KONICA MINOLTA CA-2500 Series 2D Color Analyzer Instruction Manual

June 16, 2024
KONICA MINOLTA

KONICA MINOLTA CA-2500 Series 2D Color Analyzer

Product Information

Specifications

  • Product Name: 2D Color Analyzer CA-2500 Series
  • Manufacturer: Konica Minolta Optics, Inc.
  • Measurement Type: Luminance and Color Analysis
  • Dimension: Not specified
  • Weight: Not specified
  • Power Supply: AC adapter (supplied as a standard accessory or optional AC-A312)
  • Voltage: 100-240 V
  • Frequency: 50/60 Hz
  • Website: https://manual-hub.com/

Product Usage Instructions

Safety Warnings and Cautions
To ensure correct use of this instrument, please adhere to the following points:

Warnings:

  • Do not use this instrument in places where flammable or combustible gases (gasoline, etc.) are present. Doing so may cause a fire.
  • Always use the AC adapter and power cord supplied as a standard accessory or optional (AC-A312), and connect it to an indoor AC outlet of rated voltage 100-240 V 50/60 Hz. Failure to follow either of these may result in damage to the instrument or AC adapter, fire, or electric shock. Insert the power plug fully and securely. If the power plug is not pushed completely, it may cause fire or electric shock.
  • Do not disassemble or modify this instrument or AC adapter. Doing so may cause fire or electric shock.

Cautions:

  • Use this instrument near an AC outlet for easy plugging or unplugging when using the AC adapter.
  • Do not place this instrument on an unstable or sloping surface that may cause it to drop or overturn. Dropping or overturning may injure someone around. Take care not to drop this instrument when carrying it.
  • When cleaning, disconnect the power plug. Connecting may cause electric shock.

Introduction
Thank you for purchasing our 2D Color Analyzer CA-2500 series. This instrument measures the unevenness of luminance and colour on various display units or projectors, backlights, and car instrument panels in 2 dimensions. Read this manual carefully before use. CA-2500 and CA-2500S are applied as representative examples for explanation and illustration respectively in this instruction manual. Only those with different specifications are indicated model by model.

FAQ

  • Q: Where can I find the full user manual?
    A: The full user manual can be found at https://manual-hub.com/.

  • Q: Can I use this instrument in the presence of flammable gases?
    A: No, it is not recommended to use this instrument in places where flammable or combustible gases are present as it may cause fire.

  • Q: What should I do if I accidentally disassemble the instrument?
    A: If you accidentally disassemble the instrument, please contact a Konica Minolta authorized service facility for assistance. Disassembling the instrument may cause fire or electric shock.

  • Q: How should I clean the instrument?
    A: When cleaning, always disconnect the power plug to prevent electric shock.

Before using this instrument, please read this manual.

Safety Symbols

The following symbols are used in this manual to prevent accidents that may occur as a result of incorrect use of the instrument.

Trademarks

  • “Microsoft®”, “Windows®”, “Windows® XP”, “Windows® 7”, “Windows® 8” and “Excel®” are registered trademarks of Microsoft Corporation in the U.S.A. and other countries.
  • “Intel®” and “Pentium®” are registered trademarks of Intel Corporation in the U.S.A. and other countries. Other company names and product names mentioned in this manual are the registered trademarks or trademarks of their respective companies.

Notes on this manual

  • No part of this manual may be reprinted or reproduced in any form or by any means without the permission of Konica Minolta Optics, Inc.
  • The contents of this manual are subject to change without notice.
  • Every effort has been made to ensure the accuracy of the contents of this manual. However, should you have any questions or comments, or find an error or missing section, please contact a Konica Minolta-authorized service facility.
  • Konica Minolta accepts no responsibility for consequences resulting from failure to follow the instructions outlined in this manual, the condition above notwithstanding.

Safety Warnings and Cautions

To ensure correct use of this instrument, read the following points carefully and adhere to them. After you have read this manual, keep it in a safe place where it can be referred to anytime a question arises.

Introduction

Thank you for purchasing our 2D Color Analyzer CA-2500 series. This instrument measures the unevenness of luminance and colour on various display units or projectors, backlights and car instrument panels in 2 dimensions. Read this manual carefully before use. “CA-2500” and CA-2500S are applied as representative examples for explanation and illustration respectively in this instruction manual. Only those with different specifications are indicated model by model.

Packaging material
Retain accompanying packaging materials (carton, protector, and plastic bag) for future use. This is a delicate measurement instrument. Use packaging materials supplied in purchasing in case this instrument needs to be transferred for such purpose as maintenance in a Konica Minolta-authorized service facility. These packaging materials are useful for minimizing shock or vibration to this instrument in such a situation. Should any of these packaging materials be lost or broken, contact the nearest Konica Minolta- authorized service facility.

