Megger PAT300 series Portable Appliance Testers User Manual
- June 4, 2024
- Megger
Table of Contents
PAT300 series
Portable Appliance Testers
PAT300 series
Portable Appliance Testers
PAT300 series Portable Appliance Testers
- Simple manual or automatic testing
- Bond testing at 25 A, 10 A, and 200 mA
- Differential, touch, and substitute leakage testing
- Full-color display
- Tests portable RCDs and extension leads
- Pre-set test sequences for Class I, Class II, and extension leads
- Flash Testing at 1.5 kV and 3 kV
- Tests to DIN VDE 0701-0702
DESCRIPTION
An easily portable desktop appliance tester for testing the safety of portable
electrical equipment to meet health and safety regulations. The PAT300 series
are fully featured testers with dedicated test buttons for direct access to
tests. They are designed for customers who do not require the complexity of a
fully configurable database of clients and results within the tester but do
need a complete range of functions to allow automatic or manual testing of the
widest range of electrical assets. There are two products in the range – the
PAT320 and the PAT350. The PAT350 is identical to the PAT320, with the
addition of flash-test capability for use in environments such as
manufacturing, production, or tool-hire shops.
The simple push-button operation makes the PAT300s fast and intuitive in use.
All regulatory test requirements are supported, including Class I and Class
II, IEC power leads, extension leads, and full tests for portable RCDs. An
automatic mode is available for Class I and Class II testing. In automatic
mode, the tests proceed sequentially through bond, insulation, and operation,
indicating a pass or fail at each test. If a failure occurs, testing is
stopped. When manual testing, each test is preceded by a selection screen
where the test parameters are selected, such as bond test current, insulation
test voltage, or leakage test type. These diagnostic buttons provide direct
access to any test individually, allowing single tests to be performed
following repair or a suspect result.
Accessories supplied with all models include a combined earth-bond and insulation test lead, an adaptor for testing extension leads, and a carry-case convenient product and lead storage.
APPLICATION
There is a legal requirement for any landlord, employer, or owner of a place of work or public place, to ensure that all electrical equipment accessible by tenants, employees, or the public is maintained in a safe condition, and an acceptable method of ensuring this is by routine electrical testing. This can be performed by electrical contractors, specialist PAT testing organizations, maintenance departments, or facilities management companies.
The PAT300 range is suitable for performing portable appliance testing in locations such as hotels, public houses, schools, colleges, nurseries, shops, offices, theatres, banks, restaurants, cafés, sports and leisure facilities, cinemas, factories, and hair salons, etc. The PAT350 is suitable for use in tool-hire shops where equipment is routinely tested prior to hiring and will perform a flash test in addition to standard PAT tests.
SPECIFICATIONS
Electrical supply range
230 V ±10% ±1 V @ 50 Hz ±10% ±0.1 Hz
Instrument accuracy cannot be guaranteed when operated on supply frequencies
other than those specified above
Bond test
(10 A and 25 A)
Open circuit voltage:
9 V a.c. ±10% ±0.1 V (Supply: 230 V 50 Hz)
Earth bond resistance accuracy:
±5% ±3 digits (0 to 0.49 Ω)
±5% ±5 digits (0.5 to 1.99 Ω)
Earth bond resistance resolution:
10 mΩ (0 to 1.99 Ω)
Display range: 10 to 1.99 Ω
Continuity
Continuity test voltage:
4.0 V d.c. -0% +10
Continuity resistance accuracy:
±5% ±3 digits (0 to 0.49 Ω)
±5% ±5 digits (0.5 to 19.99 Ω)
Continuity resistance resolution:
10 mΩ (1 to 19.99 Ω)
Display range: 0 to 19.99 Ω
Insulation test
(250 V and 500 V)
Insulation resistance accuracy (230 V):
±2% ±5 digits (0 to 19.99 MΩ)
±5% ±10 digits (20 to 99.99 MΩ)
Insulation resistance resolution:
0.01 MΩ
Display range: 0 to 99.99 MΩ
Substitute leakage current
Test voltage and frequency: 40 V a.c. ±10% @ nominal mains frequency
Leakage current accuracy: ±5% ±5 digits
Leakage current resolution: 0.01 mA
Display range: 0 to 19.99 mA
Leakage current
Test voltage: Nominal mains supply
Differential leakage current accuracy: ±5% ± 5 digits
Differential leakage current resolution: 0.01 mA
Display range: 0 to 19.99 mA
Touch current test
Test voltage: Nominal main supply
Touch current accuracy: ±5% ±5 digits
Touch current resolution: 0.01 mA
Display range: 0.00 to 10 mA
Operational test
Test voltage: Nominal mains supply.
