hygiena Aflatoxin B1 Rapid ELISA Kit
Product Information
Specifications
- Product Name: Helica Aflatoxin B1 Rapid ELISA assay
- Model Number: MYC-INS-5004-1.0
- Detection Target: Aflatoxin B1
- Sample Types: Grains, nuts, cottonseeds, cereals, and other commodities including animal feeds
Product Usage Instructions
-
Introduction
The Helica Aflatoxin B1 Rapid ELISA assay is a competitive enzyme-linked immunoassay designed for the quantitative detection of aflatoxin B1 in various samples such as grains, nuts, cottonseeds, cereals, and animal feeds. -
Principle of the Method
The assay utilizes a mouse anti-aflatoxin monoclonal antibody coated on a microwell plate. It works based on the competitive binding between aflatoxin B1 in the sample and a labelled aflatoxin B1 conjugate. The intensity of the colour developed is inversely proportional to the concentration of aflatoxin B1 in the sample. -
Kit Contents
The Helica Aflatoxin B1 Rapid ELISA assay kit includes the following components:- 1X Pouch containing an antibody-coated microwell plate (96 wells in a microwell holder)
- 1X Dilution plate containing 96 non-coated wells (mixing wells)
- 6X Vials of reagents (specific contents not mentioned in the manual)
-
Assay Procedure
- Prepare the samples by following the guidelines provided for each sample type.
- Take out the antibody-coated microwell plate from the pouch and place it on a clean surface.
- Add 100 μL of each sample or standard solution to the appropriate wells of the microwell plate.
- Add 50 μL of the aflatoxin B1 conjugate to each well.
- Cover the plate with a plate sealer and incubate at room temperature for a specific time (time not mentioned in the manual).
- Remove the plate sealer and wash the wells with the provided washing solution.
- Add 100 μL of substrate solution to each well and incubate at room temperature for a specific time (time not mentioned in the manual).
- Add 100 μL of stop solution to each well to stop the reaction.
- Measure the absorbance of each well at the appropriate wavelength using a microplate reader.
-
Interpretation of Results
Compare the absorbance values of the sample wells with those of the standard wells to determine the concentration of aflatoxin B1 in the samples. The manual does not provide specific guidelines for interpretation, so refer to the product datasheet or contact technical assistance for further information. -
Technical Assistance
If you have any technical questions or need assistance with the product, please contact Hygiena’s technical support team at [insert contact details].
FAQ (Frequently Asked Questions)
-
Q: Can this assay be used for human diagnostic purposes?
A: No, this assay is not approved for human diagnostic or treatment purposes. It is intended for research and development, quality assurance, and quality control under the supervision of technically qualified persons. -
Q: Are the assays approved by regulatory agencies?
A: The assays are not approved by the United States Food and Drug Administration or any other US or non-US regulatory agency for use in human diagnostics or treatment. They should not be used as the sole basis for assessing the safety of products for release to consumers. -
Q: What can aflatoxin contamination cause?
A: Aflatoxin contamination can cause various health disorders in humans, including hepatocellular carcinoma, aflatoxicosis, Reye’s syndrome, and chronic hepatitis. In animals, it can lead to death, chronic diseases, reproductive interference, immune suppression, and decreased milk and egg production.
Helica® Aflatoxin B1 Rapid ELISA
For the quantitative detection of Aflatoxin B1 in grains, nuts, cottonseeds,
cereals and other commodities including animal feeds.
Introduction
- Aflatoxins are toxic metabolites produced by a variety of molds such as Aspergillus flavus and Aspergillus parasiticus. They are carcinogenic and can be present in grains, nuts, cottonseed and other commodities associated with human food or animal feeds. Crops may be contaminated by one or more of the four sub-types of Aflatoxin: B1, B2, G1 and G2. Aflatoxin B1 is the most toxic and frequently detected form. The other types present a significant danger if at high concentrations.
- Aflatoxins have been implicated in human health disorders including hepatocellular carcinoma, aflatoxicosis, Reye’s syndrome and chronic hepatitis. Animals are exposed to aflatoxins by consuming feeds that are contaminated by aflatoxin-producing fungal strains during growth, harvest or storage. Symptoms of toxicity in animals range from death to chronic diseases, reproductive interference, immune suppression and decreased milk and egg production.