Note on use

Operating environment

  • Do not use this instrument outdoors since the standard accessory AC adapter (AC-A312) is designed for indoor use.
  • Never disassemble this instrument for being composed of delicate electronic components.
  • Use this instrument at a rated voltage of 100-240 V . Connect the AC power cord to the AC outlet with a rated voltage of 100- 240 V 50/60 Hz. Connected voltage should not be outside the range of ±10% of nominal.
  • This instrument is classified into a Pollution Degree 2 as an instrument used mainly in manufacturing plants, laboratories, warehouses or equivalents. Use this instrument in a metal dust-free and non-condensing potential environment.
  • This instrument is categorized into Overvoltage Category I as equipment that is powered by an AC adaptor connected to commercially available power.
  • If the protective earth terminal is equipped with the PC to control this instrument, connect the PC terminal to a protective earth outlet. If not, electric shock may occur due to earth leakage.
  • Take care not to enter foreign substances like water or metal in this instrument. Operating in such a state causes danger.
  • Do not use this instrument under direct sunlight or near a heater. The internal temperature of this instrument becomes much higher than the ambient temperature which may break this instrument. Also, be sure to keep good ventilation in using this instrument. Never cover the air duct of this instrument.
  • Avoid rapid change in ambient temperature which may form dew condensation.
  • Avoid using this instrument in extremely dusty or humid places.
  • Use this instrument at ambient temperature between 10ºC and 30ºC and relative humidity of 70% or less with no condensation. Operating this instrument outside the specified temperature and humidity range may unsatisfy its original performance.
  • Do not use this instrument at altitudes above 2000 meters (6560 ft).
  • Even if the product is used within the specified operating temperature/humidity range, the displayed values may be affected by long-term conditions of use. If you have any questions about specifications, please contact the nearest Konica Minolta-authorized service facility.

This instrument

  • Do not subject this instrument to strong impact or vibration.
  • Do not forcibly pull, bend, or apply strong force to the power cord for the attached AC adapter or USB cable. This may result in snapping.
  • Connect this unit to a power source with minimal noise.
  • Should breakage or abnormality be found during operation, switch power off immediately and unplug. Then, refer to “Error Check”.
  • Should this instrument break down, do not try to disassemble and repair it by yourself. Contact the nearest Konica Minolta-authorized service facility.

Lens and filter inside lens mount on body

  • Be sure that there is no dirt on the surface of the lens and filter inside the lens mount on the body before measurement. Correct measurement results cannot be obtained if dirt, dust or handsoil is left or uncleaned.
  • Do not touch the surface of the lens and filter inside the lens mount on the body with your hand.
  • The lens and filter inside the lens mount on the body could become fogged up if the ambient temperature fluctuates under high humidity. Be careful not to generate such conditions for correct measurement cannot be performed.

Storage

Body

  • Cover lens mount with standard accessory mount cover in storing lens in the removed state.
  • Do not store this instrument under direct sunlight or near a heater. The internal temperature of this instrument could become much higher than ambient temperature which may break this instrument.
  • Store this instrument at ambient temperature between 0ºC and 30ºC and a relative humidity 70% or less with no condensation or between 30ºC and 35ºC and relative humidity of 55% or less with no condensation. Storage under high temperatures and humidity may deteriorate the performance of this instrument. For added safety, we recommend storage with a drying agent at room temperature.
  • Even if the product is stored within the specified storage temperature/humidity range, the displayed values may be affected by long-term conditions of storage. If you have any questions about specifications, please contact the nearest Konica Minolta-authorized service facility.
  • Take care not to form condensation. Avoid rapid change in ambient temperature when transferring the body for storage.
  • Put body in the packaging box supplied when purchased or the optional soft case (CA-A60) to store in a safe place.

Lens

  • Cover the lens with a lens cap, or lens hood with a hood cap for storage.
  • When you remove the lens from the body for storage, cover the lens with a lens cap and mount cap.

Temperature/Humidity conditions under which the instrument is used and stored
It is recommended that the instrument be used and stored under standard conditions (Temperature: 23ºC; Relative humidity: 40%), and that areas subject to high temperature and/or humidity be avoided. In addition in order to maintain the measurement accuracy of this instrument, it is recommended that it be inspected regularly about once a year. For details on having the instrument inspected, please contact the nearest Konica Minolta-authorized service facility. Even if the product is used within the specified operating temperature/humidity range or stored within the specified storage temperature/humidity range, the displayed values may be affected by long-term conditions of use or storage.
If the instrument is left under the following high-temperature conditions for a long period of time, the displayed values may change as follows:

  • Temperature: 30ºC;
  • Relative humidity: 70%;
  • Period under these conditions: 720 hours (30 days)
  • Accuracy:
    • Luminance: ±0.4 %;
    • Chromaticity: ±0.003
  • Inter-point error:
    • Luminance: ±0.2 %;
    • Chromaticity: ±0.0003
  • Temperature: 35ºC;
  • Relative humidity: 55%;
  • Period under these conditions: 336 hours (14 days)
  • Accuracy:
    • Luminance: ±1 %;
    • Chromaticity: ±0.006
  • Inter-point error:
    • Luminance: ±0.5 %;
    • Chromaticity: ±0.001

These differences in display values are due to the instrument materials and/or components being affected by the temperature and humidity conditions of long- term use or storage. In particular, optical filters are easily affected by temperature or humidity. Although measures have been taken to improve resistance to temperature/humidity changes, the accumulated effect of long- term use or storage may affect the displayed values.

