Sonel PQM-700 Analysis 4 Computer Software User Manual
- June 13, 2024
- Sonel
Table of Contents
PQM-700 Analysis 4 Computer Software
USER MANUAL
SONEL ANALYSIS 4
computer software
Applies to instruments: PQM-700 PQM-701(Z, Zr) PQM-702(A, T) PQM-703 PQM-707
PQM-710 PQM-711
MPI-540 MPI-540-PV
SONEL S.A. Wokulskiego 11 58-100 widnica
Poland
Version 4.6.0.2 29.06.2023
Sonel Analysis 4 User Manual
Note: Due to continuous development of our software, the actual appearance of
some screens may slightly differ from that presented in this operating
manual. This manual describes the software and devices in the following
versions: Sonel Analysis v4.6.0 PQM-700 v1.16 PQM-701Z/Zr v1.18 PQM-702,
PQM-703, PQM-710, PQM-711 v1.50 PQM-707 v1.22 MPI-540, MPI-540-PV
v2.81.02
Note: The latest version of the software may be downloaded from the
manufacturer’s website. Note:
Detailed information about the analyzer functions may be found in the
operating manuals of specific analyzers.
The icon with the analyzer name (here: PQM-700) is placed next to sections of
the text that refer to specific features of the analyzer, particularly to
availability/unavailability of a given program function.
2
1 Installing and starting the program
TABLE OF CONTENTS
1 Installing and starting the program…………………………………………………..5
1.1 Minimum hardware requirements………………………………………………………………5 1.2 Installation of the program………………………………………………………………………..5 1.3 Starting the program ……………………………………………………………………………….8
2 Getting started……………………………………………………………………………… 10
2.1 Connecting with the analyzer …………………………………………………………………. 10 2.2 Startup window……………………………………………………………………………………..14
2.3
Recording settings wizard ……….. 15
3 Recording Settings ………………………………………………………………………. 19
3.1 General settings …………………………………………………………………………………… 21 3.2 Measurement
configuration…………………………………………………………………….22
3.2.1 Main Settings……………………………………………………………………………………………… 23 3.2.1.1 BASIC Tab
…………………………………………………………………………………………….. 23 3.2.1.2 ADDITIONAL
tab………………………………………………………………………………………. 25
3.2.2 Recording settings according to EN 50160 or other standard………………………………
28 3.2.3 Recording settings according to user settings …………………………………………………..
31 3.2.4 VOLTAGE card……………………………………………………………………………………………… 32
3.2.4.1 BASIC Tab …………………………………………………………………………………………….. 32 3.2.4.2 ADDITIONAL
tab………………………………………………………………………………………. 36 3.2.5 CURRENT
…………………………………………………………………………………………………….. 37 3.2.6 POWER AND
ENERGY………………………………………………………………………………………. 38 3.2.6.1 POWER tab
……………………………………………………………………………………………. 38 3.2.6.2 ADDITIONAL
tab………………………………………………………………………………………. 40 3.2.6.3 ENERGY
tab……………………………………………………………………………………………. 40 3.2.7 HARMONICS
………………………………………………………………………………………………… 41 3.2.7.1 VOLTAGE tab
…………………………………………………………………………………………. 41 3.2.7.2 CURRENT tab
………………………………………………………………………………………….. 43 3.2.7.3 ADDITIONAL
tab………………………………………………………………………………………. 43 3.2.8 INTERHARMONICS
………………………………………………………………………………………….. 44 3.2.8.1 VOLTAGE tab
…………………………………………………………………………………………. 44 3.2.8.2 CURRENT tab
………………………………………………………………………………………….. 46 3.2.8.3 MAINS SIGNALLING tab
……………………………………………………………………………… 47 3.2.9 Default configuration
profiles…………………………………………………………………………. 48 3.3 Auto-probes
support………………………………………………………………………………49
4 Analyzer Settings …………………………………………………………………………. 50
4.1 TIME AND SECURITY card …………………………………………………………………………….50 4.2 PROBES PHASE card………………………………………………………………………………….51 4.3 WIRELESS CONNECTION card……………………………………………………………………….52 4.4 SERIAL CONNECTION card…………………………………………………………………………..54 4.5 Additional settings card …………………………………………………………………………. 54
5 Analyzer database………………………………………………………………………… 56
5.1 GSM connection settings ……………………………………………………………………….58
5.2
Wi-Fi settings ……………………………………………………………….. 59
5.2.1 Configuring Wi-Fi in the Access Point mode. …………………………………………………… 60
5.2.2 Configuring Wi-Fi in Client mode……………………………………………………………………. 60
6 Live mode ……………………………………………………………………………………. 63
3
Sonel Analysis 4 User Manual
6.1 Waveforms …………………………………………………………………………………………..63 6.1.1 Synchronization of
waveforms in LIVE mode …………………………………………………… 65
6.2 Timeplots…………………………………………………………………………………………….. 66 6.3 Measurements
…………………………………………………………………………………….. 67
6.3.1 Energy costs calculator………………………………………………………………………………… 68 6.4 Phasor
diagrams ………………………………………………………………………………….. 69 6.5 Harmonics
…………………………………………………………………………………………… 70 6.6 Interharmonics
…………………………………………………………………………………….. 72
7 Data Analysis……………………………………………………………………………….. 74
7.1 Reading data from the analyzer ………………………………………………………………74
7.1.1
Reading data from MPI-540, MPI-540-PV and PQM-
707 devices ……………………………………………………………………………………………….. 75
7.2
Reading data directly from the
memory card ……………………………………………………………………………………….. 77
7.3 Selecting the time interval for analysis …………………………………………………….. 78
7.4 Analyzing obtained data ………………………………………………………………………… 80
7.4.1 GENERAL view ……………………………………………………………………………………………… 80
7.4.2 MEASUREMENTS view……………………………………………………………………………………… 81
7.4.2.1 Time plot ……………………………………………………………………………………………… 84
7.4.2.2 Harmonics/interharmonics graph ……………………………………………………………… 86
7.4.2.3 Cyclic waveform plot………………………………………………………………………………. 87
7.4.2.4 User report …………………………………………………………………………………………… 87
7.4.2.5 Generate report on compliance with standard…………………………………………….. 88
7.4.2.6 Report according to EN 50160 Standard – description …………………………………. 91
7.4.2.7 Report according to standard (recommendation) – micro-installations …………….. 97
7.4.2.8 Summary report…………………………………………………………………………………… 102
7.4.3 Events …………………………………………………………………………………………………….. 103
7.4.3.1 Value/duration graph ……………………………………………………………………………. 105
7.4.3.2 ANSI/CBEMA graph …………………………………………………………………………….. 106
7.4.3.3 Waveform/RMS1/2 graph ……………………………………………………………………….. 107 7.4.3.4 Event Report ………………………………………………………………………………………. 108
7.4.4 CONFIGURATION view ………………………………………………………………………………….. 108
7.4.5 Energy cost report …………………………………………………………………………………….. 109
7.4.5.1 Types of tariffs …………………………………………………………………………………….. 109 7.4.5.2 Energy cost report description …………………………………………………………….. 109
7.4.6 Data export ………………………………………………………………………………………………. 111
8 Other program options ……………………………………………………………….. 112
8.1 Analyzer status, starting and stopping recording………………………………………112 8.2
Deleting data ……………………………………………………………………………………… 113 8.3 Program configuration
…………………………………………………………………………. 114
8.3.1 Main Settings……………………………………………………………………………………………. 114 8.3.2 Recording
configuration ……………………………………………………………………………… 115 8.3.3 Live mode
………………………………………………………………………………………………… 120 8.3.4 Color
settings……………………………………………………………………………………………. 120 8.3.5 Data analysis
……………………………………………………………………………………………. 121 8.3.6 Report settings
…………………………………………………………………………………………. 122 8.3.7 Media settings
………………………………………………………………………………………….. 123 8.4 Upgrading software and firmware
…………………………………………………………. 126 8.4.1 Automatic software
update………………………………………………………………………….. 126
9 Manufacturer ……………………………………………………………………………… 127
4
1 Installing and starting the program
1 Installing and starting the program
“Sonel Analysis 4” is an application required to work with instruments of PQM-
7XX and MPI-540 series. It enables the user to: configure the analyzer, read
data from the device, real-time preview of the mains, delete data in the
analyzer, present data in the tabular form, present data in the form of
graphs, analyze data for compliance with EN 50160 standard (reports) and other
user-defined reference
conditions, independent operation of multiple devices, upgrade the software
and the device firmware to newer versions.
It is possible to simultaneously run multiple instances of Sonel Analysis.
1.1 Minimum hardware requirements
Table 1 presents the minimum and recommended configuration of a PC running
“Sonel Analysis 4” software.
Table 1. Minimum and recommended PC configuration.
Configuration
Processor RAM
Free space on hard disk
Graphics card
USB Internet access (for automatic
updates) Operating system
Minimum
Recommended
1.5 GHz
1 GB
200 MB 32 MB, min. resolution 1366×768
2.4 GHz
4 GB
16 GB 128 MB min. resolution 1920×1080
Windows 7, Windows 8 (8.1), Windows 10
1.2 Installation of the program
Note In order to facilitate installing drivers for the analyzer, it is
recommended to install “Sonel Analysis 4” software (including drivers as
described in the following instructions) before connecting
a USB cable.
To start the installation of “Sonel Analysis 4” software, run the installation file included in the set or downloaded from the manufacturer’s website.
5
Sonel Analysis 4 User Manual
Fig. 1. Installer – the initial screen. The screen will be displayed as shown
in Fig. 1. After clicking “Next>” button read the software license agreement
and click “I agree”. Then you see the history of revisions. In the next screen
(Fig. 2) indicate whether the application is to be installed for all users or
just for the one currently loggedin. The next screen allows you to select
installation components (Fig. 3). It is worth paying attention to the
Automatic Start function – thanks to it Sonel Analysis will start
automatically after starting the computer. The following two screens indicate
the location of the installation and location of the application with the
program name, which will be visible in the Start menu. To begin the
installation, press “Install”.
Fig. 2. Installer selection of users.
6
1 Installing and starting the program
Fig. 3. Installer choosing components. At the end of the software
installation, the window shown in Fig. 4 will be displayed. When box “Launch
Sonel Analysis 4.x.x” is ticked, then after pressing “Finish” button, the
application will be launched.
Fig. 4. Finishing the installation. At this point, the analyzer may be
connected to a PC. The system should automatically recognize the connected
device. If the installation was successful, the computer is ready to cooperate
with the analyzer.
7
Sonel Analysis 4 User Manual
1.3 Starting the program
When the program starts, the main window appears as shown in Fig. 5.
Fig. 5. The main screen. On the left side, the toolbar is displayed with the
program functions most commonly used. The toolbar may be moved by ‘gripping’
the dots at its top. The individual icons have the following meanings: STARTUP
WINDOW the window which groups the most frequently used operations
(recording
settings, data preview, reading data from the disk or analyzer). Description
in sec. 2.2. OPEN depending on the context, it enables the user to load from
the disk: the analyzer
configuration (files .settings), saved analysis (files .analysis), or saved
recording (files *.pqm7XX), SAVE depending on the context, it enables the
user to save the analyzer configuration on the disk (while editing the
configuration), save raw data or present analysis files (during the analysis),
RECORDING SETTINGS measurement configuration module of the analyzer (Sec.
3), LIVE MODE the mode of reading current, real-time values of mains (Sec.
6), ANALYSIS the module for data analysis directly from the analyzer or from
the memory card (Sec. 7), CONTROL window for controlling recording and
changing the active measurement configuration (Sec. 8.1), ANALYZER DATABASE
module for managing analyzers cooperating with Sonel Analysis (Sec.5), PROGRAM
SETTINGS module for configuring options of Sonel Analysis (Sec.8.3) ,
DISCONNECT ends the communication session with the analyzer.
8
1 Installing and starting the program Extensions of the files created by Sonel
Analysis software are as follows: .settings measurement configuration files
of the analyzer, .config Sonel Analysis configuration files, .pqm7XX
recorded data files (source format of the analyzer), .analysis analysis
files.
The user may select commands from the top menu, by clicking icons with the
mouse, or by using keyboard shortcuts (they are valid in the whole program):
F2 startup window, F3 analyzer database, F4 analyzer settings (including
time and security), F5 recording settings, F6 Live mode – reading real-
time values, F7 Control window, F8 data analysis, F9 Sonel Analysis
settings, CTRL+S saves the analysis on the disk or saves a screenshot in the
live mode, CTRL+O opening of a file stored on disk.
There are many other keyboard shortcuts. Most of them are visible in the menu
next to the name of an item.
Tip The user may select commands using a mouse / keyboard (standard operation
similar to Windows, ENTER – select option, ESC – Cancel, TAB
– move to the next button, etc.).
9
Sonel Analysis 4 User Manual
2 Getting started
2.1 Connecting with the analyzer
Before sending any data to/ from the analyzer, the application must establish
a connection with the device. The connection screen is displayed automatically
when the user wants to perform an operation that requires such a connection,
and the analyzer is not currently connected. A first-time user may start from
the Startup screen (described in sec. 2.2). When using it, the connection with
the analyzer is established when it is required.
In order to establish the connection with the analyzer: select any option that
requires an active connection e.g. LIVE MODE, ANALYSIS or CONTROL, use
ANALYZER DATABASE and connect to the indicated analyzer by selecting it from
the list, then
use the CONNECT SELECTED option.
After choosing one of the above options (when no previous connection was
established), the program displays window CONNECTION and starts scanning in
the search for available analyzers (see Fig. 6). The analyzers are searched
via cable (USB) and using wireless connections (when OR-1 is connected to a PC
and when a given analyzer offers this function). The user may also activate
the option in the program settings of searching analyzers connected via GSM
(applies to analyzers with built-in GSM modem) after enabling option TCP/IP
OVER GSM in program configuration. Analyzers operating with Wi-Fi may be found
after they are correctly configured (see sec. 5 and the manual of the
analyzer) and after selecting option WIFI in section ACTIVE MEDIA.