- Most controlling government agencies worldwide have regulations regarding the amount of aflatoxins allowable in human and animal foodstuffs. Accurate and rapid determination of the presence of aflatoxin in commodities is of paramount importance.
Intended Use
- Hygiena’s Helica® Mycotoxin ELISA kits are user-friendly, cost-effective kits for the detection of mycotoxins in a wide range of commodities including animal feeds, grains, corn and animal urine, designed to protect humans and animals from dangerous side effects of mycotoxins.
- The Helica Aflatoxin B1 Rapid ELISA assay is a competitive enzyme-linked immunoassay intended for the quantitative detection of aflatoxin B1 in grains, nuts, cottonseeds, cereals and other commodities including animal feeds.
- Data obtained from assays should not be used for human diagnostic or human treatment purposes. Assays are not approved by the United States Food and Drug Administration or any other US or non-US regulatory agency for use in human diagnostics or treatment. Assays should not be used as the sole basis for assessing the safety of products for release to consumers. The information generated is only to be used in conjunction with the user’s regular quality assurance program.
- Not approved for clinical diagnosis. Use for research and development, quality assurance and quality control under the supervision of technically qualified persons.
Principle of the Method
The Helica Aflatoxin B1 Rapid ELISA assay is a solid phase direct competitive enzyme immunoassay (ELISA). An aflatoxin-specific antibody optimized to react with B1 is coated to a polystyrene microwell. Toxins are extracted from a ground sample with 70% methanol. The extracted sample and HRP-conjugated Aflatoxin B1 are mixed and added to the antibody-coated microwell. Aflatoxin from the extracted sample and HRP-conjugated Aflatoxin B1 compete to bind with the antibody-coated to the microwell. After this incubation period, the contents of the wells are decanted, and washed and an HRP substrate (TMB) is added which develops a blue colour in the presence of an enzyme. An acidic stop solution is added which changes the chromogen color from blue to yellow. The microwells are measured optically by a microplate reader with an absorbance filter of 450nm (OD450). The optical densities of the samples are compared to the ODs of the kit standards and a result is determined by interpolation from the standard curve.
Kit Contents
Package/Number | Component | Description |
---|
1X Pouch
| Antibody- coated
microwell plate
| 96 wells (12 eight-well strips) in a microwell holder coated with a mouse
anti-aflatoxin monoclonal antibody, Ready-to-Use.
1X Plate| Dilution plate| 96 non-coated wells (12 eight-well strips) in a
microwell holder, ready to use. (Mixing wells)
6X Vials
| Standards| 1.5 mL/vial of Aflatoxin B1 at the following concentrations: 0.0,
0.2, 0.5, 1.0, 2.0 and 4.0 ng/mL in organic solution, Ready-To-Use.
2X Bottles| Conjugate| 2 x 12 mL of Aflatoxin B1 conjugated to peroxidase in
buffer with preservative, Ready-to-Use.
1X Bottle| Substrate| 12 mL stabilized tetramethylbenzidine (TMB), Ready-to-
Use.
1X Bottle| Stop solution| 12 mL acidic solution, ready to use.
1X Pouch| PBS-T powder| PBS with 0.05% Tween® 20, bring to 1 litre with
distilled water and store refrigerated. (Wash buffer)
TWEEN® 20 is a registered trademark of CRODA International Plc.
Materials Required but Not Provided
- A grinder sufficient to render sample to the particle size of fine instant coffee
- Collection container with minimum 125 mL capacity
- Balance with 20 g measuring capability
- Graduated cylinder: 100 mL
- Methanol, reagent grade: 70 mL per sample
- Distilled or deionized water: 30 mL per sample
- Filter Paper: Whatman #1 or equivalent
- Filter Funnel
- Glass tubes
- Centrifuge
- Pipettor with tips: 100 μL and 200 μL
- Timer
- Wash bottle
- Absorbent paper towels
- Microplate reader with 450 nm filter
Storage and Shelf Life
- Store reagents at 2 to 8 °C. Do not freeze.
- Reagents should be used by the expiration date stamped on the individual labels.
- HRP-labeled conjugate and TMB-substrates are photosensitive and are packaged in a light-protective opaque bottle. Store in the dark and return to storage after use.
Precautions and Waste Disposal
General Precautions:
- Bring all reagents to room temperature (19 to 25 °C) before use.
