RICE LAKE WEIGHING SYSTEMS SCT-1SX-PROFINET Indicators and Controller User Guide

June 17, 2024
RICE LAKE WEIGHING SYSTEMS

beko RDNE455E40DXBN Mount Freezer Stainless

Electrical Schematic

  • Load cells excitation: 5 V.
    Load cells output: 6 mV/V max.

  • INPUT: 12 to 48 Vdc,
    OUTPUT: 48Vac or 60Vdc, 0.5 A max

  • For UL approved models: equipment to be powered by 12 to 24 Vdc LPS or Class 2 power source.

  • CONSUMPTION: 4 W max. (without load cells).

Manuals are available from Rice Lake Weighing Systems at www.ricelake.com/manuals
Warranty information is available at www.ricelake.com/warranties

Key Functions

Configuration menu

| Decreases digit / Scrolls down.
| Increases digit / Scrolls up.
| Enters the setup. Selects digit to modify.

| Enters a step / Confirms.
| Clears / Exits a step (no save).
Weighing mode

| Clears the displayed gross weight.
| Short press: executes semiautomatic tare. Long press: allows to enter known tare.
| Activates / deactivates the function.
| Short press executes data transmission on the printer serial port.

Long press: Setpoint configuration.

| ON/Standby of the instrument.

Indicator Light Descriptions

1 Weight on zero.
2 Unstable weight.
3 A tare is active.
4 A function is active.
5 Digital output 1 is active.
6 Digital output 2 is active.

Configuration Menu

  1. Reboot the weight transmitter
  2. Press the    key when display shows the 888888 message:

HOW TO EXIT THE MENU AND SAVE YOUR CONFIGURATION

  1. Press **C** key repeatedly until save? appears; press ** to save or press C** to exit without saving

Maximum Scale Capacity, Increment and Decimal Point Settings


Set the decimal point position and the minimum scale increment*1
` **( 0,001 - 0,002 - 0,005 - 0,01 - 0,02 - 0,05 - 0,1 - 0,2 - 0,5 - 1 - 2 - 5

  • 10 - 20 - 50 ).*`
    Set the maximum scale capacity
    2 (max 999999 ).

Examples:

For a 60000 lb scale, with 2 lb
increment: div.deC = **2**
CapaC = **60000**

For a 10000 g scale, with 0.1 g increment:
**div.deC**= **0,1**
CapaC = **10000,0**

For a 3000 lb scale, with 0.05 lb increment:
div.deC = **0,05**
CapaC = **3000,00**

1 Increment = the amount that the scale will increment by as weight is added or removed.
2 Maximum capacity = the maximum weight that can be measured using the scale you are creating.

Theoretical Calibration

Set the total load cells capacity (up to 999999).
Set the load cells sensitivity (up to 999999) .
Dead load weight (from -9999.9 to 99999.9)

Zeroi Mechanic Tare (pre-tare zeroing)

Zeroing of the pre-tare (or mechanical tare).
This functionality allows to zero the weigh of the scale structure (e.g. empty silo, conveyor, etc.) without changing the calibration in memory

Calibration with Sample Weight

Zero point acquisition.
Sample weight acquisition.

Load Cell Diagnostics (µV/V)

It allows to verify signal of each channel. It must be included into the range 0 to 3 mV/V.
Signal have to be stable and it have to increase by increasing the weight on the scale.

Input Setting

Output Settings

HOW TO PROGRAM SETPOINTS

Fieldbus Settings

Set the IP address:

Weight Filter

The active weight filter is displayed, alternating with the weight value.
Press  keys to scroll through the available filters (from slowest to fastest, F1 to F11).

Programming Errors

MESSAGE DESCRIPTION SOLUTION

| Calibration error| First calibrate the zero point (zero), then proceed with sample weight acquisition (span) (section 9).
| Calibration error| Check the connection of the load cell. Verify the load cell signal is stable, valid and greater than the previously acquired point.
| Calibration error| Increase the calibration weight.
| Calibration error| Check the signal from the load cell increases when weight is incremented on the scale.
| Calibration error| Repeat calibration and verify capacity and division have been correctly set.
| Instrument not configured| Transmitter needs to be configurated.
| Calibration error| Verify the signal from the load cell is not negative.
| Calibration error| Verify the signal from the load cell is not negative.
| Weight unstable| Check in parameter that the signal is stable.
If the connection of the cells is with 4 wires, check that the sense jumpers are inserted.
| A/D converter error| Converter failure. Reboot the instrument.
| Global load cell error| Signal anomaly: check the load cells connection.

