hygiena BAX-Q7 Ready Reference for Real Time PCR
Assays

STEP 1: PREPARATION

  • Add 150 μL protease to 12 mL lysis buffer

  • Add 200 µL lysis reagent to cluster tubes

    • Lysis reagent can be stored a 2-8°C for up to two weeks
      Installation Instruction
  • Ensure thermal blocks are pre-heated to 37°C and 95°C prior to use

  • Ensure cooling blocks are stored at 2 – 8°C prior to use
    Installation Instruction

STEP 2: LYSIS

  • Transfer 5 μL* enriched samples to cluster tubes

    • *For E. coli O157:H7 and STEC, use 20 µL
      Installation Instruction
  • Heat cluster tubes (First Stage)

Installation Instruction

  * 37°C for 20 minutes:
  * Campylobacter
  * E. coli O157:H7 Exact
  * E. coli – STEC suite
  * Salmonella
  * Shigella
  * Vibrio
  • Heat cluster tubes (Second Stage)
    **95°C for 10 minutes:
    All targets

Installation Instruction
**

  • Cool cluster tubes for a minimum of 5 minutes in cooling block Unopened processed lysates can be stored at 2-8°C for up to two weeks.
    Installation Instruction

STEP 3: PCR

  • Create rack file, turn on cycler, and initialize
    Installation Instruction

  • Arrange PCR tubes in PCR cooling block with black carrying tray
    Installation Instruction

  • Hydrate PCR tablets with 30 µL lysate from cooled lysates
    For Real-Time Salmonella and E. coli O157:H7 Exact, let hydrated tablets sit in the cooling block for 10-30 minutes prior to placing tubes in Q7 Cycler.
    Installation Instruction

  • On software, click next, place PCR tubes in Q7 cycler and run program
    Installation Instruction

  • Review results on screen
      Negative
     Positive
     Indeterminate
     Signal error

Ready Reference for Real-Time Listeria PCR Assays

STEP 1: PREPARATION

  • Add 150 μL protease to 12 mL lysis buffer
  • Add 200 µL Lysing Agent 2 to protease and lysis buffer mixture
  • Add 200 µL lysis reagent to cluster tubes
  • Ensure thermal blocks are pre-heated to 55°C and 95°C prior to use
  • Ensure cooling blocks are stored at 2 – 8°C prior to use

STEP 2: LYSIS

  • Transfer 5 μL enriched samples to cluster tubes

  • Heat cluster tubes (First Stage)
    55°C for 30 minutes: Genus Listeria L. monocytogenes

  • Heat cluster tubes (Second Stage)
    95°C for 10 minutes: All targets

  • Cool cluster tubes for a minimum of 5 minutes in cooling block
    Unopened processed lysates can be stored at 2-8°C for up to two weeks

STEP 3: PCR

  • Create rack file, turn on cycler, and initialize

  • Arrange PCR tubes in PCR cooling block with black carrying tray

  • Hydrate PCR tablets with 30 µL lysate from cooled lysates
    Recommended: 10 – 30 min hold in cold block for hydrated tablets prior to placing in Q7 cycler

  • On software, click next, place PCR tubes in Q7 cycler and run program

  • Review results on screen
    Negative
    Positive
    Indeterminate
    Signal error