After a successful scan, the list of detected analyzers appears on the screen.
Analyzers are displayed with their model, serial number and the type of
communication link. Clicking on the selected analyzer and pressing SELECT icon
approves the analyzer selected from the list. The analyzer may be also
selected by double clicking on its field. Pressing SEARCH AGAIN restarts
scanning in the search for the analyzers.
It should be noted that the first listed option search for all analyzers
listed in the analyzer database and those directly connected with a USB cable,
which extends the search process. At the other hand, the second option
(connecting to the analyzer indicated in the database) is trying to find just
the indicated analyzer, which significantly shortens the connection process,
when the analyzer database is extensive. Particularly large delay is related
to the search of analyzers via the GSM network.
After selecting the analyzer, the program asks the user to enter PIN code,
which protects it against unauthorized access. It consists of three digits
(0…9). By default PIN is set in the factory at 000.
Note When wrong PIN is entered three times in a row, data transmission is
blocked for 10 minutes.
PQM-707 does not require a PIN code when connecting.
10
2 Getting started
Fig. 6. Analyzer selection window.
Notes Detecting an analyzer via OR-1 wireless module, Wi-Fi or GSM is possible
only when the unique serial number of this analyzer was earlier entered into
the analyzer database (doesn’t apply to PQM-707). Basing on this number, the
software filters other analyzers (e.g. those within the range of radio
interface) which are not owned by the owner of a given software copy. The
serial number may be entered into the database manually (section 5) or after
connecting the analyzer via USB, entering correct PIN code and selecting STORE
PIN IN DATABASE option (see Fig. 7). Then the analyzer is added to the
database of the analyzers (doesn’t apply to PQM-707). If the STORE PIN IN
DATABASE option is checked in the authorization window, the serial number and
the entered PIN will be associated, to avoid the need to entry it again during
the next connection (serial number and analyzer model are automatically added
to the analyzer database). After a successful connection, a window should
appear to confirm the connection with the analyzer see Fig. 8. This screen
displays the analyzer information, such as its serial number, firmware and
hardware versions. If automatic log-in is unsuccessful, the window shown in
Fig. 7 is displayed again.
11
Sonel Analysis 4 User Manual
Fig. 7. PIN code verification.
Fig. 8. The successful connection with the analyzer. Entering incorrect PIN
results in displaying window shown in Fig. 9. 12
2 Getting started Note
When the transmission is blocked after three unsuccessful attempts to enter
the PIN, during the next attempt to connect to the analyzer, the window will
appear with the following message “Communication blocked!
Incorrect PIN code!”
Fig. 9. Incorrect PIN. An unsuccessful attempt to connect the analyzer for
reasons not attributable to PIN will trigger the error message. Press RETRY
button to repeat the attempt, or go to the analyzer selection window and
select another analyzer, or rescan for available analyzers. When during the
communication the analyzer is switched off or the USB cable is plugged out, or
other event would prevent receiving answer from the analyzer, the message
shown in Fig. 10 will be displayed.
Fig. 10. Connection lost. 13
Sonel Analysis 4 User Manual
2.2 Startup window
Compared to its previous version, Sonel Analysis 4 offers a new tool that can
help (especially beginners) to work with the program and shorten the process
of setting up the analyzer.
The first element of the tool is the STARTUP WINDOW, which may be displayed by
clicking icon on the toolbar, or selecting it from menu ANALYZERSTARTUP
WINDOW. A window will pop-up, as shown in Fig. 11:
Fig. 11. Startup window. In the window there are four buttons: SET UP AND
RECORDING select this option to run Analyzer Settings Wizard. See more about
this tool in sec. 0. LIVE MODE is used to start the live view of parameters
measured by the analyzer. If the con-
nection with the analyzer has not been established yet, then connection window
is shown (see sec. 2.1). Selecting this option is the same as selecting LIVE
MODE in the toolbar or menu (described in sec. 6). GET RECORDED DATA allows
user to download and analyse the measurement data from the analyzer. If the
connection with the analyzer has not been established yet, then connection
window is shown (see sec. 2.1). Selecting this option is the same as selecting
ANALYSIS in the toolbar or menu (described in sec. 7). OPEN DATA FILE this
option is selected to open a file stored on the disk (extensions .analysis or
.pqmxxx) and its further analysis. Data analysis was described in detail in
sec. 7).
14
2 Getting started
2.3
Recording settings wizard
Fig. 12. Settings wizard main window.
Recording settings wizard is a tool that allows you to configure the analyzer
for the most common scenarios, applying “step-by-step” method. To start the
Wizard select from the toolbar RECORDING SETTINGS (or from menu
ANALYZERRECORDING SETTINGS), or use STARTUP WINDOW and select SET UP AND
RECORDING. The window shown in Fig. 12 will be displayed. RECORDING SETTINGS
WIZARD this option allows you to enter simplified settings using the “step-
by-step” method. The wizard is described later in this section. ADVANCED
RECORDING SETTINGS the standard method for setting the analyzer with access
to
all analyzer settings (method known from previous versions of Sonel Analysis).
A detailed description is in sec. 3. OPEN SETTINGS FILE opening a previously
saved file with the measurement settings (files with settings extension).
Settings in the wizard are simplified and divided into several steps, where
the user selects one of the recording parameters and the moves to the next
screen. ANALYZER (Fig. 13) in the first step of setting the analyzer, the
user must select the type of
analyzer to be set. This step is required due to differences in the
measurement capabilities of individual analyzers. SCENARIOS (Fig. 14) at
this stage, user selects the type of measurements, these are two mutually
exclusive possibilities: SELECT SCENARIO this option allows user to select
one of the predefined measurement
scenarios: All parameters Data Logging (V, I, Hz, THD) Power Study / Energy
audit Harmonics to Class A/S Power Quality to Class A/S
15
Sonel Analysis 4 User Manual SELECT STANDARD recording for compliance with
standard EN 50160 or other standard.
Select a relevant standard from the list. This setting will allow to generate
a compliance report for the specified standard.
Fig. 13. Settings wizard selection of analyzer. MAINS SYSTEM indicate the
type of measured system/network, NOMINAL FREQUENCY FN select frequency (50
or 60 Hz), NOMINAL VOLTAGE UN select nominal voltage. The list indicates the
applied phase voltage
(first value) and phase-to-phase voltage (second value). For measurements with
voltage transducers you may select VOLTAGE TRANSDUCERS box and define
parameters of a transducer (see also description of setting the transducers in
sec. 3.2.1.1). PROBES TYPE indicate the type of current probes used. If for
a given setting the current measurement is not required, select NONE option.
MEASUREMENT INTERVAL is a basic averaging period for measured parameters.
Parameters indicated in the scenario will be saved at the specified interval,
which directly translates into a volume of recorded data and the possible
recording time. Approximate recording time is displayed at the bottom of the
screen. TRIGGERING you can choose option IMMEDIATE (recording started by the
user after completing the set up), or ACCORDING TO SCHEDULE (the user defines
intervals, at which the analyzer records the data. The intervals are defined
on the next screen).
16
2 Getting started RECORDING SCHEDULE (optional step) in this screen the user
defines time intervals if at the
previous stage he/she selected option ACCORDING TO SCHEDULE. See also sec.
3.2.1.1 and schedule setup. WIZARD SUMMARY the last step of the wizard (Fig.
15) allowing user to send the created settings to the analyzer (click SEND
SETTINGS). If the analyzer is not connected, the selection and connection
procedure will be started (sec. 2.1), and then the settings will be sent.
Note: The entire contents of the memory will be deleted. The wizard as default
uses CONFIGURATION 1 of the analyzer; it also reserves for it the entire
analyzer memory space. The setting created with the Wizard may be saved on the
disk for later use as a file *.settings (option SAVE SETTINGS). Alternatively,
you can select ADVANCED SETTINGS, where the initial settings may be viewed and
modified and then sent to the analyzer (window ADVANCED RECORDING SETTINGS,
will be shown as in sec. 3.2).
Fig. 14. Settings wizard – Scenario selection.
17
Sonel Analysis 4 User Manual Fig. 15. Settings wizard summary. 18
3 Recording Settings
3 Recording Settings
Note: The analyzer measurement settings module is not available in MPI-540,
MPI-540-PV and PQM-707.
After selecting option ANALYZERRECORDING SETTINGS from the main menu (or
clicking the icon) SETTINGS window will be displayed as shown in Fig. 16. It
is the most important part of the analyzer configuration. Here the user
determines which parameters will be recorded by the analyzer; the mains type
and nominal values of the parameters.
The left part of the screen is divided into two parts (Fig. 16): LOCAL and
ANALYZER. Top part (LOCAL) is used by the user for parameters modification,
whereas the bottom part (ANALYZER) shows the current analyzer settings and is
of read-only type. Each part has a drop-down tree divided into four
configurations and ANALYZER SETTINGS.
PQM-700 has only one measurement configuration.
Fig. 16. Measurement configuration main screen and settings tree. Each of
the four measurement configurations is independent from others. This is the
place where the user defines the mains system, nominal voltage, frequency,
type of probes and parameters for recording and detecting events. The icons
next to the configuration may be displayed in various colors: grey – no
connection with the analyzer,
19
Sonel Analysis 4 User Manual green – present configuration is synchronized
with the analyzer configuration and with the
configuration saved on the disk. blue – current configuration is compatible
with the analyzer, but differs from the one saved to the disk, yellow – the
configuration is not compatible with the analyzer, but is compatible with the
one
saved to disk, red – the current configuration is not compatible with the
analyzer and not compatible with the
one saved to disk. RECEIVE SETTINGS button is used to read the analyzer
current settings in order to edit them in the computer. If the settings have
been previously modified by the user, a warning message will be displayed. The
correct reading is also confirmed by a relevant message. Then all icons in the
measurement configurations tree will change to blue, which means that the
settings in the application and in the analyzer are identical. The SEND
SELECTED CONFIGURATION button sends one selected configuration from the top
part (LOCAL) to the instrument. The SEND ALL CONFIGURATIONS button allows to
send all the configurations. Before sending the settings, the user is asked to
confirm the operation (Fig. 17).
Fig. 17. Confirmation of entered settings after selecting SEND ALL
CONFIGURATIONS.
Note Selecting the SEND ALL CONFIGURATIONS button will delete all data on the
memory card. This option should be used every time the meter
software is updated. This is essential to avoid data compatibility issues.
Sending the configuration no. X by pressing the SEND SELECTED CONFIGURATION
button will overwrite the current configuration no. X. In both cases, the
previously collected data should be read from the
analyzer and saved to a local disk.
20
3 Recording Settings
Note Saving new settings in the analyzer is not possible when the device is in
the recording mode (an appropriate message will be displayed to inform
about this Fig. 18).
Fig. 18. Programming impossible due to recording in progress.
3.1 General settings
The first element to be configured in the left-hand side of the configuration window is GENERAL
SETTINGS. This part is divided into three tabs:
ANALYZER TYPE: allows the user to choose the analyzer model for which the settings will be modified. As the software provides support for several models of analyzers with different capabilities, the user should pre-select the analyzer type from the list, to match the settings with analyzer performance. If the user is already connected to the analyzer, the analyzer type is selected automatically. In addition, the user may specify the type of the default analyzer in the program configuration (see section 8.3.1),
MEMORY ALLOCATION: this tab (Fig. 19) allows
user to allocate space of the analyzer’s memory card to the individual measurement
configurations. Use four sliders to allocate desired amount of memory to a given configuration
(in MB and %). In the middle part of the screen you will see a visualization of memory allocation.
All available card memory (100%) may be freely divided among the four configurations. It is
possible to allocate 100% of the memory to a single configuration; in such case recording in
other configurations is impossible – for them only “live” preview of mains parameters is available.
All measurement configurations have linear recording in the memory, when the memory
allocated to a given configuration is used up, the recording stops. Please remember that
modification of memory allocation will require deleting all data from the memory card. Thus, it
is advisable to download such data and save it on a local disk. New memory allocation is sent
to the analyzer after clicking SEND button (along with configurations of all measurement points).
GPS SYNCHRONIZATION: two parameters included here
are provided to define the behavior of the analyzer when the time measurement source changes
during recording.
21
Sonel Analysis 4 User Manual TIME RESYNCHRONIZATION THRESHOLD defines the time
difference (in seconds) between the RTC (Real Time Clock) of the analyzer and
UTC (Universal Time Clock) received from the GPS satellites. If the difference
between the two clocks is greater than the time entered in the field, then the
analyzer immediately change its time to UTC. If the difference does not exceed
the threshold value, the analyzer will gradually reach UTC without the
immediate (step) change. TIME RESYNCHRONIZATION THRESHOLD set to zero disables
the immediate time shifting and ensures a gradual resynchronization. The speed
of the re-synchronization is controlled by RESYNCHRONIZATION FACTOR. It is a
percentage value from 0 to 100, which defines how fast the internal analyzer
time will achieve UTC time during in the resynchronization process. When the
RESYNCHRONIZATION FACTOR is equal to 100%, then in every second, half of the
second time is skipped (shifted) forward or backward. At lower values of the
factor, such correction is proportionately slower, prolonging the re-
synchronization time. If the recording process is not active, then the change
in the internal time is always immediate.
Fig. 19. Memory allocation.
3.2 Measurement configuration
To drop down the “tree” of a measurement configuration, click once on the
triangle at the configuration, or double-click on a given configuration.
