ThermElc TE-02 Temperature Data Logger User Manual
- June 14, 2024
- ThermELC
Table of Contents
ThermElc TE-02 Temperature Data Logger
Product Introductions
ThermElc TE-02 is used for monitoring temperature of the sensitive goods during storage and transportation. After the recording is finished, the ThermElc TE-02 is connected to any USB port and automatically generates a PDF report with the temperature logging results. No additional software is required to read out the ThermElc TE-02.
Main feature
- Multiple use logger
- Auto PDF logger
- Auto generate CSV reports
- Logging of 32,000 values
- Interval of 10 seconds to 18 hours
- No special device driver required
- MKT alarm and Temperature Alarm
PLEASE NOTE:
After the device is configurated for the first time or after re-configuration,
please leave the device in an open environment over 30 minutes. This will
ensure the device is calibrated with the accurate current temperature
Quick start
Software Download
https://www.thermelc.com/pages/download config your parameter
Helpshttps://www.thermelc.com/pages/contact-us
Configuration of a ThermElc TE-02
The device can be configured using the free Data Management software.
- Time zone: UTC
- Temperature scales: ℃ /℉
- Screen display: Always on / Timed
- Log interval: 10 to 18 hours
- Start delays: 0/ timed
- Stop mode: Press button/ disabled
- Time format: DD/MM/YY or MM/DD/YY
- Start mode: Press button or Timed
- Alarm setting: Upper Limit and Lower Limit
- Description: Your reference which will appear on the report
Operation Functions
-
Start Recording
Press and hold the PLAY ( ) button for approximately 3sec.The ‘OK’light is on and the ( ) or ( WAIT ) indicates the logger is started. -
Mark
When the device is recording, press and hold the PLAY ( ) button for more than 3sec, and the screen will switch to the ‘MARK’ interface. The number of ‘MARK’ will increase by one, indicating data was marked successfully.
( Note :One record interval can mark one time only, the logger can mark 6 times in one recording trip. Under the status of start delay, the mark operation is disabled.) -
Stop Recording
Press and hold the STOP ( ) button for more than 3sec until the ‘ALARM’ light is on, and the STOP ( ) symbol displays on the screen, indicating stopping recording successfully. (Note: If the logger is stopped during the status of start delay, a PDF report is generated when inserted into PC but without data.) During the normal recording process, press PLAY ( ) to switch to a different display interface. -
The interfaces shown in sequence are respectively: Realtime Temperature > LOG > MARK > Temperature Upper Limit > Temperature Lower Limit.
-
Get Report
Connect logger to PC via USB, and it will auto-generate PDF and CSV file
LCD Display Instructions
Battery Replacement
Technical Specifications