sauermann KT 220 Data Logger with Universal Probe User Manual

June 15, 2024
sauermann

USER MANUAL
CLASS 220 KISTOCK
KT 220, KH 220 and KTT 220

Safety instructions

1.1 Precautions for use
Please always use the device in accordance with its intended use and within parameters described in the technical features in order not to compromise the protection ensured by the device.
1.2 Symbols used
For your safety and in order to avoid any damage of the device, please follow the procedure described in this user manual and read carefully the notes preceded by the following symbol:
The following symbol will also be used in this user manual:
Please read carefully the information notes indicated after this symbol.

Presentation of the device

2.1 Use
The class 220 KISTOCK dataloggers allow the measurement of several parameters:

  • KT 220: internal measurement of temperature and one external universal input for probe
  • KH 220: internal measurement of temperature, humidity and light and one external universal input for probe
  • KTT 220: thermocouple temperature measurement and two thermocouple external inputs
    This class of devices is available with or without display.
    The communication between the device and the computer is carried out via an USB cable with the female micro-USB connector.

2.2 Applications
The KISTOCK dataloggers are ideal for the control of several parameters (temperature, humidity, light, current, voltage, impulsion, relative pressure…). They ensure the traceability in the food industry environment as well as they  validate the proper functioning of industrial installations.

2.3 References

Device
reference| Display| Internal sensors| External sensors| Parameters| Number of recording points
---|---|---|---|---|---
Number| Type| Number| Type

KT 220 0| Yes| 1| Temperature| 1| Input for
universal
probes*
impulsion, water| Temperature, humidity,
current, voltage, impulsion,  water pressure| 1000000
KT 220 – N| No
KH 220 – 0| Yes| 3| Temperature,
humidity,
light| Temperature, humidity, current, voltage, pressure,  light
KH 220 – N| No
KTT 220 – 0| Yes| –| 2| Inputs for
thermocouple
probes| Temperature
KTT 220 – N| No

  • Input which allows to plug several compatible probes: please see the optional cables and probes page 8.

2.4 Description of the device

sauermann KT 220 Data Logger with Universal Probe -
device

Presentation of the device

2.5 Description of keys
OK key: allows to start or stop the dataset or change of scrolling group (see page 11)
Selection key: allows the functions scroll (see page 11)

2.6 Description of LEDs

sauermann KT 220 Data Logger with Universal Probe - Description of
LEDs

2.7 Connections
The communication between the device and the computer is carried out via an USB cable and with the female micro-USB connector.sauermann KT 220 Data
Logger with Universal Probe - Connections

2.8 Mounting
The class 220 KISTOCK has magnetic mountings, so you can fix it easily.

Technical features

3.1 Technical features of the devices

| KT 220| KH 220| KTT 220
---|---|---|---
Units displayed| °C, °F, °Ctd, °Ftd,%RH,
mV, V, mA, A, bar’| °C, °F, °Ctd, °Ftd,%RH, lux, fc, mV, V, mA,
A, bar’| °C, °F
Resolution| 0.1°C, 0.1°F,
0.1%RH, 1 mV, 0.001 V,
0.001 mA, 0.1 A, 0.1 bar|  01°C 0 1°F 0 1%RH 1 lux, 0. fc 1 mV
‘ ‘ ‘ ‘ • ” ‘ ”
0 001 V, 0.001 mA, 0.1 A, 0.1 bar
‘ ‘| 0.1°C, 0.1°F
External input| Female micro-USB connector
Input for probe| 1 universal input’| 1 universal input’| 2 inputs for
thermocouple probes
(K,1, T, N, S)
Internal sensor| Temperature| Temperature, humidity, light| Temperature
Type of sensor| NTC| Temperature: NTC| Thermocouple
Humidity: capacitive
Light: photodiode
Measuring range| Measuring range of the
internal sensor’:
From -40 to +70°C| Measuring range of the internal sensor’:

Temperature: from 20 to +70°C| –
K: from 200 to +1300°C
1: from -100 to +750°C
T: from -200 to +400°C
N: from -200 to +1300°C
S: from 0 to 1760°C
Humidity: from 0 to 100 % RH
Light: from 0 to +10 000 lux
Accuracies4| ±0.4°C from -20 to 70°C
±0.8°C below -20°C| Temperature. ±0.4°C from 0 to 50°C| K,J,T, N: ±0.4°C from 0 to
1300°C
±(0.3% of reading +0.4°C)
below 0°C
S: ±0.6°C
±0.8°C below 0°C or above 50°C
Humidity’ ±2%RH from 5 to 95%RH
from 15°C to 25°C
1 ight: ±10% of reading +10 lux
Setpoint alarms| 2 setpoint alarms on each channel
Frequency of measurement| From 1 second to 24 hours
Operating temperature| From -40 to +70°C| From -20 to 70°C| From -20 to 70°C
Storage temperature| From -20 to 50°C
Battery life| 4 years6
European directives| 2011/65/EU RoHS II; 2012/19/EU WEEE; 2014/30/EU EMC; 2014/35/EU

