RHOPOINT ID Imaging Transmission Appearance Meter
Product Information
Specifications
- Manufacturer: Rhopoint Instruments
- Country of Manufacture: United Kingdom
- Model: Rhopoint ID Imaging Transmission Appearance Meter
About the Rhopoint ID Imaging Transmission Appearance Meter
The Rhopoint ID Imaging Transmission Appearance Meter is a high-tech device
that allows for accurate measurement and analysis of the appearance of various
materials. It utilizes state-of-the-art optic and electronic components to
provide reliable and precise results.
Measurement Technology
The Rhopoint ID Imaging Transmission Appearance Meter uses advanced imaging transmission technology to capture and analyze the appearance of materials. It measures parameters such as ASTM Haze, Sharpness, Clarity, and Waviness to provide comprehensive insights into the visual quality of the material.
Measurement Parameters
The Rhopoint ID Imaging Transmission Appearance Meter measures the
following parameters:
- ASTM Haze
- Sharpness
- Clarity
- Waviness
Accessories
The Rhopoint ID Imaging Transmission Appearance Meter comes with the following accessories:
- Power cord
- USB cable
- Calibration standard
- Cleaning cloth
- User manual
Functional Overview
The Rhopoint ID Imaging Transmission Appearance Meter offers various functions to facilitate accurate measurements and analysis:
- Measurement screen icons
- Power control
- Unit on/off switch
- Instrument cleaning
- Taring the unit
Power
The Rhopoint ID Imaging Transmission Appearance Meter can be powered using
the provided power cord. It also has a built-in battery for portable use.
Product Usage Instructions
Storage and Handling
Proper storage and handling of the Rhopoint ID Imaging Transmission
Appearance Meter is crucial to maintain its performance and longevity. Follow
these guidelines:
- Store the instrument in a clean and dry environment.
- Avoid exposing the instrument to extreme temperatures or humidity.
- Handle the instrument with care to avoid any damage.
Switching the Unit On/Off
To switch on the Rhopoint ID Imaging Transmission Appearance Meter:
- Ensure that the power cord is connected to a power source.
- Press the power button located on the instrument.
- The instrument will start up and display the measurement screen.
To switch off the Rhopoint ID Imaging Transmission Appearance Meter:
- Press and hold the power button for a few seconds.
- The instrument will shut down.
Keeping the Instrument Clean
To maintain accurate measurements, it is important to keep the Rhopoint ID Imaging Transmission Appearance Meter clean. Follow these cleaning instructions:
- Use a soft, lint-free cloth to wipe the instrument’s surface.
- Avoid using abrasive cleaners or solvents.
- Regularly clean the instrument’s lens to remove any dust or smudges.
- Ensure that the instrument is dry before use.
Taring the Unit
Taring the unit is necessary to establish a baseline measurement. Follow these steps:
- Place a clean, unobstructed sample on the instrument’s measurement area.
- Press the “Tare” button on the instrument.
- The instrument will set the current measurement as the reference point.
Frequently Asked Questions (FAQ)
Q: Can the Rhopoint ID Imaging Transmission Appearance Meter be used with different materials?
A: Yes, the Rhopoint ID Imaging Transmission Appearance Meter can be used with various materials, including plastics, glass, coatings, and films.
Q: How can I obtain the latest software updates for the Rhopoint ID Imaging Transmission Appearance Meter?
A: To obtain the latest software updates for the Rhopoint ID Imaging Transmission Appearance Meter, please contact the Rhopoint Authorised Distributor for your region. They will provide you with the necessary information and instructions.
Q: Can I use the Rhopoint ID Imaging Transmission Appearance Meter without connecting it to a computer?
A: Yes, the Rhopoint ID Imaging Transmission Appearance Meter can be used as a standalone device without connecting it to a computer. It has a built-in display screen that allows you to view and analyze measurements directly on the instrument.
Operating Instructions
Imaging Transmission Appearance Meter Rhopoint ID
Thank you for purchasing this Rhopoint product. Please read these instructions
carefully before operating this product and retain them for future
reference. The images shown in this manual are for illustrative purposes only.
Manufactured by Rhopoint Instruments in the United Kingdom
This instruction manual contains important information about the setup and use
of the Rhopoint ID Imaging Transmission Appearance Meter. It is therefore
essential that the contents be read before powering up and using the
instrument.
If this instrument is passed to other users you must ensure that the
instruction manual is supplied with the instrument. If you have any questions
or require additional information about this product, please contact the
Rhopoint Authorised Distributor for your region.
The technology and components used in the device are based on state-of-the art
optic and electronics. As part of Rhopoint Instruments commitment to
continually improving the technologies used in their products, they reserve
the right to change information included in this document without prior
notice.
© Copyright 2020 Rhopoint Instruments Ltd. All Rights Reserved.
Rhopoint is a registered trademark or trademark of Rhopoint Instruments Ltd in
the UK and other countries.
Other product and company names mentioned herein may be trademarks of their
respective owner.
No portion of the software, documentation or other accompanying materials may
be translated, modified, reproduced, copied or otherwise duplicated (with the
exception of a backup copy), or distributed to a third party, without prior
written authorisation from Rhopoint Instruments Ltd.
