WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V-CHEM Veterinary Chemistry Analyser User Guide

May 15, 2024
WOODLEY EQUIPMENT COMPANY LTD

WOODLEY EQUIPMENT COMPANY LTD LOGO https://www.praxisdienst.com/en/Veterinary/Laboratory+Supplies/Laboratory+Technology/Analysis+Devices/InSight+V+CHEM+Veterinary+Chemistry+Analyser.html
InSight V-CHEM
Quick Reference Guide WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V CHEM
Veterinary Chemistry AnalyserInSight V-CHEM SM 0718 Ver1.0

This document is a quick reference guide. For more comprehensive information please refer to your InSight V-CHEM Operator’s Manual

External features of the analyser

WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V CHEM Veterinary Chemistry
Analyser - External features

InSight V-CHEM Function Buttons

WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V CHEM Veterinary Chemistry
Analyser - Function Buttons

Warning – To avoid scratching the screen do not use sharp or hard objects to touch the screen

Testing a sample

Sample Preparation

  • Sample type – Serum, Lithium-Heparinised Whole blood or Plasma
  • Sample Volume – 100ul per test panel
  • Testing must be conducted within 60 minutes (at room temperature)
  • To prevent haemolysis of the sample, do not refrigerate, freeze, or shake whole-blood sample.
  • Plasma or serum can be stored at Room Temp for 5 hours prior to testing or up to 48 hours if stored at 2-8C

Rotor Preparation

  • Rotors should be stored at 2-8C and can be used straight from the refrigerator without the need for warming
  • Avoid placing unopened reagent discs in temperatures greater than 25C (77F) for more than 48 hours.
  1. Before using a reagent disc, carefully check the foil pouch for any damage. Tear open the package from the upper right side of the pouch.
  2. Remove the reagent disc gently with finger and thumb
  3. After removal use your other hand to hold the disc by the edge. Avoid contact with disc surface where optical measurement takes place.
  4. Check the reagent disc for any damage – do not use reagent discs that have been dropped or damaged.
  5. After opening the pouch, the disc must be used within 20 minutes. Do not place the disc back in the refrigerator for later use.
  6. Remove the foil tab from the rotor to release the diluent by pulling the tab at a 45° angle.

Dispensing the sample

  1. Attach a new pipette tip to the end of the 100ul pipette provided. Do not touch the end of the tip.

  2. Using your thumb push down the plunger on the top of the pipette to the stop position.

  3. Immerse the pipette tip below the surface of the sample.

  4. Slowly release the plunger to draw up the sample, avoid drawing in any air, as this will cause the disc to fail.

  5. Remove the pipette from the sample tube.

  6. Insert the pipette tip into the sample port (indicated by a red triangle) on the rotor

  7. Keep the reagent disc level and hold the pipette at a 45-degree angle to the surface of the disc.

  8. Gently push down the plunger at the top of the pipette to the stop position. All of the sample should have been expelled into the sample chamber. Keeping the plunger held down gently remove the pipette tip from the sample port. The plunger may now be released, remove the disposable tip from the pipette and discard the tip.

  9. Press the Analyse icon on the touch screen

  10. Select the sample type (whole blood, serum or plasma)

  11. When indicated on the screen, Scan the QR code on the foil pack that the rotor came in

  12. Hold a reagent disc that contains a sample by its edge and the disc kept level.
    Avoid touching the surface of the reagent disc. Place the disc into the drawer until it clicks and gently remove the blue film, press Close to close the disc drawer.

  13. Press the ‘Yes’ icon to confirm the blue film has been removed.

  14. Use the touch screen to enter the species and Patient ID

  15. The fully automated analysis will now commence once the disc drawer closes

  16. When the analysis is completed, the system will store the analysis results and display them on the screen.

To cancel a test in progress, press the X icon on the screen, the system will prompt for confirmation. A reagent disc cannot be re used if a test has been cancelled.

Interpreting the results

WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V CHEM Veterinary Chemistry
Analyser - Interpreting

A typical report printout is shown above. The heading of a report printout includes information such as Sample Patient Name, Owner Name, ID Number, Gender, Age, Sample Type and rotor Lot Number. The test results section is printed in five columns: Analyte Name, Analyte Result, Indicator, Reference Range and Specified Units. The end of the report printout includes information such as the sample indices, test date and time and blank areas for the operator and reviewer to sign.

https://www.praxisdienst.com/en/Veterinary/Laboratory+Supplies/Laboratory+Technology/Analysis+Devices/InSight+V+CHEM+Veterinary+Chemistry+Analyser.html

Accessing Previous Results

The InSight V-CHEM automatically stores tests results for display, printing or Export to a computer.
To access previous results: From the Home Screen > press ‘Previous Results’ then select patients.

WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V CHEM Veterinary Chemistry
Analyser - Previous Results

Enter the patient ID or time range to search for the result.

WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V CHEM Veterinary Chemistry
Analyser - patient ID

The analyser will display all results as filtered, use the up and down arrows to scroll through the list.

WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V CHEM Veterinary Chemistry
Analyser - results as filtered

Date and Time setting

The system date and time will usually be set before installation and the system remembers the date and time between uses.
To change the date and time, from the home screen select ‘Settings’ then press ‘Date/Time’.
Use the up and down arrows to change the day, month, year, hour, minute and seconds. Press ‘Save’ once correct date and time has been set.

WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V CHEM Veterinary Chemistry
Analyser - Time setting

Troubleshooting

When there is a problem with the analyser, a warning and error message will be shown.

Error Code Problem Description Solution
101 Multi-switch fault Please contact Woodley Equipment or your authorised

dealer.
Refer to the Maintenance Manual (Authorised dealers only)
102| +12V power fault
103| -12V power fault
105| AD (AD fault)
107| Optical module fault
108| LED fault
109| Optical output is unstable
202| Drawer open fault
203| Drawer close fault
206| Motor speed fault
213| Temperature fault
214| Insufficient lamp exposure
215| Dark current
216| Light intensity saturation
220| Temperature data
transmission fault
221| PTioo fault
222| Upper NTC fault
223| Lower NTC fault
224| Upper heating film fault
225| Lower heating film fault
301| FPGA fault
302| Scan module fault
303| 6o6 board fault

Error Codes for Reagent Disc and Troubleshooting

Error Code Problem Description Solution
208 Insufficient sample According to Section 3.3, repeat the analysis with a

new reagent disc
209| Insufficient diluent
210| Sample and diluent
not mixed correctly
211| Sample interference,
severe haemolysis,
severe lipaemia or
severe jaundice| Confirm sample quality and repeat the analysis with a new reagent disc. If symptom
persists, use an alternative testing method or send the sample to a reference laboratory.
212| Disc expired| Please use the reagent disc within the validity period.

InSight V-CHEM Rotors

WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V CHEM Veterinary Chemistry
Analyser - CHEM Rotors InSight V-CHEM SM 0718 Ver1.0

Documents / Resources

| WOODLEY EQUIPMENT COMPANY LTD SM 0718 InSight V-CHEM Veterinary Chemistry Analyser [pdf] User Guide
SM 0718, SM 0718 InSight V-CHEM Veterinary Chemistry Analyser, InSight V-CHEM Veterinary Chemistry Analyser, Veterinary Chemistry Analyser, Chemistry Analyser, Analyser
---|---

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals