Emerald 61057 Serenity Cell on Cell Alternating Pressure Low Air Loss Mattress System Instruction Manual

June 16, 2024
EMERALD

** Tel:888-363-5563
Email:[email protected]
www.emeraldsupply.net
SELECTIS ALTERNATING PRESSURE PUMP AND MATTRESS
USER MANUAL
Model 61057 & 61058**

Important Safeguards

When using electrical products, especially when children are present, basic safety precautions should always be followed, including the following
READ ALL INSTRUCTIONS BEFORE USING
DANGER
To reduce the risk of electrocution:

  •  Always unplug this product immediately after use.
  • Do not use while bathing.
  • Do not place or store product where it can fall or be pulled into a tub or sink.
  • Do not place or drop into water or other liquids.
  • Do not reach for a product that has fallen into water. Unplug immediately.

WARNING

To reduce the risk of burns, electrocution, fire or injury to persons:

  • A product should never be left unattended when plugged in.
  • Close supervision is necessary when this product is used by, on, or near children or invalids.
  • Use this product only for the intended use as described in this manual.
  • Do not use attachments not recommended by the manufacturer.
  • Never operate this product if it has a damaged cord or plug, if it is not working properly, if it has been dropped or damaged, or dropped into water. Return the product to a service center for examination and repair.
  • Keep the cord away from heated surfaces.
  • Never block the air openings of the product or place it on a soft surface, such as a bed or couch, where the air openings may be blocked. Keep the air openings free of lint, hair and the like.
  • Never drop or insert any object into any opening or hose.
  • Do not use outdoors or operate where aerosol (spray) products are being used.
  • Connect this product to a properly grounded outlet only. See Grounding Instructions.

GROUNDING
Before any connection is made, make certain that this product is connected to a protective earth conductor via the three-wire power cord; the three-blade plug should be inserted only into a socket outlet that provides a protective earth contact.
SAVE THESE INSTRUCTIONS

Product Functions

Intended Use
The Selectis Serenity pump and mattress are intended to reduce the incidence of pressure ulcers while optimizing patient comfort. It also provides the following:

  • Individual home care setting and long-term care
  • Pain management as prescribed by a physician

Pump
The functions of the Selectis Serenity pump are described below. Please refer to the figures of the Selectis pump.

  • Power Switch(1)
    The power switch is at the right side of the pump.
    Turn ON.‘OFF, pump will start/stop operation.

  • Pressure adjust knob(2}
    Turn the soft/firm knob to set a comfortable pressure level.

  • Normal Pressure Indicator(3)
    A visible indicator (green) tells the pressure has reached a preset or user- defined level.

  • Low Pressure Indicator(4)
    A visible indicator (yellow or red) warns the pressure is below a preset or ser-defined level.

  • Alarm Mute(5)
    The audible/visible alarm turns on after 2.5 minutes when the pressure is low.
    To mute the audible alarm, press the mute button.

  • Static Control(6)
    Press to set the air mattress to static mode.

Mattress
The Selectis Serenity pump comes with single cell type mattress and cell on cell type mattress

Installaon & Operang Instrucons

  • Place pump on horizontal surface or hang it by means of the hook.
  • Open maress on the bed. Secure it with end flaps or straps. Air tubes must be at foot end of bed.
  • Connect air feed tubes to air outlet on the maress and pump, ensuring that the tubes are not“kinked”or twisted. Allow space for bed rails to drop freely.
  • Turn on the pump power. The indicator of the power switch will come on. The pump starts to pump air into the maress.
  • The Low Pressure indicator (yellow or red) will come on as inflaon undergoes.
  • The audible alarm is acvated, accompanied by the Low Pressure indicator, to alert that pressure is low. Press the Alarm mute buon to mute the audible alarm, and its indicator will flash.
  • For first inflaon, it is recommended that you press Stac buon to enter a quick operaon.
  • When the pressure reaches the preset level within approximately 30 minutes, the Normal Pressure indicator will come on, and the Low Pressure indicator will turn off.
  • Turn the Pressure Adjust Knob to set a comfortable pressure level from so to firm.
  • Press the Stac buon to set it in stac mode, and the Stac indicator will come on. The stac mode will be started within approximately 6 minutes. Press the Stac buon again to switch back to alternang mode.