Rhopoint ID Imaging Transmission Appearance Meter
2
Contents
1
Storage and Handling
5
2
About the Rhopoint ID Imaging Transmission Appearance Meter
6
3
Measurement Technology
6
4
Measurement Parameters
7
5
Accessories
9
6
Optional Accessories
9
7
Functional Overview
10
8
Measurement Screen Icons
11
9
Power
13
10
Switching the Unit On/Off
13
11
Warming up the Unit
13
12
Keeping the Instrument Clean
14
13
Taring the Unit
14
14
Measuring ASTM Haze, Sharpness, Clarity, and Waviness with the
16
Rhopoint IDTX
15
Rhopoint ID Transmission Analysis Software
17
16
Measuring the Checking Standard
17
17
Taking an In-Contact Measurement
18
18
Assessing Distance Dependency or Matching End Use Applications
19
19
Measuring ASTM Haze & Sharpness
20
20
Starting a Measurement – Single Reading
20
21
Starting a Measurement – Multipoint Average
21
22
Pass / Fail Tab
23
23
Alignment Tab
25
24
Connection Tab
26
25
SettingsTab
28
26
Rhopoint ID-L Transmission Appearance Analysis Software
31
27
Connecting the Instrument to the PC
32
28
Using the IDTX Software on Multiple Screens
33
Rhopoint ID Imaging Transmission Appearance Meter
3
30
Setting Up the Connection
34
31
Navigating the ID Software
38
32
ID Results Table
39
33
IDTX Software – Live-view and Captured Images
50
35
IDTX Software -Visual Analysis of Results: Sharpness
53
36
IDTX Software -Visual Analysis of Results: Haze
54
37
IDTX Software – Visual Analysis of Results: Waviness
55
38
Notes
55
Rhopoint ID Imaging Transmission Appearance Meter
Storage and Handling
To ensure safe and reliable operation of this instrument it should be placed
on a rigid, flat, level surface. Prevent exposure of the instrument to direct
sunlight Do not place the instrument near electromagnetic fields or in an
environment with excessive vibrations. The instrument’s metal body and
touchscreen are resistant to a variety of solvents; however, the recommended
cleaning method is by using a soft damp cloth. Only use approved accessories
and spare parts. All spare parts are available from Rhopoint Instruments Ltd
and approved distributors. Do not attempt to open the instrument. There are no
user serviceable parts inside. Please contact Rhopoint Instrument or an
approved distributor for your service requirements
Rhopoint ID Imaging Transmission Appearance Meter
About the Rhopoint ID Imaging Transmission Appearance Meter
The Rhopoint ID Imaging Transmission Appearance Meter (IDTX) quantifies the
transmission appearance of transparent, semi-transparent and translucent
samples such as plastic, films, glass and liquids.
The instrument uses a high specification camera to take images of a specially
designed target graticule through the sample to be analysed.
These images are analysed using digital imaging techniques to quantify the
appearance qualities of the material in metrics that are highly correlated to
human visual perception.
The images can then be analysed using the optional viewer software to compare
the transmission characteristics of materials in conditions that replicate the
final use of the material.
Uniquely the instrument can quantify the see-through quality of a sample in
contact with an object or at variable viewing distances up to 45mm.
Measurement Technology
A backlit, high accuracy, reference target graticule functions as the viewed
object, creating a highly defined pattern of light intensities with optimally
sharp transitions between the backlit and masked areas.
A sample placed in direct contact (or at the desired separation) with the
graticule is photographically imaged and analysed to quantify the reduction in
sharpness of the backlit-to-masked (i.e. light-to-dark) transitions of the
graticule compared with its reference image.
At a separation distance of 7.86mm (referred to as 8mm) the instrument
measures transmission haze by calculating the reduction of contrast in black
and white areas of the target graticule.
Rhopoint ID Imaging Transmission Appearance Meter
Measurement Parameters
Haze: Certain transparent materials exhibit an optical property known as haze;
light scattered during transmission through the material causes any object
viewed through it to have less contrast and appear grey and milky.
· When materials with high haze are observed the material becomes visually
intrusive and the viewed object becomes less distinct.
· Samples with low haze have better transmission contrast and are much less
visually intrusive, the viewed object has high contrast.
The Rhopoint IDTX measures Haze at any distance, however at 8mm distance using
the supplied haze spacer the results conform to specifications written for
traditional sphere instrument which comply to ASTMD1003.
Inter-instrument Haze agreement between traditional sphere instruments and
Rhopoint ID for commercial plastic films (0-30 HU, <1000 µm) has been verified
as <0.5% H (SD).
Inter-instrument Haze agreement between traditional sphere instruments and
Rhopoint ID for transparent plastic materials (0-30 HU, <6mm) has been
verified as <1.5% H (SD).
Sharpness: Quantifies any reduction in the optical sharpness of an object
viewed through the tested material. Poor optical sharpness is often caused by
surface roughness or irregularities in the material.
