ThermElc TE-02 PRO H Temperature and Humidity Data Logger User Manual
- June 17, 2024
- ThermELC
Table of Contents
User Manual
TE-02 PRO H
Temperature and Humidity
Data Logger
Product Introductions
ThermElc TE-02 PRO H is used for monitoring temperature and humidity of the sensitive goods during storage and transportation. TE-02 Pro H is integrated with USB & Plug-N-Play features for auto-generating secured PDF & CSV report without requiring additional Hardware (cable/interface) or software. TE-02 Pro H dataloggers are equipped with highly precise sensor with accuracy of ±0.5°C for entire range of 30°C to +60°C and ±3% rH.
Main Feature
- Multiple use temperature & humidity logger
- Auto generates PDF reports
- Auto generates CSV reports
- Logging of 32000 points
- No special device driver required
- Temperature and humidity over-limit alarm
Specification of ThermElc TE-02 PRO H
Type | Reusable / Multiple Use |
---|---|
Measurement Range | -30°C to +60°C |
Accuracy | ±0.5℃(-30℃ to +60℃) |
Humidity Range | 0% to 100% rH |
Humidity Accuracy | ±3% rH |
Memory Capacity | 32,000 Values |
Recording Interval | 10 sec to 18 Hrs (Customizable) |
Battery Type | 3V / Replaceable CR2032 |
Dimension | IP65 |
Weight | 89 mm x 36 mm x 16 mm |
Certifications | approx. 30 Grams |
Validation Certificate | EN12830, CE |
Report | Hard Copy Included |
Alarm Range | Self-Generated PDF + CSV |
Connectivity | High & Low Alarm |
Protection Class | USB 2.0 |
Quick Start
Quick start
ThermElc TE-02 Pro H
- Software Downloadhttps://www.thermelc.com/pages/download
- Press start 3 seconds
- Measure Monitoring
- Read
- Report
- Helpshttps://www.thermelc.com/pages/contact-us
Operation Functions
-
Start Recording
Press and hold the START button for approximately 3sec.The OK light is on and the ( ) or (WAIT) on screen indicates the logger is started. -
Mark
When the device is recording, press and hold the START button for more than 3sec, and the screen will switch to the “MARK’ interface. The number of MARK’ will increase by one, indicating data was marked successfully. -
Stop Recording
Press and hold the STOP button for more than 3sec until the STOP ( ****) symbol displays on the screen, indicating stopping recording successfully. -
Switch the display
Shortly press START button to switch to different display interface.
The interfaces shown in sequence are respectively:
Real-time Temperature > Real-time Humidity > LOG > MARK > Temperature Upper Limit > Temperature Lower Limit > Humidity Upper Limit > Humidity Lower Limit. -
Get Report
Connect the data logger to PC via USB, and it will auto-generate PDF and CSV reports.
LCD Display Instructions
**** | Data logger is recording |
---|---|
**** | Data logger has stopped recording |
WAIT | Data logger is in Start Delay status |
**** | Temperature and Humidity is within the limited range |
x and ↑ | Measured value exceeds its upper limit |
x and ↓ | Measured value exceeds its lower limit |