sauermann FG 110 Gas Leak Detector User Manual
- June 13, 2024
- sauermann
Table of Contents
sauermann FG 110 Gas Leak Detector
Gas leak detector
Description of the device
- Backlight key
- OK key
- On/Off/Esc key
- Select key
- Hold/min/max key
General features
Measurement units | ppm, %vol, %LEL |
---|---|
Measuring element | Semiconductor sensor |
Display | 4 lines, LCD technology. Dimensions 50 x 36 mm 2 lines of 5 digits |
with 7 segments (value) 2 lines of 5 digits with 16 segments (unit)
Probe| Flexible, 300 mm
Housing| ABS, protection IP54
Keypad| 5 keys
European directives| 2014/30/EU EMC; 2014/35/EU Low Voltage; 2011/65/EU RoHS
II; 2012/19/EU WEEE
Power supply| 4 batteries AAA LR03 1.5 V
Battery life| 20 hours
Ambience| Neutral gas
Conditions of use (°C, %RH, m)| From 0 to +50 °C. In non condensing
conditions. From 0 to 2000 m
Storage temperature| From -20 to +80 °C
Auto shut-off| Adjustable from 0 to 120 min
Weight| 295 g
Technical specifications
Accuracy(1) | Measurement range in CH4 | Resolution |
---|---|---|
20% of full scale | From 0 to 10,000 ppm From 0 to 1% vol 0 to 20% LEL | 1 ppm |
0.001% vol 0.001% LEL
% LEL: is the limiting concentrations of gas (in air) that is needed for the
gas to ignite and explode.
Example for CH4 gas: 100% LEL = 5% vol
% vol: is the concentrations of gas (in air) expressed in percentage of volume
in air.
Example for CH4 gas : 1% vol = 10 000 ppm
Perform a measurement
-
Turn on the device by pressing on “On/Off/Esc”.
The device displays its name “FG110”. When it turns on, the device emits a beep then performs a pre-heating phase. This phase lasts 60 seconds. The count of this pre-heating time displays on screen. -
During the normal functioning, in the absence of gas leak, the device emits a regular beep*.
-
Otherwise, if the device detects the presence of gas, the repetition frequency of the beep will increase as the detected concentration in
Freeze the measurement
During a measurement:
-
Press once on “Hold/min/max”.
“Hold” appears on screen and the measurement is frozen. -
Press on “On/Off/Esc” to exit the hold function. The device returns to the measurement display.
Display the minimum and the maximum
Once the measurement is frozen:
-
Press on “Hold/min/max”.
The device displays the maximum value measured since the last questioning at the top of the screen and the minimum value measured since the last questioning at the bottom of the screen. -
Press on “On/Off/Esc” to return to the measurement display
Adjust the device
Device drop-down menu
Select the measuring unit
The device is on and displays the measurements.
-
Press on “Select”.
“UNIT“ blinks on screen. -
Press on “OK”.
The unit currently used blinks on screen. -
Press on “Select” until the desired unit appears: ppm, %VOL, %LEL.
-
Press on “OK” to validate the unit selection.
-
“UNIT” blinks on screen.
-
Press on “On/Off/Esc” to return to the measurement display
Adjust the auto shut-off
The device is on and displays the measurement.
-
Press on “Select” until “AUTO OFF“ blinks on screen.
-
Press and “OK”.
The time before the device auto shut-off blinks at the bottom of the screen. -
Press on “Select” until the required time before auto shut-off appears: 15, 30, 45, 60, 75, 90, 105, 120 minutes or OFF.
-
Press on “OK” to validate.
“AUTO OFF” blinks on screen. -
Press on “On/Off/Esc” to return to the measurement
Activate or deactivate the keys beep
The device is on and displays the measurement.
-
Press on “Select” until “BEEP” blinks on screen.
-
Press on “OK”.
“OFF” or “ON” blinks on screen. -
Press on “Select” to activate the keys beep “ON” or deactivate it “OFF”.
-
Press on “OK” to validate.
“BEEP” blinks on screen. -
Press on “On/Off/Esc” to return to the measurement.
Adjust the alarm
This function allows to activate an alarm. When the device detects a leak, it
emits a beep whose frequency is increasingly fast as and when the
concentration increases and gets closer to defined alarm threshold. The beep
becomes continuous when the device reaches and exceeds this threshold. This
threshold is between 1000 and 10 000 ppm.
The device is on and displays the measurement.
-
Press on “Select” until “AL 1” blinks on screen.
-
Press on “OK”.
A succession of 5 zeros displays and the first zero blinks. -
Press on “Select” to select the first digit value : 0, 1 or Off.
-
If 1 is selected, the maximum threshold (10 000 ppm) is reached. Press on “OK” until “AL 1” blinks on screen.
-
If Off is selected, the alarm is deactivated. Press on “OK” to validate, “AL 1” blinks on screen.
-
If 0 is selected, the following digit blinks. Press on “Select” to select its value, then press on “OK”. The following digit blinks, perform the same procedure. Press on “OK” when the last digit value is selected, “AL 1” blinks on screen.
-
Press on “On/Off/Esc” to return to the measurement display
Activate the backlight
The device is on.
- Press on “Backlight” to activate the device backlight.
- Press again on this key to deactivate the backlight.
Change the batteries
- Remove the front part at the back of the device.
- Change the old batteries by AAA LR03 1.5V batteries.
- Replace the front part.