This optical effect is observed as an increase in blurriness when viewing an
object through the sample. Sharpness values are expressed as a percentage
(0-100%), 100% represents a material with perfect sharpness.
Many materials have optical properties where the amount of blurriness is
dependent on the distance between the material and the viewed object.
Anisotropic Sharpness: the process of extruding materials can cause marked
optical differences in the direction of extrusion and orthogonal to that
direction. These differences can appear as a increased blurring in a single
direction or feint lines in the material. The Rhopoint ID software quantifies
this effect as sharpnessthis can be expressed as an average S or in horizontal
SH or vertical Sv directions.
Clarity: Quantifies the blurriness of an object when viewed through a
material, results are proportionate with Sharpness, but measurement scale is
compressed and measurement resolution is reduced.
Clarity is a scale used by traditional haze and clarity meters, when measured
using the 8mm adaptor plate Rhopoint ID Clarity measurements conform to
specifications written for these meters. Inter-instrument Clarity agreement
between Rhopoint ID and traditional sphere instruments for commercial plastic
films (<1000µm) is typically <0.4% C (SD).
Rhopoint ID Imaging Transmission Appearance Meter
7
Inter-instrument Clarity agreement between traditional sphere instruments and
Rhopoint ID for thick transparent plastic materials (<6mm) is typically <0.5%
C (SD).
Waviness: this is an optical effect caused by large structures (0.1-2mm) on
the surface of the material. If the structure is homogeneous it is often
described as orange peel- the surface resembles the peel of an orange.
If the effect is anisotropic visible lines can often be seen when looking
through the material.
Unlike sharpness these larger structures can cause dynamic distortion when the
film is moved over a target image/object which has straight edges- the edges
appear to distort and wave as the material is moved.
Waviness measures the visible distortion of the ID graticule edges through the
material- W is the standard deviation of the edge deformation in µm.
Waviness homogeneity is assessed by comparing W (average waviness) with WH
(horizontal waviness) and Wv (vertical waviness) using Rhopoint ID software.
Transmission: The Rhopoint ID measures Transmission in three ways:
Visible transmission (narrow angle scatter), this is what the camera sees.
Diffuse transmission (Wide angle scatter) and is what the camera doesn’t see
due to being scattered away from the observer.
Total transmission is a calculation of Visible and Diffuse Transmission with
correlation to other haze meters.
The amount of light transmitted to the camera is dependent on the ID of the
sample and the separation distance.
Visible transmission is correlated to the perceived intensity of light
reflected from a viewed object when viewed through the sample.
Distance Dependence: The Rhopoint IDTX-L (see note) can quantify this
relationship between distance and optical performance by measuring in contact
or at defined separation distances.
The ability of the IDTX-L to measure at definable separation distances means
that the suitability of materials for applications with different air gaps can
be directly compared.
Thin films and plastic materials often demonstrate a reduction in sharpness
and increase in haze which peak with a separation distance of 2-4mm between
the sample and material.
Rhopoint instruments offer bespoke adaptor plates or 3D CAD models for local
manufacture of 3-D printed plates.
Rhopoint ID Imaging Transmission Appearance Meter
Accessories
The instrument is supplied as a standard package complete with all accessories
required to power and operate the unit.
· RHOPOINT IDTX Imaging Transmission Appearance Meter · Calibration
certificate · 1 x checking standards · IDTX Haze 8mm spacer · 5mm spacer · DC
power supply and mains cable (UK, EU & US) · USB with operating manual ·
Cleaning cloth · USB 2 to LAN adaptor · LAN cable · USB with operating manual
and software
Optional Accessories
Distance haze measurement pack (B3100-001): 2 x 5mm spacers and 5 x 2mm
spacers
Abrasion tester (B3100-002)
Wet sample adaptor (B3100-003)
Rhopoint ID Imaging Transmission Appearance Meter
9
Functional Overview
1
7
5 2
3
6 4
Label No. Function
1
Display and touch-screen
2
Camera optics
3
Illuminated graticule
4
Levelling feet
5
Power socket
6
USB socket
7
Power button
Rhopoint ID Imaging Transmission Appearance Meter
8 10
Measurement Screen Icons
Tabs
Function
Switch to Home Tab
Switch to Pass/Fail Tab
Switch to Alignment Tabb
Switch to Connection Tab
Metrics
Switch to Settings Tab
Function H-Haze ASTM When the ASTM 8mm adaptor plate is in place, measurements
comply to ASTM D1003.
H-Haze the contrast of a viewed object, cloudiness, milkiness.
T-Visible Transmission the intensity of light from a viewed object.
W-Waviness is a measure of the wavy distortion of objects viewed through the
material.
Rhopoint ID Imaging Transmission Appearance Meter
11
Buttons
S-Sharpness the blurriness of a viewed object (%).
C-Clarity the blurriness of a viewed object (%). When the ASTM 8mm adaptor
plate in place measurements comply to specifications written for Sphere Haze
and Clarity Meters (BYK). Pressing the S-Sharpness Icon toggles it.
Function Take a single or multiple measurement
Tare
Rhopoint ID Imaging Transmission Appearance Meter
12
Power
The Rhopoint IDTX has a supplied low voltage DC power supply. The user must
connect this to the appropriate power input connector.