Highlighting (by single-clicking) selected configuration will result in
displaying its main settings on the right side of the screen. Main settings
consist of two tabs: BASIC and ADDITIONAL. After dropping down selected
configuration, the following list is displayed: STANDARD settings for
recording in accordance with EN 50160 and other standards, VOLTAGE settings
of voltage-related parameters, divided into two tabs: BASIC and ADDITIONAL
, 22
3 Recording Settings CURRENT current parameters, POWER AND ENERGY power
and energy parameters divided into three tabs: POWER,
ADDITIONAL and ENERGY, HARMONICS harmonics parameters divided into three
tabs: VOLTAGE, CURRENT and
ADDITIONAL, INTERHARMONICS interharmonics parameters divided into three
tabs: VOLTAGE, CURRENT,
MAINS SIGNALLING. It is possible to transfer settings between different
configurations. To copy the settings from one configuration to another, right-
click on the header of CONFIGURATION X (where X is the number of the point) in
the tree of configurations and select COPY. Then, to transfer these settings
to another point, right-click them (as above) and select PASTE (CONFIGURATION
X) (where X is the number of the configuration). The next chapters presents
the description of the main settings screen, and its individual cards. 3.2.1
Main Settings 3.2.1.1 BASIC Tab Main settings screen of the configurations and
BASIC tab is shown in Fig. 20. It is divided into a few sections: the mains-
related settings are grouped in the top part, whereas the part located below
relates to recording according to the schedule and time intervals. In
addition, here the user may assign desired name to a configuration and assign
the hysteresis used during the event detection.
Fig. 20. Setting the mains parameters, nominal mains parameters and
transducers. 23
Sonel Analysis 4 User Manual
The following parameters are defined in the part related to the mains:
VOLTAGE UN: nominal voltage 64/110 V, 110/190 V, 115/200 V, 120/208 V, 127/220 V,
220/380 V, 230/400 V, 240/415 V, 254/440 V, 277/480 V, 290/500 V, 400/690 V (phase /phase-
to-phase, depending on the mains type; in case of split-phase network, the phase-to-phase
voltage is twice the phase voltage),
FREQUENCY FN: Mains nominal frequency 50 or 60 Hz,
MAINS SYSTEM one-phase, split-phase, wye with N, delta, wye without N, and Aron’s
measuring systems; a connection drawing for selected system is shown in the field below the
name. For mains of 3-wire type, the nominal voltage is the phase-to-phase value (the second
value specified in MAINS NOMINAL VALUES ‘UN’ field),
PROBES TYPE indicates the type of current probes used for the measurements. The user
may select C-4(A), C-5(A), C-6(A), C-7(A) CT probes or F-x(A), F-xA1 or F-xA6 flexible probes.
If the analyzer supports auto-detection of probes the option AUTO may be selected. If the current
measurement is not required, select NONE option.
VOLTAGE TRANSDUCERS they define the voltage transducers ratio (when used). Please
select one of the three possible parameters of the transducer that will be calculated using the
other two:
Selecting PRIMARY will grey out the primary voltage, which is calculated
basing on two other parameters that may be changed by the user: the voltage of
the secondary circuit –
SECONDARY and RATIO. Primary voltage is calculated as the product of secondary
voltage and ratio.
Selecting SECONDARY will grey out the secondary voltage, which is calculated by the software, basing Primary voltage and Ratio. Secondary output voltage is equal to the quotient of the primary voltage and ratio.
Selecting RATIO will grey out “Ratio” field. The division is determined by defining the
primary and secondary voltage. Ratio is calculated as the ratio of the primary and
secondary voltage.
After turning the option of voltage transducers field VOLTAGE UN is greyed out; a new nominal
voltage is the primary voltage of the transformer.
CURRENT TRANSDUCERS define the current transducer ratio. The measuring range of
selected probes may be extended using external current transducers. The same method for
determining the ratio is used as in the case of the voltage transducers. Select one of the three
parameters, which will be calculated automatically based on the other two. The resulting
measuring range (primary current which does not exceed the nominal range of used probe) is
displayed below the probe selection list.
ADDITIONAL MEASUREMENTS use the additional option fields to define possible
measurements of additional parameters, such as neutral conductor current (with another probe,
if the system type is appropriate) and the N-PE voltage (for systems with separate N and PE
conductors). Please note that the analyzer will not measure these parameters, if they are not
checked in this place.
Note: In PQM-700 analyzer the N-PE voltage measurement is not available.
AVERAGING PERIOD for measurements use this field to define the basic results averaging
period. Available times (a set of available values may be different for different types of
analyzers) are as follows: half-period (special mode, only voltage and current RMS1/2 values are
recorded), 200 ms, 1 s, 3 s, 5 s, 10 s, 15 s, 30 s, 1 min, 3 min, 5 min, 10 min, 15 min, 30 min,
60 min, 120 min,
Note: For PQM-700 and PQM-701, if the user turned on the option of standard-
compliant recording in STANDARD tab, it will result in setting the averaging
time on 10 minutes and the selection list block averaging period selection
list. To change this settings, the user must first disable the standard-
compliant recording in the tab.
24
3 Recording Settings
TRIGGERING of recording enables the recording activation mode:
IMMEDIATE immediate measurement after pressing START/STOP button or starting the
recording process from the software,
SCHEDULED scheduled measurement; after recording starts (by pressing the button or
from the PC) the analyzer compares the present time with intervals defined in the
schedule and uses them to start and stop the recording,
THRESHOLD measurement after the limit for any active event is exceeded after activating, the device analyses the mains and waits until a parameter is exceeded – this
event triggers recording – at this moment data are recorded (logged) on the memory card.
Other elements of the settings screen (Fig. 20):
SCHEDULED LOGGING (recording) four time periods may be specified. After selecting a
given period by clicking the arrow, enter the start date and time (optionally you may specify the
end date/time). If the user sets the recording start without defining its end, the device will operate
until it is manually stopped or until the memory is filled-up. Time intervals must not overlap – the
program does not allow you to configure overlapping ranges. The following time intervals must
be set chronologically.
PQM-700 analyzer allows only one time period to be defined.
CONFIGURATION NAME the user may assign a name for selected configuration.
EVENTS DETECTION HYSTERESIS it determines the size of the hysteresis used in determining
event detection thresholds. It may be set within the range of 0.01…10.00%. The typical size of
the hysteresis is 2%. More about this issue may be found in the operating manual of the
analyzer.
CURRENT LIMIT when current measurement is
enabled, then selecting this option enables user to set the threshold of RMS current, below
which the value will be zeroed with all related parameters. See additional description in the
operating manual of the analyzer.
3.2.1.2 ADDITIONAL tab
Note: In PQM-700 and PQM-701 analyzers this tab is not available.
Fig. 21 presents the screen after selecting ADDITIONAL tab at MAIN SETTINGS of
the measurement configuration. Sliders are available in this screen for
defining recording time of several types of waveforms and plots: WAVEFORMS
determining the recording time for instantaneous voltage and current (sampling
frequency of 10.24 kHz) accompanying events detected by the RMS voltage
(swells, dips and interruptions, exceeding max. threshold values of UN-PE), DC
voltage component, RMS current (exceeding minimum and maximum values) and
transient events. The user may specify: recording time from 100 ms to 1 s, in
20 ms steps, pretrigger time from 40 ms to 960 ms, in 20 ms steps.
Note: Recording time of waveforms may be different from the stated here,
depending on the type of the analyzer. Appropriate times can be found in the
user’s manual for individual analyzers.
Waveforms are recorded both at the beginning and at the end of the event.
Pretrigger time defines the portion of the total recorded time that will be
assigned as “pretriggering” period (both for starting and stopping). This time
cannot be longer than the recording time.
25
Sonel Analysis 4 User Manual
Cyclic waveforms (recorded once per every averaging period) always have fixed
time equal to 3 periods and this setting does not applies to them.
To simplify the process, recording and pretrigger times are provided in
seconds or milliseconds, but in fact the analyzer records with an accuracy of
a single period of mains. For example, setting the waveform recording time at
560 ms and pretrigger time at 380 ms, results in a recording a total of 28
periods of mains and pretrigger equal to 19 periods (for 50 Hz mains). This is
important if the mains frequency deviates significantly from the nominal 50
Hz. In case of mains with 60 Hz the situation is similar, but the number of
periods is rounded down after dividing the time set by the nominal period time
of 16.67 ms. In order to enable recording of the waveforms, select option
RECORD WAVEFORMS AND RMS1/2 at VOLTAGE card of BASIC tab. RMS1/2 determining
the recording time for voltage and current RMS1/2 values accompanying events
detected by the RMS voltage (swells, dips and interruptions, exceeding max.
threshold values of UN-PE) and RMS current (exceeding minimum and maximum
values). The following features may be set: recording time from 1 s to 30 s,
in 0.1 s steps, pretrigger time from 0.1 s to 4.9 s, 0.1 s in steps.
Note: Recording time of RMS1/2 waveforms may be different from the stated
here, depending on the type of the analyzer. Appropriate times can be found in
the user’s manual for individual analyzers.
RMS1/2 values are recorded both at the beginning and at the end of the event.
Pretrigger time defines the portion of the total recorded time that will be
assigned as “pretriggering” period (both for starting and stopping). This time
cannot be longer than the recording time.
Similarly to waveforms, recording and pre-trigger times are specified in
seconds or milliseconds, but in fact the analyzer records a specified number
of RMS1/2 values, assuming one half-period as 10 ms in 50 Hz networks and 8.33
ms in 60 Hz networks. In order to start recording RMS1/2, select option LOG
WAVEFORMS AND RMS1/2 at VOLTAGE card and BASIC tab.
TRANSIENT CHARTS for determining recording time of high speed waveforms
for voltage channels, where transient events were detected. The following features may be set:
recording time (depending on the selected sampling frequency of transients), up to 2 ms for
sampling rate 10 MHz and 200 ms for 100 kHz (recording range always includes from 2000
to 20000 samples),
pretrigger time is between 10% and 90% of the recording time.
In order to start recording transient graphs, select option LOG WAVEFORMS AND TRANSIENT
CHARTS at VOLTAGE card and BASIC tab.
MAINS SIGNALLING – for defining recording time of ripple
control signal waveforms, after detecting exceeded threshold of control signal amplitude (as
specified in settings). Consequent 10/12-period values are recorded (i.e. at approximately 200
ms) for the time indicated by slider: from 10 s to 120 s. After recording the entire waveform it is
possible to detect the following events of this type with successive time graphs.
Recording of control signals waveforms is activated after selecting LOG EVENTS (AND TIME PLOT)
at INTERHARMONICS card and MAINS SIGNALLING tab.
TEMPERATURE temperature recording of external sensor may be activated (only PQM-
702T with ST-2 temperature probe). Temperature is recorded with the same averaging time as
26
3 Recording Settings the other parameters (AVERAGING PERIOD parameter in BASIC
tab), provided that the temperature readings are performed with a frequency of
approx. 1 time per second. For averaging periods shorter than this period,
recorded temperatures will repeat.
Fig. 21. Additional settings of the measurement configuration.
27
Sonel Analysis 4 User Manual
3.2.2 Recording settings according to EN 50160 or other standard
After selecting STANDARD position from the selection tree, the user may
quickly choose analyzer settings according to EN 50160 or other Standard (Fig.
22). This standard specifies the characteristics and quality criteria that
should be met by the distribution networks of low, medium and high voltage.
Settings shown in the list are defined in the program settings in the
location: OPTIONS PROGRAM SETTINGS RECORDING SETTINGS STANDARD DEFAULT
SETTINGS.
Note The procedure of measurements for standard compliance is different for
different types of the analyzers:
· PQM-700, PQM-701(Z, Zr), PQM-707: activating standard-compliant
recording (by ticking box ENABLE LOGGING ACCORDING TO STANDARD) sets the
required configuration parameters and blocks them to prevent their
modification by the user. Only the parameters required by the selected
standard are recorded.
· PQM-702(A,T), PQM-703, PQM-710, PQM-711: activating the standard-
compliance recording does not prevent the user to set the recording parameters
in the user configuration, as the analyzer allows simultaneous recording of
parameters required by the standard and completely independent recording
according to the user configuration (dual recording). The only exception to
the rule of independence are the thresholds of voltage events, which are set
in accordance with the standard requirements and cannot be changed. The option
of such simultaneous recording is available in the specified analyzers,
starting from firmware version 1.17. When you want to record only the standard
parameters, turn off (uncheck) all parameters in the recording configuration
(leave only ENABLE LOGGING ACCORDING TO STANDARD field active and optionally
enable recording of control signals).
After choosing from the list a specific Standard, the analyzer marks on the
selection tree (of the measurement configuration) the parameters necessary to
perform measurements according to chosen Standard and locks them to prevent
inactivating them by the user. For example, if the averaging time was
previously set to 3 seconds, after selecting “Enable logging according to
standard” and choosing specific Standard from the list, it will change into
10minutes and it will be grayed out, preventing any further changes.
Additionally the following measurement parameters are marked: average voltage,
THD voltage, voltage harmonics, etc.
After selecting a specific Standard profile from the list, the program
modifies in the configuration only the thresholds of voltage events, leaving
all other options available to the user. The analyzer will automatically
record all the parameters necessary to generate a report on compliance with
the Standard (switching on the recording of control signal is left to the
user). The recorded Standard parameters will also be available for viewing.
Regardless of recording data for compliance with the standard, the user may
adjust almost all parameters for own recording (except for the above mentioned
thresholds for voltage events), according to own averaging time and types of
parameters. Such recording is called dual recording both sets of recorded
parameters may be independently analysed in the ANALYSIS module. For example,
it is possible to enable recording for compliance with the standard (e.g.
10-minute averages will be recorded for RMS voltage and harmonics, 10-second
for frequency values) and the simultaneous
28
3 Recording Settings recording for diagnostic purposes covering the specified
parameters with 1-second averaging (including the average of 1-second
voltage).