VA Accuracy:
±5% ±10 digits (0 to 99 VA)
±5% ±50 digits (100 VA – 999 VA)
±5% ±100 digits (1000 VA – 3700 VA)
Resolution: 1 VA (0 to 3700 VA)
Display range: 0 to 3.99 KVA
Ext Lead Test
Test performed: Bond insulation and polarity.
Polarity test voltage: 12 V dc
Polarity:
Lead OK
Live neutral S/C
Live neutral reversed
Live/neutral O/C
Flash test
(PAT350)
Flash test voltage:
1500 V a.c nominal for Class I
3000 V a.c nominal for Class II
Flash test current:
< 3.5 mA short circuit @ 253 V primary supply voltage
Flash test breakdown current accuracy:
±5% ±5 digits
Flash test breakdown current resolution: 0.01 mA
Display range: 0 to 3.0 mA
RCD tests
RCD test voltage and frequency: Nominal mains supply
Test current accuracy:
-8% to -2% (1/2 x I)
+2% to +8% (1 x I,5 x I)
Trip time accuracy: ±1% ±5 digits
Trip time resolution: 0.1 ms
Display range:
0 to 1999 ms (1/2 x I)
0 to 300 ms (1 x I)
0 to 40 ms (5 x I)
Fuse Test
Test voltage: 3.3 V
Warning: Audible beep if the fuse is OK
Operating temperature range and humidity| -10 to +50 °C
90% RH @ -10 to +30 °C
75% RH @ +30 to +50 °C
---|---
Storage temperature range and humidity| 90% RH @ -10 to +30 °C
75% RH @ +30 to +50 °C
Weight| PAT310 (with packaging) 3.2 kg
PAT320 (with packaging) 4.9 kg
PAT350 (with packaging) 5.5 kg
Dimensions| 210 mm (H) x 280 mm (W) x 390 mm (L)
Dimensions (instrument and case)| PAT310 120 mm (H) x 255 mm
(W) x 320 mm (L) PAT320
155mm (H) x 255 mm (W) x 320 mm (L)
PAT350 155mm (H) x 255 mm (W) x 320 mm (L)
Maximum altitude| 2,000 m
IP rating| IP40
Safety| IEC 61010-1: CAT II 300V
EMC| IEC 61326-1: 2006
IEC 61326-2-2: 2005.
ORDERING INFORMATION
Description | Order Code |
---|---|
PAT310-EU | 1000-742 |
PAT350-EU | 1000-953 |
Included accessories | |
Printed quick start guide | |
Full user guide on CD | |
Calibration certificate | |
Continuity/earth bond lead | 1001-233 |
Extension lead adaptor Schuko | 1001-235 |
Flash test lead | 5310-401 |
Carry case with lead/document pouch | 2000-962 |
Optional accessories | |
Plug adaptor IEC C6 – C13 | 1001-232 |
415 V adaptor lead (4-pin) to SC (CEE7/7) (16 A) | 1000-768 |
415 V adaptor lead (5 pin) to SC (CEE7/7) (16 A) | 1000-771 |
Roll of 1000 FAIL test labels | 1001-227 |
Roll of 1000 PASS test labels | 1000-971 |
PAT accessory pouch | 2001-044 |
PAT test certificate book | 1001-299 |
Screen protector (2) | 1002-572 |
CalPlus GmbH
Zentrale Berlin
Heerstraße 32 • 14052 Berlin
Tel.: 030 214982-0 • Fax: 030 214982-50
office@calplus.de •
www.calplus.de| CalPlus GmbH
Niederlassung Hamburg
Normannenweg 30 • 20537 Hamburg
Tel.: 040 3039595-0 • Fax: 040 3039595-50
scopeshop@calplus.de •
www.calplus.de
---|---
CERTIFICATION ISO
Registered to ISO 9001:2000 Cert. no. Q 09290
PAT300EU_DS_en_V05
www.megger.com
Megger is a registered trademark