Switching the Unit On/Off
The Rhopoint IDTX has a built in CPU which takes approximately 30 seconds from
power up to boot into measurement mode. It is important that the use follows
the power up and power down procedures below. To power up the instrument press
and hold the power button until the target graticule illuminates (3 seconds),
release the button and the instrument will continue in the boot up sequence-
please note the front screen remains OFF until the boot sequence is complete
(25 seconds). It is not advisable to switch off the power during boot up as
the instrument may enter a recovery mode. If this occurs, please contact
Rhopoint Service for instructions on re-initializing the instrument. To power
down the unit navigate to the settings tab and then press and hold the power
icon A prompt will be displayed to confirm power down, select Yes to confirm.
The instrument will the begin the power down cycle, indicated by the screen
turning off, the LED and fan will power down shortly after signaling the
instrument can be safely disconnected from power.
Warming Up the Unit
For highest accuracy measurement it is advised that the instrument light
source is left to achieve stability for 4 minutes prior to initial tare.
Rhopoint ID Imaging Transmission Appearance Meter
13
Keeping the Instrument Clean
The measurement graticule is hardwearing but can scratch with the presence of
sharp objects; always place samples and spacers with care. If moderate levels
of grease residues or dust are present on the graticule the measurement may
become affected. It is good practice to wipe the graticule before first use
daily or more often if used in a high contamination environment. We advise
using the enclosed optical cloth or a similar micro-fiber substitute. For
heavy contamination use any commercially available lens cleaning solution on
the graticule. Avoid touching the camera lens if this becomes contaminated,
please contact Rhopoint Service for further advice.
Taring the Unit
After power up it is necessary to tare (zero) the unit. The measurement
buttons are unavailable (greyed out) to the user until the unit has completed
the tare procedure.
To tare the unit press and hold the tare symbol (3 sec) on the touch screen.
Rhopoint ID Imaging Transmission Appearance Meter
14
Once taring is complete measurement buttons are available and the instrument is ready for use.
Tolerance on tared valuesAfter tare, the instrument will measure the graticule as follows. (Average of 5 readings after 4 min warm up period) H: 0.0% (+/- 0.1%) T: 100% (+/- 0.3%) S: 100% (+/- 0.2%)
Rhopoint ID Imaging Transmission Appearance Meter
15
Measuring ASTM Haze, Sharpness, Clarity, and Waviness with the Rhopoint IDTX
Place the 8mm Haze spacer on the instrument.
ASTM will be displayed after the H metric on the screen indicating that
measurements comply to ASTM D1003.
Other metrics are displayed according to the user selection in the settings
menu. The touch screen can be used to view the other metrics.
Rhopoint ID Imaging Transmission Appearance Meter
16
Measuring the Checking Standard
We recommended that the user measures the checking standard weekly or after
moving or transporting the unit. Visually inspect the standard before
measurement, it should be free of dust and contamination. Place the standard
on the instrument.
The standard locates in place by magnets; ensure that the standard can locate properly.
Press the
icon.
Compare the measured values with those recorded on the check standard, if the values are within the stated tolerance the instrument is ready to use.
Please note that the Sharpness value is sensitive to positioning, slightly adjust the position of the standard if the value is out of tolerance.
Cleaning of the checking standard is a delicate procedure and so the user should not use any cleaning cloth on the surface. We recommend if there is dust or debris is present on the standard blow the surface with an optical grade dry air aerosol.
If the standard becomes damaged or contaminated, please contact service@rhopointinstruments.com for more advice.
In case of the standard measuring out of tolerance repeat the tare process and re-measure, if the problem persists contact service@rhopointinstruments.com for more advice.
Rhopoint ID Imaging Transmission Appearance Meter
17
Taking an In-Contact Measurement
The In-contact measurement method’s primary use case is to measure the optical
impact on a viewed object when the sample is in direct contact. To perform a
measurement, place the sample directly on to the measurement graticule.
For fast easy manipulation high clarity samples with limited distance
dependency, the user may manually hold and move the sample. Wear gloves if it
is necessary to avoid contaminating the samples.
Rhopoint ID Imaging Transmission Appearance Meter
18
Assessing Distance Dependency or Matching End Use Applications
To measure distance dependency the user can place the sample spaced away from
the platen. We recommend the use of the optional Rhopoint IDTX spacers. To
achieve the distances required, the user can stack multiple spacers together.
To match an end use application the distance between the sample and platen should match a chosen value based on those in the real-world application.
Rhopoint ID Imaging Transmission Appearance Meter
19
Measuring ASTM Haze & Sharpness
To measure ASTM Haze & sharpness the user should place the ASTM Haze spacer in
position and the sample placed on top.
Measuring with the ASTM haze spacer in place is often the default method for most applications.
Starting a Measurement Single Reading
To initiate a single measurement- press the
button on the touch screen.
Rhopoint ID Imaging Transmission Appearance Meter
20
The instrument will initiate a reading and report the result.
Starting a Measurement Multipoint Average
The Rhopoint ID can calculate statistics for a measured batch of readings. To select the batch size, press on the x1 number in the top right corner of the
button.
Preset batch sizes of 3,4,5,10 or 200 can be selected.
To initiate a multipoint measurement, press the
button on the touch screen.
Rhopoint ID Imaging Transmission Appearance Meter
21
The instrument will initiate a reading and report the result.
The instrument introduces a delay before starting the first measurement and
between readings – this allows the user time to manipulate the sample. To
finish any batch early press the stop button.
After completion, the instrument will report basic statistics for the
measurements. For more complex analysis we recommend the use of the ID-L
software.
To Exit the statistics screen, press the
button.
Rhopoint ID Imaging Transmission Appearance Meter
22
Pass / Fail Tab
The Pass/Fail tab is used to enable/disable on-screen pass/fail indication and
set min/max tolerancing. Pressing this tab enters the Pass/Fail menu.
.
Pass/Fail tolerancing can be set for each/any of the metrics displayed. disabled.
selected Pass/Fail is
To enable Pass/Fail press the
corresponding to the metric required.
There are two options – Pass when greater than the value entered – Pass when less than the value entered
Select an option and enter a target value, press
to save.
When a measurement is made, if the value complies with the conditions selected the value will be displayed in Green indicating a PASS, however if the value does not comply with the conditions selected the value will be displayed in Red indicating a FAIL.
Rhopoint ID Imaging Transmission Appearance Meter
23
If a USB memory device is inserted in the USB port of the instrument (as detailed earlier in Functional Overview) the Pass/Fail tolerancing will be saved to the memory device for future use and recall otherwise they will be held in memory.
To recall the Pass /Fail limits from a memory device press error message will be displayed.
. If a memory device is not inserted an
Pass/Fail limits can be set around a sample measurement which can then be used as a standard when limits are applied.
To do this, place the sample onto the instrument and press
A measurement will be made and the screen below displayed.
Press Yes to save the values or No to continue. Tolerances can then be enabled and set by pressing Once set, the tolerances can be saved by pressing Pass/Fail Tolerancing can also be locked by pressing
for each metric. to unlock enter the code 8157
Rhopoint ID Imaging Transmission Appearance Meter
24
Alignment Tab
To access the alignment tab, press the
tab.
The alignment tab allows the user to see a live view of the graticule to aid in alignment of samples. It also allows the user to align the ROIS of the instrument automatically if the sample used is causing a linear optical offset.
To automatically align the instrument press the
button.
On a successful completion an overlay of the measurement areas will appear.
Rhopoint ID Imaging Transmission Appearance Meter
25
Connection Tab
To access the connection tab, press the
tab.
To enable remote control mode, press the
button and it will turn into
on success.
To disable remote control mode, press the
button and it will turn into
on success.
Rhopoint ID Imaging Transmission Appearance Meter
26
The following advanced connection settings should only be attempted by users
with knowledge of network connections and adapter settings
Advanced Connection Settings
The current network parameters of the device get displayed in the C: (Current)
row of the table. These values by default get automatically generated and are
of the dynamic network type and so will change with every disconnect and
reconnect based on network variables.
Settings a Static Network configuration. This will permanently set the network configuration of the instrument to static, and it will not be able to be put into an automatic network configuration thereafter
To set a Static Network configuration, network details can be entered into the respective
D: (Desired) row and are confirmed with the
button.
A keyboard is required to enter values into these fields, this can be connected into the instruments USB port (refer to Functional Overview for further information)
Rhopoint ID Imaging Transmission Appearance Meter
27
Settings Tab
To access the settings tab, press the
tab.
The settings tab provides system configuration and Power Off.
Pressing the information icon provides version information about the
instrument its last certified date and serial number.
Select Auto Exposure ON or OFF
Rhopoint ID Imaging Transmission Appearance Meter
28
Auto exposure helps improve transmission agreeability between T and T1, T2,
T3. It also has a small effect on H values especially in low transmission
sample. The instrument should be re-tared every time this setting is changed.
To set the Date and Time press A screen wil be displayed as below –
Set the correct Time and Date click on the Day, Month or Year and select using
the drop-down numeric value. Pressing the configuration icon allows displayed
metrics to be turned on or off.
Rhopoint ID Imaging Transmission Appearance Meter
29
Pressing each of the individual metrics enables or disables display. Pressing the volume icon allows the instrument volume to be increased / decreased or muted.
Factory menu used for firmware upgrades and calibration not accessible unless
Service USB key is inserted.
Pressing the power icon turns the unit off.