Fig. 22. ‘Standard’ tab for measurement configuration. Basic mains measurement
time, according to EN 50160, is one week. In this time, the following
parameters are measured: RMS voltage, frequency, Total Harmonic Distortion,
levels of individual voltage harmonics, unbalance (in three-phase mains),
flicker. Because the week-long measurements are related to gathering a large
amount of data, it was decided to use averaging algorithms. All the evaluated
parameters, excluding the frequency and long-term flicker, are averaged over
10 minutes. Thus, within a week, the user obtains 1008 values for each of
these parameters. These values are then evaluated by comparing them to the
criteria defined in the standard. If all parameters are within the specified
thresholds, it can be concluded that the measured mains meets the quality
requirements of EN 50160. Table 2 contains the current criteria for
measurements according to EN 50160 in LV networks (up to 1 kV) and MV networks
(36 kV). These settings are used by selecting the default settings for EN
50160 profile.
29
Sonel Analysis 4 User Manual
Table 2. Criteria for power quality according to standard EN 50160 for mains voltage up to 1 kV (low voltage networks).
Parameter
Slow changes in
voltage
(RMS
voltage)
Basic measurement
time
10 minutes
Criterion
For LV networks: for 95% of the measurements the deviation from the nominal
voltage shall be within the range of ±10% Unom for 100% of the measurements:
15%…+10% Unom
Frequency
10 seconds
For MV networks:
for 99% of the measurements the deviation from the nominal
voltage shall be within the range of ±10% Unom for 100% of measurements: ±15%
Unom For networks connected synchronously:
for 99.5% of the measurements the deviation should be within ±1% fnom (i.e.
for 50 Hz it is 49.5…50.5 Hz) for 100% of measurements: 6%…+4% fnom (47…52
Hz)
Flicker
Voltage unbalance
Total
Harmonic
Distortion (THD-F)
Voltage harmonics
2 hours 10 minutes 10 minutes 10 minutes
For networks connected asynchronously (e.g. island networks): for 95% of the measurements the deviation should be within ±2% fnom (i.e. for 50Hz it is 49…51 Hz) for 100% of measurements: ±15% fnom (42.5…57.5 Hz) for 95% of the measurements the long-term flicker severity PLT should be 1 for 95% of the measurements of the negative-sequence unbalance factor should be 2% for 95% of measurements, THD-F should be 8%
For 95% of the measurements, the level of each voltage harmonic
component related to the fundamental component should be less
than:
Odd order harmonics Even order harmonics
Order
Relative level
Order
Relative level
3
5.0%
2
2.0%
5
6.0%
4
1.0%
7
5.0%
6 … 24
0.5%
9
1.5%
11
3.5%
13
3.0%
15
0.5%
17
2.0%
19
1.5%
21
0.5%
23
1.5%
25
1.5%
The procedure for setting the analyzer on the measurement compliant with the
standard is as follows. On the main screen for the measurement point settings,
the user should set the correct mains type, nominal voltage, frequency, type
of probes, or select the appropriate transducers.
Then select STANDARD tab and enable ENABLE LOGGING ACCORDING TO STANDARD
option. This causes unlocking the standard version list. Then, from the list
select the appropriate element.
After completing the settings, and after sending the configuration to the
analyzer the user may trigger the recording process from the PC program or by
using START/STOP button. Before that make sure that the correct measurement
configuration is selected in the analyzer.
30
3 Recording Settings
The user has the ability to edit the standard default settings defined in the
preferences. It may be useful in situations where the criteria defined in the
Standards will change. The user will be able to adjust the settings yourself
without the need to upgrade Sonel Analysis software.
3.2.3 Recording settings according to user settings
If the user did not chose recording in accordance with the standard, then all
available parameters may be freely controlled by the user.
In PQM-702, PQM-703, PQM-710 and PQM-711 analyzers, due to the dual recording
functionality, activating the standard-compliance recording does not limit
control over recording acc. to the user settings (excluding the modification
of threshold event voltage events).
All settings are included in the main card of the measurement configuration
(BASIC and ADDITIONAL tabs) and five other cards (and their tabs): VOLTAGE,
CURRENT, POWER AND ENERGY, HARMONICS and INTERHARMONICS.
In most cases, the settings for the selected parameter may be divided into a
section of cyclical recording of parameter values according to the selected
averaging time, and into the section related to event detection. The user may
select to record the value of average, minimum, maximum and instantaneous
parameter. For voltage and current the user may define an additional averaging
time that is used for detecting minimum and maximum values (depends on
analyzer type). To clarify these issues let us use an example.
Example Global averaging time has been set to 1 minute, while min/max
determination period for voltage has been set to 5 seconds. All four voltage
options have been checked: recording of average, minimum, maximum and
instantaneous values. Question: How these values are determined and what is
recorded on the memory card?
1. Average value is determined as the RMS (root mean square) from
10/12-period measurements (approx. 200 ms). There will be approx. 300 of such
measurements per 1 minute. Average voltage value is calculated
as the square root of arithmetic mean of squared input values:
=
=1
2
where: Uavg 1-minute RMS voltage, k number of gathered 10/12-period values, Ui RMS voltage of 10/12-period type. 2. Min/max averaging equal to 5 seconds means that firstly the analyzer
averages 25 successive Ui values (using the method described above). Minimum and maximum values are selected from all 5-second average
values gathered during 1 minute. Both, minimum and maximum values
will be recorded.
3. Instantaneous value is the last 10/12-period voltage value measured
during 1-minute averaging period. This value will also be saved on the
memory card.
31
Sonel Analysis 4 User Manual
For most of parameters the user may enable detection of events. “Event” is the
term, which describes the situation, when the measured value exceeds the
threshold(s) set by the user. Depending on a parameter, the number of
thresholds may be as follows: only one (e.g. for negativesequence unbalance
factor – exceeding the maximum value), two (e.g. for frequency – exceeding the
upper and lower limits), three – for voltages (exceeding the threshold of
swell, dip and interruption).
Information about the event is recorded by the analyzer on the memory card at
the time of its completion. Then the following information is recorded: start
time of the event, end time of the event, threshold value, parameter extreme
value during the event, parameter average value during the event.
If the recording is interrupted during an event, information about this event
is also saved, but only within the following scope: start time of the event,
threshold value, extreme value of the parameter recorded by the end of the
recording process.
For the following parameters: RMS voltage and RMS current, changes in the
shape of the voltage envelope and the voltage phase, the recording of
waveforms and RMS1/2 values may be enabled at the beginning and end of the
event. Recording time of waveforms and RMS1/2 is set in ADDITIONAL tab, of the
card for the measurement configuration settings (see sec. 3.2.1.2.)
3.2.4 VOLTAGE card
VOLTAGE card is divided into two tabs: BASIC and ADDITIONAL.
3.2.4.1 BASIC Tab Fig. 23 presents basic settings for recording voltages.
Depending on the mains type, the settings
relate to phase-to-neutral voltages (single-phase, split-phase, three-phase
4-wire systems) or phase-to-phase voltages (three-phase 3-wire systems). As a
standard, the user may check the recording of the average value (AVG), minimum
(MIN), maximum (MAX) and instantaneous (INST) of the RMS voltage. Option ALL
selects all four, or remove them all. The same set of options is available for
the DC component of the voltage.
Measurement of minimum and maximum value is performed with the user-defined
averaging time (option MIN/MAX CALCULATION PERIOD). Available averaging times:
½ period, period, 200 ms, 1 s, 3 s or 5 s.
Note Averaging times 200 ms, 1, 3 and 5 sec. are actually expressed as
multiples of the mains basic period: 200 ms 10/12 periods (50/60 Hz) 1
second 50/60 periods 3 seconds 150/180 periods 5 seconds 250/300 periods
For systems with phase-to-neutral voltages (split-phase and three-phase with a
neutral conductor N), there is an additional option of recording RMS phase-to-
phase voltages. In such case only the average value is recorded. 32
3 Recording Settings
Checking option LOG WAVEFORMS AFTER EACH AVERAGING PERIOD results in recording
three voltage waveform periods after each successive averaging periods (for
example every 10 minutes).
The part related to events is located on the right side of the screen.
Checking option LOG EVENTS activate the fields for voltage threshold values.
They may be entered as percentage or absolute values. The setting ranges are
as follows: SWELLS: 0.00…+20.00% with a resolution of 0.01% or in volts within
the same range, DIPS: 0.00…-99.90% with a resolution of 0.01% or in volts,
INTERRUPTIONS: 0.00…-100.00% with a resolution of 0.01% or in volts up to 0.00
V (the dip level
cannot be lower than the interruption level).
Switching between the threshold percentage values to absolute values results
in automatic calculation to selected unit.
Note When the user changes mains nominal voltage after the event threshold
have been set, the thresholds expressed in percentage values will remain
on the same levels.
For mains with neutral conductor, it is possible to set also the maximum N-PE
voltage level option N-PE VOLTAGE OVER LIMIT. Exceeding this threshold in
volts will trigger an event. Setting the threshold to zero disables the
detection of events of this type.
After selecting LOG WAVEFORMS AND RMS1/2 option, the analyzer will record the
defined number of periods for voltage and current waveforms and RMS1/2 values,
at the beginning and the end of an event. Recording time is set in ADDITIONAL
tab, of the card for the measurement point settings, see sec. 3.2.1.2.
In addition, you can set the detection of voltage events by other methods than
exceeding the RMS voltage values that can help in network diagnostics:
· WAVESHAPE VARIATIONS: 1 … 100% with a resolution of 1%.
In this method, two adjacent periods of the voltage waveform are compared with
each other the difference between them is calculated and its maximum amplitude
is checked, which is then compared with the threshold set by the user. The
percentage threshold refers to the nominal voltage value. For example, for Un
= 230 V and the detection threshold set to 10%, such waveform changes are
detected whose amplitude is greater than or equal to 23 V. If the voltage
waveform in the next period does not fit within the specified tolerance range
(envelope), then an event is detected. The analyzer, after the event is
detected, records the time of occurrence, duration and the maximum registered
percentage difference between the compared waveforms.
It should be remembered that at low values of the detection threshold, the
analyzer can detect a very large number of events in a short time interval.
For this reason, the user has the HOLD TIME parameter in seconds. After the
event is detected, the analyzer blocks detection of subsequent events (in a
given channel) for the time specified by this parameter. It can be set in the
range from 1 s to 600 s. If the option LOG WAVEFORMS AND RMS 1/2 is also
selected, it will also be able to view oscilloscope waveforms and RMS1/2 plots
accompanying the event in addition to the recorded event.
33
Sonel Analysis 4 User Manual
· PHASE JUMPS: 1 … 360 with 1 resolution. Selecting the PHASE JUMPS field
enables the detection of sudden changes in the voltage fun-
damental phase angle. The detection algorithm compares the angles of the
fundamental voltage component of two or three neighbouring periods. If the
difference in angles is greater than the user-defined threshold expressed in
degrees, information about the detection of the event along with the measured
value of the phase angle change is recorded. If the option LOG WAVEFORMS AND
RMS VALUES is also selected, then in addition to the recorded event, you will
also be able to see the oscilloscope waveforms and RMS1/2 graphs accompanying
the event.
RAPID VOLTAGE CHANGES (RVC): Checking field LOG RAPID VOLTAGE CHANGES (RVC)
enables the detection of the
instantaneous RMS voltage disturbances, known as RVC. A more detailed
description of such events may be found in the operating instructions of the
analyzers. The user enters two parameters: THRESHOLD, which defines Umax
value, above which a RVC event is detected, as well as HYSTERESIS. The
hysteresis must be less than the threshold. If you have enabled the detection
of events such as dip/swell/interruption, then the threshold for RVC events is
limited to the sum of the thresholds for dip and swell. This is related to the
fact that RVC events may not exceed the thresholds of dip and swell – such
events are then treated as dips and swells, not as RVC events.
When the recording acc. to a standard is activated, and the standard defines
the requirements for the RVC, the user cannot modify these parameters the
change is possible only by modifying the default profile of the standard in
program configuration.
Similarly as for other events in this sheet, checking option LOG WAVEFORMS AND
RMS 1/2 will trigger the recording of oscilloscopic waveforms and RMS1/2
graphs that accompany the RVC event.
In polyphase systems, the device detects both single-phase events and
polyphase events (in accordance with IEC 61000-4-30).
PQM-700 and PQM-701 analyzers do not support recording of waveshape
variations, phase jumps and RVC events.
DC VOLTAGE: To enable the recording of events resulting from exceeding DC
component, activate option LOG
EVENTS in the VOLTAGE DC part of the screen, which relates to the DC
component. Entering a value in volts in MAX field triggers an event after
detecting a voltage DC is higher than the value (threshold) entered in volts
(the absolute value of the DC voltage is checked). Setting the threshold to
zero disables the detection of events of this type.
34
3 Recording Settings
Fig. 23. Basic settings for voltage measurements. In case of PQM-703 and
PQM-711 analyzers the card includes a set of controls enabling the user to
activate and configure the block of transient detection (fast and short-term
disturbances in the mains voltage). The detection is triggered after selecting
option LOG EVENTS (TRANSIENTS). The user may set the following parameters:
SAMPLING FREQUENCY – it is the sampling frequency of the transient detection
block. One out of five frequencies may be chosen: 100 KHZ (slowest), 500 KHZ,
1 MHZ, 5 MHZ, 10 MHZ (fastest). The sampling frequency must be chosen
depending on the expected frequency content of disturbances in the network.
Detection method: THRESHOLD or SLEW RATE. After selecting THRESHOLD option,
set the minimum amplitude of disturbances in the range from 50 V to 5000 V
that should be detected as an event and its recording. After selecting SLEW
RATE the analyzer detects disturbances with slew rate exceeding the value
specified next to the selection box. Slew rate is changed indirectly, by
selecting the sampling frequency (from 100 V/500 µs for 100 kHz to 100 V/5 µs
for 10 MHz) Recording time charts of detected transients with related 10.24
kHz waveforms may be activated by selecting option RECORD WAVEFORMS AND
TRANSIENT CHARTS. Recording time is set in ADDITIONAL tab, of the card for the
measurement configuration settings, see sec. 3.2.1.2.