Profinet

PROFINET REGISTERS
Data Byte DESCRIPTION
Gross weight 0 (MSB) Bytes 1, 2, 3 and 4 contain the Gross Weight

value.
1
2
3 (LSB)
Net weight| 4(MSB)| Bytes 5, 6, 7 and 8 contain the Net Weight value.
5
6
7 (LSB)
**Input status register| 8 (MSB)| Bit
15(MSB) Bit 14
Bit 13
Bit 12
Bit 11
Bit 10
Bit 9
Bit 8(LSB)|
Bit 15 (MSB) Active channel.
Bit 14 Active Channel
Bit 13 No function.
Bit 12 No function.
Bit 11 No function.
Bit 10 No function.
Bit 9 Status of input n. 2.
Bit 8 (LSB)** Status of input n. 1.| Bit 15| Bit 14| Active Channel
---|---|---
0| 0| Channel 1
0|  | Channel 2
1| 0| Channel 3
1| 1| Channel 4
9(LSB)| Bit 7(MSB)
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0(LSB)| 1 = Scale unloaded (gross weight = 0).
Tare PT (1 = PT tare is active).
Tare (1 = Tare is active).
Overload condition (0 = No; 1 = Overload).
Underload condition (0 = No; 1 = Underload).
Weight Stability (0 = Unstable; 1 = Stable).
Gross Weight Polarity (0 = “+”; 1 = “-”).
Net Weight Polarity (0 = “+”; 1 = “-”).
Command status register| 10(MSB)| Last received command.
11 (LSB)| Bit 7(MSB)
Bit 6
Bit 5
Bit 4
Bit 3
Bit 2
Bit 1
Bit 0(LSB)| Last command result.
Last command result.
Last command result.
Last command result.
Counting of processed commands.
Counting of processed commands.
Counting of processed commands.
Counting of processed commands.
Output status register| 12 (MSB)| No Function.
13(LSB)| Bit 7(MSB)

Bit 2
Bit 1
Bit 0(LSB)| No function.

No function.
Digital output 2 status (0 = OFF; 1 = ON).
Digital output 1 status (0 = OFF; 1 = ON).
Selected page| 14 (MSB)| Shows the value of the selected page (3001).
15 (LSB)
µV| 16 (MSB)| µV value.
17(LSB)

PROFINET REGISTERS FOR COMMAND SENDING
Data Byte DESCRIPTION
Not used 0 Always 0.
Command 1 Main available commands:
Parameter 1 2 (MSB) First parameter of the command.

Parameter is always expressed in absolute mode (no decimals, no sign).
3
4
5 (LSB)
Parameter 2| 6 (MSB)| Second parameter of the command.
Parameter is always expressed in absolute mode (no decimals, no sign).
7
8
9(LSB)
| 10(MSB)| Used in advanced configuration, refer to the complete Fieldbus manual for further information.

31 (LSB)

EXAMPLE 1

For zeroing the weight on the scale:
2. Set the command in byte 2

Byte Value
1 00 Hex
2 01 Hex

EXAMPLE 2

For setting a preset tare of 1000 lb:

  1. Set the tare value in parameter 1 (byte 3, 4, 5, 6)
  2. Set the command in byte 2
Byte Value
1 00 Hex
2 03 Hex
3 (MSB) 00 Hex
4 00 Hex
5 03 Hex
6 (LSB) E8 Hex

CUSTOMER SERVICES

Visit our website www.RiceLake.com
© Rice Lake Weighing Systems Content subject to change without notice
230 W. Coleman St.
• Rice Lake, WI 54868 • USA USA: 800-472-6703
International: +1-715-234-9171
September 29, 2023 www.ricelake.com PN 219720 Rev A