Cleaning

Body
If this unit becomes dirty, wipe it with a dry and soft cloth. Do not use organic solvents like benzine or thinner and other chemical agent for cleaning. Should none of these methods be helpful, contact the nearest Konica Minolta-authorized service facility.

Lens and filter inside lens mount on the body
Should it be gotten dirt or dust, wipe off with a dry and soft cloth or lens cleaning paper. Do not use organic solvents like benzine or thinner and other chemical agent for cleaning. Should none of these methods be helpful, contact the nearest Konica Minolta-authorized service facility.

Notes on transfer

  • Use the original packaging materials to minimize the shock and vibration to the product when transporting it.
  • Pack the product and all of its accessories together in one package before sending them to us for repair, inspection and/or calibration.

Maintenance
A periodic checkup is recommended annually to maintain the measurement accuracy of the instrument. For details on checkup, contact the nearest Konica Minolta-authorized service facility.

Disposal Method
Make sure that the CA-2500, its accessories and the packing materials are either disposed of or recycled correctly per local laws and regulations.

Standard Accessory

Attached lens cap, mount cap, lens hood and hood cap corresponding with the lens type for each model.

  • Lens Cap
    • Place on the lens for protection in not using this instrument.
    • For standard or telephoto lens
    • For wide lens
  • Mount Cap
    • Place on the lens mount for protection when the lens is removed from the body.
    • For standard or telephoto lens
    • For wide lens
  • Lens Hood
    • For standard or telephoto lens CA-A68
    • For wide lens CA-A69
    • For macro measurement CA-A70
  • Hood Cap
    • Place on the lens for protection in not using this instrument.
    • For standard lens, telephoto lens, or macro measurement CA-A71
    • For wide lens CA-A75
  • Mount Cover CA-A61
    • A place for protection of lens mount of the body when lens is removed.
  • AC Adapter AC-A312
    • Supplies power from AC outlet.
    • Input: 100-240 V 0.75 A 50/60 Hz
    • Output: 12 V 3.0 A
  • Data Management Software CA-S25w
    • For data management of this instrument from a PC.
  • USB Cable (3 m) IF-A18
    • Used for communication between this instrument and PC.
  • Calibration Data DVD
    • Supplied with each lens.

Optional Accessory

  • Soft Case CA-A60
    Used to keep this instrument and accessories or carry them with hand. Never use for transfer.

  • Tripod CS-A3

  • Pan Head CS-A4

  • Mura Evaluation Software CA-Mura
    This program is an add-on software for CA-S25w. It uses the “mura quantitative evaluation method” developed by Sony to provide the “mura indexes” and “mura index images” ref lecting human sensitivity.

System Configuration

CA-2500A (With all lenses)

CA-2500S (With standard lens)

CA-2500W (With wide lens)

CA-2500T (With telephoto lens)

CA-2500SW (With standard and wide lenses)

CA-2500ST (With standard and telephoto lenses)

CA-2500WT (With wide and telephoto lenses)

Names and Functions of Parts

Body

  • Power switch Switches this instrument on/off. [|] side for ON; [ ○ ] side for OFF. Green LED lights up while switched on.…………………………………………… (p.19)
  • AC adapter input terminal To which accessory AC adapter is connected.…………………………… (p.18)
  • USB connector To which USB cable is connected when used with PC.… ……………… (p.19)
  • Air duct Air duct to air-cool interior. Do not cover here. Air-cooling fan built in this instrument operates while switched on.
  • Lens mount To install lens on.
  • Lens attaching portion To which lens hood is attached.
  • Screw hole for fixing Used to set this instrument on a tripod or jig.

Lens

The body and lens arranged in a pair are calibrated. Pair those with identical numbers.

CA-2500A, CA-2500S, CA-2500SW, CA-2500ST

CA-2500A, CA-2500W, CA-2500SW, CA-2500WT

CA-2500A, CA-2500T, CA-2500ST, CA-2500WT

  • Lens hood Protects lens. Remove this once when replacing the lens.
  • The focus ring Adjusts focus.
  • The aperture ring is Being fixed, this should not be used. Do not operate by removing or revolving the screw for this would deteriorate the performance.
  • Macro ring Placed between telephoto lens and body when utilizing telephoto lens for macro. You cannot place 2 types of macro rings simultaneously.

Goods/Environment to Prepare on the Customer’s Side

Get prepared for the below goods/environment before using this instrument.