35
Sonel Analysis 4 User Manual 3.2.4.2 ADDITIONAL tab
In this tab (see Fig. 24) the user defines recording parameters for the mains
frequency, Crest Factor, voltage flicker and voltage unbalance. Similarly to
other parameters, the user may select recording of average, minimum, maximum
and instantaneous values. In the case of unbalance, the recorded parameters
include: three symmetrical components in volts (zero, positive and negative
sequence voltages) and two unbalance factors (in %): zero and positive
unbalance. Additionally, the user may enable event recording. The setting
ranges are as follows: frequency: 40 … 70 Hz with a resolution of 0.01 Hz or
0.01%, whereas the minimum value
cannot be higher than the defined mains reference frequency, and the maximum
value cannot be lower this reference value, crest factor: 1.00 … 10.00,
resolution 0.01, negative sequence unbalance : 0.00…20.00% every 0.01%,
flicker (short-term PST and long-term PLT): 0.00 … 20.00 in 0.01 steps.
Fig. 24. Additional settings for voltage measurements.
36
3 Recording Settings 3.2.5 CURRENT
Screen of Basic settings is shown in Fig. 25. Note
This tab is not available if in the main measurement configuration screen the
current measurement was disabled by selecting NONE option in PROBES TYPE
field.
The following elements are grouped here: recording RMS current, selection list
of averaging times to determine minimum and maximum values (similarly to
voltage, the available times include: ½ period, period, 1, 3 and 5 seconds),
recording current crest factor, recording unbalance factors and current
symmetrical components (as in the case of voltage
unbalance), DC current component (only when C-5(A) current probes are
selected).
Fig. 25. Current measurement settings. For each parameter, the user may enable
event logging: for RMS current two thresholds may be set : the maximum (upper
limit) and minimum (lower
limit) value. The settings range is from 0.00 A up to the rated current
measuring range (probes and possibly current transducers). The minimum value
cannot be greater than the maximum
37
Sonel Analysis 4 User Manual value. For mains with neutral conductor, the user
may set different thresholds for phase channels and for neutral channel (if
current measurement in N conductor is enabled). current crest factor: two
thresholds (minimum and maximum); adjustment range: 1.00…10.00; in 0.01 steps
negative sequence unbalance: one threshold (maximum), range 0.0…20.0%; in 0.1%
steps. for the DC component a current threshold can be defined. If the
absolute value of DC current exceeds the limit, the event will be recorded.
This kind of event can be enabled only when current probes with DC measurement
capability are selected, that is C-5(A). 3.2.6 POWER AND ENERGY The
configuration part relating to the power and energy settings has three tabs:
POWER ,ADDITIONAL and ENERGY.
Note These tabs are not available if in the main measurement configuration
screen the current measurement was disabled by selecting NONE option
in PROBES TYPE field. 3.2.6.1 POWER tab
Fig. 26 presents POWER tab.
Fig. 26. Power and energy Power tab. 38
3 Recording Settings Similarly to voltage and current, the user may set
recording average, minimum, maximum and
instantaneous values. The user may activate registering of the following
parameters: active power, reactive power, apparent power, distortion power /
apparent distortion power.
For each of these power types, the user may enable event recording (logging).
Two thresholds can be set for each power type (minimum and maximum) and the
analyzer will detect every case of their exceeding.
The setting range is 0.00 W … 999.9 MW (for active power), 0.00 var … 999.9
Mvar (for reactive power), 0.00 VA … 999.9 MVA (for apparent power), 0, 00 var
/ VA … 999.9 Mvar / MVA (for distortion power/ apparent distortion power).
At the bottom of the screen, the user may see the method for calculating
reactive power. Two alternative methods are available: as required by IEEE
1459-2010 standard, according to Budeanu’s power theory (this method is not
recommended).
Fig. 27. Power and energy Additional tab. This option has been added due to
the doubts about the power measurement according to traditionally used
Budeanu’s method. IEEE 1459-2010 standard suggests slightly different
calculation methods for this power, providing correct results also in
unbalanced systems with distorted current and voltage waveforms. More about
this issue may be found the operating manual of the analyzer.
39
Sonel Analysis 4 User Manual After selecting Budeanu option both reactive
power Q and distortion power are calculated by the analyzer based on the
Budeanu’s theory. After selecting IEEE-1459 option, the reactive power is the
Q1 reactive power of the fundamental component, and SN non-fundamental
apparent power (expressed in VA) is calculated instead of the distortion
power. 3.2.6.2 ADDITIONAL tab ADDITIONAL tab (Fig. 27) shows parameters
related to the power measurements: Power Factor: event threshold setting range
0.00 … 1.00 in 0.01 steps, Displacement power factor / cos: event threshold
setting range 0.00 … 1.00 in 0.01 steps, 4-quadrant tan, which is the ratio of
reactive to active power. Event threshold setting range: 0.00…10.0 in 0.01
steps. In every case both positive and negative values within the range are
checked.
PQM-700 and PQM-701 analyzers do not record 4-quadrant tan. 3.2.6.3 ENERGY tab
ENERGY tab is shown in Fig. 28. Apart from the option of enabling the
recording process of energy values with a defined averaging period, the user
may also define event detection when energy exceeds a given threshold. The
setting ranges are as follows: Active energy EP: 0.00 Wh…9.90 TWh, with
resolution of 0.01 Wh, Reactive energy EQ (4-quadrant): 0.00 varh…9.90 Tvarh,
with resolution of 0.01 varh, Apparent energy ES: 0.00 VAh…9.90 TVAh with
resolution of 0.01 VAh.
For active energy, the analyzer separately measures two parameters: consumed
(received) and supplied (delivered) energy. The indicated threshold refers to
both values (i.e. the analyzer checks the exceeding threshold for both,
consumed and supplied energy).
For reactive energy, the analyzer records and detects events separately for
each of four quadrants: EQ(L+), EQ(C-), EQ(L-), EQ(C+).
PQM-700 and PQM-701 analyzers do not record 4-quadrant reactive energy.
40
3 Recording Settings
Fig. 28. Power and energy Energy tab. 3.2.7 HARMONICS
In HARMONICS card the user may configure recording method and event detection
for voltage and current harmonics and other parameters directly related to
them. The card has three tabs: VOLTAGE , CURRENT and ADDITIONAL.
PQM-700 measures harmonics only to the 40th order. 3.2.7.1 VOLTAGE tab
Screen of this tab is shown in Fig. 29. The user may choose recording of
minimum, average, maximum and instantaneous values for THD in voltage and
voltage harmonic amplitudes. The user may also activate the event recording
for THD and harmonics amplitudes: for THD in voltage, an event will be
recorded after THD exceeds the threshold set by the user.
Event threshold setting range is 0.00 … 100.00% in 0.01% steps Detection
thresholds for events of voltage harmonics may be set independently for each
of 49
harmonics (from 2 to 50). The user may select units: volts or percentage. The
adjustment range for event thresholds: in volts 0.00…UNOM in 0.01 V steps, as
percentage values 0.00… 100.00% in 0.01% steps. Setting the threshold to zero
disables the event detection for the specified harmonic. Setting limits for
individual harmonics may be performed in two ways: completing the table below
the graph or directly on the graph. When the table is filled, after values are
approved, the graph located above presented the table visualization.
41
Sonel Analysis 4 User Manual To change the limit on the graph, double-click on
a selected harmonic and then holding down the left mouse button, set the
desired level. By clicking a harmonic in the table, the user may use the
following keys: SPACEBAR- used to enter the editing selected harmonic (the
user may also immediately start typing the threshold), ENTER – confirming
values TAB confirming values and moving to the next harmonics, UP and DOWN
arrow – changing the value of the limit, LEFT and RIGHT arrow – used for
moving to the previous or next harmonics HOME and END – used to toggle between
the first and last harmonics.
Note The level of THD and harmonics presented as a percentage value is
referenced to the fundamental component.
Fig. 29. Voltage harmonics settings. The user also has the option to decide
how many harmonics will be used to calculate THD: 40 or 50 harmonics. This
setting is automatically transferred to the second tab CURRENT (and vice
versa).
42
3 Recording Settings
3.2.7.2 CURRENT tab Screen of this tab is shown in Fig. 30. Available options
are similar to those in VOLTAGE tab:
recording THD (with event detection option) and recording current harmonics
(with setting of event thresholds for individual harmonics). Additionally the
TDD recording (average value) can be enabled.
PQM-701 analyzer does not have the capability to measure TDD.
The event threshold setting range is as follows: for THD: 0.00…200.00% every
0.01%, for harmonics: in amperes 0.00…INOM in 0.01 A steps, as a percentage
value 0.00…200.00% in
0.01% steps. Setting the threshold to zero disables the event detection for
the specified harmonic.
3.2.7.3 ADDITIONAL tab
Fig. 30. Current harmonics settings.
This tab is not available in PQM-700 analyzer.
ADDITIONAL tab (Fig. 31) includes the following parameters: K-factor: the user
may activate the recording and event detection when the threshold is
exceeded. The threshold may be set in the range of 0.00 … 50.00 in 0.01 steps.
Angles between voltage and current harmonics: the user may activate the
recording the
angles are recorded for 1..50 harmonics, Harmonics active power: cyclical
recording of active power for the 1..50 harmonics (minimum,
maximum, average and instantaneous values), Harmonics reactive power: cyclical
recording of reactive power for the 1..50 harmonics
(minimum, maximum, average and instantaneous values),
43
Sonel Analysis 4 User Manual
Fig. 31. Additional settings of harmonics. 3.2.8 INTERHARMONICS
In PQM-700 and PQM-701 analyzers the options for measuring interharmonics and
mains signalling is not available. In INTERHARMONICS tab, the user may
configure the manner of recording and event detection from current and voltage
interharmonics, TID and mains signalling. The card is divided into three tabs:
VOLTAGE, CURRENT, MAINS SIGNALLING. 3.2.8.1 VOLTAGE tab Screen of this tab is
shown in Fig. 32. The user may choose recording of minimum, average, maximum
and instantaneous values for TID in voltage and voltage interharmonic
amplitudes. Similarly as in case of harmonics, the user may also enable event
logging for TID and interharmonics amplitudes: for TID in voltage, an event
will be recorded after TID exceeds the threshold set by the user. Event
threshold setting range is 0.00 … 100.00% in 0.01% steps Detection thresholds
for events of voltage interharmonics may be set independently for each of 51
interharmonics (from 0 to 50). The user may select units: volts or percentage.
The adjustment range for event thresholds: in volts 0.00…UNOM in 0.01 V steps,
as percentage values 0.00…100.00% in 0.01% steps. Setting the threshold to
zero disables the event detection for the specified interharmonic. 44
3 Recording Settings Setting limits for individual interharmonics is carried
out in the same way as for the harmonics. Note The level of TID and
interharmonics presented as a percentage value is referenced to the
fundamental component.
Fig. 32. Voltage interharmonics settings.
45
Sonel Analysis 4 User Manual 3.2.8.2 CURRENT tab
Screen of this tab is shown in Fig. 33. Available options are identical as
those in VOLTAGE tab: recording TID (with event detection option) and
recording current interharmonics (with setting of event thresholds for
individual interharmonics). The event threshold setting ranges are as follows:
for TID: 0.00…100.00% in 0.01% steps, for interharmonics: in amperes 0.00…INOM
in 0.01 A steps, as a percentage value
0.00…100.00% in 0.01% steps. Setting the threshold to zero disables the event
detection for the specified interharmonic.
Fig. 33. Current interharmonics settings.
46
3 Recording Settings 3.2.8.3 MAINS SIGNALLING tab
MAINS SIGNALLING tab is not available for PQM-700 and PQM-701 analyzers.
Screen of this tab is shown in Fig. 34. The user may configure two monitored
frequencies of ripple control signals: UR1 and UR2. The frequency of each may
be set in the range of 5…3000 Hz. The meaning of each option is as follows:
AVERAGE selecting it activates recording of the average value of a given
control signal in
the averaging interval, MAXIMUM – selecting it activates recording of the
maximum value of a given control signal
in the averaging interval (out of 10/12-period values) LOG EVENTS (AND TIME
PLOT) activates recording of an event when specified threshold is
exceeded by the control signal (as specified in volts or percentage values of
the nominal voltage). The exceeded value triggers recording of the signal time
plot. Recording time is set at ADDITIONAL card in main settings of the
measurement configuration (see sec. 3.2.1.2). In case of recording in
accordance with EN 50160 (and related), in addition to averages, all the
3-second mean values are also recorded.
Fig. 34. Mains signaling settings.
47
Sonel Analysis 4 User Manual
3.2.9 Default configuration profiles. After installing, the application
applies the default settings of the analyzer, where the following
measurement configuration profiles are saved: Voltage , Voltage and current ,
Power , Power and harmonics.
Names of these profiles are entered in the names of measurement configuration
(K1 – Voltage, K2 – Voltage and current, K3 – Power, K4 – Power and
harmonics). Parameters recorded in individual profiles are shown in Table 3.
Table 3. The list of recorded parameters for the profiles of default measurement configurations.