In stac mode, the maress providesa firm surface that makes it easier for the paent to transfer or reposion. The stac mode prevents the paent from booming out when in a sing posion.

Environment Requirements

  • Operaton Temperature: 50℉~95℉ (10℃~35℃)
  • Storage Temperature: 5℉~122℉ (-15℃~50℃)
  • Operaton Humidity: 20%~80% non-condensing
  • Storage Humidity: 10%~90% non-condensing

Cleaning Guidelines

  • Can be cleaned by disinfectant soluon
  • Do not immerse the pump in water
  • Do not heat or steam in autoclave
  • The matress and cover are aerated unl they are fully dry

Maintenance

General

  • Check the power cord and plug to see if there are abrasions or excessive wear.
  • Check the maress cover for signs of wear or damage. Ensure the maress cover and tubes are connected together correctly.
  • Plug in the pump unit and check the airflow from the hose connecon port. The airflow should alternate between ports every half-cycle me.
  • Check the air hoses to see if there are any kinks or breaks. For replacement, please contact your local agent or dealer.
  • Make sure the maress tube is well connected.
  • Check the pump unit and make sure both power indicators are off when the switch is turned off.

Low pressure
Examine if there is any air leakage between the pump and the maress connecons or from the air maress tubes:

  • Check connectors between the air maress and pump. If there is any disconnecon, please reconnect it.
  • Check the CPR Valves. Ensure their outlets are sealed.
  • Check the air-connecng tubes. Ensure each single cell is not broken.
  • Set the pressure at Max firm. Keep the tubes fully inflated and inspect for air leakage.
  • Check if there is any air leakage from cells. Ensure no leakage occurs. If any leakage occurs, please contact your local agent or dealer.

Technical Specifications

Pump Mattress
Item: Selectis Serenity Item:Selectis Serenity =61057

Selectis Serenity cell on cell= 61058
Power supply: 120V/60Hz, 1A| Size :80″ (L) x 36″ (W) x 8″ (H)
Air output:8 liter/min.| Top cover: Quilted bi-elastic cover with zipper
Pressure range: 30 mmHg – 60 mmHg| [Air cells:Nylon PVC
Selectis Serenity =61057 =8″ single chamber
Selectis Serenity cell on cell= 61058 =5″ upper layer alternating and 3″ lower layer static
Cycle time: 10 min.| Base:
Selectis Serenity =61057- Nylon / PVC 70D
Selectis Serenity cell on cell= 61058 =Nylon / PVC 840D
Static Mode| 20 air cells
Alarm Reset Mode| CPR valve for emergency procedures
With normal / low pressure light| Transport function
Size: 28cm(L) x 20cm(W) x 10cm(H)| Quick connectors
Weight: 2.2 kg| Pillow function

Warranty

Your Emerald Supply branded product is under warranty to be free of defects in materials and workmanship for 18 months from the date of purchase for the original consumer purchaser.
This device was built to exacting standards and carefully inspected prior to shipment. This Limited 18 Month Warranty is an expression of our confidence in the materials and workmanship of our products and our assurance to the consumer of years of dependable service.
This warranty does not cover device failure due to owner misuse or negligence, or normal wear and tear. The warranty does not extend to non-durable components, such as rubber accessories, casters, and grips, which are subject to normal wear and need periodic replacement.
If you have a question about your Emerald Supply device or this warranty, please contact an authorized Emerald Supply dealer.

Tel:888-363-5563
Email:[email protected]
www.emeraldsupply.net

Documents / Resources

| Emerald 61057 Serenity Cell on Cell Alternating Pressure Low Air Loss Mattress System [pdf] Instruction Manual
61057 Serenity Cell on Cell Alternating Pressure Low Air Loss Mattress System, 61057, Serenity Cell on Cell Alternating Pressure Low Air Loss Mattress System, Cell on Cell Alternating Pressure Low Air Loss Mattress System, Alternating Pressure Low Air Loss Mattress System, Pressure Low Air Loss Mattress System, Low Air Loss Mattress System, Loss Mattress System, Mattress System
---|---