InSight V5
Quick Reference
Guide
10.18SMVer1
This document is a quick reference guide. For more comprehensive information
please refer to your InSight V5 Operator’s Manual
InSight V5 Functions Buttons
How to Run a Sample Using the InSight V5
- Use EDTA K3 anticoagulated blood collection tube
- Fill 2/3’s of the tube with blood
- Ensure a background count and auto clean has been performed before analysis
- Use the touch screen to select ‘Sample Analysis’ screen
- Select ‘Next Sample’
- Enter Patient details including sample ID and species, press ‘OK’ to save
- Thoroughly mix blood on roller mixer or by slowly inverting the tube 10 times
- REMOVE SAMPLE CAP and place sample under the probe
- Push the sampling trigger
- Remove the sample when the analyser ‘beeps’
- Results will automatically display and print out once analysis is complete
How to Enter New QC Normal Ranges
When a new haematology vial is received please check the control lot number.
If it is a new control batch (i.e. it has a different lot number & expiry date
to previous vial) the new normal ranges need to be entered into the InSight V5
control file.
-
Select ‘QUALITY CONTROL’
-
Select ‘QC settings’
-
Select ‘New’
-
Using the information supplied on the new QC assay sheet, enter the following information:
– The new QC batch lot number (i.e. B057)
– Expiry date (dd/mm/yyyy)
– Target values and limits
Press ‘Save’ key to save QC information -
Run the QC
NOTE: Keep QC vials refrigerated and allow at least 15 minutes to reach room temperature (18-30 o C) before use. After sampling wipe the threads of both vial and cap with a lint free tissue. Replace cap and return to refrigerator for maximum open vial stability which is up to 14-28 days if stored at 2-8° C. If you require assistance, please contact your Woodley Equipment Sales Representative or Woodley Equipment Company Technical Support on +44 (0)1204 669033 and press option 1.
How to Run a QC
Woodley Equipment Company recommends a Quality Control (QC) should be analysed on the InSight V5 at the start of each working day.
Why Run a QC?
- To assess the quality and performance of the practice laboratory.
- To ensure that results generated by the InSight V5 are correct.
- QC monitors staff, equipment, reagents and result reporting.
Running A QC
- Navigate to QC test screen by selecting ‘QUALITY CONTROL’
- Select QC Analysis
- Ensure that the QC has been stored correctly, is in date and check that the vial lot number and expiry date correspond to those on screen. If a new batch, enter the new QC reference ranges.
- Thoroughly mix blood by slowly inverting tube 10 times. DO NOT use a rotary mixer. Make sure the sample is warmed to room temperature.
- REMOVE SAMPLE CAP and place QC sample under the probe
- Press back plate to start analysis
- Results will automatically display once analysis is complete
- Check that the QC is within the ranges
- Press ‘QC Graph’ to open Levey-Jennings graph screen
- Sign the QC log sheet under the appropriate date and return QC to the fridge
NOTE: Keep QC vials refrigerated and allow at least 15 minutes to reach
room temperature (18-30 o C) before use. After sampling wipe the threads of
both vial and cap with a lint free tissue. Replace cap and return to
refrigerator for maximum open vial stability which is up to 14-28 days if
stored at 2-8° C.
In the event of a QC failure
- Reprocess the sample
- If it’s still out of limits, repeat with a fresh QC
- Contact your Woodley Sales Representative or Woodley Equipment Company Technical Support on +44 (0)1204 669033 and press option 1.
How to Change a Reagent Pack
Changing reagents when error ‘Insufficient reagent’ shown
- Remove the cap from the new reagent and connect reagent tubes from old reagent bottle to the new bottle.
- Select ‘Remove error’
- Scan RFID card
- The analyser will automatically prime reagents.
Changing reagents through the Reagent Management
- Remove the cap from the new reagent and connect reagent tubes from old reagent bottle to the new bottle.
- Select ‘Reagent Management’ tab at the top of the screen
- Select ‘Setup’
- When indicated place RFID card on the RFID reader on the front of the analyser for each reagent bottle
- Analyser will indicate that the reagent information has been set successfully
- The analyser will prime the reagents and the consumption counters will be reset
- After reagent change, run a Blank measurement to ensure no bubbles remain in the system
**Instrument waste can be safely disposed from the waste container into a sink if accompanied by running water
Standby and Startup
Weekdays
Standby: The InSight V5 is automatically programmed to standby each day
and does not need to be switched off manually
Wakeup/Start up cycle: Touch the screen to exit the standby mode and push
the back plate to start the startup cycle.
Weekend
Shutdown: At the end of the week the InSight V5 needs to be shutdown. Select
the ‘Off’ icon on the main screen. Click ‘Yes’ to perform shutdown. Analyser
will prompt to place cleanser under probe. When prompted the analyser can be
switched off at the back.
Startup Cycle : If the InSight V5 is powered off, firstly switch the
power on at the back of the analyser. Login as user: admin password: admin The
analyser will automatically complete the startup cycle and a background count.
Check that the background measurement is acceptable, then the system is ready
to process a QC and then blood samples.
