Megger PAT300 series Portable Appliance Testers User Manual

June 4, 2024
Megger

PAT300 series
Portable Appliance Testers

PAT300 series
Portable Appliance Testers

PAT300 series Portable Appliance Testers

  • Simple manual or automatic testing
  • Bond testing at 25 A, 10 A, and 200 mA
  • Differential, touch, and substitute leakage testing
  • Full-color display
  • Tests portable RCDs and extension leads
  • Pre-set test sequences for Class I, Class II, and extension leads
  • Flash Testing at 1.5 kV and 3 kV
  • Tests to DIN VDE 0701-0702

DESCRIPTION

An easily portable desktop appliance tester for testing the safety of portable electrical equipment to meet health and safety regulations. The PAT300 series are fully featured testers with dedicated test buttons for direct access to tests. They are designed for customers who do not require the complexity of a fully configurable database of clients and results within the tester but do need a complete range of functions to allow automatic or manual testing of the widest range of electrical assets. There are two products in the range – the PAT320 and the PAT350. The PAT350 is identical to the PAT320, with the addition of flash-test capability for use in environments such as manufacturing, production, or tool-hire shops.
The simple push-button operation makes the PAT300s fast and intuitive in use. All regulatory test requirements are supported, including Class I and Class II, IEC power leads, extension leads, and full tests for portable RCDs. An automatic mode is available for Class I and Class II testing. In automatic mode, the tests proceed sequentially through bond, insulation, and operation, indicating a pass or fail at each test. If a failure occurs, testing is stopped. When manual testing, each test is preceded by a selection screen where the test parameters are selected, such as bond test current, insulation test voltage, or leakage test type. These diagnostic buttons provide direct access to any test individually, allowing single tests to be performed following repair or a suspect result.

Accessories supplied with all models include a combined earth-bond and insulation test lead, an adaptor for testing extension leads, and a carry-case convenient product and lead storage.

APPLICATION

There is a legal requirement for any landlord, employer, or owner of a place of work or public place, to ensure that all electrical equipment accessible by tenants, employees, or the public is maintained in a safe condition, and an acceptable method of ensuring this is by routine electrical testing. This can be performed by electrical contractors, specialist PAT testing organizations, maintenance departments, or facilities management companies.

The PAT300 range is suitable for performing portable appliance testing in locations such as hotels, public houses, schools, colleges, nurseries, shops, offices, theatres, banks, restaurants, cafés, sports and leisure facilities, cinemas, factories, and hair salons, etc. The PAT350 is suitable for use in tool-hire shops where equipment is routinely tested prior to hiring and will perform a flash test in addition to standard PAT tests.

SPECIFICATIONS

Electrical supply range
230 V ±10% ±1 V @ 50 Hz ±10% ±0.1 Hz
Instrument accuracy cannot be guaranteed when operated on supply frequencies other than those specified above

Bond test
(10 A and 25 A)
Open circuit voltage:
9 V a.c. ±10% ±0.1 V (Supply: 230 V 50 Hz)
Earth bond resistance accuracy:
±5% ±3 digits (0 to 0.49 Ω)
±5% ±5 digits (0.5 to 1.99 Ω)
Earth bond resistance resolution:
10 mΩ (0 to 1.99 Ω)
Display range: 10 to 1.99 Ω

Continuity
Continuity test voltage:
4.0 V d.c. -0% +10
Continuity resistance accuracy:
±5% ±3 digits (0 to 0.49 Ω)
±5% ±5 digits (0.5 to 19.99 Ω)
Continuity resistance resolution:
10 mΩ (1 to 19.99 Ω)
Display range: 0 to 19.99 Ω

Insulation test
(250 V and 500 V)
Insulation resistance accuracy (230 V):
±2% ±5 digits (0 to 19.99 MΩ)
±5% ±10 digits (20 to 99.99 MΩ)
Insulation resistance resolution:
0.01 MΩ
Display range: 0 to 99.99 MΩ

Substitute leakage current
Test voltage and frequency: 40 V a.c. ±10% @ nominal mains frequency
Leakage current accuracy: ±5% ±5 digits
Leakage current resolution: 0.01 mA
Display range: 0 to 19.99 mA