Controlling PC

  • OS:
    Windows® XP Professional SP3, Windows® XP Professional x64 Edition SP2 Windows® 7 Professional 32bit (x86), Windows® 7 Professional 64bit (x64) Windows® 8 Pro 32bit (x86), Windows® 8 Pro 64bit (x64)
    (English, Japanese, or Chinese (simplified font) version of the above OS) The PC must meet the system requirements specified by the applicable OS or listed below, whichever is higher.

  • CPU Memory Hard disk drive

    • Equivalent to Pentium® 4, 2.8 GHz or better
    • 1024 MB or more
    • At least 80 MB of available disk space is required on the system drive (the drive where the OS is installed).
    • To install the calibration data, the following disk space is required for each lens:
    • For a standard lens: approximately 540 MB
    • For a wide-angle lens: approximately 470 MB
    • For a telephoto lens: approximately 1.3 GB
    • To save measurement data on the hard disk, you need to reserve the following additional space on the hard disk:
    • For example, approximately 110 MB is required to save 10 pieces of measurement data with 980 × 980 resolution.
    • About 1 MB is required for configuration files such as the spot setting file and measurement condition file.
  • Display
    Display hardware capable of displaying 1280 × 768 pixels/65,536 colours (High Color: 16-bit colour) or better

  • Other
    CD-ROM drive Required for installation 1 DVD-ROM drive Required for installation of calibration data
    In place of the above two drives, you can use a single drive that is capable of reading both CD-R and DVD-R media.
    A USB port is required for the connection of the instrument.

In addition, the following items are useful.

Mounting stage
The table on which this instrument is installed. Below are the recommended conditions.

  • Can change measurement distance according to the measurement object.
  • Painted black to prevent the entrance of light from diffusely reflecting.

Blackout curtain (quickly prepared darkroom)/Darkroom

  • Needed to block this instrument and measurement object from outside light.
  • Get prepared for a blackout curtain for a quickly prepared dark room or dark room.

Installing

  • Warning for safety
    Do not use this instrument in places where flammable or combustible gases (gasoline etc.) are present. Doing so may cause a fire.

  • Caution for safety
    Do not place this instrument on an unstable or sloping surface which may drop or overturn it. Dropping or overturning may injure someone around. Take care not to drop this instrument when carrying it.

  • To set this instrument on a tripod or jig, use a screw hole for fixing at the bottom of this instrument. 2 types of holes are available as follows;

  • Tripod screw hole: To set on a tripod. The screw depth is 6.5 mm.

  • ISO screw hole: To set on the jig. User ISO 5 mm screw and screw depth is 6.5 mm.

Placing Lens

Below is the lens type accompanying the respective CA-2500 series. The lens hood serves as the standard corresponding to the lens type.

  • CA-2500A Standard, wide and telephoto lenses, and 2 macro rings
  • CA-2500S Standard lens
  • CA-2500W Wide lens
  • CA-2500T Telephoto lens and 2 macro rings
  • CA-2500SW Standard and wide lenses
  • CA-2500ST Standard and telephoto lenses, and 2 macro rings
  • CA-2500WT Wide and telephoto lenses, and 2 macro rings

Operation procedure

  1. Remove the mount cover from the lens mount on the body.

  2. Place the lens on the lens mount before rotating clockwise.

    • Be sure to insert the lens straight. If forcibly put aslant, the lens mount could be broken.
    • Revolve the lens until fixed.
      For CA-2500A, CA-2500T, CA-2500ST, or CA-2500WT
      In using a macro ring for either of CA-2500A, CA-2500T, CA-2500ST, or CA- 2500WT, sandwich it between lens and lens mount.
  3. Align 2 pins of the lens hood and 2 holes of the body. Then, insert the lens hood as if it covers the lens. Finally, rotate the screw portion of the lens hood clockwise.

    • Select the appropriate lens hood depending on the lens type. An inappropriate lens hood may cause the lens head to be positioned back from the lens hood, which leads to incorrect measurement.
    • Hold the circumference of the lens hood.
    • Rotate the screw portion of the lens hood until fixed.
    • Check whether the body and lens have identical serial numbers. Serial numbers can be found on labels of the body and lens side respectively. To pursue high-accuracy measurement, calibration was done on the maker’s side for each unit before shipment combined with each accessory lens for the CA-2500 series. This means that the attached calibration coefficient data for each unit is valid exclusively for the packaged combination of unit and lens.
    • Set lens type to use in software CA-S25w. (For details, see the instruction manual for data management software CA-S25w.) This enables the application of appropriate calibration coefficient data depending on usage conditions which would finally give correct measurement results.

Replacing lens
Follow the reverse procedure above to replace the lens. First, remove the lens hood and then the lens.

Connecting AC Adapter

Use an accessory AC adapter for a power source.

Warning
(Failure to adhere to the following points may result in death or serious injury.)

  • Always use the AC adapter and power cord supplied as a standard accessory or optional (AC-A312), and connect it to an indoor AC outlet of rated voltage 100-240 V 50/60 Hz. Failure to follow either of these may
    result in damage to the unit, fire or electric shock.