Voltage Voltage U Crest Factor U Frequency Symmetrical components and unbalance factors U
Voltage and current Voltage U Crest Factor U Frequency Symmetrical components and unbalance factors U Current I Crest Factor I Symmetrical components and unbalance factors I
Power Voltage U Crest Factor U Frequency Symmetrical components and unbalance factors U Current I Crest Factor I Symmetrical components and unbalance factors I Active power P Reactive power Q Apparent power S Distortion power D Cos Power Factor Tan
Power and harmonics Voltage U Crest Factor U Frequency Symmetrical components and unbalance factors U Current I Crest Factor I Symmetrical components and unbalance factors I Active power P Reactive power Q Apparent power S Distortion power D Cos Power Factors Tan K factor THD U THD I Harmonics U Harmonics I Flicker Pst and Plt
Recording in default profiles is done without saving the waveforms after the averaging period. Only average values of parameters are recorded and event detection is disabled. Network type is set as 3-phase wye system with a neutral conductor and the averaging time is 10 seconds. For profiles that require current measurements, flexible probes F-x(A) are selected.
48
3 Recording Settings
3.3 Auto-probes support
The Sonel Analysis application from version 4.4.2 supports current probes with
automatic type recognition. This option is available for analyzers PQM-702
(A/T), PQM-703, PQM-710 and PQM711 with hardware revision HWg (or newer) and
firmware v1.40 or newer, when using probes that have the letter A in the
designation, e.g. F-3A or C-4A. When configuring current probes for
measurement instead of indicating their specific type, you can select AUTO
option. Several aspects related to this functionality are described below.
To start the recording with the auto-probe option enabled in the
configuration, it is necessary to correctly connect the auto-probes in all
required current channels. If any incompatibilities are detected, start of
recording will not be possible. In the case of Sonel Analysis an appropriate
error message will be displayed. In the case of manual operation from the
analyzer level, an error message will also be displayed on the LCD screen
recording can not be started.
The current status of the detected probes can be checked in the PROBES PHASE
window (ANALYZER SETTINGS PROBES PHASE, see chapter 4.2). For each current
channel, the type of set (if a specific type was indicated in the
configuration) or detected probe (if the AUTO option was selected) is given.
On the LCD screen of the analyzer (screen no. 9) if the configuration of the
probes is correct, in the Probe line, the symbol of detected automatic probes
is displayed, e.g. Probes: AUTO (F3A). If there is a detection error or
mismatch, a question mark is displayed, e.g. Probes: AUTO (?).
It is required that all current channels used have exactly the same type of
connected autoprobes. For example, in a 1-phase system with IN current
measurement, in channel I1 and IN it is necessary to connect the same type of
auto-probes, e.g. F-3A. Combinations of different auto-probes are not allowed
(e.g. F-3A in I1 and C-4A in IN).
The type of probes recognized at the start of recording is latched during the
entire recording period, until its completion. The current path in the
analyzer is set accordingly to the detected probes and its change is blocked
during the recording. This means that if at the start of the recording, the
F-3A probe was detected and blocked, then during the recording other automatic
probes were connected to the analyzer, e.g. C-4A – the currents from
connecting the C-4A probes will not be correctly measured, because the
measurement path has been configured to work with flexible probes. Restoration
of F-3A probes will result in correct measurement of currents from the moment
of their reconnection. The fact of blocking a given type of probes during
recording is indicated on the analyzer’s screen with an additional asterisk.
The type of probes detected and assigned to recording (latched) is possible to
view during data read-out and analysis using Sonel Analysis. See Fig. 57 and
Fig. 75 with their descriptions.
49
Sonel Analysis 4 User Manual
4 Analyzer Settings
Note: Analyzer Settings module is not available in MPI-540, MPI-540-PV and
PQM-707 meters.
After selecting from the main menu: ANALYZERANALYZER SETTINGS (or use a
shortcut key F4) the window shown in Fig. 35 is displayed. The following cards
are available:
TIME AND SECURITY , PROBES PHASE , WIRELESS CONNECTION (only PQM-702(A/T),
PQM-703, PQM-710, PQM-711), SERIAL CONNECTION (only PQM-701Zr), ADDITIONAL
SETTINGS.
4.1 TIME AND SECURITY card
Here (Fig. 35) the user may set the time and parameters important for the
analyzer security. Current date and time of the analyzer is shown on an on-
going basis in DATE AND TIME IN ANALYZER field. In order to synchronize a PC
system time with the analyzer time press the SET button next to DATE AND TIME
IN PC button.
Note: To ensure that the local time on PC and analyzer be the same, the time
zone and DST (Daylight Saving Time) option should be set appropriately. See
also description below.
In the upper left part of the window, an icon and status of GPS signal is
displayed. When correct time from the GPS satellite was received, ‘GPS signal’
icon turns green and message “YES (2D + Time)” is displayed. In the absence of
GPS signal, the icon is grey and message “NO SIGNAL” is shown.
Date and time of the analyzer for devices with built-in GPS (PQM-702 (A/T),
PQM-703, PQM710 PQM-711) are displayed in orange if there is no
synchronization with GPS time or green, if the synchronization is achieved.
Note: If the memory card of the analyzer contains any data saved with a date
later than the set date, the time change is not possible appropriate message
is displayed. To change time in this situation, you must first delete all the
data on the memory card (use the formatting options, or send a new
configuration to the device; first read the previously recorded data).
TIME ZONE field allows user to specify the time shift between the local time
and the UTC. For example, in case of Poland, select UTC +01:00. Many countries
in summer months use additional shift of one hour. For compatibility of the
analyzer clock with local time, when applying Daylight Saving Time, check DST
box. Time change to daylight saving time takes place at different times in
different countries. Analyzer does not automatically adjust DST. Checking DST
box adds 1 extra hour to the displayed time (local time). Its internal clock
is always UTC clock.
50
4 Analyzer Settings
Fig. 35. Setting the time and security measures. Checking KEYBOARD LOCK box,
results in immediate key locking after the recording is activated. In this
case, the user will have to enter the 3-digit code (using three buttons of the
analyzer) to unlock the keys. This code is independent from PIN number and may
be changed by pressing CHANGE button, next to KEYBOARD LOCK box. “Keyboard
Lock” function prevents stopping an active recording process by unauthorized
persons.
In meters with a built-in GPS module and GSM modem (PQM702(T), PQM-703,
PQM-710, PQM-711), the ANTI-THEFT option is available. Check this box will
enable this mode. More information on this issue may be found in the
instruction manual of these analyzers. SLEEP MODE makes the screen blank 10
seconds after activating the recording process.
4.2 PROBES PHASE card
In case of a reverse probe connection to the measured circuit (right direction
is the one where the arrow on the probe indicates the load/receiver), select
the option from menu ANALYZERANALYZER SETTINGS and selecting PROBES PHASE
(Fig. 36). Click a probe icon in the selected channel to reverse the phase by
180, which is equivalent to reverse placement on the probe on the test lead.
The change of direction applies to all measurement configurations and occurs
immediately. The reversal effect can be verified on the phasor diagram screen,
waveforms screen or by change of active power sign in the channel. The probe
phase cannot be reversed while the recording is in progress.
An additional function of this window is the preview of the currently
configured type of current probes, which is displayed above the picture of the
probe. If the AUTO option is selected in the configuration as the probe type,
i.e. automatic detection of the type of connected probes, then the indicated
type of probe detected for each channel is displayed in the indicated place.
The analyzer’s manual contains information about the hardware and firmware
versions that are required to support this functionality.
51
Sonel Analysis 4 User Manual
Fig. 36. Reversing the probe phase.
4.3 WIRELESS CONNECTION card
This tab is not available in PQM-700 and PQM-707 analyzers. PQM-701(Z, Zr) has
neither GSM communication options nor Wi-Fi. WIRELESS CONNECTION (Fig. 37)
includes the following elements: OR-1 COMMUNICATION ENABLED button used for
enabling/disabling radio interface to communicate with the receiver OR-1. GSM
COMMUNICATION ENABLED button used for enabling/disabling the built-in GSM
modem. Disabling the modem is recommended when GSM communication is not used.
WI-FI COMMUNICATION ENABLED button it enables / disables the built-in Wi-Fi
module (valid for PQM-710 and PQM-711). Disabling the module is recommended
when Wi-Fi communication is not used.
Note Be careful when disabling interfaces, as the user may turn off both
interfaces when connected to the analyzer via selected interface. In such
case, a warning is displayed informing that further communication is
impossible, and re-activation will be possible by connecting the analyzer
with other active medium.
52
4 Analyzer Settings
Fig. 37. Analyzer settings Wireless connection. SMS ALERTS – here the user
may enter up to four telephone numbers, to which SMS Alerts
will be sent. It is recommended that the phone numbers start with “plus’ and
the country code. SMS will be sent in the following situations:
recording starts, recording ends, analyzer is switched-off due to discharged
battery, analyzer is switched-on due to power supply return, analyzer re-
location was detected (when the anti-theft function is activated). GPS signal
is lost for longer than 5 minutes (when the anti-theft function is active),
GPS signal is available after interruption longer than 5 minutes (when the
anti-theft
function is active). ACTIVE button, next to the field with number is used to
activate and deactivate of a given number. After entering the numbers and
choosing activities, data must be sent to the analyzer by clicking SEND
button. SMS messages contain the time of the event (acc. to the analyzer time,
adjusted to local time) and the serial number of the analyzer. SMS alerts
require active GSM modem for proper operation (with a SIM card).
53
Sonel Analysis 4 User Manual
4.4 SERIAL CONNECTION card
This card is available only in PQM-701Zr. This window allows user to change
the settings of the serial port in PQM-701Zr. The user may change: parity
(even, odd, none), flow control (disabled or hardware) the number of stop bits
(1 or 2), bit rate (range from 2400 to 921600 bit/s). The change is made
immediately after CHANGE button is pressed and the connection is not
interrupted (as is the case with the RS-232 connection). Settings cannot be
changed if the connection to the analyzer is made via an external GSM modem.
Fig. 38. Changing the settings of the serial port in PQM-701Zr.
4.5 Additional settings card ADDITIONAL SETTINGS card includes two elements:
LANGUAGE selection list that allows user to set the language used on the
analyzer screen.
Language selection of the analyzer’s interface is not available for PQM-700,
PQM-701(Z, Zr) and PQM-707. You can select from the following languages:
Polish, English, English American (with date format MM/DD/YYYY), German,
Russian, Ukrainian, Chinese Simplified, Chinese Traditional.
54
4 Analyzer Settings HEATER ENABLED button is used for switching on/off the
heater that heats the interior of the analyzer when ambient temperatures are
below 0C. When the analyzer is activated in temperature below zero, it
controls the temperature inside the analyzer by enabling the heater to
maintain the temperature inside above 0C. This results primarily from the need
to provide positive temperature to built-in lithium-ion battery, as charging
the battery in temperatures below zero Celsius is not allowed.
Fig. 39. Language and heater settings. Note: The heater should always be
activated. The heater may be turned off only when the analyzer is operated at
a temperature above 0°C or when it is necessary to reduce the power consumed
by the analyzer, e.g. during measurements with voltage transformers, from
which the analyzer is to be supplied.
55
Sonel Analysis 4 User Manual
5 Analyzer database
Select from the main menu OPTIONSANALYZER DATABASE – here you may add or edit
existing analyzers (Fig. 40).
Only analyzers entered by the user into the database will be displayed in the
selection box when searching analyzers via radio. Serial number entered into
the database must match the serial number of the analyzer, which you want to
connect with. Analyzers with other serial numbers are rejected by the program
and not shown in the selection window (it does not apply to USB search).
Column AUTO PIN informs, whether the program requires PIN code before
transmission. If the box is disabled, the user will be required to enter PIN.
In addition, the user may activate the option of reminding about the need for
re-calibration. The manufacturer specifies the time after which the user
should perform a calibration of the instrument, in order to maintain its
declared accuracy. The accuracy degrades due to the ageing of components. To
unlock this feature, select option CERTIFICATE OF CALIBRATION. If it is
checked, then the following lists are active VALIDITY PERIOD and REMIND
BEFORE. They are used for: VALIDITY PERIOD (may be set at 6, 12, 18 or 24
months) it defines the frequency of instrument
calibration, REMIND BEFORE (may be set at week, two weeks, month) it
specifies the period for which the
program is required to generate the reminding message. Checking the list of
analyzers for validity of their calibration certificate occurs during start-up
of Sonel Analysis a window is displayed appear with a reminder.
Fig. 40. Analyzer database.
The remaining columns in the table: CALIBRATION DATE indicates the date of
last calibration, DESCRIPTION – brief text description related to the
analyzer, DATE OF CALIBRATION the date of analyzer’s calibration as set by
the user,
56
5 Analyzer database AUTO PIN? if checked it means that the user doesn’t have
to enter the analyzer’s PIN each time when connecting (the PIN is stored in
database), SIM ACTIVE? indicates whether the SIM card is correctly
configured in the analyzer (applies to analyzers with a built-in GSM modem and
is set after correct connection of the analyzer via GSM), GSM IP address – the
IP address of the analyzer in GSM network is displayed (in analyzers with
built-in GSM modem and PQM-701Zr), Wi-Fi IP address the IP address of the
analyzer in Wi-Fi network (local or Internet) (applies only to PQM-710 and
PQM-711 analyzers),
Clicking ADD enables user to add an analyzer – as shown in Fig. 41. Please
select a model analyzer, enter its serial number and optional description. The
calibration date may be also entered and the program will use it to alert the
user before the expiry date of the calibration certificate. Changing PIN code
and all of communication settings (Wi-Fi and GSM) is possible only after
connecting to the analyzer.
REMOVE icon allows the user to delete the selected analyzer from the database
(after confirming it by the user).
EDIT icon allows the user to change the parameters of the analyzer and
configuration of the communication parameters.
When editing selected analyzer click CHANGE PIN button to change its PIN. The
new code must be entered twice and then accepted.
Fig. 41. Adding an analyzer to the database.
57
Sonel Analysis 4 User Manual
5.1 GSM connection settings
In case of analyzers with built-in GSM modem, it is necessary to properly
configure the parameters to connect the analyzer to a GSM network and allow
connection of Sonel Analysis with a remote analyzer. To do this, click CHANGE
GSM SETTINGS button. Two main applications of this function are available:
Configuring the analyzer to operate in a GSM network (first activation of this
function in the analyzer requires to connect the analyzer via USB)
Adding a remote analyzer to the database (when analyzer has already configured
connection to GSM network (offline)).