  1. Some units are available only with optional probes.
  2. Input which allows to plug different compatible probes: see optional probes and cables page 8.
  3. Other measuring ranges are available according to the connected probe: see optional probes and cables page 8.
  4. All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurement carried out in the same conditions, or carried out with calibration compensation.
  5. Factory calibration uncertainty: ±0.88%RH. Temperature dependence: ±0.04 x (T-20) %RH (if T<15°C or T>25°C)
  6. Non-contractual value. Based on 1 measurement each 15 minutes at 25 °C. A correct operation of the device and the storage conditions must be respected.

3.2 Features of the housing

Dimensions 93.2 x 65.2 x 30.5 mm
Weight 115 g
Display 2 lines LCD screen (for models with display) Screen size: 39 x 34 mm

2 indication LEDs (red and green)
Control| 1 OK key
1 Selection key
Material| Compatible with food industry environment ABS housing
Protection| IP 65: KT 220 IP 54: KTT 220* IP 40: KH 220
PC communication| Female micro-USB connector USB cable
Battery power supply| 1 AA lithium 3.6 V battery
Environmental conditions of use| Air and neutral gases
Hygrometry: in non-condensing condition Maximum altitude: 2000 m

3.3 Optional probes and cables

Reference Description Measuring range

Temperature and humidity probe
KTHA| Interchangeable hygrometry and ambient temperature probe| Hygrometry: from 0 to 100°/0RH Temperature: from -20 to +70°C
KTHD| Remote interchangeable hygrometry and temperature probe
NTC Temperature probe
KSI-50 / KSI-150| IP65 immersion probe| From -40 to +120°C
KSA-150| Ambient use probe| From -40 to +120°C
KSF-2| Wire probe| From -20 to 100°C
KSPP-150| IP68 penetration probe| From -40 to +120°C
KSP-150| IP65 penetration probe| From -40 to +120°C
KCV-220| Probe with velcro| From -20 to +90°C
Current and voltage input cab/es and pulse input cable
KCTD-10-B| Voltage input cable| 0-5 V or 0-10 V
KCCD-02-B| Current input cable| 0-20 mA or 4-20 mA
KCTD-I-B| Pulse input cable| Maximal voltage: 5 V
Type of input: TTL frequency counting Maximal frequency: 10 kHz
Maximum number of recordable
points: 20 000 points

For more details, please see the “Measuring probes for KISTOCK class 220” and “Thermocouple probes” datasheets.
Connect a probe:
➢ Open the mini-DIN connection cap on the bottom of the KISTOCK.
➢ Connect the probe in such a way the mark on the probe is in front of the user.sauermann KT 220 Data Logger with Universal Probe - Connect a
probe

3.4 Dimensions (in mm)
3.4.1 Devices

sauermann KT 220 Data Logger with Universal Probe -
Devices

3.4.2 Wall mountsauermann KT 220 Data Logger with Universal Probe -
Wall mount

4.1 Display

| DATASET is finished.
---|---
| Indicates that one value is being recorded.
It flashes: the DATASET did not start already.
| Flashing slowly: DATASET is between 80 and 90 % of the storage capacity.
Flashing quickly: DATASET is between 90 and 100 % of the storage capacity.
Constant: storage capacity full.
| Constant: indicates that the batteries have to be replaced.
| Screen actualization of measured values.
| The displayed values are the maximum/minimum values recorded for the channels displayed.
| Indication of the direction of exceeding the threshold in the recorded measurement
| Temperature in °Celsius.
Temperature in °Fahrenheit .
| Relative humidity (KH 220).

The selected values to display during the configuration with the KILOG software will scroll on the screen every 3 seconds.
The display can be activated or deactivated via the KILOG software.
At extreme temperatures, the display can become hardly readable and its display speed can slow down at temperatures below 0°C. This has no incidence on the measurement accuracy.

4.2 Functions of LEDs

sauermann KT 220 Data Logger with Universal Probe - Functions of
LEDs

Alarm LED
If the red “Alarm” LED has been activated, it has 3 states:

  • Always OFF: no setpoint alarms has been exceeded
  • Flashing quickly (5 seconds): a threshold is currently exceeded on one channel at least
  • Flashing slowly (15 seconds): at least one threshold has been exceeded during the dataset

Operating LED
If the green “ON” LED has been activated,  it flashes every 10 seconds during therecording period.