  • If this instrument is not used for a long time, disconnect the AC adapter from the AC outlet. Accumulated dirt or water on prongs of the AC adapter plug may cause fire and should be removed. Be sure to remove it before use.

  • Do not forcibly pull any part of the power cord when unplugging since this may cause fire or electric shock. Gently disconnect by holding the plug. Also, do not handle the power plug with wet hands. Doing so may cause electric shock.

  • Do not forcibly bend, twist or pull the power cord. Also, do not place heavy objects on the power cord, or damage or modify one. Any of these may cause fire or electric shock due to damage to the power cord.

  • Do not disassemble or modify this instrument or AC adapter. Doing so may cause fire or electric shock.

  • Should this instrument or AC adapter be damaged or smoke or odd smell be generated, do not keep using one without correction. Doing so may cause a fire. Switch the power off immediately, and unplug the AC adapter from the AC outlet in such a case.

  • If operating on a battery, remove the battery and contact the nearest Konica Minolta-authorized service facility.

Caution
(Failure to adhere to the following points may result in injury or damage to the instrument or other property.)

  • Use this instrument near the AC outlet for easy plugging or unplugging using an AC adapter.

Connection procedure

  1. Be sure that the power switch is OFF ([ ] mark pressed).
  2. Connect the AC adapter plug to the AC adapter input terminal of the body.
  3. Plug the AC adapter to the outlet (AC 100-240 V, 50/60 Hz).
    • Insert the AC adapter plug all the way seated in the AC outlet.

ON (|) / OFF ( ) of Power Switch

To ON

  1. Press [|] side of the power switch to ON.
    To OFF

  2. Press [ ****] side of the power switch to OFF after completion of measurement.

Connecting with PC

  • Before connecting the CA-2500 body with the PC, install software CA-S25w on the PC. (For details, see the Installation Guide for data management software CA-S25w.)
  • Use an accessory USB cable (3 m) IF-A18 for this purpose.
  • No problem would occur if the USB cable is plugged or unplugged while power is on, but instruction here indicates the case when power is off.

Operation procedure

  1. Switch power off [ ****].
  2. Connect the USB cable to the USB connector on the body.
  3. Insert to the end. Check whether the connection is secure.
  • The communication interface of this instrument complies with USB2.0.
  • Always hold plug portion of the USB cable in unplugging. Never hold the cord portion to unplug.
  • Match the USB cable plug shape with the slot of either the body or PC before inserting.
  • Any USB port is usable on a PC if it has multiple, but abnormal operation could be detected on the PC in case it is used together with another USB device other than CA-2500.

Starting Up and Ending System

Starting up
Switch power on as instructed. If not, an erroneous operation could occur.

  1. Switch the power on for the body. (Press [|] side.)
    • Motor sounds indicate the controller inside the machine starts initialization.
    • The fan inside the body starts to run.
    • A 15-minute warm-up at least (to leave in this state after switching on.) is recommended for accurate measurement.
  2. Start up Windows by turning on the controlling PC.
  3. Connect the PC and CA-2500 with a USB cable.
    • PC recognizes that CA-2500 has been connected.
  4. Select “CA-S25w” from Start → [Program].
    • CA-S25w starts to run.

Connection recognition error

Sometimes, the PC fails to recognize its connection with CA-2500 if CA-2500 was switched on after connecting with a USB cable to the PC that had already started up. Such conditions could not allow control of CA-2500 from software. Start up this instrument according to the procedure above again.

  • In case the PC has already started, unplug the USB cable once and then connect after switching CA-2500 on.
  • Switch CA-2500 on before the PC if the USB cable cannot be unplugged for some reason concerning the setting. Should a connection error occur, unplug the USB cable once and connect again. Then, both can start up normally.

Ending

  1. Select “End CA-S25w” from the “File” of the menu bar.
    • CA-S25w ends.
    • You can finish through standard application operation of OS.
  2. Finish Windows and switch off the PC.
  3. Switch off the body. (Press [ ****] side.)

Setting the measurement distance and the pint ring range index

Setting measurement distance
Five types of objective lenses are available for CA-2500 depending on the size of measuring objects. The relation between the size of measuring objects and the measurement distance for each lens is as follows:

CA-2500A, CA-2500S, CA-2500SW, CA-2500ST

Standard lens

Size of measurement object (One-side length of square)| Measurement distance
98 mm approx.| 250 mm
210 mm approx.| 500 mm
440 mm approx.| 1000 mm
890 mm approx.| 2000 mm

CA-2500A, CA-2500W, CA-2500SW, CA-2500WT

Wide lens

Size of measurement object (One-side length of square)| Measurement distance
145 mm approx.| 200 mm
410 mm approx.| 500 mm
850 mm approx.| 1000 mm
1770 mm approx.| 2000 mm