In the first case, enter the following data in the appropriate fields: Static
IP number assigned to your SIM card, The name of the access point (APN), User
name and password (not always required). After confirming the settings, a
window will be displayed to enter the PIN code of SIM card.
Then, if the SIM card is inserted, the analyzer will initiate the procedure to
log into GSM network, applying specified settings. All parameters should be
provided by the SIM card provider.
In the second case, the analyzer may be in another location. In order to
connect to the analyzer, the user must know its IP address. Adding the
analyzer to the database is carried out as follows:
Click CHANGE GSM SETTINGS button. “Search window” will be displayed – click
‘Cancel’. Message “The analyzer is not connected” will be displayed with a
panel for GSM settings
in offline mode, in order to enter the IP number of the analyzer (Fig. 42).
Enter the IP address of the remote analyzer and click ‘OK’. You may try to
connect to the analyzer (e.g. by using option CONNECT SELECTED or by
applying the standard method). Please note that the application searches
analyzers via the Internet only if you selected TCP/IP OVER GSM option in
program settings.
Using the procedure shown above, the user may add PQM-701Zr (with an external
GSM modem) to the database of analyzers. Modem configuration is performed
using the software provided by the manufacturer of the modem.
58
5 Analyzer database
Fig. 42. Entering GSM settings in the analyzer database.
5.2
Wi-Fi settings
The PQM-710 and PQM-711 analyzers are equipped with Wi-Fi module that can be
used for wireless communication with the analyzer. In the analyzer database
the user can input all the needed parameters for setting a proper connection
of Sonel Analysis software with the analyzer.
This is the purpose of the configuration panel that appears after selecting
CHANGE WI-FI SETTINGS button.
There are two distinct modes for this panel:
online mode when analyzer is connected to the application while editing
options in this panel. In this mode all available parameters can be set. This
mode is used to change the configuration that is saved by the analyzer, and
also the parameters used by Sonel Analysis software for communication. It is
mainly used for initial configuration of analyzer for operation in Wi-Fi
network.
offline mode when analyzer is not connected (e.g. it is in remote location).
In this mode no new configuration data is sent to the analyzer, and only
parameters used by Sonel Analysis can be modified. This mode is used when user
wants to add a new analyzer to the database, which already is configured for
Wi-Fi operation. This mode is indicated by the “(offline)” word on top of the
configuration panel. If the user wants to change parameters in offline mode,
and is already connected to the analyzer, the connection has to be terminated
first, and then CHANGE WI-FI SETTINGS button has to be clicked. In the opened
window with found analyzers the cancel button must be selected. A message
“Analyzer is not connected” will be displayed and then a configuration panel
in offline mode will be opened.
Offline mode is applicable only when the analyzer operates in the mode of Wi-
Fi client.
59
Sonel Analysis 4 User Manual 5.2.1 Configuring Wi-Fi in the Access Point mode.
To configure the analyzer for operation in the Access Point mode, press CHANGE
WI-FI SETTINGS and set the operating mode switch to ACCESS POINT. The
displayed window will look like the one in Fig. 43. In field NETWORK NAME
(SSID) enter the name of the Access Point, which will be distributed by the
analyzer. If for some reason you are not satisfied with the default Wi-Fi
channel (channel 10), you may change it by ending SSID name with “_chXX”,
where XX is the channel number in the range of 1…13. Then it is necessary to
set the PASSWORD for the access point it should be entered twice in the
fields below. The password must contain at least 8 characters.
Pressing RESTORE DEFAULTS button restores the factory settings of the
analyzer, specified in its manual.
After approving the settings with OK button, the analyzer restarts the Wi-Fi
module and activates the Access Point with the specified settings. It is now
possible to connect to a PC with a Wi-Fi network distributed by the analyzer,
and then establish a connection with Sonel Analysis software. 5.2.2
Configuring Wi-Fi in Client mode
To configure the analyzer for operation in the Wi-Fi client mode, press CHANGE
WI-FI SETTINGS and set the operating mode switch to CLIENT. If the analyzer is
connected to Sonel Analysis, then the displayed window will look as shown in
Fig. 44 (online mode). In the online mode configuration can be performed in
the following way:
Enter the following data: the access point name (SSID field) and in case of a
secured network tick MODIFY box and enter the password (KEY field). In case of
a non-secured network, KEY field remains empty, but MODIFY box must be ticked.
Fig. 43. Wi-Fi setup screen in online mode (access point in the analyzer).
60
5 Analyzer database
Fig. 44. Wi-Fi configuration screen in online mode (client mode).
Select the method for assigning IP address. In case of manual mode – enter
appropriate values into fields: IP ADDRESS , NET MASK and GATEWAY. In
automatic mode, select DHCP.
PORT field is not editable, it is always 4002. In case of operating in other
subnetworks (the Internet), fill-in EXTERNAL IP ADDRESS and
EXTERNAL PORT fields. For a direct connection (tablet analyzer), and for
working in a local network (tablet Wi-Fi router analyzer) these fields must be
left inactive. These settings are used only by Sonel Analysis when connecting
to the analyzer they are not stored in the analyzer memory. Confirm the
settings by pressing OK. This will send new data entered in fields: SSID, Key,
IP configuration, to the analyzer. If the correct data have been entered, the
analyzer will try to connect to Wi-Fi Access Point. After disconnecting, the
connection status may be followed on screen <10/10>. The following procedure
applies to offline mode, when the analyzer was previously configured, and
there is a need to add it to the database and enter IP address and port, under
which the analyzer Wi-Fi interface is available. These parameters are required
to connect with the analyzers working in a different subnetwork than the
tablet (computer), especially for connecting to the Internet. It is not
necessary to enter these parameters if the analyzer is connected to the same
network as the PC. To properly configure the connection, the following
elements will be required: EXTERNAL IP ADDRESS and EXTERNAL PORT. Configure
the analyzer in the following manner: Select the appropriate analyzer from the
database and click EDIT or use ADD button to add it to the database. In case
of operating in other subnetworks (the Internet), fill-in EXTERNAL IP ADDRESS
and EXTERNAL PORT fields. Confirm the settings by pressing OK.
61
Sonel Analysis 4 User Manual
Fig. 45. Example configuration when the analyzer is connected via another
network (offline mode).
62
6 Live mode
6 Live mode
Note: Live mode is not available in MPI-540 and MPI-540-PV. To enter the live
mode, click LIVE MODE icon. If the connection to the analyzer has not been
previously established, the program goes first to the connection screen (see
chapter 2.1). In this mode, the screen is divided into a few tabs: WAVEFORMS
it presents waveforms of current and voltages as in a oscilloscope, TIMEPLOT
presents RMS values of voltage and current as a timeline, MEASUREMENTS
instantaneous values of measured parameters (phase and total values), PHASOR
DIAGRAM graphical representation of phasors of fundamental voltage and
current components, HARMONICS view of harmonic components of individual
phases for currents and voltages, INTERHARMONICS view of interharmonic
components of individual phases for currents and voltages, The more detailed
description of individual tabs is given below.
6.1 Waveforms
The voltage and current waveforms presents two mains periods, as they would be
seen on the oscilloscope screen. A sample screen is shown in Fig. 46. The
graphs are refreshed instantaneously, allowing user to view the mains
parameters in real time. All channels are presented in a single graph.
Fig. 46. Live mode screen waveforms. 63
Sonel Analysis 4 User Manual
The buttons
above the diagram are used to enable/disable displaying of a chosen
voltage/current channel. Buttons next to the boxes are used to change the color of presented
waveforms.
The graph may be freely expanded in both axes. In the corners of the window there are four
buttons , which activate displaying a window thumbnail with actually displayed area marked in
orange. The thumbnail is displayed in one of the four corners of the window, depending on the
button used. Four buttons marked by the magnifying glass icon are used for zooming-in and –
out . Buttons are used to select the maximum or minimum zoom of a selected axis. Mouse
scrolling wheel may be also used for zooming-in and out. Using the scrolling wheel closer to the
vertical edge of the window results in zooming-in of Y-axis, whereas using the scrolling wheel closer
to horizontal edge – zooming-in of X-axis.
Zoom the selected area by holding down SHIFT button (the mouse pointer changes to a cross)
and select the desired area with the mouse.
It is also possible to move the selected area within the graph. To do so, hold CTRL key (the
cursor changes to a hand) and holding the left mouse button, move the display area.
The time scale (X-axis) is updated basing on the measured frequency value (e.g. from 0 to
40 ms for 50 Hz).
Two independent descriptions of Y-axis values are presented. For example, the user may
display values in volts (for voltage channels) on the left axis, and in amperes (for current channels)
on the right axis. Descriptions of axes are controlled by selection lists . These lists may include
the following items:
ALL V when more than one voltage channel is measured. In this situation, one scale is
selected for all the voltage waveforms, and the waveforms are scaled,
ALL A when more than one current channel is measured. In this situation, one scale is
selected for all the current waveforms, and the waveforms are scaled,
a separate entry for each measured channels. If the user selects this element from the list,
then the screen displays a scale for the selected channel (it may be adjusted automatically or
manually if the range was set by the user).
Note: If the graph presents other waveforms (voltage or current depending on the
selected scale) then their actual amplitude does not correspond to the indicated scale.
The scale applies only to the selected waveform.
The scale is selected: statically – the upper scale range is set at the value
close to the nominal voltage or maximum
probe range, dynamically – the graphs are automatically scaled to the highest
instantaneous value of the
waveform.
manually to do this use SET SCALE button, and then choose waveform(s) for
which the scale is to be set manually, disabling the automatic mode and this
activity will unlock the options of
editing the minimum and maximum values of the axis. After clicking SET button,
the scale of a given waveform is updated.
Switching between static and dynamic scaling mode is available after clicking the right mouse button on the graph: the context menu includes the following options: SET SCALE TO NOMINAL VALUE (for voltage) or SET SCALE TO PROBE RANGE (for current) and SET SCALE AUTOMATICALLY, both for voltage and current values. After turning the program on, auto-scaling is turned off.
64
6 Live mode 6.1.1 Synchronization of waveforms in LIVE mode
If the analyzer is able to receive a GPS signal (in unfavourable locations it
is recommended to use a dedicated 10m external antenna, offered as an
additional accessory), there is a possibility of a very precise
synchronization of the analyzer with the clock of GPS system. The
synchronization is indicated by green colour of the clock on the LCD of the
analyzer or by green colour of time in ANALYZER SETTINGS window. Time
synchronization is an essential condition and basis for a waveforms
synchronization function in LIVE mode. Every 30 seconds in every minute i.e.
at xx:00 and xx:30 of each minute, the phase shifts of waveforms are ‘frozen’,
which is additionally indicated by a characteristic start of the waveform
marked by a vertical marker (see Fig. 47; first sample of voltage waveforms
has a large amplitude equal to the peak value this sample is only an
indication not the actual value of the voltage). This indicator clearly
informs about the status of waveform image synchronization with the GPS.
After activating two separate copies of Sonel Analysis and connecting with two
different analyzers in LIVE mode, the images of waveforms shown on the screen
at the same time for 30 seconds are synchronized with each other. By comparing
the positions of L1 voltage signals, it may be unambiguously determined
whether U L1 test leads are connected to the same phase (assuming, of course,
that both analyzers are connected to the same supply/distribution network)
the same applies to L2 and L3 voltages. Assuming that one of the analyzers
connected in accordance with the description of line conductors is considered
as the reference analyzer, then the second analyzer synchronized basing on the
comparison of indications identifies unambiguously any errors in connections
or descriptions.
Fig. 47. LIVE mode waveform synchronization the marker of GPS
synchronization is visible.
Only after adjusting the connections and obtaining full compliance of waveform
images, the full compatibility of the connection between two analyzers may be
guaranteed. This is a necessary
65
Sonel Analysis 4 User Manual condition for the simultaneous multipoint
diagnosis that allows user to compare individual parameter values between the
measuring points.
If the GPS signal is not available and the time of the analyzer is not
synchronized with it, the behaviour of the graph of waveforms returns to
normal operation, i.e. the graph starts with a rising crossing of the zero
point of the reference phase.
6.2 Timeplots
Fig. 48 presents the screen with current and voltage timeplot. The
configuration elements are identical as those in the waveform screen: its
upper part includes controls for switching on/off the display of
current/voltage; changing the color of waveforms; zooming-in and the vertical
axis.
The time scale is shown in the graph from the moment to entering the Live
mode. The time format is mm:ss. After reaching the end of the window, the
entire graph shifts to the left by 30 seconds. There is also an option to
clear the graph tab in FILECLEAR. After clearing the screen, the timescale on
the plot is shown from the moment of this operation.
The graph is updated regardless of whether the tab is enabled or not, i.e.
waveforms are generated in the background, despite viewing for example
harmonics.
As in case of waveforms, the user may apply the static scaling (according to
the general settings of the measurement configuration) or dynamic scaling.
Fig. 48. Graph with values in real time.
