.
- Position the device on the patch you would like to measure and press the device down to the positioning target.
- Press the measurement button on the device. Status indicator light switches off during a measurement and pulsates green once the measurement has been successfully finished.
- To remove the i1Pro 3 device from the positioning target, gently pull the mounting connector with the positioning target out of the mounting interface at the rear of the i1Pro 3 device.
Display Measurement
To prepare your i1Pro 3 device for a display measurement you need to mount the i1Pro 3 device on the display holder. Make sure your device is calibrated and the status indicator is pulsating white before you mount the device.
- You can adjust the length of the counter weight of the display holder for the size of your screen. The counter weight is equipped with hook-and-loop fasteners on both ends which allow you to shorten and lengthen the display holder to the correct length.
- Mount the i1Pro 3 measurement device on the display holder by inserting the mounting connector on the display holder in the mounting interface at the rear of the i1Pro 3 device. Make sure that the nose of the i1Pro 3 device seats firmly in the display holder.
- Fix the limiting aperture of the i1Pro 3 measurement device in the middle of your screen.
- Set the software for display measurement and start the measurement process in the application. Status indicator lights switch off during measurement.
- To remove the i1Pro 3 device from the display holder, gently pull the mounting connector with the display holder out of the mounting interface at the rear of the i1Pro 3 device.
Ambient Light Measurement
Before you start measuring, please attach the ambient light measuring head on the limiting aperture of the i1Pro 3 device. If you own multiple devices make sure that the serial number on the ambient light measuring head matches the serial number of your i1Pro 3 device. The ambient light measuring head can also be used in conjunction with the tripod holder.
Projector Measurement
To prepare your i1Pro 3 device for projector measurements you can use the device on the tripod holder, or mount the tripod holder on a tripod with one of the two standard screws you supply (1). Next you need to mount the i1Pro 3 device on the tripod holder. Make sure that your device is calibrated and the status indicator is pulsating white before you mount the device.
- If you are mounting the tripod holder on a tripod, follow the instructions for mounting a tripod head in the user manual of your tripod.
- Mount the i1Pro 3 measurement device on the tripod holder by inserting the mounting connector on the tripod holder in the mounting interface at the rear of the i1Pro 3 device.
- To remove the i1Pro 3 device from the tripod holder, gently pull the mounting connector with the tripod holder out of the mounting interface at the rear of the i1Pro 3 device.
Setup the i1Pro 3 device for projector measurements.
- Place the i1Pro 3 device on the tripod holder next to the projector or at a distance in front of your screen.
- Open the locking screw on the tripod holder and adjust the angle of the i1Pro 3 device. The limiting aperture of the i1Pro 3 should point towards the center of your screen. Close the locking screw once the device is in the right position.
- Adjust the lighting of your room to the same conditions as you will use for the projector and presentation.
- Configure your software for projector measurement and follow the instructions on the screen.
General Maintenance
To maintain the measurement performance of your i1Pro 3 over its life time, it is important to observe some rules and to periodically clean certain parts that are critical for measurement performance. A clean white tile and protective glass are absolutely critical for measurement accuracy.
- The i1Pro 3 device is sensitive to mechanical shocks. To avoid damage during transport the i1Pro 3 must be always shipped in its original packaging.
- Do not store or operate the i1Pro 3 device in dirty, greasy or dusty environments. Do not use the i1Pro 3 in environments with temperatures higher than 35°C (95ºF) or less than 10°C (50ºF). Do not store the i1Pro 3 in environments with temperatures higher than 50°C (122ºF) or less than -10°C (14ºF).The plastic housing of the device may be cleaned using a cloth dampened in water with a soap solution if necessary.
- The protection glass in the limiting aperture of the i1Pro 3 device may become dusty after some time. You can remove the protective glass to clean it and the inside of the limiting aperture:
- Keep the white reference ceramic tile clean at all times. You can clean the white reference ceramic tile with isopropyl alcohol on a soft clean cloth if required.
- Do not touch the diffuser disc of the ambient light measuring head. If necessary, clean it with compressed air or isopropyl alcohol.
- The guide ruler including the “zebra stripe” can be cleaned with isopropyl alcohol when required.
Troubleshooting
Diagnostic Software
Use the i1Diagnostics software if you believe your i1Pro 3 is not functioning
properly.
If the status indicator turns solid red after calibration the i1Pro 3’s
internal sensor diagnostic has discovered a problem.
In this case please perform the following steps:
- Download the i1Diagnostics software at www.xrite.com/i1diagnostics and install the software on your computer.
- Launch i1Diagnostics Software.
- Perform a functionality check by following the instructions given from the i1Diagnostics Software.
- If i1Diagnostics software reports an error store the test report and contact your nearest X-Rite Support Center. As an alternative you can also take a look at our online support resources http://xrite.com/support/i1Profiler to troubleshoot your issue.
Warranty
X-Rite warrants this Product against defects in material and workmanship for a period of twelve (12) months from the date of shipment from X-Rite’s facility, unless mandatory law provides for longer periods. During such time, X-Rite will either replace or repair at its discretion defective parts free of charge.
X-Rite’s warranties herein do not cover failure of warranted goods resulting from: (i) damage after shipment, accident, abuse, misuse, neglect, alteration or any other use not in accordance with X-Rite’s recommendations, accompanying documentation, published specifications, and standard industry practice; (ii) using the device in an operating environment outside the recommended specifications or failure to follow the maintenance procedures in X-Rite’s accompanying documentation or published specifications; (iii) repair or service by anyone other than X-Rite or its authorized representatives; (iv) the failure of the warranted goods caused by use of any parts or consumables not manufactured, distributed, or approved by X-Rite; (v) any attachments or modifications to the warranted goods that are not manufactured, distributed or approved by X-Rite. Consumable parts and Product cleaning are also not covered by the warranty.
X-Rite‘s sole and exclusive obligation for breach of the above warranties shall be the repair or replacement of any part, without charge, which within the warranty period is proven to X-Rite‘s reasonable satisfaction to have been defective. Repairs or replacement by X-Rite shall not revive an otherwise expired warranty, nor shall the same extend the duration of a warranty.
Customer shall be responsible for packaging and shipping the defective product to the service center designated by X-Rite. X-Rite shall pay for the return of the product to Customer if the shipment is to a location within the region in which the X-Rite service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for products returned to any other locations. Proof of purchase in the form of a bill of sale or receipted invoice which is evidence that the unit is within the Warranty period must be presented to obtain warranty service. Do not try to dismantle the Product. Unauthorized dismantling of the equipment will void all warranty claims. Contact the X-Rite Support or the nearest X-Rite Service Center, if you believe that the unit does not work anymore or does not work correctly.
THESE WARRANTIES ARE GIVEN SOLELY TO BUYER AND ARE IN LIEU OF ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIMITED TO THE IMPLIED WARRANTIES OF MERCHANTABILITY, FITNESS FOR A PARTICULAR PURPOSE OR APPLICATION, AND NON-INFRINGEMENT. NO EMPLOYEE OR AGENT OF X-RITE, OTHER THAN AN OFFICER OF X-RITE, IS AUTHORIZED TO MAKE ANY WARRANTY IN ADDITION TO THE FOREGOING.
IN NO EVENT WILL X-RITE BE LIABLE FOR ANY OF BUYER’S MANUFACTURING COSTS, OVERHEAD, LOST PROFITS, GOODWILL, OTHER EXPENSES OR ANY INDIRECT, SPECIAL, INCIDENTAL OR CONSEQUENTIAL DAMAGES BASED UPON BREACH OF ANY WARRANTY, BREACH OF CONTRACT, NEGLIGENCE, STRICT TORT, OR ANY OTHER LEGAL THEORY. IN ANY EVENT OF LIABILITY, X-RITE’S MAXIMUM LIABILITY HEREUNDER WILL NOT EXCEED THE PRICE OF THE GOODS OR SERVICES FURNISHED BY X-RITE GIVING RISE TO THE CLAIM.
Service and Support
RECERTIFICATION
X-Rite recommends an annual recertification of i1Pro 3 spectrophotometer.
Contact your X-Rite dealer or the nearest X-Rite Service Center to find out
more about recertification.
SERVICE CENTERS
Visit our i1Pro 3 website at http://xrite.com to locate the nearest X-Rite
Service Center or contact your X-Rite dealer for more information.
SUPPORT OPTIONS
We are confident you will be pleased with your new X-Rite product. At the same
time, we understand your need for on-demand support and training in today’s
complex and rapidly evolving workflows. Because of this we are making the
following on-line and free support tools available to registered users.
These support tools are found at: www.xrite.com/support/i1Profiler
- Free and unlimited access to the i1Profiler interactive training video; available within the i1Profiler software too.
- Free and unlimited access to the X-Rite on-line Help Desk, which includes detailed and helpful support articles related to frequently asked questions.
- Free access to X-Rite’s e-mail support and call center during the warranty period for issues resulting from general product use provided:
– You are the original registered user of the product
– You are using the current version of i1Profiler software
– You are using currently supported hardware and supported platform (operating system)
Because we recognize the need for advanced support related to using i1Profiler and X-Rite products in various workflows, as well as support in post warranty time periods, we are happy to offer fee based support programs and training options delivered by X-Rite’s highly experienced Color Experts.
For complete information, please visit www.xrite.com/i1Profiler/Supportoptions
Specification