To return to measurement screen press the
icon,
Rhopoint ID Imaging Transmission Appearance Meter
30
Rhopoint ID-L Transmission Appearance Analysis Software
Minimum Processor
RAM Storage Graphics card Display
System Requirements
Detail 1 gigahertz (GHz) or faster with 2 or more with a 64 bit processor or
System on a Chip (SoC) 4 gigabyte (GB) 64 GB or larger storage device
Compatible with DirectX 12 or later with WDDM 2.0 driver High definition
(720p) display that is greater than 9″ diagonally, 8 bits per colour channel
Optical Transfer Function
Recommended Processor
RAM Storage Graphics card Display
Detail
3 gigahertz (GHz) or faster with 4 or more with a 64 bit processor or System
on a Chip (SoC)
8 gigabyte (GB)
64 GB or larger storage device
Compatible with DirectX 12 or later with WDDM 2.0 driver
High definition (1080p) display that is greater than 24″ diagonally, 8 bits
per colour channel
Installation
The software is located on the supplied USB drive, the required files are
located in the IDTX_Lab folder. Copy the IDTX_Lab folder to your desktop. Open
the IDTX_Lab folder. Double click on the IDTX_Lab.exe to start the installer.
Please follow the installer instructions. PLEASE NOTE THAT DURING INSTALLATION
YOUR PC MAY RESTART WITH NO WARNING, ENSURE ALL WORK IS SAVED BEFORE PROCEDING
Once installed the software will automatically run and the IDTX_Lab folder can
be deleted from the desktop.
Rhopoint ID Imaging Transmission Appearance Meter
31
Buttons
Overview
Function
The Home Tab is the main tab of the program, here you can control the
instrument to take readings and review results.
The Graph Tab allows the user to view data that has been selected for
graphing.
The Allignment Tab allows the user to use an auto alignment procedure to
attempt to correct linear distortion from materials.
The Settings Tab allows the user to scan and connect to discovered
instruments, as well as change configuration settings within the program.
Connecting the Instrument to the PC
Direct connection: use the provided USB 2.0 (or similar) adaptor to connect
the Rhopoint ID to the PC using the provided ethernet cable.
USB 2.0 to Ethernet Connector
Rhopoint ID Imaging Transmission Appearance Meter
32
LAN connection: the software can operate multiple instruments which are
connected to a local network. Connect the instrument directly to the required
network using the provided ethernet cable (or similar).
LAN
Using the IDTX Software on Multiple Screens
Rhopoint IDTX Software supports the use of multiple screens for operation,
viewing and manipulating data, comparing images, and viewing graphs.
For effective data analysis it is preferable to run the software on a PC with
multiple monitors.
Rhopoint ID Imaging Transmission Appearance Meter
33
Setting Up the Connection
Power up the ID instrument and click the
tab on the instrument touch screen.
Click on the
button it will turn to
when a connection is established.
Now navigate to the
tab on the PC software.
There are two methods of discovering an instrument, they are Automatic discovery and manual discovery.
Rhopoint ID Imaging Transmission Appearance Meter
34
Method 1: Automatic Discovery
Automatic discovery will probe the network adapters on the computer using UDP
packets. Any compliant responses are then listed in the discovered device
list.
Press the
button to begin automatically scanning for compliant instruments. The button should
animate for the duration of the scan.
When the scan is complete the
will stop animating.
Rhopoint ID Imaging Transmission Appearance Meter
35
Method 2: Manual Discovery
Manual discovery will only send a UDP packet to the input address allowing for a connection to be made faster or large networks where Method 1 may take a while to complete.
The user can add an instrument manually using its network information found on the instrument.
tab of the
Press the
button to confirm the entry.
Adding a device
On a successful discovery of either of the above methods the instrument will
be displayed in the discovered device list as shown below. If no instruments
appear in the discovered device list, please check your connection.
Rhopoint ID Imaging Transmission Appearance Meter
36
To establish a connection with the Instrument, Click the desired instrument row in the list and then click the button as shown below to add the device to the Primary Instrument slot.
To remove the instrument from the Primary slot, press the
button.
Once the instrument and software connect, the TARE button on the instrument and software will turn from grey to blue.
Rhopoint ID Imaging Transmission Appearance Meter
37
Navigating the ID Software
ID Instrument View
The left panel in the instrument software replicates the touch screen of the instrument and has similar functionality. Interacting with software instrument view screen will change the on-screen display and viceversa, the information remains synchronized during operation.
1. Buttons a. Click the
buttons to begin single or multiple measurement operations, respectively.
b. Click the (X5) on the top right of the
button to change the auto-measurement batch size,
this value can be typed and is any value between 1 and 200 as the batch limit is two hundred measurements.
c. Click and hold the
button to tare the instrument (wait until the button has stopped
spinning before releasing the mouse button).
Rhopoint ID Imaging Transmission Appearance Meter
38
2. Measurement Parameters
a. Click on the
or
b. Click on the
or
icon to switch between transmission and waviness values. icon to switch between sharpness and clarity values.
3. Live view/measurement image
The Instrument View includes a live view from the camera or an image from the
selected/last taken measurement.
To select a measurement, see ID Results Table View a measurement. To switch
on/off live view see ID Settings & Live View Console Toggle Live View.
ID Results Table
The center Panel of the software contains the results for all measurements taken in the current batch. The bottom section contains statistics for all the results listed in the table.
Rhopoint ID Imaging Transmission Appearance Meter
39
View a Measurement
Clicking a previously taken measurement in the table will update the ID
Instrument View and the ID Setting and Live View with the results and image
for the selected measurement. If the instrument is in Live View mode, the
results will update but the camera will continue to show a live view of the
graticule.