66
6 Live mode
6.3 Measurements MEASUREMENTS tab provides a view of a range of parameters of
the measured mains. Fig. 49
presents an example of a screen that shows a summary table of values read from
the analyzer and displayed in real time. In the following rows the individual
parameters are grouped into several categories: VOLTAGE RMS voltage U, DC
component UDC, frequency f, CURRENT RMS current I, DC component IDC , POWER
active power P, reactive power QB (according to the Budeanu theory) or non-
fundamental apparent power Q1 (acc. to IEEE 1459), apparent power S,
distortion power D (acc. to Budeanu) or distortion apparent power SN (acc. to
IEEE 1459), ENERGY consumed active energy EP+ and reactive active energy
EP-, reactive energies EQ (acc. to Budeanu or IEEE 1459, 4-quadrant or
consumed and supplied), apparent energy ES , FACTORS Power Factor PF,
Displacement Power Factor/cos, tan factor (or 4-quadrant factors), THD for
voltage and current, Crest Factors for voltage and current, flicker PST and
PLT , UNBALANCE for voltage: zero sequence component U0, positive sequence
component U1 and negative sequence component U2, negative sequence unbalance
u2, zero sequence unbalance u0; or current: zero sequence component I0,
positive sequence component I1 and negative sequence component I2, negative
sequence unbalance i2, zero sequence unbalance i0. TEMPERATURE it indicates
the temperature of the external sensor (Sonel ST-2) in degrees Celsius. The
temperature sensor may be connected only to PQM-702T.
Fig. 49. Indications of phase and total values 67
Sonel Analysis 4 User Manual
Each group may be enabled/disabled independently of the other groups and
column background color may be changed using the buttons in the top of the
screen.
When the program is started for the first time, the following parameters are
shown: voltage, current, power and factor.
Using preferences the user may change the colors of groups and select groups
to be displayed (see sec. 8.3.3).
Parameter values are displayed in the consecutive columns, marked as follows:
L1 L1 phase in circuits with a neutral conductor N L2 L2 phase in circuits
with a neutral conductor N, L3 L3 phase in circuits with neutral conductor
N, N-PE values of parameters in the N-PE voltage channel or in IN current
channel, L1-2 L1 phase in circuits without neutral conductor (with phase-to-
phase voltages), L2-3 L2 phase in circuits without neutral conductor (with
phase-to-phase voltages), L3-1 L3 phase in circuits without neutral
conductor (with phase-to-phase voltages), TOTAL the total value of the
entire system or the average value of the phase values.
For example, for a three-phase system with N conductor, the phase parameters
values are displayed in columns L1, L2, L3. If the measurement of N-PE voltage
and/or IN current is enabled in the configuration, then in N-PE column the
analyzer displays also parameter values calculated for this case. TOTAL column
displays the total values for the whole system.
In case of delta systems, the phase values are displayed in columns L1-2,
L2-3, L3-1 and TOTAL.
If the current limit function is activated for currents below the threshold
(CURRENT LIMIT option), the zeroed values are marked with an asterisk (*).
Note In some mains configurations it is not possible to calculate (measure)
the values of some parameters. For example, in a delta three-phase system, it
is impossible to calculate the cos displacement power factor between current
and voltage, as the measured voltage is phase-to-phase voltage,
and the measured current is line current which is a vector sum of two phase
currents.
When in a given mains configuration a selected parameter is not calculated,
the screen displays “—” instead of its value.
In the right part of the tab, the user may additionally define the calculation
method for THD and TID (list THD AND HARMONICS VALUES RELATIVE TO): in
relation to fundamental component (THDF/ TIDF) or RMS value (THDR/ TIDR).
6.3.1 Energy costs calculator
On the right side, in MEASUREMENT tab (see Fig. 49), a button is available for
ENERGY COST CALCULATOR. After clicking the lower part of the window, a table
with TARIFF CALCULATOR is displayed, which shows energy values of the current
session (from opening tab MEASUREMENTS) and energy costs by three tariffs
configured in the program settings. Energy is counted by the calculator as
long as window LIVE MODE is open. The basis for calculating is total active
energy input (EP+).
68
6 Live mode To analyze and create reports on energy consumption based on
recorded data use the module
for reporting energy costs which is in the data analysis section (see sec
7.4.5).
6.4 Phasor diagrams
Phasor diagram (Fig. 50) is used to present the vectors of fundamental voltage
and current components. Angles between the vectors correspond to the angles of
phase shift between individual phases. A table with numerical values is
displayed on the left side of the graph. Individual rows of the table present:
UH1 amplitudes of voltage fundamental components, U_H1 angles of voltage
fundamental components in relation to L1 voltage fundamental
component (in angle degrees), IH1 amplitudes of current fundamental
components, I_H1 angles of current fundamental components in relation to L1
voltage fundamental
components (in angle degrees), U,I_H1 angles between voltage and current
vectors (U I) for a given phase (not available
in 3-wire systems), TYPE the type of a receiver, indicated by a coil symbol
for inductive load or a capacitor symbol
for capacitive load (unavailable in 3-wire systems). Above the diagram are the
on/off buttons for VOLTAGE , CURRENT, angle U,I and a load TYPE. In addition,
the user may change the default background color of the table for these
groups.
Fig. 50. Phasor diagram for wye connection.
69
Sonel Analysis 4 User Manual
The vector angles are referenced to UL1 vector angle which is always = 0°. The
vector amplitude scaling takes place automatically in relation to the highest
value, independently for voltage and current.
On the diagram two buttons are present for turning the phasor diagram by 90
clockwise or counter-clockwise.
In the bottom right part, button PROBES SETUP is shown, which initiates a
window for changing phase of probes (shown in Fig. 50).
Below the PROBES SETUP button there is a control panel for checking correct
connection of the analyzer to the tested network/system. This panel displays
five indicators that user may use to quickly perform a visual check of the
analyzer connection. The displayed indicators correspond to the indicators
displayed on the screen of the analyzer (this applies to the
PQM702/703/707/710/711; a detailed description is in the manuals of those
analyzers). The indicators are as follows:
URMS RMS voltage, IRMS RMS currents,
U angles of voltage vectors (fundamental components), I angles of current
vectors (fundamental components),
f frequency. Green icon indicates the correctness of the parameter, red
probable error in connection or configuration. The icon with a question mark
indicate ambiguous situation correctness or incorrectness cannot be
determined with certainty.
6.5 Harmonics
PQM-700 and PQM-707 analyzers do not measure the K factor. Additionally, the
PQM-700 analyzer has no function for measuring active and reactive power
harmonics this analyzer measures harmonics only up to the 40-th.
Harmonics tab enables user to view harmonic values in voltage and current,
active and reactive harmonics power, THD and K-factor. A typical screen is
shown in Fig. 51. Parameters are displayed for one phase, which is selected by
buttons on the right.
A bar graph with harmonics is displayed in the central part of the screen:
from DC component, through the first component up to the 50th.
MEASUREMENT list on the right side of the window allows selecting one of two
options: VOLTAGE, CURRENT displays the voltage and current harmonics,
HARMONICS POWER displays active and reactive harmonics power.
On the left and right side of the graph, axes are displayed and marked with
units corresponding to the chosen mode: VOLTAGE, CURRENT – left side is
assigned to voltage axis, while the right side is to current axis. In
HARMONICS POWER mode, on the left side, the axis of active power is located,
while on the right the axis of reactive power. This arrangement may be
changed, using the lists above the axes.
The tables below the bar graph present (from the left): THD factors, K-factors
(only in VOLTAGE, CURRENT mode),
70
6 Live mode
Fig. 51. Harmonics on 2D graph.
a table with numerical values showing the following parameters:
o U – voltage harmonics magnitudes (including DC) (only in the VOLTAGE,
CURRENT view),
o I – current harmonics magnitudes (including DC) (only in the VOLTAGE,
CURRENT view),
o P – harmonics active powers (only in HARMONICS POWER view),
o Q – harmonics reactive powers (only in HARMONICS POWER view), o cos()U,I –
displacement power factors of harmonics, o U,I – phase angles between voltage
and current harmonics (U – I), o U – absolute phase angles of voltage
harmonics in reference to the voltage fundamental
angle (harmonics of order 2 and higher). Fundamental angle is shown in
reference to the L1 voltage phase angle, similar as in the phasor screen. o I
– absolute phase angles of current harmonics in reference to the voltage
fundamental angle (harmonics of order 2 and higher). Fundamental angle is
shown in reference to the L1 voltage phase angle, similar as in the phasor
screen.
The user has the access to two settings:
THD AND HARMONICS RELATIVE TO two options are available: relative to RMS or to the first
harmonics (this option is active only after selecting the percentage unit see below),
HARMONICS UNIT here also two options are available: in absolute units [V], [A], or as
percentage values.
A right-click on the graph in VOLTAGE, CURRENT mode, displays a context menu
with the following options: SET SCALE ACCORDING TO NOMINAL (for voltage) or
SET SCALE ACCORDING TO PROBES RANGE
(for current). Use this option to re-scale the harmonics bars and adapts them
to the graph window height, taking into account the nominal value,
71
Sonel Analysis 4 User Manual SET SCALE AUTOMATICALLY. This feature
automatically adjusts the scale. VOLTAGE LIMITS (option available only for
voltage harmonics and after selecting percentage
display). This option results in imposing harmonics limits defined in the
default standard (the default standard and its limits are determined in
PROGRAM SETTINGS, in section RECORDING SETTINGS STANDARD DEFAULT SETTINGS, see
also sec. 8.3.2). HIDE FIRST HARMONIC. This option turns on/off the plot
containing the fundamental harmonic.
Fig. 52. The screen with harmonic powers.
6.6 Interharmonics
PQM-700, PQM-701(Z, Zr) and PQM-707 have no option to measure interharmonic
components. INTERHARMONICS tab enables the view of interharmonics values of
voltage and current and TID factors. A sample screen is shown in Fig. 53.
Parameters are displayed for one phase, which is selected by buttons on the
right. In the central part of the window, a bar graph with interharmonics is
presented: it ranges from the subharmonic group (0) through all other groups
interharmonics up to the 50-th. On the left and right side of the graph, axes
are displayed and marked with units corresponding to the chosen mode: VOLTAGE,
CURRENT – left side is assigned to voltage axis, while the right side is to
current axis. This arrangement may be changed, using the lists above the axes
.
72
6 Live mode
The tables below the bar graph present (from the left): TID factors, a table
with numerical values of interharmonics amplitudes for voltage and current.
Fig. 53. Interharmonics graph.
The user has the access to two settings:
TID AND INTERHARMONICS VALUES RELATIVE TO two options are available: relative to RMS
or to the first (fundamental) harmonics,
INTERHARMONICS VALUES UNIT here also two options are available: in absolute units [V],
[A], or as percentage values.
A right-click on the graph, displays a context menu with the following
options:
SET SCALE ACCORDING TO NOMINAL (for voltage) or SET SCALE ACCORDING TO PROBES
RANGE (for current). Use this option to re-scale the interharmonics bars and
adapts them to the graph window height, taking into account the nominal value,
SET SCALE AUTOMATICALLY. This feature automatically adjusts the scale.
73
Sonel Analysis 4 User Manual
7 Data Analysis
The analysis block allows the user to read data collected by the analyzer
during the recording, view numerical data and graphical visualization. The
user may generate reports that in a clear manner describe the tested system.
described in sec. 7.1.1.
Reading data from MPI-540, MPI-540-PV and PQM-707 meters is
7.1 Reading data from the analyzer
After selecting analysis from the upper menu or clicking ANALYSIS icon, when
no analyzerprogram connection is established, a standard connection procedure
is launched to connect the analyzer and the program (sec. 2.1).
After correct connecting the analyzer to a PC, a window is displayed – as
shown in Fig. 54 and data may be selected for downloading and further
analysis. To do this, select the measurement configurations from which you
want to download the data. On the right side of the bar representing the
amount of stored data, the actual amount of data is displayed. To start
downloading data from
the selected measuring configuration, click READ DATA button.
Fig. 54. Selecting data for read-out.
Select DELETE DATA AFTER READING option to delete data from the configuration,
from which the data is read. Therefore, after pressing READ DATA button before
the transmission, an appropriate warning message is displayed
Data may be deleted from the card at any time by entering ANALYZER menu and
choosing the option of ANALYZERDELETE DATA. This action may be used to delete
data of chosen measurement configurations (without changing their parameters)
or to completely format the card.
During the data read-out, the status bar at the bottom of the window shows the
download progress and CANCEL button, which may be used for stooping the
process.
74
7 Data Analysis
PQM-702, PQM-703, PQM-710 and PQM-711 analyzers use the built-in mass storage
mode to accelerate the readout of data via the USB connection. This mode is
used when the amount of data to be downloaded exceeds 1 MB. If the amount of
data is below that limit, a standard (slower) communication mode is used.
Other analyzers use only standard communication mode.
Note When other, additional windows are also opened for displaying on-going
(live)
data from the analyzer (e.g. LIVE MODE or CONTROL), the standard (slower)
communication mode is used for data download. Before pressing READ DATA
button, close these additional windows in order to use the high-speed data
reading mode.
After data read-out, a window is opened for saving the file .pqm7xx (the file
extension depends on the analyzer model). In a file with this extension, the
software saves identical copy of data read from the analyzer (not processed by
the program). Saving this data is not required but highly recommended. Without
this file performing a next analysis, basing on recorded data will require
another data read-out from the analyzer (this is particularly important if the
user reads the data from several measuring configurations – saving later file
.analysis concerns only one measurement configuration selected for the
analysis; data from other configurations need to be re-read). Saving file
.pqm7xx will also allow you to create multiple analyses and .analysis files
from a selected (reduced) period of time, which will reduce the size of these
files.
Note When the user selected DELETE DATA AFTER READING option without saving
data before proceeding to data analysis, the source data will be irreversibly
lost. It will be possible to perform analysis and save data to a file with
*.analysis extension only from one measuring configuration (selected for
analysis), but you will be unable to fully use the source data for further
analysis (e.g.
from different time range), or from another measuring configuration.
7.1.1
Reading data from MPI-540, MPI-540-PV and
PQM-707 devices
A method of reading data differs in a few aspects from the description in sec. 7.1. After selecting ANALYSIS option the menu, similarly to other analyzers, the procedure of connection with the analyzer is launched (if the connection is not established earlier). After a successful connection the window shown in Fig. 55 is di
References
Read User Manual Online (PDF format)
Read User Manual Online (PDF format) >>