CA-2500A, CA-2500T, CA-2500ST, CA-2500WT

Telephoto lens| With low-magnification macro ring (Macro 1)|
---|---|---
Size of measurement object (One-side length of square)| Measurement distance| Size of measurement object (One-side length of square)| Measurement distance|
115 mm approx.| 900 mm approx.| 57 mm approx.| 500 mm|
275 mm approx.| 2000 mm approx.| With high-magnification macro ring (Macro 2)|
420 mm approx.| 3000 mm approx.| Size of measurement object (One-side length of square)| Measurement distance|
 |  |
27 mm approx.| 300 mm|

Optical lens features varying light intensity and illuminance distribution of the imaging surface if the focus ring is rotated. Therefore, lens calibration has been done lens by lens according to the distance from the lens to the measurement object (distance indicator of focus ring). Each has its calibration coefficient. Correct calibration coefficient can be applied by selecting what kind of lens is used for which focusing distance, in software CA-S25w.

CA-2500A, CA-2500S, CA-2500SW, CA-2500ST

Standard lens

Distance| With distance label| On indicator
0.25 m| ○| Decimal point indicator position (See Figure 1-1.)
 |  | Intermediate indicator between previous and next reference positions (Center) (See Figure 1-2.)
0.3 m| ○| Decimal point indicator position
 |  | Intermediate indicator between previous and next reference positions (Center)
0.5 m| ○| Decimal point indicator position
 |  | Intermediate indicator between previous and next reference positions (Center) (See Figure 1-3.)
1 m| ○| Center of “1”
 |  | Intermediate indicator between previous and next reference positions (Center)
****| ○| End of focus-ring travel ([∞] side)

CA-2500A, CA-2500S, CA-2500SW, CA-2500ST

Wide lens

Distance| With distance label| On indicator
0.2 m| ○| Indicator position (See Figure 3-1.)
0.24 m| ○| Indicator position
0.3 m| ○| Indicator position (See Figure 3-2.)
0.5 m| ○| Indicator position
1 m| ○| Indicator position
| ○| Indicator position

CA-2500A, CA-2500S, CA-2500SW, CA-2500ST

Wide lens

Distance| With distance label| On indicator
0.9 m| ○| Decimal point indicator position (See Figure 2-1.)
 |  | Intermediate indicator between previous and next reference positions (Center)
1 m| ○| Center of “1”
 |  | Intermediate indicator between previous and next reference positions in 1:2 (short-distance side) (See Figure 2-2.)
 |  | Intermediate indicator between previous and next reference positions in 2:1 (long-distance side) (See Figure 2-3.)
1.5 m| ○| Decimal point indicator position
 |  | Intermediate indicator between previous and next reference positions in 1:2 (short-distance side)
 |  | Intermediate indicator between previous and next reference positions in 2:1 (long-distance side)
3 m| ○| Center of “3”
 |  | Intermediate indicator between previous and next reference positions in 1:2 (short-distance side)
 |  | Intermediate indicator between previous and next reference positions in 2:1 (long-distance side)
**| ○| End of focus-ring travel ( [ ] side)
With low-magnification macro ring (Macro 1)
Distance| With distance label| Precise position on the focus ring
| ○| End of focus-ring travel ( [ ] side)
With high-magnification macro ring (Macro 2)
Distance| With distance label| Precise position on the focus ring
| ○| End of focus-ring travel ( [ ∞** ] side)

Caution
The CA-2500 has 28 calibration points as shown in the table above, at which the measuring accuracy is ensured. When using the instrument at a point other than the calibration points, use the calibration coefficient of the closest calibration point. Doing so may result in a decrease in measuring accuracy.

Example of reference position to read distance
How to read distance in case of standard lens or telephoto lens

The ring (focus ring) on which number values are written rotates in the case of the standard lens or telephoto lens. If the “distance indicator” includes a decimal point, “.” is determined as the reference position; and if it does not include a decimal point, the centre of the number or [ ] is determined as a reference position.

Standard lens

Telephoto lens

How to read distance in case of wide lens
The ring (focus ring) on which the indicator position (the white circle in the figures below) is written rotates in case of a wide lens. Set the indicator position to the line of number value.

Wide lens

Outer Dimensions

(Unit: mm)
When standard lens and lens hood are placed

Error Check

Should an error be found in this instrument, try the corrective actions shown in the following table. If this does not help, this instrument has possibly been broken. Please contact the nearest Konica Minolta-authorized service facility with the indicated error number.

Error

No.

| Symptom| Item to Check| Corrective Action
---|---|---|---
1| The power does not turn on.

Does not normally start up.

| Has AC adapter been plugged in

outlet?

| Connect AC adapter.
Has the AC adapter been connected

with body?

| Connect AC adapter.
Have you connected the correct AC

adapter?

| Use our specified standard accessory AC adapter and power cord or optionally offered AC adapter AC-A312.