Leakage current
Test voltage: Nominal mains supply
Differential leakage current accuracy: ±5% ± 5 digits
Differential leakage current resolution: 0.01 mA
Display range: 0 to 19.99 mA

Touch current test
Test voltage: Nominal main supply
Touch current accuracy: ±5% ±5 digits
Touch current resolution: 0.01 mA
Display range: 0.00 to 10 mA

Operational test
Test voltage: Nominal mains supply.
VA Accuracy:
±5% ±10 digits (0 to 99 VA)
±5% ±50 digits (100 VA – 999 VA)
±5% ±100 digits (1000 VA – 3700 VA)
Resolution: 1 VA (0 to 3700 VA)
Display range: 0 to 3.99 KVA

Ext Lead Test
Test performed: Bond insulation and polarity.
Polarity test voltage: 12 V dc
Polarity:
Lead OK
Live neutral S/C
Live neutral reversed
Live/neutral O/C

Flash test
(PAT350)
Flash test voltage:
1500 V a.c nominal for Class I
3000 V a.c nominal for Class II
Flash test current:
< 3.5 mA short circuit @ 253 V primary supply voltage
Flash test breakdown current accuracy:
±5% ±5 digits
Flash test breakdown current resolution: 0.01 mA
Display range: 0 to 3.0 mA

RCD tests
RCD test voltage and frequency: Nominal mains supply
Test current accuracy:
-8% to -2% (1/2 x I)
+2% to +8% (1 x I,5 x I)
Trip time accuracy: ±1% ±5 digits
Trip time resolution: 0.1 ms
Display range:
0 to 1999 ms (1/2 x I)
0 to 300 ms (1 x I)
0 to 40 ms (5 x I)

Fuse Test
Test voltage: 3.3 V
Warning: Audible beep if the fuse is OK

Operating temperature range and humidity| -10 to +50 °C
90% RH @ -10 to +30 °C
75% RH @ +30 to +50 °C
---|---
Storage temperature range and humidity| 90% RH @ -10 to +30 °C
75% RH @ +30 to +50 °C
Weight| PAT310 (with packaging) 3.2 kg
PAT320 (with packaging) 4.9 kg
PAT350 (with packaging) 5.5 kg
Dimensions| 210 mm (H) x 280 mm (W) x 390 mm (L)
Dimensions (instrument and case)| PAT310 120 mm (H) x 255 mm
(W) x 320 mm (L) PAT320
155mm (H) x 255 mm (W) x 320 mm (L)
PAT350 155mm (H) x 255 mm (W) x 320 mm (L)
Maximum altitude| 2,000 m
IP rating| IP40
Safety| IEC 61010-1: CAT II 300V
EMC| IEC 61326-1: 2006
IEC 61326-2-2: 2005.

ORDERING INFORMATION

Description Order Code
PAT310-EU  1000-742
PAT350-EU  1000-953
Included accessories
Printed quick start guide
Full user guide on CD
Calibration certificate
Continuity/earth bond lead 1001-233
Extension lead adaptor Schuko 1001-235
Flash test lead 5310-401
Carry case with lead/document pouch 2000-962
Optional accessories
Plug adaptor IEC C6 – C13 1001-232
415 V adaptor lead (4-pin) to SC (CEE7/7) (16 A) 1000-768
415 V adaptor lead (5 pin) to SC (CEE7/7) (16 A) 1000-771
Roll of 1000 FAIL test labels 1001-227
Roll of 1000 PASS test labels 1000-971
PAT accessory pouch 2001-044
PAT test certificate book 1001-299
Screen protector (2) 1002-572

CalPlus GmbH
Zentrale Berlin
Heerstraße 32 • 14052 Berlin
Tel.: 030 214982-0 • Fax: 030 214982-50
[email protected]www.calplus.de| CalPlus GmbH
Niederlassung Hamburg
Normannenweg 30 • 20537 Hamburg
Tel.: 040 3039595-0 • Fax: 040 3039595-50
[email protected]www.calplus.de
---|---

CERTIFICATION ISO
Registered to ISO 9001:2000 Cert. no. Q 09290
PAT300EU_DS_en_V05
www.megger.com
Megger is a registered trademark