InSight V5 Haematology Result Flags
Flag Type | Flag information |
---|---|
WBC | Abnormal |
Leucopenia
Neutrophilia
Neutropenia
Lymphocytosis
Lymphopenia
Monocytosis
Eosinophilia
Basophilia
Suspicious| WBC abnormal
Abnor. WBC scattergram
Abnor. WBC histogram
Left Shift?
Immature Cell?
RBC Lyse Resistant?
Abn./Atypical Lym?
Abnormal WBC Channel
Abnormal DIFF Channel
RBC/HGB| Abnormal| Erythrocytosis
Anisocytosis
Macrocytosis
Microcytosis
Anaemia
Hypochromia
Suspicious| Abnor. RBC Distr.
Dimorphologic
Iron Deficiency?
HGB Abnor./Interfere?
RBC Clump?
Abnormal RBC Channel
Abnormal HGB Channel
PLT| Abnormal| Thrombocytosis
Thrombocytopenia
Suspicious| Abnor. PLT Distr.
PLT Clump?
If you require assistance or are unsure of any result or flags, please contact your Woodley Equipment Account Manager or Woodley Equipment Company Technical Support on 0044 (0)1204 669033 and press option 1.
Haematology Sample Handling
Your InSight V5 haematology analyser can only produce results as good as the
sample provided. Sample handling can be the limiting factor in producing
accurate results. Check QC’s regularly, if results are within range then any
aberrant results often indicate improper sample handling.
Venipuncture: A quick non-traumatic venipuncture is critical. Use a 22 gauge
needle or larger to minimise the likelihood of cell damage (haemolysis) and
formation of clots. Avoid repositioning of the needle and/or excessive suction
on the syringe. Fractious cats are likely to have platelet aggregates. If in
doubt consult a blood film.
Sample transfer: Immediately transfer the blood sample into an EDTA tube.
Remove needle and slowly express the sample down the side of the tube to
prevent haemolysis and clots. DO NOT use heparin tubes.
Sample volume: Fill the tube adequately (at least ½ full). Under filled blood
tubes (incorrect ratio EDTA: Blood) will increase the EDTA concentration and
can cause cell shrinkage leading to falsely decreased MCV and HCT.
Sample mixing: Thoroughly mix sample either on a mixer or 8-10 inversions by
hand up until placing the tube in the sample holder (especially horses). Check
the tube for clots prior to sampling, DO NOT test a clotted sample- take a
repeat sample.
Time delay: Blood samples should be processed as soon as possible after
collection. Platelets are affected by time delay as they have a short half
life and can clump even in the presence of an anti-coagulant. Platelet counts
performed after 4-6 hours should be considered suspect. Prolonged sample
storage before processing especially when not refrigerated can lead to
swelling of the RBC leading to a falsely elevated MCV. If there is a time
delay between sample collection and testing store the sample 2-8° C.
Excessive temperature can lead to cell autolysis (self breakdown).
How to Perform a Cleanser Soak
Cleanser Soak should be performed when the background results exceed the
range, bad differential and if a clot remains in the system following other
maintenance procedures. You may also be instructed to perform this by your
Woodley Equipment Representative or by Woodley Equipment Technical Support
when troubleshooting.
Select ‘Service’
Maintenance- Cleanser Soak.
Select ‘Yes’ to perform Cleanser Soak Present the Cleanser under the sample probe and press
‘Aspirate’
The cleanser Soak will take 20minutes
Interpreting Blood Cell Histograms
Histograms demonstrate the distribution of blood cell populations providing a
graphic representation of cell size versus cell number. As cells pass through
the aperture the different sized pulses generated by impedance changes are
sorted and stored in the analyser memory. The memory creates a volumetric
distribution histogram for each blood cell population (red blood cells, white
blood cells and platelets). Histograms give a visual representation of cell
size and number and although not considered to be diagnostic for any
particular condition, they can be used to identify abnormal cell populations
or sampling problems. They also work as a quality check when flags are
produced to provide a useful tool to determine if further investigations, such
as the viewing of a blood film are required. The user must be familiar with
the appearance of normal histograms to make a useful assessment.
As each WBC passes through the flow cell, the laser beam shines through the
cell and the light is scattered depending on the contents of the white cell.
The sizes of the peaks vary depending on the frequency and size of each cell
type.
Therefore, it is possible to observe microcytic (small red cells) and
macrocytic (large red cells) red blood cell populations from the size and
position of the peaks.
Red Blood Cells
White Blood Cells
The intensity of scattered light reflects the blood cell size and
intracellular density.
The low-angle scattered light signal (LS) shows cell size, while the middle-
angle (MS) and high-angle scattered light signal show intracellular
information (nucleus and cytoplasm information). The optical detector receives
this scattered signal and converts it into electrical pulses.
Platelets
Platelet histograms can be used to determine the volume distribution of
platelet populations. Thrombocytopenia is caused by either the failure of
production or by increased consumption. Platelet distribution width (PDW) is
increased in the latter group producing large platelets which are depicted on
the right side of the histogram.