  • When you disconnect the wrong adapter and reconnect an AC-A312 adapter, wait at least 5 minutes before turning the

power switch on after turning it off.

Is the AC power within the rating?| Be sure the AC power is within ±10% of the

nominal voltage.

2| The time for machine initialization after switched on varies.| It could vary depending on the machine’s position when switched on. It should not be regarded as a failure if it takes only a few seconds.| If a long time is needed, switch the power off and contact the nearest Konica Minolta- -authorised service facility.
3| PC cannot communicate with CA-2500.| Check whether the PC correctly recognizes CA-2500.| PC could not properly recognize this instrument depending on PC type or how CA-2500 was started up. Restart the PC, software, and CA-2500 according to each

instruction.

Check whether the device driver normally operates with the device driver manager.| If not, uninstall the software once for reinstallation. See the instruction manual for CA-S25w.
Has the USB cable been connected

securely?

| Connect this instrument and PC securely.
Check the standard for USB.| Cannot be used with those supporting except for USB2.0. Use a PC with which USB2.0 is usable.

Main Specification

Model CA-2500S CA-2500W CA-2500T
Light receptor CCD image sensor (monochrome); 2/3-inch; Effective number of

pixels: 1,000 x 1,000 pixels; Equipped with XYZ filter (closely matches CIE 1931 color-matching function) and ND filter
Lens| Interchangeable       Standard, wide, and telephoto lenses; low- magnification and high-magnification macro rings (for use with telephoto lens)
Measurement points (Resolution)| 980 x 980 (490 x 490 or 196 x 196 selectable by using Data Management Software CA-S25w)
Color indication modes| XYZ, Lvxy, Lvu’v’,    uv, Dominant wavelength, Excitation purity, Lv contrast
Display modes| Pseudocolor, Chromaticity diagram, Spot, Cross section, Color difference




Measurement sizes

(length per side of square) (*1)

| ****

Standard lens

| ****

Wide lens

|
Telephoto lens| With low-magnification

macro ring

| With high-magnification

macro ring

Approx. 98 mm or more

(depending on the

distance)

| Approx. 145 mm or more

(depending on the

distance)

| Approx. 115 mm or more

(depending on the

distance)

| ****

Approx. 57mm (Fixed)

| ****

Approx. 27mm (Fixed)




Measurable size for typical measurement distances (size/ distance)

| 98 mm / 250 mm

Approx.

| 145 mm / 200 mm

Approx.

| 115 mm / 900 mm

Approx.

| ****



57 mm / 500 mm Approx. (Fixed)

| ****



27 mm / 300 mm Approx. (Fixed)

210 mm / 500 mm

Approx.

| 410 mm / 500 mm

Approx.

| 275 mm / 2,000 mm

Approx.

440 mm / 1,000 mm

Approx.

| 850 mm / 1,000 mm

Approx.

| ****

420 mm / 3,000 mm

Approx.

890 mm / 2,000 mm

Approx.

| 1,770 mm/ 2,000 mm

Approx.

Measurement luminance

range

(including ND filter use)

| ****

0.05 – 100,000 cd/m2

| ****

0.05 – 100,000 cd/m2

| ****

0.25 – 100,000 cd/m2

| ****

0.25 – 100,000 cd/m2

| ****

0.5 – 100,000 cd/m2


Measurement time (*2)

| Single : Approx. 5 sec. or more; 4-time integration: Approx. 6 sec. or more; 16-time integration: Ap- prox. 10 sec. or more; 64-time integration : Approx. 25 sec. or more; 256-time integration : Approx. 80 sec. or more



Accuracy (*3)

| Luminance| ±3 %| ±3 %| ±3 %| ±3 %| ±3 %
Chromaticity| ±0.005| ±0.005| ±0.005| ±0.005| ±0.005
 | Temperature/humidity drift (within the operating temperature/humidity range)

Luminance: ±2% of change compared to reference temperature and relative humidity of 23°C and 40% Chromaticity: ±0.004 of change compared to reference temperature and relative humidity of 23°C and 40%

Repeatability

(*4)

| Luminance| 0.5 %| 0.5 %| 0.5 %| 0.5 %| 0.5 %
Chromaticity| 0.001| 0.001| 0.001| 0.001| 0.001


Inter-point error (*5)

| Luminance (6)| ±2 %| ±2 %| ±2 %| ±2 %| ±2 %
Chromaticity (
6)| ±0.002| ±0.002| ±0.002| ±0.002| ±0.002
Luminance (7)| ±3 %| ±3 %| ±3 %| ±3 %| ±3 %
Chromaticity (
7)| ±0.003| ±0.003| ±0.003| ±0.003| ±0.003
Other functions| Measurement sync (Synchronization frequency selectable), User calibration Integration function
Interface| USB 2.0 or higher
Operating temperature and humidity

range (*8)

| 10-30°C, Relative humidity 70% or less/No condensation
Storage temperature and humidity range (*8)| 0-30°C, Relative humidity 70% or less/No condensation, 30-35°C, Relative humidity 55% or less/No

condensation


Size

| Body only| 160 (W) × 164 (H) × 192 (D) mm (Height including handle: 211 mm)
When the lens and lens hood are

attached

| 224 (D) mm| 219 (D) mm| 224 (D) mm| 230 (D) mm| 237 (D) mm
Weight| 3.5 kg approx. (when the standard lens and lens hood are attached)
Power source| AC adapter 100-240, 0.75 A, 50/60 Hz