Deleting a Measurement
Pressing delete on a keyboard will delete the currently highlighted
measurement. Click in the tick box for each measurement will select it for
global deletion, clicking the delete all “ticked” measurements.
button will
Customize the Results Table Parameters
The Results Table can be customized to any selected results or comments from the Full Data Table.
To view the Full Data Table, click on the
button.
This will bring up the expanded table window for use with multiple screens and the ability to add hidden columns back to the main table.
Rhopoint ID Imaging Transmission Appearance Meter
40
Clicking on the tick box in the top corner of any parameter will include/exclude it from the ID Results Table. Parameters can be dragged and dropped to rearrange columns as required and results can be copied to the clipboard by pressing Ctrl + C. This is available in both tables
Parameter Description
S – Sharpness the blurriness of a viewed object (%). H – Haze the contrast of
a viewed object, cloudiness, milkiness. T – Visible transmission VT (narrow
angle scatter) and is what the camera sees DT – Diffuse transmission (Wide
angle scatter) and is what the camera doesn’t see due to being scattered away
from the observer. TT – is a calculation of Total transmission to be compared
to other haze meters and is the sum of VT and DT
Rhopoint ID Imaging Transmission Appearance Meter
41
Paremeter Description
Sharpness, Clarity & Haze parameters are calculated from image analysis of
seven areas of the target graticule.
Symbol H HH HV
HH1 HH2 HH3 HH4
HV1 HV2 HV3 HV4
S SH SV SH1 SH2 SH3 SH4
SV1 SV2 SV3 SV4
W WH WV WH1 WH2 WH3 WH4
WV1 WV2 WV3 WV4
C CH CV CH1 CH2 CH3 CH4
CV1 CV2 CV3 CV4
Parameter Haze
Horizontal Haze Vertical Haze
Horizontal Haze H1, H2, H3, H4 Vertical Haze
V1, V2, V3
Sharpness Horizontal Sharpness
Vertical Sharpness Horizontal Sharpness
H1, H2, H3, H4 Vertical Sharpness
V1, V2, V3
Waviness Horizontal Waviness
Vertical Waviness Horizontal Waviness
H1, H2, H3, H4 Vertical Waviness
V1, V2, V3
Clarity Horizontal Clarity
Vertical Clarity Horizontal Clarity
H1, H2, H3, H4 Vertical Clarity
V1, V2, V3
Calculation Average of HH & HV Average of HH1 HH2 HH3 HH4 Average of HV1 HV2
HV3 HV4 Calculated using Michelson contrast of B/W zones adjacent to edge
transition Calculated using Michelson contrast of B/W zones adjacent to edge
transition
Average of SH & SV Average of SH1 SH2 SH3 SH4 Average of SV1 SV2 SV3 SV4
Calculated using edge spread function of vertical
graticule edges Calculated using edge spread function of horizon-
tal graticule edges
Average of WH & WV Average of WH1 WH2 WH3 WH4 Average of WV1 WV2 WV3 WV4
Calculated using edge deformation analysis
Calculated using edge deformation analysis
Average of HH & HV Average of CH1 CH2 CH3 CH4 Average of CV1 CV2 CV3 CV4
Calculated using regression analysis- C% (BYK)
vs. ID Sharpness Calculated using regression analysis- C% (BYK)
vs. ID Sharpness
Rhopoint ID Imaging Transmission Appearance Meter
42
Transmission
Transmission values are calculated using the following zones.
T1 T
T2
T3
Symbol T T1 T2 T3
Parameter Transmission Transmission zone 1 Transmission zone 2 Transmission zone 3
Calculation
Pixel intensity full graticule area/ Pixel intensity tare image
Pixel intensity area T1/ Pixel intensity T1 tare image
Pixel intensity area T2/ Pixel intensity T2 tare image
Pixel intensity area T3/ Pixel intensity T3 tare image
Rhopoint ID Imaging Transmission Appearance Meter
43
User Defined Columns
The Rhopoint ID software has eleven fields where customers can input their own
process data to improve traceability during data analysis. The first column
has a fixed name “Comments” all other fields can be named by the user- these
names will be transferred to the data output .csv. Click into the small blue
box under the speech bubble to input a field name.
Statistics are shown for all measurements in the data table.
Add measurement distance to add a measurement distance to the measurement data the distance text box should be used.
Rhopoint ID Imaging Transmission Appearance Meter
44
Add Measurement Distance
To add a measurement distance to the measurement data the distance text box
should be used. As each new measurement is taken it is added to the bottom of
the table.
Display Graph
Select the parameters for graphing by clicking the header in either of the
results tables.
The Header should highlight as shown below and its color is displayed in the
header.
To change the color of the column, click the colored rectangle and a color
picker will appear. You can select your color here and it will be applied.
Rhopoint ID Imaging Transmission Appearance Meter
45
Click on the Graph tab. A graph of the selected parameters is displayed
The yellow indicator displays the current selected result row in the table.
The
button will bring out a window containing the graph for multiple screen use.