- Do not interchange reagents between kits of different lot numbers.
- Do not use solutions if cloudy or precipitate is present.
- Do not return unused reagents to their original bottles. The assay procedure details the volumes required.
- Adhere to all time and temperature conditions stated in the procedure.
- During the sample extraction, avoid cross-contamination.
- Devices, such as a blender, must be cleaned after each sample preparation.
- Samples tested should have a pH of 7.0 (± 1.0). Excessive alkaline or acidic conditions may affect the test results.
Safety Precautions:
Mycotoxins (aflatoxins, trichothecenes and others) are well-known human
carcinogens and are considered highly toxic. Because mycotoxins can cause
human illness, appropriate safety precautions must be taken and personal
protective equipment worn when handling samples, reagents, glassware and other
supplies and equipment that potentially could be contaminated with mycotoxins.
- Before using this assay, please review the Safety Data Sheet(s) available at www.hygiena.com.
- Refer to your site practices for safe handling of materials.
- It is strongly advised that protective gloves, a lab coat and safety glasses be worn at all times while handling mycotoxin kits and their respective components. Consider all materials, containers and devices that are exposed to samples or standards to be contaminated with mycotoxins.
- Never pipette reagents or samples by mouth.
- Standards are flammable. Caution should be taken in the use and storage of these reagents.
- The stop solution contains sulfuric acid, which is corrosive. Please refer to the SDS. Do not allow to contact skin or eyes. If exposed, flush with water.
Disposal:
Decontaminate materials and dispose of waste per your site practices and as
required by local regulations. Do not dispose of these materials down the
drain. Please note that there is a potential for mycotoxin contamination in or
on any of the kit components provided.
- Dispose of all materials, containers and devices in the appropriate receptacle after use. Before conducting the assay, prepare a waste container to act as a receptacle for your kit waste. Eject contaminated pipette tips and discard all other related materials into this container.
- Once the assay is completed, the container should be treated with a sufficient amount of 5 – 6% sodium hypochlorite (NaOCl) to saturate the contents of the container (approximately 1/10th the volume of the container). The NaOCl will denature the mycotoxins and neutralize the waste, making it safe for disposal. Invert the container several times to thoroughly coat all waste.
- In the case of an accidental toxin spill, treat the spill surface with 5 – 6% NaOCl for a minimum of 10 minutes, followed by 5% aqueous acetone. Wipe dry with absorbent paper towels.
Preparation of Samples
Note:
The sample must be collected according to the appropriate established sampling
techniques.
Extraction Procedure
-
Prepare the extraction solution (70% Methanol) by adding 30 mL of distilled or deionized water to 70 mL of methanol (reagent grade) for each sample to be tested.
-
Grind a representative sample to the particle size of fine instant coffee (95% passes through a 20-mesh screen).
-
Weigh out a 20 g ground portion of the sample and add 100 mL of the extraction solvent (70%methanol).
Note: The ratio of sample to extraction solvent is 1:5 (w/v). -
Mix by shaking in a sealed container or in a blender for a minimum of 2 minutes.
-
Allow the particulate matter to settle, then filter 5 – 10 mL of the extract through a Whatman #1 filter paper (or equivalent) and collect the filtrate to be tested. The sample is now ready for testing.
Assay Procedure
Note:
It is recommended that a multi-channel pipettor be utilized to perform the
assay. If a single channel pipettor is used, it is recommended that no more
than a total of 16 samples and standards (2 test strips) are run.
-
Bring all the reagents to room temperature before use. Prepare wash buffer by reconstituting the PBS-T powder packet by washing out the contents with a gentle stream of distilled or deionized water into a 1- Liter container. Fill to 1 Liter with distilled or deionized water and store refrigerated when not in use.
-
Place one mixing well in a microwell holder for each standard and sample to be tested. Place an equal number of antibody-coated microtiter wells in another microwell holder. Return unused wells to the foil pouch with desiccant and reseal.
-
Mix each reagent by swirling the reagent bottle before use.
-
Dispense 200 μL of the conjugate into each Dilution Well.
-
Using a fresh pipette tip for each, dispense 100 μL aliquots of each standard and sample to appropriate dilution well-containing conjugate. Mix by priming pipettor at least 3 times.