Standard accessories

| Lens Hood| CA-A68| CA-A69| CA-A68| CA-A70
Hood Cap| CA-A71| CA-A75| CA-A71
Calibration data DVD| Included with each lens.
Other| Mount Cover, AC Adapter AC-A312, AC cable, USB Cable IF-A18, Data Management Software CA-S25w, SDK CA2D-SDK
Optional accessories| Soft Case CA-A60, Tripod CS-A3, Pan Head CS-A4, Calibration certificate, Mura Evaluation Software CA-Mura

  1. Error in angle of view: 7%

  2. Measurement time differs depending on the brightness of the measurement object, PC performance, and data processing contents.
    The specifications above were obtained under Konica Minolta’s measurement conditions shown below:

    • PC CPU: Intel® Core™ i7-3770 3.40 GHz
    • Memory: 8 GB
    • Data processing: Pseudocolor display of luminance data
    • Resolution: 980 x 980
    • Shutter speed: Y measurement: 1/64 sec., XZ measurement: 1/32 sec.
    • Measurement subject brightness:
    • Standard/wide lens: Approx. 80 cd/m2, Telephoto lens: Approx. 300 cd/m2
    • Low-magnification macro ring and telephoto lens: Approx. 400 cd/m2,
    • High-magnification macro ring and telephoto lens: Approx. 600 cd/m2
    • The measurement time becomes longer when the object is dark. The longest measurement time is approx. 14 seconds with 1-time integration, approx. 45 seconds with 4-time integration, approx. 3 minutes with 16-time integration, approx. 11 minutes with 64- time integration and approx. 44 minutes with 256-time integration
  3. The specifications above were obtained under Konica Minolta’s measurement conditions shown below:
    Measurement subject brightness:

    • Standard/wide lens: Approx. 35 cd/m2, Telephoto lens: Approx. 140 cd/ m2
    • Low-magnification macro ring and telephoto lens: Approx. 250 cd/m2,
    • High-magnification macro ring and telephoto lens: Approx. 250 cd/m2
    • Distance: Minimum distance of each lens
    • Evaluation: Based on the average obtained within 10% range at the centre of the screen
    • Temperature: 23°C±2°C
    • Relative humidity: 40%±10%
    • Measuring light: White, reference light source
    • Integration: 64 times (Normal mode)
  4. The specifications above were obtained under Konica Minolta’s measurement conditions shown below:

    • Resolution: 196 x 196
    • Shutter speed: Y measurement: 1/64 sec., XZ measurement: 1/32 sec.
    • Gain: Normal (x1),
    • Light intensity level: Midpoint of appropriate light intensity range
    • Evaluation: Based on the maximum repeatability (2 ) of all pixels
    • Temperature: 23°C±2°C
    • Relative humidity: 40%±10%
    • Measurement subject: White, reference light source
    • Integration: 64 times (Normal mode)
  5. The specifications above were obtained under Konica Minolta’s measurement conditions shown below:

    • Measurement subject brightness:
    • Standard/wide lens: Approx. 40 cd/m2, Telephoto lens: Approx. 160 cd/m2
    • Low-magnification macro ring and telephoto lens: Approx. 200 cd/m2,
    • High-magnification macro ring and telephoto lens: Approx. 350 cd/m2
    • Distance: Calibration distance of each lens
    • Resolution: 196 x 196
  6. Evaluation (*6): Based on the maximum/minimum values obtained in a square at the centre of the screen measuring 60% of the height and width of the entire screen

  7. (*7): Based on the maximum/minimum values obtained in the entire screen

    • Temperature: 23°C±2°C
    • Relative humidity: 40%±10%
    • Measurement subject: White, reference light source
    • Integration: 64 times (Normal mode)
  8. Even if the instrument is stored within the specified usage (or storage) temperature-humidity range, the displayed value may change depending on long-period usage or storage conditions.

    • Windows®, Excel® and Word® are registered trademarks or trademarks of Microsoft Corporation in the United States and other countries.
    • Intel® and Pentium® are registered trademarks or trademarks of Intel Corporation in the United States and other countries.
    • The specifications and drawings given here are subject to change without prior notice.
    • For specifications, please contact your nearest Konica Minolta sales office or dealer.

© 2013 KONICA MINOLTA, INC.

References

Read User Manual Online (PDF format)

Loading......

Download This Manual (PDF format)

Download this manual  >>

KONICA MINOLTA User Manuals

Related Manuals