4.3 Functions of keys
OK key: allows to start, stop the dataset or change of scrolling group like described in the following tables.
Selection key: allows the scroll values in the scrolling group like described in the following tables.

sauermann KT 220 Data Logger with Universal Probe - Figure
11sauermann KT 220 Data Logger with
Universal Probe - Figure 12

  • Please see the summary table of the groups organisation page 13.
    ** %RH: only the KT 220 and KH 220.
    *** Lux: only the KH 220.

4.3.1 Groups organisation
The table below summarises the groups organisation and measured values available during a measurement dataset.

sauermann KT 220 Data Logger with Universal Probe - Figure
13

4.3.2 Measurements scroll
According to the selected parameters during the configuration and according to the type of device, the measurement scroll is carried out like following:
Temperature Hygrometry Light Parameter 1 of probe Parameter 2 of probe
4.4 PC communication
➢ Insert the CD-ROM in the reader and follow the installation procedure of the KILOG software.

  1. Plug the male USB connector of the cable to an USB connection on your computer**.
  2. Open the USB cap on the right side of the datalogger.
  3. Connect the male micro-USB connector of the cable to the female micro-USB connector of the device.

sauermann KT 220 Data Logger with Universal Probe - PC
communication

4.5 Configuration, datalogger download and data processing with the KILOG software
Please see the KILOG software user manual: “KILOG-classes-50-120-220-320”.
The date and time updates automatically when a new configuration is loaded.

5.1 Replace the battery
With 4 years* battery life, KISTOCK guarantees long-term measurement.
The icon appears when the battery has to be replaced.
To replace the battery:

  1. Unscrew the 4 screws on the back side of the device with a screwdriver.
  2. Remove the back side and the old batteries.
  3. Insert the new battery and respect the polarity.
    BAT
    ➢ Replace the back side and screw it.

sauermann KT 220 Data Logger with Universal Probe -
battery

Only use trademark or high quality batteries in order to guarantee the announced autonomy.
After the battery replacement, the device must be reconfigured.
5.2 Device cleaning
Please avoid any aggressive solvent.
Please protect the device and probes from any cleaning produce containing formalin, that may be used for cleaning rooms and ducts.
5.3 Safety lock wall mount with padlock
➢ Mount the safety lock support on the required place.

  1. Present the KISTOCK datalogger on the support starting with the inferior part
  2. Clip the KISTOCK on the support by falling back the superior part
  3. Insert the padlock to ensure the safety lock function
    sauermann KT 220 Data Logger with Universal Probe -
padlock

➢ To remove the datalogger from the support, proceed on reverse order.
The padlock can be replaced by a fail-safe sealed
The datalogger can be placed on the screw-mount without the safety lock function

Accessories

A calibration certificate is available as option in paper format.
We recommend to carry out a yearly checking.

Accessories References Illustrations
1 M lithium 3.6 V KBL-AA
Safety lock wall mount with padlock KAV-220

Wired extension for class 220 KISTOCK probes
In polyurethane, 5 m length with male and female mini-DIN connectors
Note• several extensions can be wired in order to obtain up to| KRB-220|
25 m cable length
KILOG software: Configuration and data processing software KILOG software allows to configure, save and process your data in a very simple way.| Software only: KILOG-3-N|
Complete set (software +
1 USB cable): KIC-3-N
Data collector
Collects up to 20 000 000 points from one or several KISTOCK directly on-site. Results restitution on PC of realised datasets| KNT-320|
USB micro-USB cable which allows to plug your KISTOCK datalogger to your PC| CK-50|

Only the accessories supplied with the device must be used.

Troubleshooting

Problem Probable cause and possible solution
No value is displayed, only the icons are present. The display is configured

on “OFF”. Configure it on “ON” with the KILOG software (see page 13).
“hi” or “lo” is displayed| The measurement range is exceeded. There is a problem with the sensing element.
The display is completely off* and there is no communication with the computer.| The battery has to be replaced. (see page 14).
The display indicates “- – – -” instead of the measured value.| The probe is disconnected. Plug it again to the datalogger.

BE CAREFUL! Material damages can happen, so please apply the precautionary measures indicated.

sauermanngroup.com

Documents / Resources

| sauermann KT 220 Data Logger with Universal Probe [pdf] User Manual
KT 220 Data Logger with Universal Probe, KT 220, Data Logger with Universal Probe, Logger with Universal Probe, Universal Probe
---|---

Read User Manual Online (PDF format)

Read User Manual Online (PDF format)  >>

Download This Manual (PDF format)

Download this manual  >>

Related Manuals