Note: The operator must record the location of each standard and sample throughout the test. -
Using a new pipette tip for each, transfer 100 μL from each dilution well to a corresponding antibody-coated microtiter. Incubate at room temperature for 15 minutes.
Note: The mixing wells contain enough solution to run each standard and/or sample in duplicate (recommended). If running each standard or sample in singlets or if more replicates are needed, the volumes of assay diluent and sample/ standard should be scaled accordingly. -
Decant the contents from microwells into a discard basin. Wash the microwells by filling each with PBS-T wash buffer, then decanting the wash into a discard basin. Repeat wash for a total of 5 washes.
-
Tap the microwells (face down) on a layer of absorbent towels to remove residual buffer.
-
Measure the required volume of substrate reagent (1mL/strip or 120 μL/ well) and place it in a separate container. Add 100 μL to each microwell. Incubate at room temperature for 5 minutes. Cover to avoid direct light.
-
Measure the required volume of stop solution (1mL/strip or 120μL well) and place it in a separate container. Add 100 μL in the same sequence and at the same pace as the substrate was added.
-
Read the optical density (OD) of each microwell with a microtiter plate reader using a 450nm filter
(using an air blank or a differential filter of 630 nm). Record the optical density (OD) of each microwell. -
Setting the zero standard as 100% binding (Bo), calculate % binding (%B) for each standard and sample as a percentage of the zero binding (%B/Bo).
Interpretation of Results
Construct a dose-response curve using the OD values expressed as a percentage of the OD of the zero standard (0.0 ng/mL) against the aflatoxin content of the standard. Unknowns are measured by interpolation from the standard curve. The information contained on the label of a standard vial refers to the contents of that vial. However, the sample has been diluted at a 5:1 ratio with 70% methanol, so the level of aflatoxin shown by the standard must be multiplied by 5 in order to indicate the ng of aflatoxin per gram of commodity (ppb) as follows.
Standard (ng/mL)
|
Commodity (ng/mL)
---|---
0.0| 0.0
0.2| 1.0
0.5| 2.5
1.0| 5.0
2.0| 10.0
4.0| 20.0
The sample dilution results in a standard curve from 1 ppb to 20 ppb. If a sample contains aflatoxin at greater than the highest standard, it should be diluted appropriately in 70% methanol and retested. The extra dilution step should be taken into consideration when expressing the final result.
Assay Characteristics
Performance Data – Precision
Intra-assay Variation
A typical example of the Helica Aflatoxin B1 assay run in duplicate yielded the following standard curve and within assay variation.
Sample (ng/mL)| Mean OD| %B/B 0| CV (%)
0.0| 1.835| 100| 1.9
1.0| 1.626| 88.6| 1.5
2.5| 1.283| 69.9| 1.0
5.0| 0.720| 39.2| 2.4
10.0| 0.338| 18.4| 1.0
20.0| 0.147| 8.0| 1.0
Inter-assay Variation
Between assays, variation is expressed as a percentage of B0 for each standard. n= 6 assays
Sample (ng/mL)| %B/B 0| CV (%)
1.0| 88.5| 4.2
2.5| 75.0| 3.3
5.0| 53.9| 4.8
10.0| 28.1| 8.7
20.0| 12.0| 11.9
The below figure is a representative standard curve for aflatoxin B1 based on the above data table (inter-assay). Data from seven (7) separate standard curves was used to generate this graph.
The limit of detection (LOD) is defined as the mean plus two standard
deviations of multiple determinations of an Aflatoxin-free commodity extract.
As different commodities generate somewhat different zeros due to
‘matrix inhibition’ effects, it follows that the LOD is commodity-specific and
should be measured empirically for each different commodity.
Using the Helica Aflatoxin B1 assay:
- LOD for corn in <1.0 ppb, n=10
- LOD for peanuts is <1.0 ppb, n= 10
The Helica Aflatoxin B1 ELISA has been tested at 1mg/mL (1 million ppb) without evidence of anomalous binding behaviour (high-dose hook effect). Therefore, it may be used to assess gross environmental contamination.
Technical Assistance
For questions or comments, please contact your local distributor. You can also
email techsupport@hygiena.com, visit our
Contact Us page for regional phone numbers or request technical support at
https://www.hygiena.com/hygiena/technical-support-request.html.