Save/Export Data To save the readings and all associated images, click the SAVE Project
button (PRJ) or to save the data only press the
SAVE CSV button.
Rhopoint ID Imaging Transmission Appearance Meter
46
When Saving a CSV use the file explorer to select a location and type your
file name as normal.
When saving a project (Images and data) you will be presented with a create
new folder window, simply name the new folder, click the folder icon, and then
press OK. The folder will contain all of the images and the data in csv
format.
When successfully saved there will be an image folder and a data.csv within please do not alter the names of these as the project will fail to load.
Rhopoint ID Imaging Transmission Appearance Meter
47
Delimiter The data will be saved with the selected delimiter this can be changed using the drop-down menu in the settings tab to suit regional PC settings.
Load measurements – To load additional data into the table, press the LOAD Project
button or Load CSV
button, this data is added to the bottom of the data table, use this
feature to combine and compare batches. To view saved data in isolation first clear the table.
Rhopoint ID Imaging Transmission Appearance Meter
48
To adjust the delay between readings, adjust the slider in the viewing window.
Rhopoint ID Imaging Transmission Appearance Meter
49
IDTX Software Live-view and Captured Images
To Access the Live view buttons, click the
button, to close the panel just click the
button.
The Home Tab should transform into the below, click the
button to start the live view.
Rhopoint ID Imaging Transmission Appearance Meter
50
To disable the live view, click the
button.
The live view is useful for positioning samples before measurement.
Note: With live view switched on it is not possible to view images of saved measurements.
Click on the
button to expand the image window for use with multiple screens.
This will expand the image attributed to the measurement selected in the data table. Note: With live view switched on it will display the live image instead. Haze, sharpness, and waviness are calculated from the average of seven individual zones. It is possible to view the individual regions using the toggle buttons on the right. If horizontal values are different from vertical values, it is an indication of anisotropy (visible directionality) in the material
Rhopoint ID Imaging Transmission Appearance Meter
51
The on-instrument T value are calculated using the transmitted light through
all black and white areas, this area can be viewed by toggling the T button.
Three smaller white only transmission areas are also available by toggling
T1-T3.
Click on the
Comparing Multiple Images
button in the expanded image window.
The software will replicate and fix the current view.
Multiple windows can be opened to visually compare different materials.
Rhopoint ID Imaging Transmission Appearance Meter
52
IDTX Software Visual Analysis of Results: Sharpness
Click in the data table to select images of previous measurements.
Once a measurement is selected the keyboard arrow keys can be used to quickly
review all the images and measurements.
The images recorded by the IDTX meter are correlated to the viewing sensation when observing an object through the test material.
Rhopoint ID Imaging Transmission Appearance Meter
53
Sharpness – The graticule has a visual scale for comparison and visualisation
of Sharpness. High Sharpness – Sharp view of object (S-95%)
In the (above) captured image the lines in frequency pattern 6 are visible and
undistorted. Low Sharpness – Fuzzy- out of focus object. (S 30%)
In the (above) captured image the lines in frequency pattern four are visible
and undistorted.
IDTX Software Visual Analysis of Results: Haze
Low HAZE (1.2%)- Material is less visible, viewed object has high contrast.
The black and white squares of the graticule have high contrast when compared.
High Haze (28.35%)- The material has a white or milky appearance; the viewed
object has low contrast. The black and white graticule squares have less
contrast and appear greyer.
Rhopoint ID Imaging Transmission Appearance Meter
54
IDTX Software Visual Analysis of Results: Waviness
Plastic materials can exhibit local surface effects such as orange peel which
can be seen as small waves when a pattern or object is viewed through the
material. The picture above shows the visual evaluation line pairs on the
graticule- orange peel is causing the graticule to appear wavy (19µm SD
waviness). The picture below shows the visual evaluation line pairs on the
graticule- there is very little waviness for this sample- (3.5µm SD waviness).
Notes
Rhopoint ID Imaging Transmission Appearance Meter
55
Rhopoint Instruments Ltd Rhopoint House, Enviro 21 Park, Queensway Avenue
South, St Leonards on Sea, TN38 9AG, UK
T: +44 (0)1424 739 622 E:
sales@rhopointinstruments.com
www.rhopointinstruments.com
Rhopoint Americas Inc. 1000 John R Road, Suite 209, Troy, MI 48083, USA
T: 1.248.850.7171 E:
sales@rhopointamericas.com
www.rhopointamericas.com
Rhopoint ID Imaging Transmission Appearance Meter
Rhopoint Instruments GmbH Seebauer Office Center, Am Weiglfeld 24, 83629
Weyarn, Deutschland T: +49 8020 9214-988 E:
info@rhopointinstruments.de
www.rhopointinstruments.de
1019-02
56
References
- Rhopoint Americas - Appearance & Measurement Experts
- Rhopoint Instruments - Appearance & Measurement Experts
- Rhopoint Instruments GmbH - Experten für die Oberflächenmessung
- Rhopoint Americas - Appearance & Measurement Experts
- Rhopoint Instruments - Appearance & Measurement Experts
- Novo-Curve 4 Glossmeter - Rhopoint Instruments
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>