The Model 2305™ Eco-Scale is assembled and calibrated at the factory, then partially disassembled for shipment. Some reassembly and set-up is required before the scale can be operated.
Base Anchoring, Installation & Wiring
Step 1: Remove all parts from carton and inspect for damage – bent metal, broken wires, cracks in digital controller, etc. Any shipping damage must be reported to carrier!
Parts Bag contains
-
- 33” chain, 1) tool tray, & 6) ‘S’ hooks for cylinder restraint and tool hangers
-
- black plastic strain relief & nut
-
- Flat washer for anchor screw under load cell
-
- leveling shims
-
- masonry screws for mounting scale base to floor
-
- silicone grease for sealing load cell to mounting socket
The customer must mount the digital controller, tool rack w/ cylinder restraint chains and hooks on the wall. The base must be leveled for accurate readings by placing shims under the base.
Step 2: Place the scale on a solid, dry, and even surface. Clear the area
of all debris where the scale base will be anchored.
Step 3: Refer to 2305 anchoring, installation & wiring drawings below:
PROCEDURE TO INSTALL MODEL 2305TS CALE BASE:
- Remove 3) screws that hold theload cell in the load cell pocket on TH bar being careful not to pull on the load cell.
- Place scale base where it is to be installed. Drill holes in floor at the three spots marked for anchors. Install anchor bolts using the supplied washer under the load cell – do not tighten yet.
- Level the base in both directions using shims as required. Adjust leveling screw on top of scale, making sure there is at least.040 of an inch clearance between “T” bar and scale platform. This is necessary for scaleto function correctly. When level, tighten 13) anchor bolts, check level again.
- Apply grease supplied, to thoroughly coat the load cell top, bottom and sides, pocket in “T” bar, contact surface and holes in load cell.
- Re-in-stall load cell with (3) screws.
- Feed the wire through the strain relief and finger-tighten the nut.
- Lower the load plate onto the load cell gently, do not drop, and adjust leveling screw again if necessary.
Wire Color Code
- Green = + Signal
- White = – Signal
- Black = – Supply Voltage
- Red = + Supply Voltage
- Violet = Shield
See controller manual for rest of set-up and operating instructions.
Replacement Parts
PN: 2305LC – Canister load cell for Model 2305
©2021 Scaletron Industries Ltd., Plumsteadville, PA Scaletron,