Platelet histograms are also useful to assess if a thrombocytopenia is genuine
or due to a sample quality problem such as aggregates. If platelets are
aggregated an atypical picture is seen with uneven distribution.
InSight V5 Quality Control and Maintenance Log
MONTH:………………………………..
REVIEWED BY:………………………
DATE:………………………………….
Quality Control
NOTE: Process a QC at the start of each working day, compare your results
with results printed on the InSight V5 control data sheet.
In the event of a QC failure 1. Reprocess the sample 2. If still out of
limits, repeat with a fresh QC 3. Contact your Woodley Sales Representative or
Woodley Equipment Company Technical Support on 0044 (0) 1204 669033
Maintenance
Weekly Procedures| WEEK 1| WEEK 2| WEEK 3|
WEEK 4| WEEK 5
---|---|---|---|---|---
Cleanser Soak| | | | |
Clean the probe| | | | |
Monthly| Date Performed & Initials
---|---
Clean Fluidics|
This maintenance log is not intended to indicate when a procedure should be undertaken; it is a record of when it has been undertaken. To confirm that a procedure has been completed please initial the appropriate box. If you require assistance with any of the above procedures please contact your Woodley Equipment Company Sales Representative or Woodley Equipment Company Technical Support on 01204 669033.
InSight V5 Maintenance Schedule
Maintenance Schedule | Maintenance |
---|---|
Daily Maintenance Customer Maintenance | Startup/wakeup cycle (Refer to section |
- Shut down/sleep mode (Refer to section 6)
1. Program for each evening
2. Perform manual shutdown at the weekend then switch off InSight V5 Visually check reagent levels
Clean cover
Blank Measurement
Auto Cleaning Cycle
Weekly Maintenance Customer Maintenance| Cleanser Soak
Clean Probe
Monthly Maintenance Customer Maintenance| Clean Fluidics
Yearly Maintenance
Woodley Service Maintenance
Please contact Woodley Equipment Company Ltd for information about a fully comprehensive service contrac
Tel: +44 (0) 1204 669033
Fax: +44 (0) 1204 669034
Email: sales@woodleyequipment.com| Full Service
NB: Refer to InSight V5 User Manual for detailed instructions of maintenance procedures.
WOODLEY EQUIPMENT COMPANY LTD
nSight V5 INSTALLATION & TRAINING DOCUMENTATION
Name of Practice:………………….
Address:……………………………….
Primary Contact:……………………
InSight V5 Instrument……………
Serial Number:………………………
Phone:…………………………………..
Software Version:……………………
Installation and Training Topics
Name of Installer/Trainer:……………………….
Name of Trainee:…………………………………….
Date of Installation and Training:……………..
The above mentioned trainee has undergone a period of formal training and has
demonstrated competence in the following areas:
- Instrument Overview
a.| Power supply|
---|---|---
b.| Power switch|
c.| LCD display screen|
d.| Function buttons|
e.| External connection|
f.| Open side panel|
g.| Reagent bottles|
h.| Fluidic connections|
i.| Waste|
j.| Built in printer| - Instrument Operation
a.| Correct positioning & working temperature|
---|---|---
b.| Startup/wakeup|
c.| Shutdown|
d.| Background Measurement|
e.| Reagents/RFID card|
f.| Reagent levels|
g.| Reagent expiry|
h.| Priming|
i.| Waste|
j.| Print results| - Instrument Menu
a.| Home|
---|---|---
b.| Sample Analysis|
c.| Review|
d.| QC|
e.| Reagent Managemen|
f.| Setup|
g.| Service|
h.| Calibration| - Sample processing
a.| Basic understanding of sampling process|
---|---|---
b.| Correct sample material and volume|
c.| Define species|
d.| Define sample ID|
e.| Sample mixing|
f.| Remove the sample cap prior to processing|
g.| Process a sample|
h.| Access, review and print results|
i.| Process a ‘background’ measurement|
j.| Can identify testing errors & take appropriate action| - Control processing
a.| Correct control material and assay sheet|
---|---|---
b.| Run control samples in correct control bank|
c.| Input reference ranges into control files|
d.| Correct procedure in the event of a control|
e.| Correct control storage (2-8o C, middle shelf)|
f.| Take from fridge 20 minutes before processing|
g.| Expiry date|
h.| Open-vial stability time (14-28 days)|
i.| Correct control preparation technique|
j.| Wipe vial|
k.| QC recommendations|
l.| Importance of keeping control records| - System Maintenance
a.| Auto clean|
---|---|---
b.| Cleanser soak|
c.| Clean Sample Probe|
d.| Clean fluidics| - System preparation for Service
a.| Preparing for shipment|
---|---|---
b.| Pack instrument appropriately in InSight V5 box| - Troubleshooting
a.| Demonstrate basic knowledge of potential sampling problems and system errors|
---|---|---
b.| Troubleshooting accordingly|
The trainee named above is deemed to be fully competent in the procedures
described and able to complete InSight V5 tests
Signed Trainer:………………….
Signed Trainee:…………………
Name in Full:……………………
Name in Full:……………………
Date:………………………………..
This completed and signed documentation serves as a certificate of InSight V5
installation and training